DQR ID | Subject | Data Streams Affected |
---|---|---|
D000504.5 | TWP/SMET/C1 - Wind Direction alignment method | twpsmet60sC1.00, twpsmet60sC1.a1, twpsmet60sC1.b1 |
D000621.1 | TWP/SMET/C1 - Incorrect calibration for ARCS1 Wind speed | twpsmet60sC1.00, twpsmet60sC1.a1, twpsmet60sC1.b1, twpsmetC1.00 |
D011211.3 | TWP/SMET/C1 - precip, wind, temperature | twpsmet60sC1.a1, twpsmet60sC1.b1 |
D020214.9 | TWP/SMET/C1 - Sensor replacement | twpsmet60sC1.a1, twpsmet60sC1.b1 |
D020620.61 | TWP/C1 - Incorrect elevation reported in Manus Data | twpgndrad20sC1.a1, twpgndrad60sC1.a1, twpgndrad60sC1.b1, twpsmet60sC1.a1, twpsmet60sC1.b1 |
D040220.1 | TWP/MWR/C1 - wrong azimuth | twpmwrlosC1.b1 |
D040813.11 | TWP/SMET/C1 - RESET14 activities | twpsmet60sC1.a1, twpsmet60sC1.b1 |
D050421.6 | TWP/SMET/C1 - Negative rainfall rates | twpsmet60sC1.b1 |
D060817.3 | TWP/SMET/C1 - Reprocess: Barometric Data Changed from hPa to kPa | twpsmet60sC1.b1 |
D061011.2 | TWP/SMET/C1 - Reprocess: Tipping bucket rain gauge added | twpsmet60sC1.b1 |
D070105.1 | TWP/MWR/C1 - Missing data | twpmwrlosC1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/25/1999 | 0450 | 04/02/2000 | 0000 |
Subject: | TWP/SMET/C1 - Wind Direction alignment method |
DataStreams: | twpsmet60sC1.00, twpsmet60sC1.a1, twpsmet60sC1.b1 |
Description: | During RESET10 we discovered a problem with wind direction from WIND1 (upper anemometer). During RESET8, WIND1 was corrected by rotating the alignment ring instead of the anemometer. This resulted in a better agreement between the 2 anemometers, but is not the correct way to align an anemometer. The alignment ring was readjusted for true north and the anemometer reset in the proper slot (the original problem). |
Measurements: | twpsmet60sC1.b1:
twpsmet60sC1.00:
twpsmet60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/23/1999 | 0000 | 04/01/2000 | 2359 |
Subject: | TWP/SMET/C1 - Incorrect calibration for ARCS1 Wind speed |
DataStreams: | twpsmet60sC1.00, twpsmet60sC1.a1, twpsmet60sC1.b1, twpsmetC1.00 |
Description: | On 23 OCT 99, during Reset 8, an incorrect slope of was entered for the upper, windspeed1, wind speed sensor resulting in a 1% overestimate of the wind speed. The correct calibration was entered on 1 APR 2000 during RESET 10. |
Measurements: | twpsmet60sC1.b1:
twpsmetC1.00:
twpsmet60sC1.00:
twpsmet60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/23/2001 | 0445 | 10/23/2001 | 0645 |
Subject: | TWP/SMET/C1 - precip, wind, temperature |
DataStreams: | twpsmet60sC1.a1, twpsmet60sC1.b1 |
Description: | On 10/23/01 from 4:45 to 6:45 precipitation, wind, and temperature measurements were directly affected by RESET14 visit activities. |
Measurements: | twpsmet60sC1.b1:
twpsmet60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/24/2001 | 0040 | 10/24/2001 | 0100 |
Subject: | TWP/SMET/C1 - Sensor replacement |
DataStreams: | twpsmet60sC1.a1, twpsmet60sC1.b1 |
Description: | On 10/24/01 from 00:40 to 01:00 the T/RH probe was being replaced. |
Measurements: | twpsmet60sC1.b1:
twpsmet60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/10/1996 | 0001 | 06/22/2000 | 2359 |
Subject: | TWP/C1 - Incorrect elevation reported in Manus Data |
DataStreams: | twpgndrad20sC1.a1, twpgndrad60sC1.a1, twpgndrad60sC1.b1, twpsmet60sC1.a1, twpsmet60sC1.b1 |
Description: | (From Bill Clements) We have been using 6m MSL as the station elevation of Momote (Manus). This was based on information posted near the station barometer in the Momote Weather Service Office. It turns out that the official WMO station elevation for Momote is 4m MSL and the upper air elevation is 5m MSL. All Manus instruments except the sonde station are at 4m MSL. The sonde launching station is at 5m MSL. |
