Data Quality Reports for Session: 112154 User: liusy Completed: 05/23/2008


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D010213.6SGP/EBBR/E19 - EBBR E19 Wind Speed Icedsgp15ebbrE19.a1, sgp30ebbrE19.a1, sgp5ebbrE19.a0
D010213.7SGP/EBBR/E20 - EBBR E20 Wind Speed Icedsgp15ebbrE20.a1, sgp30ebbrE20.a1, sgp5ebbrE20.a0
D010427.5SGP EBBR E19 - E19 EBBR Left Aspirator Housing Offsgp15ebbrE19.a1, sgp30ebbrE19.a1, sgp5ebbrE19.a0
D010427.8SGP EBBR E19 - E19 EBBR Wind Direction Incorrectsgp15ebbrE19.a1, sgp30ebbrE19.a1, sgp5ebbrE19.a0
D010427.9SGP EBBR E19 - E19 EBBR Soil Heat Flow #1 Sign Reversedsgp15ebbrE19.a1, sgp30ebbrE19.a1
D010517.3SGP EBBR E20 - E20 EBBR Right Air T Incorrectsgp15ebbrE20.a1, sgp30ebbrE20.a1, sgp5ebbrE20.a0
D010605.11SGP EBBR E20 - E20 EBBR Soil Temperatures Incorrectsgp15ebbrE20.a1, sgp30ebbrE20.a1
D010605.12SGP EBBR E20 - E20 EBBR Data Incorrect During STP #3 Replacementsgp15ebbrE20.a1, sgp30ebbrE20.a1, sgp5ebbrE20.00, sgp5ebbrE20.a0
D010605.13SGP EBBR E20 - E20 EBBR Data Incorrect After Program Restartsgp15ebbrE20.a1, sgp30ebbrE20.a1, sgp5ebbrE20.00, sgp5ebbrE20.a0
D010605.14SGP EBBR E20 - E20 EBBR Some Data Incorrectsgp30ebbrE20.a1
D010605.8SGP EBBR E20 - E20 EBBR Data Incorrectsgp15ebbrE20.a1, sgp30ebbrE20.a1, sgp5ebbrE20.00, sgp5ebbrE20.a0
D010605.9SGP EBBR E20 - E20 EBBR Net Radiation Data Incorrectsgp15ebbrE20.a1, sgp30ebbrE20.a1, sgp5ebbrE20.a0
D011102.6SGP EBBR E19 - E19 EBBR Data Missingsgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.00,
sgp5ebbrE19.a1
D011102.7SGP EBBR E19 - E19 EBBR TS3 Intermittent Incorrectsgp15ebbrE19.a1, sgp30ebbrE19.a1
D011102.8SGP EBBR E19 - E19 EBBR Wind Speed Incorrectsgp15ebbrE19.a1, sgp30ebbrE19.a1, sgp5ebbrE19.a0
D011127.30SGP EBBR E19 - E19 EBBR Data Incorrect During 6 Month Checkssgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
D011127.31SGP EBBR E20 - E20 EBBR Data Incorrect During 6 Month Checkssgp15ebbrE20.a1, sgp15ebbrE20.b1, sgp30ebbrE20.a1, sgp30ebbrE20.b1, sgp5ebbrE20.a1
D011127.4SGP EBBR E20 - E20 EBBR Soil Temperature #3 Incorrectsgp15ebbrE20.a1, sgp30ebbrE20.a1
D011218.10SGP EBBR E20 - E20 EBBR Soil Heat Flow #2 Incorrectsgp15ebbrE20.a1, sgp30ebbrE20.a1
D011218.11SGP EBBR E20 - E20 EBBR Battery Voltage Low, Incorrect Datasgp15ebbrE20.a1, sgp30ebbrE20.a1
D011218.13SGP EBBR E20 - E20 EBBR Battery Voltage Low, Data Incorrectsgp15ebbrE20.a1, sgp15ebbrE20.b1, sgp30ebbrE20.a1, sgp30ebbrE20.b1
D011218.14SGP/EBBR/E20 - Reprocess: Soil Temperature #3 Incorrectsgp15ebbrE20.a1, sgp15ebbrE20.b1, sgp30ebbrE20.a1, sgp30ebbrE20.b1
D011218.9SGP EBBR E20 - E20 EBBR Soil Heat Flow #4 Incorrectsgp15ebbrE20.a1, sgp30ebbrE20.a1
D020716.15SGP/EBBR/E19 - 6 Month Checkssgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
D020716.16SGP/EBBR/E20 - 6 Month Checkssgp15ebbrE20.a1, sgp15ebbrE20.b1, sgp30ebbrE20.a1, sgp30ebbrE20.b1, sgp5ebbrE20.a1
D020717.26SGP/EBBR/E19 - Reprocess: Soil Heat Flow #3 Incorrectsgp30ebbrE19.a1, sgp30ebbrE19.b1
D020717.27SGP/EBBR/E19 - Reprocess: Soil Temperature #3 Incorrectsgp30ebbrE19.a1, sgp30ebbrE19.b1
D020717.29SGP/EBBR/E19 - Most Measurements Incorrectsgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.00,
sgp5ebbrE19.a1
D020717.30SGP/EBBR/E19 - All Measurements Zerosgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.00,
sgp5ebbrE19.a1
D020717.31SGP/EBBR/E19 - Pressure Incorrectsgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
D020717.32SGP/EBBR/E19 - Measurements Incorrect After CR10 Replacementsgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.00,
sgp5ebbrE19.a1
D020717.33SGP/EBBR/E19 - Low Battery Voltagesgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.00,
sgp5ebbrE19.a1
D020717.40SGP/EBBR/E20 - AEM Hungsgp15ebbrE20.a1, sgp15ebbrE20.b1, sgp30ebbrE20.a1, sgp30ebbrE20.b1, sgp5ebbrE20.a1
D020717.42SGP/EBBR/E20 - Battery Voltage Lowsgp15ebbrE20.a1, sgp15ebbrE20.b1, sgp30ebbrE20.a1, sgp30ebbrE20.b1, sgp5ebbrE20.00,
sgp5ebbrE20.a1
D020717.43SGP/EBBR/E20 - Pressure Almost Constantsgp30ebbrE20.a1, sgp30ebbrE20.b1, sgp5ebbrE20.a1
D020717.45SGP/EBBR/E20 - Reprocess: Soil Heat Flow #3 Incorrect During Sensor Changeoutsgp30ebbrE20.a1, sgp30ebbrE20.b1
D020717.46SGP/EBBR/E20 - Reprocess: Soil Heat Flow #2 Has a Negative Offsetsgp30ebbrE20.a1, sgp30ebbrE20.b1
D020717.47SGP/EBBR/E20 - Reprocess: Soil Heat Flow #4 Incorrectsgp30ebbrE20.a1, sgp30ebbrE20.b1
D020717.48SGP/EBBR/E20 - Measurements Zerosgp15ebbrE20.a1, sgp15ebbrE20.b1, sgp30ebbrE20.a1, sgp30ebbrE20.b1, sgp5ebbrE20.00,
sgp5ebbrE20.a1
D020717.49SGP/EBBR/E20 - Reprocess: Soil Temperature #3 Incorrectsgp30ebbrE20.a1, sgp30ebbrE20.b1
D020717.50SGP/EBBR/E20 - Measurements Incorrect During Sensor Changeoutsgp30ebbrE20.a1, sgp30ebbrE20.b1
D020717.51SGP/EBBR/E20 - Measurements Incorrectsgp15ebbrE20.a1, sgp15ebbrE20.b1, sgp30ebbrE20.a1, sgp30ebbrE20.b1, sgp5ebbrE20.00,
sgp5ebbrE20.a1
D020726.1SGP/EBBR/E20 - Reprocess: Soil Heat Flow #2 and #4 incorrectsgp30ebbrE20.a1, sgp30ebbrE20.b1
D021113.4SGP/EBBR - Incorrect Units for Soil Heat Capacity (cs1, cs2, cs3, cs4, cs5) in 30EBBR
Header
sgp30ebbrE12.a1, sgp30ebbrE13.a1, sgp30ebbrE15.a1, sgp30ebbrE20.a1, sgp30ebbrE22.a1,
sgp30ebbrE26.a1, sgp30ebbrE4.a1, sgp30ebbrE7.a1, sgp30ebbrE8.a1, sgp30ebbrE9.a1
D021113.5SGP/EBBR - Incorrect Units for Soil Heat Capacity (cs1, cs2, cs3, cs4, cs5) in 30EBBR
Header
sgp30baebbrE18.c1, sgp30baebbrE19.c1, sgp30baebbrE2.c1, sgp30ebbrE18.a1, sgp30ebbrE19.a1,
sgp30ebbrE2.a1
D030110.24SGP/EBBR/E19 - Data Incorrect during 6 month checkssgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
D030110.25SGP/EBBR/E20 - Data Incorrect during 6 month checkssgp15ebbrE20.a1, sgp15ebbrE20.b1, sgp30ebbrE20.a1, sgp30ebbrE20.b1, sgp5ebbrE20.a1
D030116.10SGP/EBBR/E19 - REPROCESS: Soil Heat Flow #3 Incorrect Calibrationsgp30ebbrE19.a1, sgp30ebbrE19.b1
D030116.11SGP/EBBR/E19 - Water in Net Radiometer Domesgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
D030116.6SGP/EBBR/E19 - Incorrect data after power interruptionsgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
D030116.7SGP/EBBR/E19 - Most Measurements Zero following Restartsgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
D030116.8SGP/EBBR/E19 - REPROCESS: Soil Temperature #3 Incorrectsgp30ebbrE19.a1, sgp30ebbrE19.b1
D030116.9SGP/EBBR/E19 - REPROCESS: Soil Heat Flow #1 Incorrectsgp30ebbrE19.a1, sgp30ebbrE19.b1
D030117.1SGP/EBBR/E19 - Pressure Lowsgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
D030117.10SGP/EBBR/E20 - Data Zero following logger restartssgp15ebbrE20.a1, sgp15ebbrE20.b1, sgp30ebbrE20.a1, sgp30ebbrE20.b1, sgp5ebbrE20.a1
D030117.11SGP/EBBR/E20 - REPROCESS: Soil Temperature #2 Incorrectsgp30ebbrE20.a1, sgp30ebbrE20.b1
D030117.12SGP/EBBR/E20 - Change of Energy Storage Incorrectsgp30ebbrE20.a1, sgp30ebbrE20.b1
D030117.13SGP/EBBR/E20 - Battery voltage lowsgp15ebbrE20.a1, sgp15ebbrE20.b1, sgp30ebbrE20.a1, sgp30ebbrE20.b1, sgp5ebbrE20.a1
D030117.14SGP/EBBR/E20 - Wind speed and direction frozensgp15ebbrE20.a1, sgp15ebbrE20.b1, sgp30ebbrE20.a1, sgp30ebbrE20.b1, sgp5ebbrE2.a1
D030117.18SGP/EBBR/E20 - AEM Hung between Home Positionssgp15ebbrE20.a1, sgp15ebbrE20.b1, sgp30ebbrE20.a1, sgp30ebbrE20.b1, sgp5ebbrE20.a1
D030117.19SGP/EBBR/E20 - REPROCESS: Heat Flow #3 and #5 Incorrectsgp30ebbrE20.a1, sgp30ebbrE20.b1
D030117.2SGP/EBBR/E20 - Data incorrect during instrument maintenancesgp15ebbrE20.a1, sgp15ebbrE20.b1, sgp30ebbrE20.a1, sgp30ebbrE20.b1, sgp5ebbrE20.a1
D030117.20SGP/EBBR/E20 - REPROCESS: Soil Heat Flow #5 Incorrectsgp30ebbrE20.a1, sgp30ebbrE20.b1
D030117.22SGP/EBBR/E20 - REPROCESS: Soil Temperature #1 Incorrectsgp30ebbrE20.a1, sgp30ebbrE20.b1
D030117.25SGP/EBBR/E20 - REPROCESS: Soil Temperature #3 Incorrectsgp30ebbrE20.a1, sgp30ebbrE20.b1
D030117.3SGP/EBBR/E20 - Resistance-based Measurements Incorrectsgp15ebbrE20.a1, sgp15ebbrE20.b1, sgp30ebbrE20.a1, sgp30ebbrE20.b1, sgp5ebbrE20.a1
D030117.4SGP/EBBR/E19 - Soil Temperatures #2 and #3 Intermittently Badsgp30ebbrE19.a1, sgp30ebbrE19.b1
D030117.6SGP/EBBR/E19 - Measurements zero following logger restartsgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
D030117.7SGP/EBBR/E19 - Change of Energy Storage Incorrectsgp30ebbrE19.a1, sgp30ebbrE19.b1
D030117.8SGP/EBBR/E20 - REPROCESS: Net Rad Coefficients Incorrectsgp30ebbrE20.a1, sgp30ebbrE20.b1, sgp5ebbrE20.a1
D030121.1SGP/EBBR/E20 - REPROCESS: Soil Temperature #1 and Soil Heat Flow #5 Incorrectsgp30ebbrE20.a1, sgp30ebbrE20.b1
D030711.10SGP/EBBR/E20 - RHtop Zerosgp30ebbrE20.a1, sgp30ebbrE20.b1
D030711.11SGP/EBBR/E19 - Soil Temperature #2 Incorrectsgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1
D030711.12SGP/EBBR/E19 - All Data Zerossgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
D030711.13SGP/EBBR/E19 - Several Soil Measurements Incorrectsgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1
D030711.14SGP/EBBR/E19 -Several Soil Measurements Were Incorrectsgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1
D030711.16SGP/EBBR/E19 - AEM Hung in home30 positionsgp30ebbrE19.a1, sgp30ebbrE19.b1
D030711.17SGP/EBBR/E19 - Soil Heat Flow #3 Incorrectsgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1
D030711.19SGP/EBBR/E19 - Soil Heat Flow#1 Incorrectsgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1
D030711.20SGP/EBBR/E19 - Many Measurements Incorrectsgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
D030711.8SGP/EBBR/E20 - Soil Heat Flow#5 Incorrectsgp15ebbrE20.a1, sgp15ebbrE20.b1, sgp30ebbrE20.a1, sgp30ebbrE20.b1
D030711.9SGP/EBBR/E20 - Soil Storage Incorrect After Startupsgp30ebbrE20.a1, sgp30ebbrE20.b1
D030712.20SGP/EBBR/E19 - 6 Month Checkssgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
D030712.21SGP/EBBR/E20 - 6 Month Checkssgp15ebbrE20.a1, sgp15ebbrE20.b1, sgp30ebbrE20.a1, sgp30ebbrE20.b1, sgp5ebbrE20.a1
D030919.10SGP/EBBR/E19 - Water in Net Radiometer; Surface Fluxes are Incorrectsgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
D030922.6SGP/EBBR/E19 - Tair Right Offscalesgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
D031031.7SGP/EBBR/E20 - Water in Net Radiometer, Surface Fluxes Incorrectsgp15ebbrE20.a1, sgp15ebbrE20.b1, sgp30ebbrE20.a1, sgp30ebbrE20.b1, sgp5ebbrE20.a1
D031124.11SGP/EBBR/E19 - Data Missing During Recalibration Changeoutsgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30baebbrE19.c1, sgp30ebbrE19.a1, sgp30ebbrE19.b1,
sgp5ebbrE19.a1, sgpebbrE19.00, sgpebbrE19.01
D031204.17SGP/EBBR/E19 - 6 Month Checkssgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
D031204.2SGP/EBBR/E19 - Data All Zerossgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
D040525.3SGP/EBBR/E19 - Metadata errorssgp15ebbrE19.a1, sgp30baebbrE19.c1, sgp30ebbrE19.a1


