DQR ID | Subject | Data Streams Affected |
---|---|---|
D040420.4 | SGP/SIRS/E27 - incorrect tracking | sgpsirs20sE27.a0, sgpsirsE27.00, sgpsirsE27.b1 |
D050106.4 | SGP/SIRS/C1 - Reprocess: UIR/DIR Incorrect calibration coefficients | sgpsirsC1.00, sgpsirsC1.b1 |
D060630.10 | SGP/SIRS/E2 - Reprocessed: Longwave Calibration error | sgpsirsE2.b1 |
D060630.11 | SGP/SIRS/E3 - Reprocessed: Longwave Calibration error | sgpsirsE3.b1 |
D060630.12 | SGP/SIRS/E4 - Reprocessed: Longwave Calibration error | sgpsirsE4.b1 |
D060630.13 | SGP/SIRS/E5 - Reprocessed: Longwave Calibration error | sgpsirsE5.b1 |
D060630.14 | SGP/SIRS/E6 - Reprocessed: Longwave Calibration error | sgpsirsE6.b1 |
D060630.15 | SGP/SIRS/E7 - Reprocessed: Longwave Calibration error | sgpsirsE7.b1 |
D060630.16 | SGP/SIRS/E8 - Reprocessed: Longwave Calibration error | sgpsirsE8.b1 |
D060630.17 | SGP/SIRS/E9 - Reprocessed: Longwave Calibration error | sgpsirsE9.b1 |
D060630.18 | SGP/SIRS/E10 - Reprocessed: Longwave Calibration error | sgpsirsE10.b1 |
D060630.19 | SGP/SIRS/E11 - Reprocessed: Longwave Calibration error | sgpsirsE11.b1 |
D060630.20 | SGP/SIRS/E12 - Reprocessed: Longwave Calibration error | sgpsirsE12.b1 |
D060630.21 | SGP/SIRS/E13 - Reprocessed: Longwave Calibration error | sgpsirsE13.b1 |
D060630.22 | SGP/SIRS/E15 - Reprocessed: Longwave Calibration error | sgpsirsE15.b1 |
D060630.23 | SGP/SIRS/E16 - Reprocessed: Longwave Calibration error | sgpsirsE16.b1 |
D060630.24 | SGP/SIRS/E18 - Reprocessed: Longwave Calibration error | sgpsirsE18.b1 |
D060630.25 | SGP/SIRS/E19 - Reprocessed: Longwave Calibration error | sgpsirsE19.b1 |
D060630.26 | SGP/SIRS/E20 - Reprocessed: Longwave Calibration error | sgpsirsE20.b1 |
D060630.27 | SGP/SIRS/E21 - Reprocessed: Longwave Calibration error | sgpsirsE21.b1 |
D060630.28 | SGP/SIRS/E22 - Reprocessed: Longwave Calibration error | sgpsirsE22.b1 |
D060630.29 | SGP/SIRS/E24 - Reprocessed: Longwave Calibration error | sgpsirsE24.b1 |
D060630.30 | SGP/SIRS/E27 - Reprocessed: Longwave Calibration error | sgpsirsE27.b1 |
D060630.8 | SGP/SIRS/C1 - Reprocessed: Longwave Calibration error | sgpsirsC1.b1 |
D060630.9 | SGP/SIRS/E1 - Reprocessed: Longwave Calibration error | sgpsirsE1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/14/2004 | 1700 | 09/15/2004 | 1650 |
10/27/2004 | 1500 | 11/05/2004 | 1900 |
Subject: | SGP/SIRS/E27 - incorrect tracking |
DataStreams: | sgpsirs20sE27.a0, sgpsirsE27.00, sgpsirsE27.b1 |
Description: | The solar tracker at E27 was not functioning properly. The problem affected tracker mounted instruments: downwelling diffuse 8-48 and direct normal pyrheliometer NIP. After tracker adjustment on 4/14/2004 1700GMT the tracker stopped tracking properly during afternoon hours. Attempts to correct tracker operation throughout the failure period resulted in variations on incorrect tracker operation. No replacement tracker was available. After adjustment on 9/15/2004 the tracker problem was resolved. It is unclear what, other than typical az and zen tracker software changes by field techs, corrected the problem. This problem period corresponds to the first time range specified (4/14 to 9/15). On 10/27/2004 a replacement tracker was installed at E27. Technical difficulties kept tracker out of service until 11/05/2004. The replacement tracker at E27 was put into serice by the field techs on 11/05/2004 1830GMT. The tracker has been tracking correctly since this time. This problem period corresponds to the second time range (10/27 to 11/05). |
Measurements: | sgpsirsE27.00:
sgpsirsE27.b1:
sgpsirs20sE27.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/27/2004 | 1429 | 01/07/2005 | 1623 |
Subject: | SGP/SIRS/C1 - Reprocess: UIR/DIR Incorrect calibration coefficients |
DataStreams: | sgpsirsC1.00, sgpsirsC1.b1 |
Description: | It appears the datalogger program at SIRS/C1 was automatically updated, unbeknownst to anyone, on 9/27/20041423GMT with an older program with incorrect UIR and DIR calibration coefficients. The correct and incorrect cal coeffs for the UIR and DIR pyrgeometers during this DQR period were: incorrectUIR correctUIR K0 -12.84 -13.00 K1 .2250 0.2573 K2 1.024 1.024 K3 -1.9280 -1.8050 incorrectDIR correctDIR K0 -13.15 -26.89 K1 0.2374 0.2967 K2 1.0260 1.0560 K3 -2.478 -3.3080 |
Measurements: | sgpsirsC1.b1:
sgpsirsC1.00:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/12/2004 | 1630 | 02/23/2005 | 1615 |
Subject: | SGP/SIRS/E2 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE2.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in February 2004. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instanteous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in September 2006. |
Measurements: | sgpsirsE2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/10/2004 | 1715 | 02/08/2006 | 1730 |
Subject: | SGP/SIRS/E3 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE3.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in March 2004. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in August 2006. |
Measurements: | sgpsirsE3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/10/2004 | 1720 | 02/22/2006 | 1615 |
Subject: | SGP/SIRS/E4 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE4.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in March 2004. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in October 2006. |
Measurements: | sgpsirsE4.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/10/2004 | 2030 | 02/22/2006 | 1817 |
Subject: | SGP/SIRS/E5 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE5.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in March 2004. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in October 2006. |
Measurements: | sgpsirsE5.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/11/2004 | 1610 | 02/09/2006 | 1553 |
Subject: | SGP/SIRS/E6 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE6.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in March 2004. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in October 2006. |
Measurements: | sgpsirsE6.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/20/2004 | 2000 | 02/07/2006 | 2005 |
Subject: | SGP/SIRS/E7 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE7.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in February 2004. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in October 2006. |
