Data Quality Reports for Session: 111876 User: ql_data Completed: 05/09/2008


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D030301.1NSA/GNDRAD/C1 - No power to IRTnsagndrad60sC1.a1, nsagndrad60sC1.b1
D031216.1NSA/GNDRAD/C1 - IRT cable failurensagndrad60sC1.b1
D031216.2NSA/GNDRAD/C1 - IRT returned to vendornsagndrad60sC1.b1
D040919.1NSA/GNDRAD/C1 - IRT power failurensagndrad60sC1.b1
D060209.4NSA/GNDRAD/C1 - power outage affecting some GNDRAD sensorsnsagndrad20sC1.a0, nsagndrad60sC1.b1, nsagndradC1.00
D060630.39NSA/GNDRAD/C1 - Reprocess: Longwave Calibration errornsagndrad60sC1.b1
D071025.3NSA/GNDRAD/C1 - Reprocessed: Incorrect calibration coefficientnsagndrad20sC1.a0, nsagndrad60sC1.b1


DQRID : D030301.1
Start DateStart TimeEnd DateEnd Time
09/24/2002172511/16/20020003
Subject:
NSA/GNDRAD/C1 - No power to IRT
DataStreams:nsagndrad60sC1.a1, nsagndrad60sC1.b1
Description:
The data logger failed to supply power to the IRT.
Measurements:nsagndrad60sC1.b1:
  • Surface Infra-Red Temperature Maxima(sfc_ir_temp_max)
  • Surface Infra-Red Temperature Minima(sfc_ir_temp_min)
  • Effective Surface (Skin) Temperature(sfc_ir_temp)
  • Surface Infra-Red Temperature, Standard Deviation(sfc_ir_temp_std)

nsagndrad60sC1.a1:
  • Surface Infra-Red Temperature Maxima(sfc_ir_temp_max)
  • Effective Surface (Skin) Temperature(sfc_ir_temp)
  • Surface Infra-Red Temperature Minima(sfc_ir_temp_min)
  • Surface Infra-Red Temperature, Standard Deviation(sfc_ir_temp_std)


Back To Table of Contents

DQRID : D031216.1
Start DateStart TimeEnd DateEnd Time
05/27/2003140009/11/20031906
Subject:
NSA/GNDRAD/C1 - IRT cable failure
DataStreams:nsagndrad60sC1.b1
Description:
The IRT signal became very noisy and the problem was traced to the cable (after the 
instrument was returned to the vendor for diagnosis). The problem started occurring when ice 
surrounding the cable began to melt. The cable was repaired and the instrument reinstalled.
Measurements:nsagndrad60sC1.b1:
  • Surface Infra-Red Temperature Maxima(sfc_ir_temp_max)
  • Surface Infra-Red Temperature Minima(sfc_ir_temp_min)
  • Effective Surface (Skin) Temperature(sfc_ir_temp)
  • Surface Infra-Red Temperature, Standard Deviation(sfc_ir_temp_std)


Back To Table of Contents

DQRID : D031216.2
Start DateStart TimeEnd DateEnd Time
09/11/2003190610/15/20030015
Subject:
NSA/GNDRAD/C1 - IRT returned to vendor
DataStreams:nsagndrad60sC1.b1
Description:
The IRT was returned to the vendor for diagnosis. It was found to be operating as expected 
and the signal noise problem was traced to the cable. The cable was repaired and the 
instrument reinstalled.
Measurements:nsagndrad60sC1.b1:
  • Surface Infra-Red Temperature Maxima(sfc_ir_temp_max)
  • Surface Infra-Red Temperature Minima(sfc_ir_temp_min)
  • Effective Surface (Skin) Temperature(sfc_ir_temp)
  • Surface Infra-Red Temperature, Standard Deviation(sfc_ir_temp_std)


