DQR ID | Subject | Data Streams Affected |
---|---|---|
D021211.4 | NSA/GNDRAD/C1 - IRT removed for testing | nsagndrad60sC1.a1 |
D030510.2 | NSA/SKYRAD/C1 - IRT measurements during precipitation events | nsaskyrad60sC1.a1 |
D990924.1 | NSA/GNDRAD/C1 - IRT insufficiently heated | nsagndrad60sC1.a1 |
D990924.2 | NSA/SKYRAD/C1 - IRT insufficiently heated | nsaskyrad60sC1.a1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/29/1999 | 1900 | 05/19/2000 | 0000 |
Subject: | NSA/GNDRAD/C1 - IRT removed for testing |
DataStreams: | nsagndrad60sC1.a1 |
Description: | The IRT instrument was at PNNL for cold temperature modifications and testing. Instruments returned for the summer season. Data collected after the instrument return looks reasonable. IRT data recorded during the period the instrument was removed is bad. |
Measurements: | nsagndrad60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/14/1998 | 2330 | 03/03/2003 | 2359 |
Subject: | NSA/SKYRAD/C1 - IRT measurements during precipitation events |
DataStreams: | nsaskyrad60sC1.a1 |
Description: | The IRT measurement of sky/cloud temperature was in question during precipitation events because rain or snow could accumulate in the upward-pointed lens barrel. This was corrected by reorienting the IRT to measure the sky temperature as reflected by a gold mirror. |
Measurements: | nsaskyrad60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/06/1998 | 2156 | 12/22/2000 | 2020 |
Subject: | NSA/GNDRAD/C1 - IRT insufficiently heated |
DataStreams: | nsagndrad60sC1.a1 |
Description: | IRT had insufficient heating to maintain an internal reference temperature above 0 degrees C. Measurements of sky temperature were over-estimated when instrument was below freezing. |
Measurements: | nsagndrad60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/14/1998 | 0130 | 12/21/2000 | 1925 |
Subject: | NSA/SKYRAD/C1 - IRT insufficiently heated |
DataStreams: | nsaskyrad60sC1.a1 |
Description: | IRT had insufficient heating to maintain an internal reference temperature above 0 degrees C. Measurements of sky temperature were over-estimated when instrument was below freezing. |
Measurements: | nsaskyrad60sC1.a1:
|