DQR ID | Subject | Data Streams Affected |
---|---|---|
D070416.3 | NSA/MWRP/C1 - Incorrect surface measurements | nsamwrpC1.b1 |
D070709.1 | NSA/MWRP/C1 - Reprocessed: 6-channel retrieval replaced by 5-channel retrieval | nsamwrpC1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/28/2007 | 0000 | 03/29/2007 | 2300 |
Subject: | NSA/MWRP/C1 - Incorrect surface measurements |
DataStreams: | nsamwrpC1.b1 |
Description: | The MWRP surface temperature and pressure readings were in disagreement with those from other instruments. The possible cause of the disagreement was ice buildup around the sensors. Readings returned to normal March 29. The temperature readings were more affected. Since the readings are erratic, with large oscillations, it is not possible to correct them. Disregard the surface temperature and humidity readings during this time. The wrong readings will affect the first layer of the temperature and humidity profiles as well. |
Measurements: | nsamwrpC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/19/2004 | 0214 | 06/02/2007 | 2359 |
Subject: | NSA/MWRP/C1 - Reprocessed: 6-channel retrieval replaced by 5-channel retrieval |
DataStreams: | nsamwrpC1.b1 |
Description: | Due to repeated failure of the V-band receiver and calibration problems, the 6-channel retrieval was retired and a new 5-channel retrieval which uses only the K-band channels was implemented. All data previously processed with the 6-channel retrieval have been reprocessed. |
Measurements: | nsamwrpC1.b1:
|