Measurements: | sgpsirsE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/10/2004 | 1845 | 02/21/2006 | 1840 |
Subject: | SGP/SIRS/E8 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE8.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in February 2004. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instanteous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in September 2006. |
Measurements: | sgpsirsE8.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/17/2003 | 1640 | 02/07/2006 | 1730 |
Subject: | SGP/SIRS/E9 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE9.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in December 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instanteous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in October 2006. |
Measurements: | sgpsirsE9.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/20/2004 | 1730 | 02/07/2006 | 2130 |
Subject: | SGP/SIRS/E10 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE10.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in February 2004. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in August 2006. |
Measurements: | sgpsirsE10.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/17/2004 | 1753 | 02/14/2006 | 2016 |
Subject: | SGP/SIRS/E11 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE11.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in February 2004. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in August 2006. |
Measurements: | sgpsirsE11.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/22/2003 | 1640 | 02/14/2006 | 1710 |
Subject: | SGP/SIRS/E12 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE12.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in July 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in October 2006. |
Measurements: | sgpsirsE12.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/11/2003 | 1935 | 02/16/2006 | 1905 |
Subject: | SGP/SIRS/E13 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE13.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in December 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in October 2006. |
Measurements: | sgpsirsE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/22/2003 | 1650 | 02/14/2006 | 1806 |
Subject: | SGP/SIRS/E15 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE15.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in July 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in October 2006. |
Measurements: | sgpsirsE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/23/2003 | 1605 | 02/15/2006 | 1533 |
Subject: | SGP/SIRS/E16 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE16.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in July 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in October 2006. |
Measurements: | sgpsirsE16.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/22/2003 | 2200 | 02/14/2006 | 2100 |
Subject: | SGP/SIRS/E18 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE18.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in July 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in October 2006. |
Measurements: | sgpsirsE18.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/24/2003 | 1600 | 02/16/2006 | 1558 |
Subject: | SGP/SIRS/E19 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE19.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in July 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in October 2006. |
Measurements: | sgpsirsE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/23/2003 | 2100 | 02/15/2006 | 2030 |
Subject: | SGP/SIRS/E20 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE20.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in July 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in November 2006. |
Measurements: | sgpsirsE20.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/23/2003 | 1520 | 02/15/2006 | 1515 |
Subject: | SGP/SIRS/E21 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE21.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in July 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in November 2006. |
Measurements: | sgpsirsE21.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/23/2003 | 1830 | 02/15/2006 | 1741 |
Subject: | SGP/SIRS/E22 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE22.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in July 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in August 2006. |
Measurements: | sgpsirsE22.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/24/2003 | 1640 | 02/16/2006 | 1705 |
Subject: | SGP/SIRS/E24 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE24.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in July 2004. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in October 2006. |
Measurements: | sgpsirsE24.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/16/2003 | 0002 | 02/15/2006 | 1750 |
Subject: | SGP/SIRS/E27 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE27.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in May 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in October 2006. |
Measurements: | sgpsirsE27.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/11/2003 | 1950 | 02/16/2006 | 1901 |
Subject: | SGP/SIRS/C1 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsC1.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in December 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in October 2006. |
Measurements: | sgpsirsC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/10/2004 | 2055 | 02/21/2006 | 2007 |
Subject: | SGP/SIRS/E1 - Reprocessed: Longwave Calibration error |
DataStreams: | sgpsirsE1.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in February 2004. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place have been reprocessed to remove the calibration bias. The reprocessed 60 second averaged data are based on 3 instantaneous 20 second data records rather than on 60 1 second instantaneous data records. Still, these data are considered far superior to the originally processed data. The reprocessed data were archived in October 2006. |
Measurements: | sgpsirsE1.b1:
|