DQRID : D010213.6
Start DateStart TimeEnd DateEnd Time
12/25/2000124512/28/20001815
Subject:
SGP/EBBR/E19 - EBBR E19 Wind Speed Iced
DataStreams:sgp15ebbrE19.a1, sgp30ebbrE19.a1, sgp5ebbrE19.a0
Description:
The anemometer was frozen stopped by freezing rain.  Indicated data is incorrect.
Measurements:sgp5ebbrE19.a0:
  • vector wind speed(res_ws)
  • scalar wind speed(wind_s)

sgp15ebbrE19.a1:
  • scalar wind speed(wind_s)

sgp30ebbrE19.a1:
  • scalar wind speed(wind_s)
  • vector wind speed(res_ws)


Back To Table of Contents

DQRID : D010213.7
Start DateStart TimeEnd DateEnd Time
12/25/2000154512/28/20002145
Subject:
SGP/EBBR/E20 - EBBR E20 Wind Speed Iced
DataStreams:sgp15ebbrE20.a1, sgp30ebbrE20.a1, sgp5ebbrE20.a0
Description:
The anemometer was frozen stopped by freezing rain.  Indicated data is incorrect.
Measurements:sgp30ebbrE20.a1:
  • vector wind speed(res_ws)
  • scalar wind speed(wind_s)

sgp5ebbrE20.a0:
  • vector wind speed(res_ws)
  • scalar wind speed(wind_s)

sgp15ebbrE20.a1:
  • scalar wind speed(wind_s)


Back To Table of Contents

DQRID : D010427.5
Start DateStart TimeEnd DateEnd Time
12/02/2000150012/08/20001946
Subject:
SGP EBBR E19 - E19 EBBR Left Aspirator Housing Off
DataStreams:sgp15ebbrE19.a1, sgp30ebbrE19.a1, sgp5ebbrE19.a0
Description:
The left aspirator housing fell off and thus the indicated measurements are incorrect.
Measurements:sgp5ebbrE19.a0:
  • top vapor pressure(vp_top)
  • Bottom humidity(hum_bot)
  • Top humidity(hum_top)
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • bottom air temperature(tair_bot)
  • top air temperature(tair_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)

sgp15ebbrE19.a1:
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Left air temperature(tair_l)
  • Left relative humidity(mv_hum_l)

sgp30ebbrE19.a1:
  • Temperature of the top humidity chamber(thum_top)
  • top air temperature(tair_top)
  • top vapor pressure(vp_top)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • latent heat flux(e)
  • bottom air temperature(tair_bot)
  • Top humidity(hum_top)
  • h(h)
  • bottom vapor pressure(vp_bot)
  • Bottom humidity(hum_bot)


Back To Table of Contents

DQRID : D010427.8
Start DateStart TimeEnd DateEnd Time
12/12/2000203012/13/20000731
Subject:
SGP EBBR E19 - E19 EBBR Wind Direction Incorrect
DataStreams:sgp15ebbrE19.a1, sgp30ebbrE19.a1, sgp5ebbrE19.a0
Description:
The wind direction sensor was frozen stopped during the indicated period.
Measurements:sgp5ebbrE19.a0:
  • vector wind speed(res_ws)
  • wind direction (relative to true north)(wind_d)
  • standard deviation of wind direction (sigma theta)(sigma_wd)

sgp15ebbrE19.a1:
  • Wind direction (relative to true north)(mv_wind_d)

sgp30ebbrE19.a1:
  • wind direction (relative to true north)(wind_d)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • vector wind speed(res_ws)


Back To Table of Contents

DQRID : D010427.9
Start DateStart TimeEnd DateEnd Time
07/24/1999044501/25/20011531
Subject:
SGP EBBR E19 - E19 EBBR Soil Heat Flow #1 Sign Reversed
DataStreams:sgp15ebbrE19.a1, sgp30ebbrE19.a1
Description:
The sign was reversed (heat flow plate #1 was installed upside down), causing incorrect 
measurements, as indicated.

The average soil heat flow can be recalculated by using an average of soil sensor sets 
2-5.  The recalculated average soil heat flow can then be used to recalculate latent and 
sensible heat fluxes.
Measurements:sgp15ebbrE19.a1:
  • Soil heat flow 1(mv_hft1)

sgp30ebbrE19.a1:
  • latent heat flux(e)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • average surface soil heat flow(ave_shf)
  • 5 cm soil heat flow, site 1(shf1)
  • h(h)


Back To Table of Contents

DQRID : D010517.3
Start DateStart TimeEnd DateEnd Time
11/02/2000234502/21/20011946
Subject:
SGP EBBR E20 - E20 EBBR Right Air T Incorrect
DataStreams:sgp15ebbrE20.a1, sgp30ebbrE20.a1, sgp5ebbrE20.a0
Description:
The right air temperature probe was incorrect frequently.  Inspection of
the right air temperature data will quickly show where it went offscale.
Bowen ratio, sensible heat flux, and latent heat flux were incorrect as a result.
Measurements:sgp30ebbrE20.a1:
  • top air temperature(tair_top)
  • latent heat flux(e)
  • top vapor pressure(vp_top)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • h(h)
  • bottom air temperature(tair_bot)
  • bottom vapor pressure(vp_bot)

sgp5ebbrE20.a0:
  • top air temperature(tair_top)
  • bottom air temperature(tair_bot)
  • bottom vapor pressure(vp_bot)
  • top vapor pressure(vp_top)

sgp15ebbrE20.a1:
  • Right air temperature(tair_r)


Back To Table of Contents

DQRID : D010605.11
Start DateStart TimeEnd DateEnd Time
11/01/2000193011/01/20001931
Subject:
SGP EBBR E20 - E20 EBBR Soil Temperatures Incorrect
DataStreams:sgp15ebbrE20.a1, sgp30ebbrE20.a1
Description:
All soil temperatures were incorrect for the indicated period.  
Latent and sensible heat fluxes were therefore incorrect also.
Measurements:sgp30ebbrE20.a1:
  • soil heat flow, site 1(g1)
  • Soil heat capacity 3(cs3)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • Soil heat capacity 2(cs2)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • soil heat flow, site 3(g3)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Soil heat capacity 5(cs5)
  • h(h)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • average surface soil heat flow(ave_shf)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • soil heat flow, site 5(g5)
  • soil heat flow, site 2(g2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • latent heat flux(e)
  • Soil heat capacity 4(cs4)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • Soil heat capacity 1(cs1)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)

sgp15ebbrE20.a1:
  • Soil temperature 5(rr_ts5)
  • Soil temperature 4(rr_ts4)
  • Soil temperature 1(rr_ts1)
  • Soil temperature 3(rr_ts3)
  • Soil temperature 2(rr_ts2)


