DQR ID | Subject | Data Streams Affected |
---|---|---|
D030527.2 | SGP/SIRS/E27 - Instrument problem, Shadowband Misalignment | sgpsirsE27.00, sgpsirsE27.b1, sgpsirs20sE27.a0 |
D030715.2 | SGP/SIRS/E5 - Downwelling direct normal (NIP) reporting low | sgpsirsE5.00, sgpsirsE5.b1, sgpsirs20sE5.a0 |
D040712.2 | SGP/SIRS/E24 - NIP shaded near sunrise | sgpsirsE24.a2, sgpsirsE24.b1, sgpsirs20sE24.a0, sgpsirs60sE24.a1 |
D040712.3 | SGP/SIRS/E24 - NIP shaded near sunrise | sgpsirsE24.b1, sgpsirs20sE24.a0 |
D060630.23 | SGP/SIRS/E16 - Reprocessed: Longwave Calibration error | sgpsirsE16.b1 |
D060630.29 | SGP/SIRS/E24 - Reprocessed: Longwave Calibration error | sgpsirsE24.b1 |
D060630.30 | SGP/SIRS/E27 - Reprocessed: Longwave Calibration error | sgpsirsE27.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/16/2003 | 0000 | 06/18/2003 | 1845 |
Subject: | SGP/SIRS/E27 - Instrument problem, Shadowband Misalignment |
DataStreams: | sgpsirsE27.00, sgpsirsE27.b1, sgpsirs20sE27.a0 |
Description: | Downwelling and Upwelling longwave values are somewhat noisy due to noisy PIR (pygeometer) case and dome thermistor measurements on both the upwelling and downwelling PIRs. This problem dates back to the beginning of the SIRS E27 data stream. |
Measurements: | sgpsirsE27.00:
sgpsirsE27.b1:
sgpsirs20sE27.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/11/2003 | 1630 | 07/16/2003 | 1840 |
Subject: | SGP/SIRS/E5 - Downwelling direct normal (NIP) reporting low |
DataStreams: | sgpsirsE5.00, sgpsirsE5.b1, sgpsirs20sE5.a0 |
Description: | There were large differences between the measured downwelling shortwave (DS) and the downwelling shortwave calculated from direct (NIP) and diffuse (DD) readings. It appears that the direct normal shortwave readings were much too low, particularly on 7/14 and 7/15. It appears that the diffuse instrument was being shaded properly, but the NIP was not tracking the sun. The problem was corrected during the preventative maintenance visit on 7/16/03 at 1840GMT. The technician reported doing the quarterly connection corrosion checks and ventilator fan cleaning, the NIP was described as having "dry-light dust" on the window, no mention of tracker alignment was made. Although uncertain, a bad connection at some point in the instrument wiring was likely to blame and was eliminated during the corrosion inspection process. |
Measurements: | sgpsirsE5.b1:
sgpsirs20sE5.a0:
sgpsirsE5.00:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/08/2002 | 1200 | 05/20/2002 | 1300 |
07/26/2002 | 1300 | 09/06/2002 | 1200 |
04/08/2003 | 1200 | 04/16/2003 | 1300 |
Subject: | SGP/SIRS/E24 - NIP shaded near sunrise |
DataStreams: | sgpsirsE24.a2, sgpsirsE24.b1, sgpsirs20sE24.a0, sgpsirs60sE24.a1 |
Description: | The NIP is shaded by the EF laptop computer enclosure near sunrise during parts of the Spring and Summer. This has occurred from 04/08 to 05/20 and from 07/26 to 09/06 every year since 2001 and will occur during these date ranges in the future as well. The shading occurs for a few minutes each day between 1200 and 1300 GMT depending on the time of the year. For the early dates, shading occurs a little after 1200 GMT and by 05/20 shading happens nearer to 1300 GMT. In late July the shading occurs near 1300 GMT and closer to 1200 GMT by 09/06. |
Measurements: | sgpsirs20sE24.a0:
sgpsirs60sE24.a1:
sgpsirsE24.b1:
sgpsirsE24.a2:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/17/2003 | 1200 | 05/20/2003 | 1300 |
07/26/2003 | 1300 | 09/06/2003 | 1200 |
04/08/2004 | 1200 | 05/20/2004 | 1300 |
07/26/2004 | 1300 | 09/06/2004 | 1200 |
Subject: | SGP/SIRS/E24 - NIP shaded near sunrise |
DataStreams: | sgpsirsE24.b1, sgpsirs20sE24.a0 |
Description: | The NIP is shaded by the EF laptop computer enclosure near sunrise during parts of the Spring and Summer. This has occurred from 04/08 to 05/20 and from 07/26 to 09/06 every year since 2001 and will occur during these date ranges in the future as well. The shading occurs for a few minutes each day between 1200 and 1300 GMT depending on the time of the year. For the early dates, shading occurs a little after 1200 GMT and by 05/20 shading happens nearer to 1300 GMT. In late July the shading occurs near 1300 GMT and closer to 1200 GMT by 09/06. |
Measurements: | sgpsirs20sE24.a0:
sgpsirsE24.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/23/2003 | 1605 | 02/15/2006 | 1533 |
Subject: | SGP/SIRS/E16 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE16.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in July 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in October 2006. |
Measurements: | sgpsirsE16.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/24/2003 | 1640 | 02/16/2006 | 1705 |
Subject: | SGP/SIRS/E24 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE24.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in July 2004. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in October 2006. |
Measurements: | sgpsirsE24.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/16/2003 | 0002 | 02/15/2006 | 1750 |
Subject: | SGP/SIRS/E27 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE27.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in May 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in October 2006. |
Measurements: | sgpsirsE27.b1:
|