Data Quality Reports for Session: 111679 User: gnk Completed: 04/23/2008


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D050201.1SGP/SONDE/C1 - Possibly bad pressure sensor or operator data entry errorsgpsondewnpnC1.b1
D060928.3SGP/SONDE/C1 - Increased precision for alt, dp, pres, rh, tdrysgpsondewnpnC1.b1


DQRID : D050201.1
Start DateStart TimeEnd DateEnd Time
01/19/2005053101/19/20050706
Subject:
SGP/SONDE/C1 - Possibly bad pressure sensor or operator data entry error
DataStreams:sgpsondewnpnC1.b1
Description:
There is a discrepency between the surface pressure and temperature data entered by the 
operator and that recorded by the THWAPS.  Review of the raw data from the radiosonde shows 
that the pressure was periodically varying by as much as 12 hPa (~988 to 1000)in the 
time period before launch.  The operator entry of 998.2 is at the high end of this range - 
the concurrent THWAPS reading was about 988.4 at the low end.  The operator entered 
temperature 6.2 degC is considerably higher than the THWAPS value of 1.8.  The pre-flight log 
entry for this sounding is missing from the OMIS database.  If not for the extremely 
unstable pressure reading I would have assumed this was an operator error.  It will take some 
more research to figure out what is going on.
Measurements:sgpsondewnpnC1.b1:
  • Surface dew point temperature(dp)
  • base time(base_time)
  • Mean Wind Speed(wspd)
  • U-component(u_wind)
  • Relative humidity inside the instrument enclosure(rh)
  • lat(lat)
  • Ascent Rate(asc)
  • Retrieved pressure profile(pres)
  • Wind Direction(deg)
  • Wind Status(wstat)
  • lon(lon)
  • Dry bulb temperature(tdry)
  • Time offset of tweaks from base_time(time_offset)
  • V-component(v_wind)
  • Dummy altitude for Zeb(alt)


Back To Table of Contents

DQRID : D060928.3
Start DateStart TimeEnd DateEnd Time
09/28/2006172812/31/99992359
Subject:
SGP/SONDE/C1 - Increased precision for alt, dp, pres, rh, tdry
DataStreams:sgpsondewnpnC1.b1
Description:
The output precision of the following variables was increased 20060928

alt  changed from 1m to 0.1m
dp   changed from 0.1degC to 0.01degC
pres changed from 0.1 hPa to 0.01 hPa
RH   changed from 1%RH to 0.01%RH
tdry changed from .1degC to 0.01degC

Note that the increase in precision does not necessarily imply an increase in accuracy.  
Users are cautioned to continue to use their best scientific judgement when using these or 
any other data.
Measurements:sgpsondewnpnC1.b1:
  • Surface dew point temperature(dp)
  • Dry bulb temperature(tdry)
  • Relative humidity inside the instrument enclosure(rh)
  • Retrieved pressure profile(pres)
  • Dummy altitude for Zeb(alt)


Back To Table of Contents



END OF DATA