Back To Table of Contents

DQRID : D010605.12
Start DateStart TimeEnd DateEnd Time
11/15/2000203111/15/20002101
Subject:
SGP EBBR E20 - E20 EBBR Data Incorrect During STP #3 Replacement
DataStreams:sgp15ebbrE20.a1, sgp30ebbrE20.a1, sgp5ebbrE20.00, sgp5ebbrE20.a0
Description:
All data was incorrect during replacement of soil temperature 
probe #3.
Measurements:sgp5ebbrE20.00:
  • null(Raw data stream - documentation not supported)

sgp30ebbrE20.a1:
  • Bottom humidity(hum_bot)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • volumetric soil moisture, site 1(sm1)
  • 5 cm soil heat flow, site 1(shf1)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • volumetric soil moisture, site 5(sm5)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • net radiation(q)
  • 5 cm soil heat flow, site 2(shf2)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • 5 cm soil heat flow, site 5(shf5)
  • Top humidity(hum_top)
  • soil heat flow, site 5(g5)
  • Retrieved pressure profile(pres)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • volumetric soil moisture, site 3(sm3)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • volumetric soil moisture, site 2(sm2)
  • bottom vapor pressure(vp_bot)
  • soil heat flow, site 3(g3)
  • h(h)
  • Soil heat capacity 5(cs5)
  • wind direction (relative to true north)(wind_d)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • bottom air temperature(tair_bot)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • soil heat flow, site 1(g1)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • top air temperature(tair_top)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • top vapor pressure(vp_top)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • average surface soil heat flow(ave_shf)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • soil heat flow, site 2(g2)
  • Soil heat capacity 4(cs4)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • vector wind speed(res_ws)
  • Reference Thermistor Temperature(tref)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • Soil heat capacity 3(cs3)
  • volumetric soil moisture, site 4(sm4)
  • Temperature of the top humidity chamber(thum_top)
  • Soil heat capacity 2(cs2)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • 5 cm soil heat flow, site 4(shf4)
  • latent heat flux(e)
  • Soil heat capacity 1(cs1)
  • scalar wind speed(wind_s)
  • 5 cm soil heat flow, site 3(shf3)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)

sgp5ebbrE20.a0:
  • vector wind speed(res_ws)
  • Retrieved pressure profile(pres)
  • bottom vapor pressure(vp_bot)
  • Top humidity(hum_top)
  • top vapor pressure(vp_top)
  • Temperature of the top humidity chamber(thum_top)
  • Bottom humidity(hum_bot)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • bottom air temperature(tair_bot)
  • wind direction (relative to true north)(wind_d)
  • net radiation(q)
  • Reference Thermistor Temperature(tref)
  • top air temperature(tair_top)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Home signal(home)
  • scalar wind speed(wind_s)

sgp15ebbrE20.a1:
  • Soil moisture 4(r_sm4)
  • Right air temperature(tair_r)
  • Soil temperature 5(rr_ts5)
  • Left air temperature(tair_l)
  • Soil moisture 3(r_sm3)
  • Right relative humidity(mv_hum_r)
  • Home signal(mv_home)
  • Signature(signature)
  • Soil heat flow 2(mv_hft2)
  • scalar wind speed(wind_s)
  • Left relative humidity(mv_hum_l)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 5(mv_hft5)
  • Soil moisture 1(r_sm1)
  • Reference temperature(rr_tref)
  • Soil heat flow 3(mv_hft3)
  • Soil moisture 1(rr_sm1)
  • Soil heat flow 4(mv_hft4)
  • Soil moisture 5(r_sm5)
  • Soil moisture 2(r_sm2)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Net radiation(mv_q)
  • Soil temperature 1(rr_ts1)
  • Soil moisture 2(rr_sm2)
  • Atmospheric pressure(mv_pres)
  • Battery(bat)
  • Soil moisture 5(rr_sm5)
  • Soil temperature 4(rr_ts4)
  • Soil moisture 3(rr_sm3)
  • Soil temperature 3(rr_ts3)
  • Soil temperature 2(rr_ts2)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 4(rr_sm4)
  • Temperature of left humidity sensor chamber(rr_thum_l)


Back To Table of Contents

DQRID : D010605.13
Start DateStart TimeEnd DateEnd Time
11/29/2000183111/29/20001901
Subject:
SGP EBBR E20 - E20 EBBR Data Incorrect After Program Restart
DataStreams:sgp15ebbrE20.a1, sgp30ebbrE20.a1, sgp5ebbrE20.00, sgp5ebbrE20.a0
Description:
All data was incorrect after a datalogger program restart.
Measurements:sgp5ebbrE20.00:
  • null(Raw data stream - documentation not supported)

sgp30ebbrE20.a1:
  • Bottom humidity(hum_bot)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • volumetric soil moisture, site 1(sm1)
  • 5 cm soil heat flow, site 1(shf1)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • volumetric soil moisture, site 5(sm5)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • net radiation(q)
  • 5 cm soil heat flow, site 2(shf2)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • 5 cm soil heat flow, site 5(shf5)
  • Top humidity(hum_top)
  • soil heat flow, site 5(g5)
  • Retrieved pressure profile(pres)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • volumetric soil moisture, site 3(sm3)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • volumetric soil moisture, site 2(sm2)
  • bottom vapor pressure(vp_bot)
  • soil heat flow, site 3(g3)
  • h(h)
  • Soil heat capacity 5(cs5)
  • wind direction (relative to true north)(wind_d)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • bottom air temperature(tair_bot)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • soil heat flow, site 1(g1)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • top air temperature(tair_top)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • top vapor pressure(vp_top)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • average surface soil heat flow(ave_shf)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • soil heat flow, site 2(g2)
  • Soil heat capacity 4(cs4)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • vector wind speed(res_ws)
  • Reference Thermistor Temperature(tref)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • Soil heat capacity 3(cs3)
  • volumetric soil moisture, site 4(sm4)
  • Temperature of the top humidity chamber(thum_top)
  • Soil heat capacity 2(cs2)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • 5 cm soil heat flow, site 4(shf4)
  • latent heat flux(e)
  • Soil heat capacity 1(cs1)
  • scalar wind speed(wind_s)
  • 5 cm soil heat flow, site 3(shf3)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)

sgp5ebbrE20.a0:
  • vector wind speed(res_ws)
  • Retrieved pressure profile(pres)
  • bottom vapor pressure(vp_bot)
  • Top humidity(hum_top)
  • top vapor pressure(vp_top)
  • Temperature of the top humidity chamber(thum_top)
  • Bottom humidity(hum_bot)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • bottom air temperature(tair_bot)
  • wind direction (relative to true north)(wind_d)
  • net radiation(q)
  • Reference Thermistor Temperature(tref)
  • top air temperature(tair_top)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Home signal(home)
  • scalar wind speed(wind_s)

sgp15ebbrE20.a1:
  • Soil moisture 4(r_sm4)
  • Right air temperature(tair_r)
  • Soil temperature 5(rr_ts5)
  • Left air temperature(tair_l)
  • Soil moisture 3(r_sm3)
  • Right relative humidity(mv_hum_r)
  • Home signal(mv_home)
  • Signature(signature)
  • Soil heat flow 2(mv_hft2)
  • scalar wind speed(wind_s)
  • Left relative humidity(mv_hum_l)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 5(mv_hft5)
  • Soil moisture 1(r_sm1)
  • Reference temperature(rr_tref)
  • Soil heat flow 3(mv_hft3)
  • Soil moisture 1(rr_sm1)
  • Soil heat flow 4(mv_hft4)
  • Soil moisture 5(r_sm5)
  • Soil moisture 2(r_sm2)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Net radiation(mv_q)
  • Soil temperature 1(rr_ts1)
  • Soil moisture 2(rr_sm2)
  • Atmospheric pressure(mv_pres)
  • Battery(bat)
  • Soil moisture 5(rr_sm5)
  • Soil temperature 4(rr_ts4)
  • Soil moisture 3(rr_sm3)
  • Soil temperature 3(rr_ts3)
  • Soil temperature 2(rr_ts2)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 4(rr_sm4)
  • Temperature of left humidity sensor chamber(rr_thum_l)


Back To Table of Contents

DQRID : D010605.14
Start DateStart TimeEnd DateEnd Time
11/29/2000193011/29/20001931
Subject:
SGP EBBR E20 - E20 EBBR Some Data Incorrect
DataStreams:sgp30ebbrE20.a1
Description:
Quantities requiring a starting value were incorrect; sensible and latent heat 
fluxes were therefore incorrect also.
Measurements:sgp30ebbrE20.a1:
  • soil heat flow, site 1(g1)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • soil heat flow, site 5(g5)
  • soil heat flow, site 2(g2)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • latent heat flux(e)
  • soil heat flow, site 3(g3)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • h(h)
  • soil heat flow, site 4(g4)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • average surface soil heat flow(ave_shf)


Back To Table of Contents

DQRID : D010605.8
Start DateStart TimeEnd DateEnd Time
07/20/2000193008/02/20000331
Subject:
SGP EBBR E20 - E20 EBBR Data Incorrect
DataStreams:sgp15ebbrE20.a1, sgp30ebbrE20.a1, sgp5ebbrE20.00, sgp5ebbrE20.a0
Description:
The data was incorrect because battery voltage was low.
Measurements:sgp5ebbrE20.00:
  • null(Raw data stream - documentation not supported)

sgp30ebbrE20.a1:
  • Bottom humidity(hum_bot)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • volumetric soil moisture, site 1(sm1)
  • 5 cm soil heat flow, site 1(shf1)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • volumetric soil moisture, site 5(sm5)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • net radiation(q)
  • 5 cm soil heat flow, site 2(shf2)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • 5 cm soil heat flow, site 5(shf5)
  • Top humidity(hum_top)
  • soil heat flow, site 5(g5)
  • Retrieved pressure profile(pres)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • volumetric soil moisture, site 3(sm3)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • volumetric soil moisture, site 2(sm2)
  • bottom vapor pressure(vp_bot)
  • soil heat flow, site 3(g3)
  • h(h)
  • Soil heat capacity 5(cs5)
  • wind direction (relative to true north)(wind_d)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • bottom air temperature(tair_bot)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • soil heat flow, site 1(g1)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • top air temperature(tair_top)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • top vapor pressure(vp_top)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • average surface soil heat flow(ave_shf)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • soil heat flow, site 2(g2)
  • Soil heat capacity 4(cs4)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • vector wind speed(res_ws)
  • Reference Thermistor Temperature(tref)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • Soil heat capacity 3(cs3)
  • volumetric soil moisture, site 4(sm4)
  • Temperature of the top humidity chamber(thum_top)
  • Soil heat capacity 2(cs2)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • 5 cm soil heat flow, site 4(shf4)
  • latent heat flux(e)
  • Soil heat capacity 1(cs1)
  • scalar wind speed(wind_s)
  • 5 cm soil heat flow, site 3(shf3)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)

