DQR ID | Subject | Data Streams Affected |
---|---|---|
D061213.1 | NSA/METTWR/C1 - CMH Sensor Calibrated Incorrectly | nsamettwr4hC1.b1 |
D061213.2 | NSA/METTWR/C1 - 40m wind sensor heater failure | nsamettwr4hC1.b1 |
D070226.1 | NSA/METTWR/C1 - 40m Wind Direction Data Suspect | nsamettwr4hC1.b1 |
D070413.1 | NSA/METTWR/C1 - PWS Data Questionable | nsamettwr4hC1.b1 |
D070529.1 | NSA/METTWR/C1 - 20m RH Data Questionable | nsamettwr4hC1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/12/2006 | 1500 | 03/14/2007 | 0234 |
Subject: | NSA/METTWR/C1 - CMH Sensor Calibrated Incorrectly |
DataStreams: | nsamettwr4hC1.b1 |
Description: | The CMH temp and DP were higher than the co-located T/RH probe at the 2m level. An incorrect calibration along with a failing thermistor caused the readings to be incorrect for both measured values Temp and Dew Point. |
Measurements: | nsamettwr4hC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/01/2006 | 0000 | 05/22/2007 | 2347 |
Subject: | NSA/METTWR/C1 - 40m wind sensor heater failure |
DataStreams: | nsamettwr4hC1.b1 |
Description: | The 40m wind direction became frozen at times. The values from the sensor were flat-lined (stdev = 0) much of the time and did not match sensors from other levels. |
Measurements: | nsamettwr4hC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/20/2006 | 0030 | 11/27/2006 | 0254 |
12/08/2006 | 0125 | 12/14/2006 | 2229 |
12/16/2006 | 1545 | 12/21/2006 | 1105 |
12/23/2006 | 1612 | 01/18/2007 | 2357 |
01/20/2007 | 1357 | 01/29/2007 | 2140 |
Subject: | NSA/METTWR/C1 - 40m Wind Direction Data Suspect |
DataStreams: | nsamettwr4hC1.b1 |
Description: | The 40m wind direction data was suspect due to icing on the vane or a failing shafter heater or a combination of both. The affected data can be easily seen on comparison plots with other levels data or by checking the standard deviation of the 40m wind direction. Any standard deviation of less than 0.10 degrees for the 40m level suggest a sluggish/ice affected sensor. |
Measurements: | nsamettwr4hC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/09/2007 | 2326 | 10/10/2007 | 0004 |
Subject: | NSA/METTWR/C1 - PWS Data Questionable |
DataStreams: | nsamettwr4hC1.b1 |
Description: | There was a degradation in the PWS lenses. New lenses were installed on 08/28/2007 but a proper calibration was not conducted until 10/10/2007. Data during the listed time period should be used with caution. |
Measurements: | nsamettwr4hC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/18/2006 | 0000 | 05/22/2007 | 2330 |
Subject: | NSA/METTWR/C1 - 20m RH Data Questionable |
DataStreams: | nsamettwr4hC1.b1 |
Description: | The 20m RH data was questionable at times. When the data appeared to be poor was when the 2m and 20m agreed better than the 20m and the 40m and/or 10m. |
Measurements: | nsamettwr4hC1.b1:
|