DQR ID | Subject | Data Streams Affected |
---|---|---|
D030819.2 | NSA/SKYRAD/C1 - IRT failed | nsaskyrad60sC1.a1, nsaskyrad60sC1.b1 |
D030820.1 | NSA/SKYRAD/C1 - IRT incorrectly configured | nsaskyrad60sC1.a1, nsaskyrad60sC1.b1 |
D030821.2 | NSA/SKYRAD/C1 - IRT incorrect min/max value | nsaskyrad60sC1.b1 |
D031007.2 | NSA/SKYRAD/C1 - IRT measurements during precipitation events | nsaskyrad60sC1.b1 |
D040213.4 | NSA/SKYRAD/C1 - Downwelling IRT bias | nsaskyrad60sC1.b1 |
D040421.2 | NSA/SKYRAD/C1 - Tracker Inoperative | nsaskyrad60sC1.b1 |
D040421.3 | NSA/SKYRAD/C1 - itermittant IRT bias | nsaskyrad60sC1.b1 |
D051101.2 | NSA/SKYRAD/C1 - IRT failure | nsaskyrad60sC1.b1 |
D060209.5 | NSA/SKYRAD/C1 - Power outage affecting some SKYRAD instruments | nsaskyrad20sC1.a0, nsaskyrad60sC1.b1, nsaskyradC1.00 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/05/2002 | 2330 | 06/27/2002 | 1920 |
Subject: | NSA/SKYRAD/C1 - IRT failed |
DataStreams: | nsaskyrad60sC1.a1, nsaskyrad60sC1.b1 |
Description: | |
Measurements: | nsaskyrad60sC1.a1:
nsaskyrad60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/27/2002 | 1920 | 07/10/2002 | 1900 |
Subject: | NSA/SKYRAD/C1 - IRT incorrectly configured |
DataStreams: | nsaskyrad60sC1.a1, nsaskyrad60sC1.b1 |
Description: | The IRT was removed for repair and replaced with a spare that was not configured properly. Data are incorrect. |
Measurements: | nsaskyrad60sC1.a1:
nsaskyrad60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/01/2001 | 0000 | 03/03/2003 | 2359 |
Subject: | NSA/SKYRAD/C1 - IRT incorrect min/max value |
DataStreams: | nsaskyrad60sC1.b1 |
Description: | The values of valid_min and valid_max for field sky_ir_temp were incorrectly assigned to -49.7 and +50.0 C, respectively. The actual values should be -60 and +40 C. |
Measurements: | nsaskyrad60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/01/1999 | 0000 | 03/28/2003 | 0200 |
Subject: | NSA/SKYRAD/C1 - IRT measurements during precipitation events |
DataStreams: | nsaskyrad60sC1.b1 |
Description: | The IRT measurement of sky/cloud temperature was in question during precipitation events because rain or snow could accumulate in the upward-pointed lens barrel. This was corrected by reorienting the IRT to measure the sky temperature as reflected by a gold mirror. |
Measurements: | nsaskyrad60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/11/2003 | 1200 | 01/09/2004 | 2120 |
Subject: | NSA/SKYRAD/C1 - Downwelling IRT bias |
DataStreams: | nsaskyrad60sC1.b1 |
Description: | The downwelling IRT was measuring higher sky/cloud temperatures than the AERI. The problem was corrected when the mirror was replaced. |
Measurements: | nsaskyrad60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/10/2003 | 0000 | 02/16/2004 | 0230 |
Subject: | NSA/SKYRAD/C1 - Tracker Inoperative |
DataStreams: | nsaskyrad60sC1.b1 |
Description: | A blown fuse resulted in the tracker not shading the PIR and B&W and the NIP not following the sun. |
Measurements: | nsaskyrad60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/28/2003 | 0200 | 06/21/2006 | 0215 |
Subject: | NSA/SKYRAD/C1 - itermittant IRT bias |
DataStreams: | nsaskyrad60sC1.b1 |
Description: | A problem with the SKYRAD IRT that itermittantly produced a gradual increase in the sky temperature measurement then a sudden correction, as compared to AERI, appears to have been caused by a combination of a failing IR detector, a poor signal cable, a contaminated (with condensation or frost) mirror or lens, and low ambient temperatures. The positive bias was apparently not consistant in either frequency or duration but on days when the unusual behavior is observed, it usually occurred from about 00:00 to 06:00 local time. The instrument was repaired and recalibrated and the signal cable replaced. |
Measurements: | nsaskyrad60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/31/2005 | 0600 | 01/18/2006 | 0032 |
Subject: | NSA/SKYRAD/C1 - IRT failure |
DataStreams: | nsaskyrad60sC1.b1 |
Description: | The IRT failed and was replaced with a spare. |
Measurements: | nsaskyrad60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/07/2006 | 0910 | 02/07/2006 | 2350 |
Subject: | NSA/SKYRAD/C1 - Power outage affecting some SKYRAD instruments |
DataStreams: | nsaskyrad20sC1.a0, nsaskyrad60sC1.b1, nsaskyradC1.00 |
Description: | During the power outage the downwelling global measurement from the PSP instrument and downwelling diffuse 8-48 instrument shows a slight increase from -5 to +0 W/m2. This is due to removal of thermal offsets induced by the AC heater during the power outage. This heater is installed within the PSP ventilator - it is necessary for the NSA/C1 climate. The downwelling longwave PIR instrument is unaffected due to case and dome temperature corrections. The IRT is AC powered and was not operational during this time period. |
Measurements: | nsaskyradC1.00:
nsaskyrad60sC1.b1:
nsaskyrad20sC1.a0:
|