sgp5ebbrE20.a0:
  • vector wind speed(res_ws)
  • Retrieved pressure profile(pres)
  • bottom vapor pressure(vp_bot)
  • Top humidity(hum_top)
  • top vapor pressure(vp_top)
  • Temperature of the top humidity chamber(thum_top)
  • Bottom humidity(hum_bot)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • bottom air temperature(tair_bot)
  • wind direction (relative to true north)(wind_d)
  • net radiation(q)
  • Reference Thermistor Temperature(tref)
  • top air temperature(tair_top)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Home signal(home)
  • scalar wind speed(wind_s)

sgp15ebbrE20.a1:
  • Soil moisture 4(r_sm4)
  • Right air temperature(tair_r)
  • Soil temperature 5(rr_ts5)
  • Left air temperature(tair_l)
  • Soil moisture 3(r_sm3)
  • Right relative humidity(mv_hum_r)
  • Home signal(mv_home)
  • Signature(signature)
  • Soil heat flow 2(mv_hft2)
  • scalar wind speed(wind_s)
  • Left relative humidity(mv_hum_l)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 5(mv_hft5)
  • Soil moisture 1(r_sm1)
  • Reference temperature(rr_tref)
  • Soil heat flow 3(mv_hft3)
  • Soil moisture 1(rr_sm1)
  • Soil heat flow 4(mv_hft4)
  • Soil moisture 5(r_sm5)
  • Soil moisture 2(r_sm2)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Net radiation(mv_q)
  • Soil temperature 1(rr_ts1)
  • Soil moisture 2(rr_sm2)
  • Atmospheric pressure(mv_pres)
  • Battery(bat)
  • Soil moisture 5(rr_sm5)
  • Soil temperature 4(rr_ts4)
  • Soil moisture 3(rr_sm3)
  • Soil temperature 3(rr_ts3)
  • Soil temperature 2(rr_ts2)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 4(rr_sm4)
  • Temperature of left humidity sensor chamber(rr_thum_l)


Back To Table of Contents

DQRID : D010605.9
Start DateStart TimeEnd DateEnd Time
10/04/2000193010/04/20001931
Subject:
SGP EBBR E20 - E20 EBBR Net Radiation Data Incorrect
DataStreams:sgp15ebbrE20.a1, sgp30ebbrE20.a1, sgp5ebbrE20.a0
Description:
The net radiation data was incorrect while the top dome was being replaced.
Sensible and latent heat fluxes are therefore also incorrect.
Measurements:sgp30ebbrE20.a1:
  • net radiation(q)
  • latent heat flux(e)
  • h(h)

sgp5ebbrE20.a0:
  • net radiation(q)

sgp15ebbrE20.a1:
  • Net radiation(mv_q)


Back To Table of Contents

DQRID : D011102.6
Start DateStart TimeEnd DateEnd Time
08/14/2001020508/23/20011430
09/08/2001180509/09/20010200
Subject:
SGP EBBR E19 - E19 EBBR Data Missing
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.00,
sgp5ebbrE19.a1
Description:
Data is missing for this period. Reason for missing data: collection problem
'Can't open device /dev/ttySI4: Device or resource busy'
Measurements:sgp5ebbrE19.00:
  • null(Raw data stream - documentation not supported)

sgp15ebbrE19.a1:
  • Reference temperature(rr_tref)
  • Soil temperature 5(rr_ts5)
  • scalar wind speed(wind_s)
  • Soil temperature 4(rr_ts4)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 1(r_sm1)
  • Home signal(mv_home)
  • Soil moisture 3(r_sm3)
  • Soil moisture 2(r_sm2)
  • Left air temperature(tair_l)
  • Left relative humidity(mv_hum_l)
  • Soil heat flow 4(mv_hft4)
  • Soil heat flow 5(mv_hft5)
  • Right relative humidity(mv_hum_r)
  • Soil moisture 4(r_sm4)
  • Soil heat flow 3(mv_hft3)
  • Soil moisture 5(r_sm5)
  • Signature(signature)
  • Dummy altitude for Zeb(alt)
  • base time(base_time)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Atmospheric pressure(mv_pres)
  • lat(lat)
  • Right air temperature(tair_r)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 2(mv_hft2)
  • Battery(bat)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil temperature 1(rr_ts1)
  • Soil temperature 2(rr_ts2)
  • Time offset of tweaks from base_time(time_offset)
  • lon(lon)
  • Soil temperature 3(rr_ts3)
  • Net radiation(mv_q)

sgp15ebbrE19.b1:
  • Right relative humidity(mv_hum_r)
  • Soil heat flow 1(mv_hft1)
  • Soil temperature 1(rr_ts1)
  • Soil heat flow 3(mv_hft3)
  • Soil temperature 2(rr_ts2)
  • lon(lon)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Reference temperature(rr_tref)
  • Battery(bat)
  • Soil temperature 3(rr_ts3)
  • Home signal(mv_home)
  • Right air temperature(tair_r)
  • Dummy altitude for Zeb(alt)
  • Soil heat flow 4(mv_hft4)
  • Soil heat flow 5(mv_hft5)
  • lat(lat)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Atmospheric pressure(mv_pres)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Soil moisture 5(r_sm5)
  • Soil temperature 4(rr_ts4)
  • Net radiation(mv_q)
  • scalar wind speed(wind_s)
  • Time offset of tweaks from base_time(time_offset)
  • Soil moisture 2(r_sm2)
  • Soil temperature 5(rr_ts5)
  • Left relative humidity(mv_hum_l)
  • Wind direction (relative to true north)(mv_wind_d)
  • Signature(signature)
  • Soil moisture 4(r_sm4)
  • Soil moisture 3(r_sm3)
  • base time(base_time)
  • Soil moisture 1(r_sm1)

sgp5ebbrE19.a1:
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • Bottom humidity(hum_bot)
  • base time(base_time)
  • wind direction (relative to true north)(wind_d)
  • bottom air temperature(tair_bot)
  • top air temperature(tair_top)
  • scalar wind speed(wind_s)
  • top vapor pressure(vp_top)
  • lat(lat)
  • lon(lon)
  • vector wind speed(res_ws)
  • Top humidity(hum_top)
  • Dummy altitude for Zeb(alt)
  • Reference Thermistor Temperature(tref)
  • Retrieved pressure profile(pres)
  • Time offset of tweaks from base_time(time_offset)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Home signal(home)
  • net radiation(q)

sgp30ebbrE19.a1:
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • vector wind speed(res_ws)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • soil heat flow, site 2(g2)
  • soil heat flow, site 1(g1)
  • net radiation(q)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • volumetric soil moisture, site 3(sm3)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • average surface soil heat flow(ave_shf)
  • volumetric soil moisture, site 4(sm4)
  • Soil heat capacity 3(cs3)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • Soil heat capacity 5(cs5)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • base time(base_time)
  • Soil heat capacity 2(cs2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • volumetric soil moisture, site 2(sm2)
  • Soil heat capacity 1(cs1)
  • latent heat flux(e)
  • bottom vapor pressure(vp_bot)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 3(g3)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Top humidity(hum_top)
  • lat(lat)
  • soil heat flow, site 5(g5)
  • scalar wind speed(wind_s)
  • bottom air temperature(tair_bot)
  • volumetric soil moisture, site 1(sm1)
  • Dummy altitude for Zeb(alt)
  • 5 cm soil heat flow, site 1(shf1)
  • 5 cm soil heat flow, site 4(shf4)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • Reference Thermistor Temperature(tref)
  • 5 cm soil heat flow, site 5(shf5)
  • top vapor pressure(vp_top)
  • lon(lon)
  • Retrieved pressure profile(pres)
  • Temperature of the top humidity chamber(thum_top)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 2(shf2)
  • soil heat flow, site 4(g4)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • Time offset of tweaks from base_time(time_offset)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • Soil heat capacity 4(cs4)
  • h(h)
  • volumetric soil moisture, site 5(sm5)
  • Bottom humidity(hum_bot)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 5 cm soil heat flow, site 3(shf3)
  • top air temperature(tair_top)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)

sgp30ebbrE19.b1:
  • Dummy altitude for Zeb(alt)
  • Reference Thermistor Temperature(tref)
  • 5 cm soil heat flow, site 2(shf2)
  • volumetric soil moisture, site 1(sm1)
  • volumetric soil moisture, site 2(sm2)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • volumetric soil moisture, site 4(sm4)
  • h(h)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • top vapor pressure(vp_top)
  • average surface soil heat flow(ave_shf)
  • volumetric soil moisture, site 5(sm5)
  • latent heat flux(e)
  • lat(lat)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 2(g2)
  • Soil heat capacity 1(cs1)
  • base time(base_time)
  • soil heat flow, site 5(g5)
  • soil heat flow, site 3(g3)
  • Soil heat capacity 2(cs2)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 5(shf5)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • bottom air temperature(tair_bot)
  • net radiation(q)
  • 5 cm soil heat flow, site 4(shf4)
  • Retrieved pressure profile(pres)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Soil heat capacity 5(cs5)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • Soil heat capacity 3(cs3)
  • Temperature of the top humidity chamber(thum_top)
  • Soil heat capacity 4(cs4)
  • Bottom humidity(hum_bot)
  • soil heat flow, site 4(g4)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • lon(lon)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • vector wind speed(res_ws)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Time offset of tweaks from base_time(time_offset)
  • volumetric soil moisture, site 3(sm3)
  • soil heat flow, site 1(g1)
  • 5 cm soil heat flow, site 1(shf1)
  • bottom vapor pressure(vp_bot)
  • Top humidity(hum_top)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 5 cm soil heat flow, site 3(shf3)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)


Back To Table of Contents

DQRID : D011102.7
Start DateStart TimeEnd DateEnd Time
04/01/2001000004/01/20011631
Subject:
SGP EBBR E19 - E19 EBBR TS3 Intermittent Incorrect
DataStreams:sgp15ebbrE19.a1, sgp30ebbrE19.a1
Description:
Soil Temperature probe #3 was intermittently incorrect.  Examine the
data and compare G3 with G1-2 and G4-5 to see when TS3 is incorrect.
Measurements:sgp15ebbrE19.a1:
  • Soil temperature 3(rr_ts3)

sgp30ebbrE19.a1:
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • average surface soil heat flow(ave_shf)
  • soil heat flow, site 3(g3)
  • h(h)


Back To Table of Contents

DQRID : D011102.8
Start DateStart TimeEnd DateEnd Time
03/01/2001030003/01/20011651
Subject:
SGP EBBR E19 - E19 EBBR Wind Speed Incorrect
DataStreams:sgp15ebbrE19.a1, sgp30ebbrE19.a1, sgp5ebbrE19.a0
Description:
The wind speed sensor was iced stopped for a while on this day.
Measurements:sgp5ebbrE19.a0:
  • vector wind speed(res_ws)
  • scalar wind speed(wind_s)

sgp15ebbrE19.a1:
  • scalar wind speed(wind_s)

sgp30ebbrE19.a1:
  • scalar wind speed(wind_s)
  • vector wind speed(res_ws)