Measurements: | twpgndrad20sC1.a1:
twpgndrad60sC1.b1:
twpsmet60sC1.b1:
twpgndrad60sC1.a1:
twpsmet60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/11/1996 | 0000 | 02/18/2004 | 2250 |
Subject: | TWP/MWR/C1 - wrong azimuth |
DataStreams: | twpmwrlosC1.b1 |
Description: | The MWR was initially installed at an azimuth angle defined as 180 degrees but the value in the configuration file was not changed from the default of 0 degrees. In examining photos taken during the installation of the AWS tower, I noticed that the MWR was rotated opposite the normal orientation. The value in the configuration file was changed to reflect the actual azimuth of the instrument. |
Measurements: | twpmwrlosC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/09/2001 | 0000 | 10/26/2001 | 2359 |
Subject: | TWP/SMET/C1 - RESET14 activities |
DataStreams: | twpsmet60sC1.a1, twpsmet60sC1.b1 |
Description: | RESET14 activities may have affected data in subtle ways during this period. |
Measurements: | twpsmet60sC1.b1:
twpsmet60sC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/09/1996 | 0000 | 03/09/2004 | 2243 |
Subject: | TWP/SMET/C1 - Negative rainfall rates |
DataStreams: | twpsmet60sC1.b1 |
Description: | Due to limitations of the Zeno dataloggers all voltage measurements from the optical rain gauge (ORG) were converted to rainfall rates. This resulted in negative values for rain rates. Any value below 0.10 mm/hr should be considered to be 0 mm/hr. New dataloggers alleviated this problem. New loggers were installed 03/09/2004 @ 2243 GMT. |
Measurements: | twpsmet60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/09/1996 | 0000 | 07/13/2006 | 2225 |
Subject: | TWP/SMET/C1 - Reprocess: Barometric Data Changed from hPa to kPa |
DataStreams: | twpsmet60sC1.b1 |
Description: | Barometric pressure data was converted from hPa to kPa in order to standardize the measurement units among ARM sites and to conform to accepted standard units determined by the scientific community. |
Measurements: | twpsmet60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/09/1996 | 0000 | 10/17/2006 | 0000 |
Subject: | TWP/SMET/C1 - Reprocess: Tipping bucket rain gauge added |
DataStreams: | twpsmet60sC1.b1 |
Description: | A tipping bucket rain gauge was added to the TWP.C1 SMET suite of instruments on 20061017. Effective 20060926, two new variables (count and rain amount) were added to the end of the twpsmet60sC1.b1 datastream. Between 9/26 and 10/17, the data for these two variables are filled with zeros. Users should not use the data during this period as the zeros are not representative of the rain fall at the site. When reprocessing of this datastream occurs these variables will be added for the 19961009-20060926 and data from 19961009-20061017 will be filled with -9999 (missing). |
Measurements: | twpsmet60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/09/1996 | 2338 | 12/11/1996 | 0003 |
02/28/1997 | 0118 | 06/27/1997 | 2048 |
11/06/1997 | 1959 | 11/12/1997 | 1800 |
01/05/1998 | 0204 | 01/08/1998 | 2048 |
04/21/1998 | 0017 | 04/29/1998 | 1803 |
09/13/1998 | 0759 | 09/20/1998 | 0900 |
11/20/1998 | 2226 | 11/27/1998 | 1026 |
02/27/1999 | 0600 | 03/02/1999 | 0159 |
03/06/1999 | 0800 | 03/09/1999 | 1807 |
09/03/1999 | 0300 | 09/10/1999 | 0400 |
10/01/1999 | 0900 | 10/18/1999 | 0000 |
10/20/1999 | 0500 | 10/22/1999 | 0600 |
01/14/2000 | 0800 | 01/16/2000 | 0342 |
01/14/2006 | 1258 | 01/16/2006 | 0638 |
Subject: | TWP/MWR/C1 - Missing data |
DataStreams: | twpmwrlosC1.b1 |
Description: | Data are missing and unrecoverable. |
Measurements: | twpmwrlosC1.b1:
|