Back To Table of Contents

DQRID : D040919.1
Start DateStart TimeEnd DateEnd Time
09/15/2004000009/22/20040552
Subject:
NSA/GNDRAD/C1 - IRT power failure
DataStreams:nsagndrad60sC1.b1
Description:
After performing corrective maintenance which involved turning off power to the GNDRAD 
instruments and data logger, power to the IRT was not restored. The problem was corrected by 
replacing the data logger.
Measurements:nsagndrad60sC1.b1:
  • Surface Infra-Red Temperature Maxima(sfc_ir_temp_max)
  • Surface Infra-Red Temperature Minima(sfc_ir_temp_min)
  • Effective Surface (Skin) Temperature(sfc_ir_temp)
  • Surface Infra-Red Temperature, Standard Deviation(sfc_ir_temp_std)


Back To Table of Contents

DQRID : D060209.4
Start DateStart TimeEnd DateEnd Time
02/07/2006091002/07/20062350
Subject:
NSA/GNDRAD/C1 - power outage affecting some GNDRAD sensors
DataStreams:nsagndrad20sC1.a0, nsagndrad60sC1.b1, nsagndradC1.00
Description:
During the power outage the upwelling global measurement from the PSP instrument shows a 
slight increase from -5 to +0 W/m2.  This is due to removal of thermal offsets induced by 
the AC heater during the power outage.  This heater is installed within the PSP 
ventilator - it is necessary for the NSA/C1 climate.
The upwelling longwave PIR instrument is unaffected due to case and dome temperature 
corrections.
The IRT is AC powered and was not operational during this time period.
Measurements:nsagndradC1.00:
  • null(Raw data stream - documentation not supported)

nsagndrad20sC1.a0:
  • Instantaneous Surface Infra-Red Signal(inst_sfc_ir_signal)

nsagndrad60sC1.b1:
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Pyranometer, Maxima(up_short_hemisp_max)
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Pyranometer, Minima(up_short_hemisp_min)
  • Surface Infra-Red Temperature Maxima(sfc_ir_temp_max)
  • Surface Infra-Red Temperature Minima(sfc_ir_temp_min)
  • Effective Surface (Skin) Temperature(sfc_ir_temp)
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Pyranometer, Standard
    Deviation(up_short_hemisp_std)
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Pyranometer(up_short_hemisp)
  • Surface Infra-Red Temperature, Standard Deviation(sfc_ir_temp_std)


Back To Table of Contents

DQRID : D060630.39
Start DateStart TimeEnd DateEnd Time
11/04/2004160012/21/20051855
Subject:
NSA/GNDRAD/C1 - Reprocess: Longwave Calibration error
DataStreams:nsagndrad60sC1.b1
Description:
Modified pyrgeometer calibration procedures were implemented beginning in December 2003.  
These modified procedures introduced a calibration bias in the longwave data.  The 
previous procedures were re-implemented at all sites between December 2005 and February 2006 to 
restore proper calibrations.                                                             
                The data collected while the incorrect procedures were in place are 
being reprocessing to remove the calibration bias.  Note, this also affected standard 
deviation, maximum and minimum data fields.
Measurements:nsagndrad60sC1.b1:
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Pyrgeometer(up_long_hemisp)


Back To Table of Contents

DQRID : D071025.3
Start DateStart TimeEnd DateEnd Time
10/04/2007013010/25/20072149
Subject:
NSA/GNDRAD/C1 - Reprocessed: Incorrect calibration coefficient
DataStreams:nsagndrad20sC1.a0, nsagndrad60sC1.b1
Description:
Installed Gndrad PSP (upwelling shortwave hemispheric irradiance) had calibration 
coefficient RS applied when it should have used Rsc.  ARM is using IR corrected coefficients 
Rsc@45 and CFc@45 for PSPs deployed at all facilities.  

PSP #32012F3 should use RSc@45=109.39 and not Rs@45=110.97.  

Reprocessed to correct upwelling shortwave hemispheric irradiance (US) by:
US(new) = (US(old) / 110.97 ) * 109.39

Reprocessing completed March 2008.
Measurements:nsagndrad20sC1.a0:
  • Instantaneous Upwelling Shortwave Hemispheric Irradiance, Pyranometer(inst_up_short_hemisp)

nsagndrad60sC1.b1:
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Pyranometer, Maxima(up_short_hemisp_max)
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Pyranometer, Minima(up_short_hemisp_min)
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Pyranometer, Standard
    Deviation(up_short_hemisp_std)
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Pyranometer(up_short_hemisp)


Back To Table of Contents



END OF DATA