Back To Table of Contents

DQRID : D011127.30
Start DateStart TimeEnd DateEnd Time
04/19/2001170004/19/20012001
10/03/2001141310/03/20011506
Subject:
SGP EBBR E19 - E19 EBBR Data Incorrect During 6 Month Checks
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
Description:
The data was incorrect during the 6 month checks.
Measurements:sgp15ebbrE19.a1:
  • Right air temperature(tair_r)
  • Reference temperature(rr_tref)
  • Soil temperature 5(rr_ts5)
  • scalar wind speed(wind_s)
  • Soil temperature 4(rr_ts4)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 1(r_sm1)
  • Soil moisture 3(r_sm3)
  • Soil moisture 2(r_sm2)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Left relative humidity(mv_hum_l)
  • Soil heat flow 4(mv_hft4)
  • Right relative humidity(mv_hum_r)
  • Soil heat flow 5(mv_hft5)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil heat flow 3(mv_hft3)
  • Soil moisture 4(r_sm4)
  • Soil temperature 1(rr_ts1)
  • Soil temperature 2(rr_ts2)
  • Soil moisture 5(r_sm5)
  • Soil temperature 3(rr_ts3)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Atmospheric pressure(mv_pres)
  • Net radiation(mv_q)

sgp15ebbrE19.b1:
  • Right relative humidity(mv_hum_r)
  • Soil heat flow 1(mv_hft1)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil temperature 1(rr_ts1)
  • Soil heat flow 3(mv_hft3)
  • Atmospheric pressure(mv_pres)
  • Soil temperature 2(rr_ts2)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Soil moisture 5(r_sm5)
  • Soil temperature 4(rr_ts4)
  • Net radiation(mv_q)
  • scalar wind speed(wind_s)
  • Reference temperature(rr_tref)
  • Soil moisture 2(r_sm2)
  • Soil temperature 5(rr_ts5)
  • Left relative humidity(mv_hum_l)
  • Soil temperature 3(rr_ts3)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 4(r_sm4)
  • Right air temperature(tair_r)
  • Soil moisture 3(r_sm3)
  • Soil heat flow 4(mv_hft4)
  • Soil heat flow 5(mv_hft5)
  • Soil moisture 1(r_sm1)

sgp5ebbrE19.a1:
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • Bottom humidity(hum_bot)
  • Top humidity(hum_top)
  • vector wind speed(res_ws)
  • wind direction (relative to true north)(wind_d)
  • bottom air temperature(tair_bot)
  • Reference Thermistor Temperature(tref)
  • top air temperature(tair_top)
  • scalar wind speed(wind_s)
  • Retrieved pressure profile(pres)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • top vapor pressure(vp_top)
  • net radiation(q)

sgp30ebbrE19.a1:
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • vector wind speed(res_ws)
  • 5 cm soil heat flow, site 1(shf1)
  • soil heat flow, site 2(g2)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • soil heat flow, site 1(g1)
  • 5 cm soil heat flow, site 4(shf4)
  • net radiation(q)
  • Reference Thermistor Temperature(tref)
  • volumetric soil moisture, site 3(sm3)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • 5 cm soil heat flow, site 5(shf5)
  • average surface soil heat flow(ave_shf)
  • top vapor pressure(vp_top)
  • volumetric soil moisture, site 4(sm4)
  • Soil heat capacity 3(cs3)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • Soil heat capacity 5(cs5)
  • Retrieved pressure profile(pres)
  • Temperature of the top humidity chamber(thum_top)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 2(shf2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Soil heat capacity 2(cs2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • volumetric soil moisture, site 2(sm2)
  • Soil heat capacity 1(cs1)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • Soil heat capacity 4(cs4)
  • h(h)
  • bottom vapor pressure(vp_bot)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 3(g3)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Top humidity(hum_top)
  • volumetric soil moisture, site 5(sm5)
  • Bottom humidity(hum_bot)
  • soil heat flow, site 5(g5)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 5 cm soil heat flow, site 3(shf3)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • bottom air temperature(tair_bot)
  • volumetric soil moisture, site 1(sm1)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)

sgp30ebbrE19.b1:
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • Reference Thermistor Temperature(tref)
  • 5 cm soil heat flow, site 2(shf2)
  • volumetric soil moisture, site 1(sm1)
  • volumetric soil moisture, site 2(sm2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • Soil heat capacity 5(cs5)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • volumetric soil moisture, site 4(sm4)
  • Soil heat capacity 3(cs3)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • h(h)
  • Temperature of the top humidity chamber(thum_top)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • Soil heat capacity 4(cs4)
  • Bottom humidity(hum_bot)
  • soil heat flow, site 4(g4)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • top vapor pressure(vp_top)
  • vector wind speed(res_ws)
  • average surface soil heat flow(ave_shf)
  • volumetric soil moisture, site 5(sm5)
  • latent heat flux(e)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 2(g2)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Soil heat capacity 1(cs1)
  • volumetric soil moisture, site 3(sm3)
  • soil heat flow, site 5(g5)
  • soil heat flow, site 1(g1)
  • soil heat flow, site 3(g3)
  • bottom vapor pressure(vp_bot)
  • 5 cm soil heat flow, site 1(shf1)
  • Soil heat capacity 2(cs2)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 5(shf5)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • Top humidity(hum_top)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • bottom air temperature(tair_bot)
  • net radiation(q)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 5 cm soil heat flow, site 3(shf3)
  • 5 cm soil heat flow, site 4(shf4)
  • Retrieved pressure profile(pres)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)


Back To Table of Contents

DQRID : D011127.31
Start DateStart TimeEnd DateEnd Time
04/04/2001170004/04/20011736
10/17/2001150010/17/20011601
Subject:
SGP EBBR E20 - E20 EBBR Data Incorrect During 6 Month Checks
DataStreams:sgp15ebbrE20.a1, sgp15ebbrE20.b1, sgp30ebbrE20.a1, sgp30ebbrE20.b1, sgp5ebbrE20.a1
Description:
The data was incorrect during the 6 month checks.
Measurements:sgp30ebbrE20.b1:
  • 5 cm soil heat flow, site 2(shf2)
  • 5 cm soil heat flow, site 3(shf3)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • Reference Thermistor Temperature(tref)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • Soil heat capacity 3(cs3)
  • 5 cm soil heat flow, site 5(shf5)
  • average surface soil heat flow(ave_shf)
  • Bottom humidity(hum_bot)
  • Soil heat capacity 1(cs1)
  • volumetric soil moisture, site 2(sm2)
  • Soil heat capacity 4(cs4)
  • top vapor pressure(vp_top)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • scalar wind speed(wind_s)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • top air temperature(tair_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • net radiation(q)
  • bottom air temperature(tair_bot)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Soil heat capacity 2(cs2)
  • volumetric soil moisture, site 1(sm1)
  • soil heat flow, site 1(g1)
  • Top humidity(hum_top)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • volumetric soil moisture, site 5(sm5)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • volumetric soil moisture, site 4(sm4)
  • Temperature of the top humidity chamber(thum_top)
  • soil heat flow, site 5(g5)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • latent heat flux(e)
  • h(h)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 4(g4)
  • Retrieved pressure profile(pres)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • soil heat flow, site 2(g2)
  • vector wind speed(res_ws)
  • 5 cm soil heat flow, site 4(shf4)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • volumetric soil moisture, site 3(sm3)
  • 5 cm soil heat flow, site 1(shf1)
  • soil heat flow, site 3(g3)
  • bottom vapor pressure(vp_bot)
  • Soil heat capacity 5(cs5)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)

sgp5ebbrE20.a1:
  • net radiation(q)
  • scalar wind speed(wind_s)
  • vector wind speed(res_ws)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • top air temperature(tair_top)
  • Top humidity(hum_top)
  • Bottom humidity(hum_bot)
  • Temperature of the top humidity chamber(thum_top)
  • top vapor pressure(vp_top)
  • wind direction (relative to true north)(wind_d)
  • bottom air temperature(tair_bot)
  • Reference Thermistor Temperature(tref)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Retrieved pressure profile(pres)
  • bottom vapor pressure(vp_bot)

sgp15ebbrE20.b1:
  • Soil heat flow 4(mv_hft4)
  • Soil temperature 2(rr_ts2)
  • Soil heat flow 5(mv_hft5)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil moisture 3(r_sm3)
  • Reference temperature(rr_tref)
  • Soil temperature 3(rr_ts3)
  • Soil temperature 5(rr_ts5)
  • Soil temperature 4(rr_ts4)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Left air temperature(tair_l)
  • Soil temperature 1(rr_ts1)
  • Net radiation(mv_q)
  • Atmospheric pressure(mv_pres)
  • Soil moisture 4(r_sm4)
  • Soil moisture 5(r_sm5)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 2(mv_hft2)
  • Left relative humidity(mv_hum_l)
  • Soil heat flow 3(mv_hft3)
  • scalar wind speed(wind_s)
  • Right air temperature(tair_r)
  • Right relative humidity(mv_hum_r)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 2(r_sm2)
  • Soil moisture 1(r_sm1)

sgp30ebbrE20.a1:
  • Bottom humidity(hum_bot)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • volumetric soil moisture, site 1(sm1)
  • 5 cm soil heat flow, site 1(shf1)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • volumetric soil moisture, site 5(sm5)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • net radiation(q)
  • 5 cm soil heat flow, site 2(shf2)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • Top humidity(hum_top)
  • 5 cm soil heat flow, site 5(shf5)
  • soil heat flow, site 5(g5)
  • Retrieved pressure profile(pres)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • volumetric soil moisture, site 3(sm3)
  • volumetric soil moisture, site 2(sm2)
  • bottom vapor pressure(vp_bot)
  • soil heat flow, site 3(g3)
  • Soil heat capacity 5(cs5)
  • h(h)
  • wind direction (relative to true north)(wind_d)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • bottom air temperature(tair_bot)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • soil heat flow, site 1(g1)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • top air temperature(tair_top)
  • top vapor pressure(vp_top)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • average surface soil heat flow(ave_shf)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • soil heat flow, site 2(g2)
  • Soil heat capacity 4(cs4)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • vector wind speed(res_ws)
  • Reference Thermistor Temperature(tref)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • Soil heat capacity 3(cs3)
  • volumetric soil moisture, site 4(sm4)
  • Temperature of the top humidity chamber(thum_top)
  • Soil heat capacity 2(cs2)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • 5 cm soil heat flow, site 4(shf4)
  • latent heat flux(e)
  • Soil heat capacity 1(cs1)
  • scalar wind speed(wind_s)
  • 5 cm soil heat flow, site 3(shf3)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)

sgp15ebbrE20.a1:
  • Soil moisture 4(r_sm4)
  • Right air temperature(tair_r)
  • Soil temperature 5(rr_ts5)
  • Left air temperature(tair_l)
  • Soil moisture 3(r_sm3)
  • Right relative humidity(mv_hum_r)
  • Soil heat flow 2(mv_hft2)
  • scalar wind speed(wind_s)
  • Left relative humidity(mv_hum_l)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 5(mv_hft5)
  • Soil moisture 1(r_sm1)
  • Reference temperature(rr_tref)
  • Soil heat flow 3(mv_hft3)
  • Soil moisture 1(rr_sm1)
  • Soil heat flow 4(mv_hft4)
  • Soil moisture 5(r_sm5)
  • Soil moisture 2(r_sm2)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Net radiation(mv_q)
  • Soil temperature 1(rr_ts1)
  • Soil moisture 2(rr_sm2)
  • Atmospheric pressure(mv_pres)
  • Soil moisture 5(rr_sm5)
  • Soil temperature 4(rr_ts4)
  • Soil temperature 3(rr_ts3)
  • Soil moisture 3(rr_sm3)
  • Soil temperature 2(rr_ts2)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 4(rr_sm4)
  • Temperature of left humidity sensor chamber(rr_thum_l)


