Data Quality Reports for Session: 108056 User: troyan Completed: 10/01/2007


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D050725.1SGP/SONDE/C1 - Bad RH DatasgpsondewnpnC1.b1
D050811.5SGP/SONDE/C1 - Bad T/RH and surface P.sgpsondewnpnC1.b1
D050912.2SGP/SONDE/C1 - Temperature/RH incorrectsgplssondeC1.c1, sgpsondewnpnC1.b1
D051027.1SGP/SONDE/C1 - Bad RH DatasgpsondewnpnC1.b1
D060928.3SGP/SONDE/C1 - Increased precision for alt, dp, pres, rh, tdrysgpsondewnpnC1.b1


DQRID : D050725.1
Start DateStart TimeEnd DateEnd Time
07/17/2005234707/18/20050141
Subject:
SGP/SONDE/C1 - Bad RH Data
DataStreams:sgpsondewnpnC1.b1
Description:
The RH values below 6 km are incorrect.  The problem relates to a failure of the GC25 
ground check set used before this particular sounding.  The failure seems to have affected 
the sensor calibration with the effect that one of the two sensors was reading artificially 
high.  Because the internal processing software uses the highest of the two sensor 
readings, the bad sensor value is chosen whenever it is not heated.
Measurements:sgpsondewnpnC1.b1:
  • Surface dew point temperature(dp)
  • Relative humidity inside the instrument enclosure(rh)


Back To Table of Contents

DQRID : D050811.5
Start DateStart TimeEnd DateEnd Time
08/01/2005113208/01/20050116
Subject:
SGP/SONDE/C1 - Bad T/RH and surface P.
DataStreams:sgpsondewnpnC1.b1
Description:
Apparently multiple failures.  The radiosonde temperature sensor broke on launch causing 
an approximate 35 degC negative offset and as a result, the RH is incorrect.  Furthermore, 
it appears that the pressure sensor was off by approximately 4 hPa, though this would 
not affect the altitude calculation, the absolute pressure is wrong.
Measurements:sgpsondewnpnC1.b1:
  • Dry bulb temperature(tdry)
  • Relative humidity inside the instrument enclosure(rh)
  • Retrieved pressure profile(pres)


Back To Table of Contents

DQRID : D050912.2
Start DateStart TimeEnd DateEnd Time
09/09/2005113209/09/20051332
Subject:
SGP/SONDE/C1 - Temperature/RH incorrect
DataStreams:sgplssondeC1.c1, sgpsondewnpnC1.b1
Description:
The temperature throughout this sounding is too high and as a result the RH values will be 
incorrect as well.  I'm not sure what is wrong here but the operators noted reasonable 
values (18.1 degC, 87%) in the log before launch.  Review of the raw data, however, shows 
the sonde reporting temperatures nearly 20 degC higher.  I suppose it is possible that 
the operator entered into the log the reference temperature from the GC25 ground check set 
rather than the temperature reported by the sonde - I can't think of any other 
explanation - there is no doubt that the temperature sensor on this sonde was bad.
Measurements:sgpsondewnpnC1.b1:
  • Surface dew point temperature(dp)
  • Dry bulb temperature(tdry)
  • Relative humidity inside the instrument enclosure(rh)

sgplssondeC1.c1:
  • Relative humidity inside the instrument enclosure(rh)
  • Dry bulb temperature(tdry)


Back To Table of Contents

DQRID : D051027.1
Start DateStart TimeEnd DateEnd Time
10/24/2005114010/24/20051204
Subject:
SGP/SONDE/C1 - Bad RH Data
DataStreams:sgpsondewnpnC1.b1
Description:
The RH values during this sounding oscillate between reasonable values and near-zero 
values.  The oscillation is due to failure of the RH sensor heating process.
Measurements:sgpsondewnpnC1.b1:
  • Surface dew point temperature(dp)
  • Relative humidity inside the instrument enclosure(rh)


Back To Table of Contents

DQRID : D060928.3
Start DateStart TimeEnd DateEnd Time
09/28/2006172812/31/99992359
Subject:
SGP/SONDE/C1 - Increased precision for alt, dp, pres, rh, tdry
DataStreams:sgpsondewnpnC1.b1
Description:
The output precision of the following variables was increased 20060928

alt  changed from 1m to 0.1m
dp   changed from 0.1degC to 0.01degC
pres changed from 0.1 hPa to 0.01 hPa
RH   changed from 1%RH to 0.01%RH
tdry changed from .1degC to 0.01degC

Note that the increase in precision does not necessarily imply an increase in accuracy.  
Users are cautioned to continue to use their best scientific judgement when using these or 
any other data.
Measurements:sgpsondewnpnC1.b1:
  • Surface dew point temperature(dp)
  • Dry bulb temperature(tdry)
  • Relative humidity inside the instrument enclosure(rh)
  • Retrieved pressure profile(pres)
  • Dummy altitude for Zeb(alt)


Back To Table of Contents



END OF DATA