Data Quality Reports for Session: 111818 User: borr Completed: 05/06/2008


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D070618.2SGP/SMOS/E13 - RH sensor failuresgp1440smosE13.b1, sgp1smosE13.b1, sgp30smosE13.b1


DQRID : D070618.2
Start DateStart TimeEnd DateEnd Time
04/16/2007000006/18/20072000
Subject:
SGP/SMOS/E13 - RH sensor failure
DataStreams:sgp1440smosE13.b1, sgp1smosE13.b1, sgp30smosE13.b1
Description:
The T/RH sensor relative humidity data is suspect during this time.  Values from this 
sensor failed the 6-month check and subsequent bi-weekly checks.  The RH values were lower by 
larger than the manufacturer suggested uncertainty of 3%.
Measurements:sgp1440smosE13.b1:
  • Time of Maximum Relative Humidity(time_max_rh)
  • Time of Maximum Vapor Pressure(time_max_vap_pres)
  • Minimum Relative Humidity(min_rh)
  • Time of Minimum Vapor Pressure(time_min_vap_pres)
  • Maximum Relative Humidity(max_rh)
  • Time of Minimum Relative Humidity(time_min_rh)
  • Maximum Vapor Pressure(max_vap_pres)
  • Minimum Vapor Pressure(min_vap_pres)

sgp30smosE13.b1:
  • Standard Deviation of Relative Humidity(sd_rh)
  • Standard Deviation of Vapor Pressure(sd_vap_pres)
  • Vapor Pressure (kiloPascals)(vap_pres)
  • Relative humidity inside the instrument enclosure(rh)

sgp1smosE13.b1:
  • Relative humidity inside the instrument enclosure(rh)
  • Vapor Pressure (kiloPascals)(vap_pres)


Back To Table of Contents