Back To Table of Contents

DQRID : D011127.4
Start DateStart TimeEnd DateEnd Time
11/01/2000230011/15/20002031
Subject:
SGP EBBR E20 - E20 EBBR Soil Temperature #3 Incorrect
DataStreams:sgp15ebbrE20.a1, sgp30ebbrE20.a1
Description:
The indicated measurements were incorrect while soil temperature probe #3
was not working.
Measurements:sgp30ebbrE20.a1:
  • latent heat flux(e)
  • soil heat flow, site 3(g3)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • Soil heat capacity 3(cs3)
  • h(h)
  • average surface soil heat flow(ave_shf)
  • 0-5 cm integrated soil temperature, site 3(ts3)

sgp15ebbrE20.a1:
  • Soil temperature 3(rr_ts3)


Back To Table of Contents

DQRID : D011218.10
Start DateStart TimeEnd DateEnd Time
02/21/2001200003/07/20011901
Subject:
SGP EBBR E20 - E20 EBBR Soil Heat Flow #2 Incorrect
DataStreams:sgp15ebbrE20.a1, sgp30ebbrE20.a1
Description:
Broken leads for SHF #2 caused measurements from it to be
incorrect during the indicated period.
Measurements:sgp30ebbrE20.a1:
  • latent heat flux(e)
  • 5 cm soil heat flow, site 2(shf2)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • soil heat flow, site 2(g2)
  • h(h)
  • average surface soil heat flow(ave_shf)

sgp15ebbrE20.a1:
  • Soil heat flow 2(mv_hft2)


Back To Table of Contents

DQRID : D011218.11
Start DateStart TimeEnd DateEnd Time
03/30/2001061503/30/20011431
03/31/2001111503/31/20011401
Subject:
SGP EBBR E20 - E20 EBBR Battery Voltage Low, Incorrect Data
DataStreams:sgp15ebbrE20.a1, sgp30ebbrE20.a1
Description:
Some measurements were incorrect because the battery voltage was low.
Measurements:sgp30ebbrE20.a1:
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • soil heat flow, site 1(g1)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • 5 cm soil heat flow, site 1(shf1)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • soil heat flow, site 3(g3)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • h(h)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • average surface soil heat flow(ave_shf)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 2(shf2)
  • 5 cm soil heat flow, site 5(shf5)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • soil heat flow, site 5(g5)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • soil heat flow, site 2(g2)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • 5 cm soil heat flow, site 4(shf4)
  • latent heat flux(e)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • 5 cm soil heat flow, site 3(shf3)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)

sgp15ebbrE20.a1:
  • Soil heat flow 2(mv_hft2)
  • Soil temperature 5(rr_ts5)
  • Soil temperature 4(rr_ts4)
  • Soil heat flow 1(mv_hft1)
  • Soil temperature 1(rr_ts1)
  • Soil heat flow 5(mv_hft5)
  • Soil temperature 3(rr_ts3)
  • Soil temperature 2(rr_ts2)
  • Soil heat flow 3(mv_hft3)
  • Soil heat flow 4(mv_hft4)


Back To Table of Contents

DQRID : D011218.13
Start DateStart TimeEnd DateEnd Time
04/02/2001124504/02/20011701
04/03/2001021504/03/20011501
04/06/2001111504/06/20011501
04/07/2001034504/07/20011401
04/23/2001104504/23/20011331
Subject:
SGP EBBR E20 - E20 EBBR Battery Voltage Low, Data Incorrect
DataStreams:sgp15ebbrE20.a1, sgp15ebbrE20.b1, sgp30ebbrE20.a1, sgp30ebbrE20.b1
Description:
The indicated data was incorrect because the battery voltage was low.
Measurements:sgp30ebbrE20.b1:
  • 5 cm soil heat flow, site 2(shf2)
  • soil heat flow, site 1(g1)
  • 5 cm soil heat flow, site 3(shf3)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • soil heat flow, site 5(g5)
  • latent heat flux(e)
  • h(h)
  • 5 cm soil heat flow, site 5(shf5)
  • average surface soil heat flow(ave_shf)
  • soil heat flow, site 4(g4)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • soil heat flow, site 2(g2)
  • 5 cm soil heat flow, site 4(shf4)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • 5 cm soil heat flow, site 1(shf1)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • soil heat flow, site 3(g3)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)

sgp15ebbrE20.b1:
  • Soil heat flow 4(mv_hft4)
  • Soil temperature 2(rr_ts2)
  • Soil heat flow 5(mv_hft5)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 2(mv_hft2)
  • Soil temperature 3(rr_ts3)
  • Soil temperature 5(rr_ts5)
  • Soil temperature 4(rr_ts4)
  • Soil heat flow 3(mv_hft3)
  • Soil temperature 1(rr_ts1)

sgp30ebbrE20.a1:
  • soil heat flow, site 1(g1)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • 5 cm soil heat flow, site 1(shf1)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • average surface soil heat flow(ave_shf)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 2(shf2)
  • 5 cm soil heat flow, site 5(shf5)
  • soil heat flow, site 5(g5)
  • soil heat flow, site 2(g2)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • soil heat flow, site 3(g3)
  • h(h)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 5 cm soil heat flow, site 4(shf4)
  • latent heat flux(e)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • 5 cm soil heat flow, site 3(shf3)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)

sgp15ebbrE20.a1:
  • Soil temperature 5(rr_ts5)
  • Soil temperature 1(rr_ts1)
  • Soil heat flow 2(mv_hft2)
  • Soil temperature 4(rr_ts4)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 5(mv_hft5)
  • Soil temperature 3(rr_ts3)
  • Soil temperature 2(rr_ts2)
  • Soil heat flow 3(mv_hft3)
  • Soil heat flow 4(mv_hft4)


Back To Table of Contents

DQRID : D011218.14
Start DateStart TimeEnd DateEnd Time
04/15/2001150010/17/20010201
Subject:
SGP/EBBR/E20 - Reprocess: Soil Temperature #3 Incorrect
DataStreams:sgp15ebbrE20.a1, sgp15ebbrE20.b1, sgp30ebbrE20.a1, sgp30ebbrE20.b1
Description:
During the indicated period, TS #3 was large negative or positive or offscale.
Average soil heat flow, sensible heat flux, and latent heat flux were incorrect
when TS #3 was incorrect. Inspect the 15 and 30 minute TS #3 data to determine 
the times of incorrect data.
Measurements:sgp30ebbrE20.b1:
  • latent heat flux(e)
  • h(h)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • average surface soil heat flow(ave_shf)
  • soil heat flow, site 3(g3)

sgp15ebbrE20.b1:
  • Soil temperature 3(rr_ts3)

sgp30ebbrE20.a1:
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • latent heat flux(e)
  • soil heat flow, site 3(g3)
  • h(h)
  • average surface soil heat flow(ave_shf)

sgp15ebbrE20.a1:
  • Soil temperature 3(rr_ts3)


Back To Table of Contents

DQRID : D011218.9
Start DateStart TimeEnd DateEnd Time
02/01/2001000002/22/20010531
03/01/2001123003/01/20011231
03/07/2001180503/07/20011901
Subject:
SGP EBBR E20 - E20 EBBR Soil Heat Flow #4 Incorrect
DataStreams:sgp15ebbrE20.a1, sgp30ebbrE20.a1
Description:
The SHF #4 produced large negative values during much of the
period because of a broken lead wire. Inspect the data to 
determine the times of incorrect SHF #4 values.
Measurements:sgp30ebbrE20.a1:
  • latent heat flux(e)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • h(h)
  • soil heat flow, site 4(g4)
  • 5 cm soil heat flow, site 4(shf4)
  • average surface soil heat flow(ave_shf)

sgp15ebbrE20.a1:
  • Soil heat flow 4(mv_hft4)


Back To Table of Contents

DQRID : D020716.15
Start DateStart TimeEnd DateEnd Time
05/16/2002143005/16/20021501
Subject:
SGP/EBBR/E19 - 6 Month Checks
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
Description:
The indicated measurements were incorrect during the 6 month checks.
Measurements:sgp15ebbrE19.a1:
  • Right air temperature(tair_r)
  • Left air temperature(tair_l)

sgp15ebbrE19.b1:
  • Right air temperature(tair_r)
  • Left air temperature(tair_l)

sgp5ebbrE19.a1:
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • bottom vapor pressure(vp_bot)
  • top vapor pressure(vp_top)
  • Temperature of the top humidity chamber(thum_top)
  • Bottom humidity(hum_bot)
  • bottom air temperature(tair_bot)
  • Top humidity(hum_top)
  • top air temperature(tair_top)

sgp30ebbrE19.a1:
  • top vapor pressure(vp_top)
  • latent heat flux(e)
  • h(h)
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • top air temperature(tair_top)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • bottom air temperature(tair_bot)
  • Top humidity(hum_top)
  • Bottom humidity(hum_bot)

sgp30ebbrE19.b1:
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • h(h)
  • Temperature of the top humidity chamber(thum_top)
  • top air temperature(tair_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Top humidity(hum_top)
  • Bottom humidity(hum_bot)
  • bottom air temperature(tair_bot)
  • top vapor pressure(vp_top)
  • bottom vapor pressure(vp_bot)
  • latent heat flux(e)


Back To Table of Contents

DQRID : D020716.16
Start DateStart TimeEnd DateEnd Time
05/01/2002191005/01/20021946
Subject:
SGP/EBBR/E20 - 6 Month Checks
DataStreams:sgp15ebbrE20.a1, sgp15ebbrE20.b1, sgp30ebbrE20.a1, sgp30ebbrE20.b1, sgp5ebbrE20.a1
Description:
The indicated measurements were incorrect during the 6 month checks.
Measurements:sgp30ebbrE20.b1:
  • Bottom humidity(hum_bot)
  • Top humidity(hum_top)
  • top vapor pressure(vp_top)
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • top air temperature(tair_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • latent heat flux(e)
  • h(h)
  • bottom air temperature(tair_bot)

sgp15ebbrE20.b1:
  • Left air temperature(tair_l)
  • Right air temperature(tair_r)

sgp5ebbrE20.a1:
  • Temperature of the top humidity chamber(thum_top)
  • top vapor pressure(vp_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • bottom vapor pressure(vp_bot)
  • top air temperature(tair_top)
  • bottom air temperature(tair_bot)
  • Top humidity(hum_top)
  • Bottom humidity(hum_bot)

sgp30ebbrE20.a1:
  • Bottom humidity(hum_bot)
  • Temperature of the top humidity chamber(thum_top)
  • bottom vapor pressure(vp_bot)
  • top air temperature(tair_top)
  • top vapor pressure(vp_top)
  • h(h)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • Top humidity(hum_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • latent heat flux(e)
  • bottom air temperature(tair_bot)

sgp15ebbrE20.a1:
  • Right air temperature(tair_r)
  • Left air temperature(tair_l)


Back To Table of Contents

DQRID : D020717.26
Start DateStart TimeEnd DateEnd Time
02/08/2002163002/21/20021601
Subject:
SGP/EBBR/E19 - Reprocess: Soil Heat Flow #3 Incorrect
DataStreams:sgp30ebbrE19.a1, sgp30ebbrE19.b1
Description:
Soil heat flow 3 was incorrect, with values too large positive and too small negative.  
Since the calibration in the program is the one supplied by the vendor, it appears 
that the calibration may be wrong.

