Subject: | TWP/VCEIL/C3 - Reduced sensitivity to high clouds |
DataStreams: | twpvceil25kC3.b1 |
Description: | The instrument had reduced sensitivity to high clouds. The instrument produced a significant number of optical cross-talk warnings and displayed a reduced capacity to detect high clouds. The window was cleaned and the instrument was power-cycled but with no improvement. Diagnostics indicated the problem was with the compensation fiber cable. The problem was corrected by replacing this standard-window instrument (that was tilted at 16 degrees) with a tropics unit that became available. |