Average soil heat flow, sensible heat flux and latent heat flux can be recalculated
using the remaining good sets of soil sensors as follows:
   ave_shf = (g1 + g2 + g4 + g5)/4
   e = -(q + ave_shf)/(1 + bowen)
   h = -(q + e + ave_shf)
Measurements:sgp30ebbrE19.a1:
  • latent heat flux(e)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • average surface soil heat flow(ave_shf)
  • 5 cm soil heat flow, site 3(shf3)
  • soil heat flow, site 3(g3)
  • h(h)

sgp30ebbrE19.b1:
  • soil heat flow, site 3(g3)
  • average surface soil heat flow(ave_shf)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • h(h)
  • latent heat flux(e)
  • 5 cm soil heat flow, site 3(shf3)


Back To Table of Contents

DQRID : D020717.27
Start DateStart TimeEnd DateEnd Time
11/01/2001000006/30/20022356
Subject:
SGP/EBBR/E19 - Reprocess: Soil Temperature #3 Incorrect
DataStreams:sgp30ebbrE19.a1, sgp30ebbrE19.b1
Description:
Soil Temperature 3 was incorrect for the period indicated, causing average soil flow, 
latent heat flux, and sensible heat flux to be incorrect also.

Average soil heat flow, sensible heat flux, and latent heat flux can be recalculated using 
the remaining good sets of soil sensors as follows:

ave_shf = (g1 + g2 + g4 + g5)/4

e = -(q + ave_shf)/(1 + bowen)

h = -(q + e + ave_shf)
Measurements:sgp30ebbrE19.a1:
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • average surface soil heat flow(ave_shf)
  • soil heat flow, site 3(g3)
  • h(h)

sgp30ebbrE19.b1:
  • soil heat flow, site 3(g3)
  • average surface soil heat flow(ave_shf)
  • h(h)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 0-5 cm integrated soil temperature, site 3(ts3)


Back To Table of Contents

DQRID : D020717.29
Start DateStart TimeEnd DateEnd Time
04/08/2002043004/08/20020431
05/19/2002040005/19/20020401
Subject:
SGP/EBBR/E19 - Most Measurements Incorrect
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.00,
sgp5ebbrE19.a1
Description:
Most measurements were incorrect for the indicated periods.
Measurements:sgp5ebbrE19.00:
  • null(Raw data stream - documentation not supported)

sgp15ebbrE19.a1:
  • Right air temperature(tair_r)
  • Reference temperature(rr_tref)
  • Soil temperature 5(rr_ts5)
  • scalar wind speed(wind_s)
  • Soil temperature 4(rr_ts4)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 1(r_sm1)
  • Home signal(mv_home)
  • Soil moisture 3(r_sm3)
  • Soil moisture 2(r_sm2)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Left relative humidity(mv_hum_l)
  • Battery(bat)
  • Soil heat flow 4(mv_hft4)
  • Right relative humidity(mv_hum_r)
  • Soil heat flow 5(mv_hft5)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil heat flow 3(mv_hft3)
  • Soil moisture 4(r_sm4)
  • Soil temperature 1(rr_ts1)
  • Soil temperature 2(rr_ts2)
  • Soil moisture 5(r_sm5)
  • Soil temperature 3(rr_ts3)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Atmospheric pressure(mv_pres)
  • Net radiation(mv_q)

sgp15ebbrE19.b1:
  • Right relative humidity(mv_hum_r)
  • Soil heat flow 1(mv_hft1)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil temperature 1(rr_ts1)
  • Soil heat flow 3(mv_hft3)
  • Atmospheric pressure(mv_pres)
  • Soil temperature 2(rr_ts2)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Soil moisture 5(r_sm5)
  • Soil temperature 4(rr_ts4)
  • Net radiation(mv_q)
  • scalar wind speed(wind_s)
  • Reference temperature(rr_tref)
  • Soil moisture 2(r_sm2)
  • Soil temperature 5(rr_ts5)
  • Battery(bat)
  • Left relative humidity(mv_hum_l)
  • Soil temperature 3(rr_ts3)
  • Home signal(mv_home)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 4(r_sm4)
  • Right air temperature(tair_r)
  • Soil moisture 3(r_sm3)
  • Soil heat flow 4(mv_hft4)
  • Soil heat flow 5(mv_hft5)
  • Soil moisture 1(r_sm1)

sgp5ebbrE19.a1:
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • Bottom humidity(hum_bot)
  • Top humidity(hum_top)
  • vector wind speed(res_ws)
  • wind direction (relative to true north)(wind_d)
  • bottom air temperature(tair_bot)
  • Reference Thermistor Temperature(tref)
  • top air temperature(tair_top)
  • scalar wind speed(wind_s)
  • Retrieved pressure profile(pres)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Home signal(home)
  • top vapor pressure(vp_top)
  • net radiation(q)

sgp30ebbrE19.a1:
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • vector wind speed(res_ws)
  • 5 cm soil heat flow, site 1(shf1)
  • soil heat flow, site 2(g2)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • soil heat flow, site 1(g1)
  • 5 cm soil heat flow, site 4(shf4)
  • net radiation(q)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • Reference Thermistor Temperature(tref)
  • volumetric soil moisture, site 3(sm3)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • 5 cm soil heat flow, site 5(shf5)
  • average surface soil heat flow(ave_shf)
  • top vapor pressure(vp_top)
  • volumetric soil moisture, site 4(sm4)
  • Soil heat capacity 3(cs3)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • Soil heat capacity 5(cs5)
  • Retrieved pressure profile(pres)
  • Temperature of the top humidity chamber(thum_top)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 2(shf2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Soil heat capacity 2(cs2)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • volumetric soil moisture, site 2(sm2)
  • Soil heat capacity 1(cs1)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • Soil heat capacity 4(cs4)
  • h(h)
  • bottom vapor pressure(vp_bot)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 3(g3)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Top humidity(hum_top)
  • volumetric soil moisture, site 5(sm5)
  • Bottom humidity(hum_bot)
  • soil heat flow, site 5(g5)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 5 cm soil heat flow, site 3(shf3)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • bottom air temperature(tair_bot)
  • volumetric soil moisture, site 1(sm1)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)

sgp30ebbrE19.b1:
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • Reference Thermistor Temperature(tref)
  • 5 cm soil heat flow, site 2(shf2)
  • volumetric soil moisture, site 1(sm1)
  • volumetric soil moisture, site 2(sm2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • Soil heat capacity 5(cs5)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • volumetric soil moisture, site 4(sm4)
  • Soil heat capacity 3(cs3)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • h(h)
  • Temperature of the top humidity chamber(thum_top)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • Soil heat capacity 4(cs4)
  • Bottom humidity(hum_bot)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • soil heat flow, site 4(g4)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • top vapor pressure(vp_top)
  • vector wind speed(res_ws)
  • average surface soil heat flow(ave_shf)
  • volumetric soil moisture, site 5(sm5)
  • latent heat flux(e)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 2(g2)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Soil heat capacity 1(cs1)
  • volumetric soil moisture, site 3(sm3)
  • soil heat flow, site 5(g5)
  • soil heat flow, site 1(g1)
  • soil heat flow, site 3(g3)
  • bottom vapor pressure(vp_bot)
  • 5 cm soil heat flow, site 1(shf1)
  • Soil heat capacity 2(cs2)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 5(shf5)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • Top humidity(hum_top)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • bottom air temperature(tair_bot)
  • net radiation(q)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 5 cm soil heat flow, site 3(shf3)
  • 5 cm soil heat flow, site 4(shf4)
  • Retrieved pressure profile(pres)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)


Back To Table of Contents

DQRID : D020717.30
Start DateStart TimeEnd DateEnd Time
04/18/2002173004/18/20021731
05/05/2002140005/05/20021401
06/27/2002150006/27/20021501
Subject:
SGP/EBBR/E19 - All Measurements Zero
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.00,
sgp5ebbrE19.a1
Description:
Measurements are often zero the first period after a CR10 restart.
Measurements:sgp5ebbrE19.00:
  • null(Raw data stream - documentation not supported)

sgp15ebbrE19.a1:
  • Right air temperature(tair_r)
  • Reference temperature(rr_tref)
  • Soil temperature 5(rr_ts5)
  • scalar wind speed(wind_s)
  • Soil temperature 4(rr_ts4)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 1(r_sm1)
  • Home signal(mv_home)
  • Soil moisture 3(r_sm3)
  • Soil moisture 2(r_sm2)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Left relative humidity(mv_hum_l)
  • Battery(bat)
  • Soil heat flow 4(mv_hft4)
  • Right relative humidity(mv_hum_r)
  • Soil heat flow 5(mv_hft5)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil heat flow 3(mv_hft3)
  • Soil moisture 4(r_sm4)
  • Soil temperature 1(rr_ts1)
  • Soil temperature 2(rr_ts2)
  • Soil moisture 5(r_sm5)
  • Soil temperature 3(rr_ts3)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Atmospheric pressure(mv_pres)
  • Net radiation(mv_q)

sgp15ebbrE19.b1:
  • Right relative humidity(mv_hum_r)
  • Soil heat flow 1(mv_hft1)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil temperature 1(rr_ts1)
  • Soil heat flow 3(mv_hft3)
  • Atmospheric pressure(mv_pres)
  • Soil temperature 2(rr_ts2)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Soil moisture 5(r_sm5)
  • Soil temperature 4(rr_ts4)
  • Net radiation(mv_q)
  • scalar wind speed(wind_s)
  • Reference temperature(rr_tref)
  • Soil moisture 2(r_sm2)
  • Soil temperature 5(rr_ts5)
  • Battery(bat)
  • Left relative humidity(mv_hum_l)
  • Soil temperature 3(rr_ts3)
  • Home signal(mv_home)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 4(r_sm4)
  • Right air temperature(tair_r)
  • Soil moisture 3(r_sm3)
  • Soil heat flow 4(mv_hft4)
  • Soil heat flow 5(mv_hft5)
  • Soil moisture 1(r_sm1)

sgp5ebbrE19.a1:
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • Bottom humidity(hum_bot)
  • Top humidity(hum_top)
  • vector wind speed(res_ws)
  • wind direction (relative to true north)(wind_d)
  • bottom air temperature(tair_bot)
  • Reference Thermistor Temperature(tref)
  • top air temperature(tair_top)
  • scalar wind speed(wind_s)
  • Retrieved pressure profile(pres)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Home signal(home)
  • top vapor pressure(vp_top)
  • net radiation(q)

sgp30ebbrE19.a1:
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • vector wind speed(res_ws)
  • 5 cm soil heat flow, site 1(shf1)
  • soil heat flow, site 2(g2)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • soil heat flow, site 1(g1)
  • 5 cm soil heat flow, site 4(shf4)
  • net radiation(q)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • Reference Thermistor Temperature(tref)
  • volumetric soil moisture, site 3(sm3)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • 5 cm soil heat flow, site 5(shf5)
  • average surface soil heat flow(ave_shf)
  • top vapor pressure(vp_top)
  • volumetric soil moisture, site 4(sm4)
  • Soil heat capacity 3(cs3)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • Soil heat capacity 5(cs5)
  • Retrieved pressure profile(pres)
  • Temperature of the top humidity chamber(thum_top)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 2(shf2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Soil heat capacity 2(cs2)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • volumetric soil moisture, site 2(sm2)
  • Soil heat capacity 1(cs1)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • Soil heat capacity 4(cs4)
  • h(h)
  • bottom vapor pressure(vp_bot)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 3(g3)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Top humidity(hum_top)
  • volumetric soil moisture, site 5(sm5)
  • Bottom humidity(hum_bot)
  • soil heat flow, site 5(g5)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 5 cm soil heat flow, site 3(shf3)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • bottom air temperature(tair_bot)
  • volumetric soil moisture, site 1(sm1)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)

sgp30ebbrE19.b1:
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • Reference Thermistor Temperature(tref)
  • 5 cm soil heat flow, site 2(shf2)
  • volumetric soil moisture, site 1(sm1)
  • volumetric soil moisture, site 2(sm2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • Soil heat capacity 5(cs5)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • volumetric soil moisture, site 4(sm4)
  • Soil heat capacity 3(cs3)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • h(h)
  • Temperature of the top humidity chamber(thum_top)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • Soil heat capacity 4(cs4)
  • Bottom humidity(hum_bot)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • soil heat flow, site 4(g4)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • top vapor pressure(vp_top)
  • vector wind speed(res_ws)
  • average surface soil heat flow(ave_shf)
  • volumetric soil moisture, site 5(sm5)
  • latent heat flux(e)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 2(g2)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Soil heat capacity 1(cs1)
  • volumetric soil moisture, site 3(sm3)
  • soil heat flow, site 5(g5)
  • soil heat flow, site 1(g1)
  • soil heat flow, site 3(g3)
  • bottom vapor pressure(vp_bot)
  • 5 cm soil heat flow, site 1(shf1)
  • Soil heat capacity 2(cs2)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 5(shf5)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • Top humidity(hum_top)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • bottom air temperature(tair_bot)
  • net radiation(q)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 5 cm soil heat flow, site 3(shf3)
  • 5 cm soil heat flow, site 4(shf4)
  • Retrieved pressure profile(pres)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)


Back To Table of Contents

DQRID : D020717.31
Start DateStart TimeEnd DateEnd Time
04/24/2002160004/24/20022201
05/25/2002210005/25/20022101
05/25/2002220005/25/20022231
05/29/2002173005/29/20021731
05/29/2002183005/29/20021831
05/29/2002200005/29/20022001
05/29/2002213005/29/20022131
05/30/2002150005/30/20021501
06/15/2002190006/15/20021901
Subject:
SGP/EBBR/E19 - Pressure Incorrect
DataStreams:sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
Description:
The pressure measurement was incorrect.  This is used to determine the Bowen ratio, 
thereby affecting the sensible and latent heat fluxes also.
Measurements:sgp5ebbrE19.a1:
  • Retrieved pressure profile(pres)

sgp30ebbrE19.a1:
  • latent heat flux(e)
  • h(h)
  • Retrieved pressure profile(pres)

sgp30ebbrE19.b1:
  • Retrieved pressure profile(pres)
  • h(h)
  • latent heat flux(e)


Back To Table of Contents

DQRID : D020717.32
Start DateStart TimeEnd DateEnd Time
05/16/2002200005/17/20020531
Subject:
SGP/EBBR/E19 - Measurements Incorrect After CR10 Replacement
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.00,
sgp5ebbrE19.a1
Description:
Replacement of the CR10 datalogger caused most measurements to be incorrect, reason 
unknown.
A different logger was installed at the end of the specified period and this resolved the 
problem.
Measurements:sgp5ebbrE19.00:
  • null(Raw data stream - documentation not supported)

sgp15ebbrE19.a1:
  • Right air temperature(tair_r)
  • Reference temperature(rr_tref)
  • Soil temperature 5(rr_ts5)
  • scalar wind speed(wind_s)
  • Soil temperature 4(rr_ts4)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 1(r_sm1)
  • Home signal(mv_home)
  • Soil moisture 3(r_sm3)
  • Soil moisture 2(r_sm2)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Left relative humidity(mv_hum_l)
  • Battery(bat)
  • Soil heat flow 4(mv_hft4)
  • Right relative humidity(mv_hum_r)
  • Soil heat flow 5(mv_hft5)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil heat flow 3(mv_hft3)
  • Soil moisture 4(r_sm4)
  • Soil temperature 1(rr_ts1)
  • Soil temperature 2(rr_ts2)
  • Soil moisture 5(r_sm5)
  • Soil temperature 3(rr_ts3)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Atmospheric pressure(mv_pres)
  • Net radiation(mv_q)

sgp15ebbrE19.b1:
  • Right relative humidity(mv_hum_r)
  • Soil heat flow 1(mv_hft1)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil temperature 1(rr_ts1)
  • Soil heat flow 3(mv_hft3)
  • Atmospheric pressure(mv_pres)
  • Soil temperature 2(rr_ts2)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Soil moisture 5(r_sm5)
  • Soil temperature 4(rr_ts4)
  • Net radiation(mv_q)
  • scalar wind speed(wind_s)
  • Reference temperature(rr_tref)
  • Soil moisture 2(r_sm2)
  • Soil temperature 5(rr_ts5)
  • Battery(bat)
  • Left relative humidity(mv_hum_l)
  • Soil temperature 3(rr_ts3)
  • Home signal(mv_home)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 4(r_sm4)
  • Right air temperature(tair_r)
  • Soil moisture 3(r_sm3)
  • Soil heat flow 4(mv_hft4)
  • Soil heat flow 5(mv_hft5)
  • Soil moisture 1(r_sm1)

sgp5ebbrE19.a1:
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • Bottom humidity(hum_bot)
  • Top humidity(hum_top)
  • vector wind speed(res_ws)
  • wind direction (relative to true north)(wind_d)
  • bottom air temperature(tair_bot)
  • Reference Thermistor Temperature(tref)
  • top air temperature(tair_top)
  • scalar wind speed(wind_s)
  • Retrieved pressure profile(pres)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Home signal(home)
  • top vapor pressure(vp_top)
  • net radiation(q)

sgp30ebbrE19.a1:
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • vector wind speed(res_ws)
  • 5 cm soil heat flow, site 1(shf1)
  • soil heat flow, site 2(g2)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • soil heat flow, site 1(g1)
  • 5 cm soil heat flow, site 4(shf4)
  • net radiation(q)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • Reference Thermistor Temperature(tref)
  • volumetric soil moisture, site 3(sm3)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • 5 cm soil heat flow, site 5(shf5)
  • average surface soil heat flow(ave_shf)
  • top vapor pressure(vp_top)
  • volumetric soil moisture, site 4(sm4)
  • Soil heat capacity 3(cs3)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • Soil heat capacity 5(cs5)
  • Retrieved pressure profile(pres)
  • Temperature of the top humidity chamber(thum_top)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 2(shf2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Soil heat capacity 2(cs2)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • volumetric soil moisture, site 2(sm2)
  • Soil heat capacity 1(cs1)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • Soil heat capacity 4(cs4)
  • h(h)
  • bottom vapor pressure(vp_bot)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 3(g3)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Top humidity(hum_top)
  • volumetric soil moisture, site 5(sm5)
  • Bottom humidity(hum_bot)
  • soil heat flow, site 5(g5)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 5 cm soil heat flow, site 3(shf3)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • bottom air temperature(tair_bot)
  • volumetric soil moisture, site 1(sm1)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)

sgp30ebbrE19.b1:
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • Reference Thermistor Temperature(tref)
  • 5 cm soil heat flow, site 2(shf2)
  • volumetric soil moisture, site 1(sm1)
  • volumetric soil moisture, site 2(sm2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • Soil heat capacity 5(cs5)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • volumetric soil moisture, site 4(sm4)
  • Soil heat capacity 3(cs3)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • h(h)
  • Temperature of the top humidity chamber(thum_top)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • Soil heat capacity 4(cs4)
  • Bottom humidity(hum_bot)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • soil heat flow, site 4(g4)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • top vapor pressure(vp_top)
  • vector wind speed(res_ws)
  • average surface soil heat flow(ave_shf)
  • volumetric soil moisture, site 5(sm5)
  • latent heat flux(e)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 2(g2)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Soil heat capacity 1(cs1)
  • volumetric soil moisture, site 3(sm3)
  • soil heat flow, site 5(g5)
  • soil heat flow, site 1(g1)
  • soil heat flow, site 3(g3)
  • bottom vapor pressure(vp_bot)
  • 5 cm soil heat flow, site 1(shf1)
  • Soil heat capacity 2(cs2)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 5(shf5)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • Top humidity(hum_top)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • bottom air temperature(tair_bot)
  • net radiation(q)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 5 cm soil heat flow, site 3(shf3)
  • 5 cm soil heat flow, site 4(shf4)
  • Retrieved pressure profile(pres)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)


Back To Table of Contents

DQRID : D020717.33
Start DateStart TimeEnd DateEnd Time
05/27/2002000006/02/20021031
Subject:
SGP/EBBR/E19 - Low Battery Voltage
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.00,
sgp5ebbrE19.a1
Description:
A battery voltage of less than 10.5 causes many of the sensors to produce incorrect 
measurements.
The 15 minute data contains the battery voltage and should be used to determine whether 
the measurements are usable, based on battery voltage.
Measurements:sgp5ebbrE19.00:
  • null(Raw data stream - documentation not supported)

sgp15ebbrE19.a1:
  • Right air temperature(tair_r)
  • Reference temperature(rr_tref)
  • Soil temperature 5(rr_ts5)
  • scalar wind speed(wind_s)
  • Soil temperature 4(rr_ts4)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 1(r_sm1)
  • Home signal(mv_home)
  • Soil moisture 3(r_sm3)
  • Soil moisture 2(r_sm2)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Left relative humidity(mv_hum_l)
  • Soil heat flow 4(mv_hft4)
  • Right relative humidity(mv_hum_r)
  • Soil heat flow 5(mv_hft5)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil heat flow 3(mv_hft3)
  • Soil moisture 4(r_sm4)
  • Soil temperature 1(rr_ts1)
  • Soil temperature 2(rr_ts2)
  • Soil moisture 5(r_sm5)
  • Soil temperature 3(rr_ts3)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Atmospheric pressure(mv_pres)
  • Net radiation(mv_q)

sgp15ebbrE19.b1:
  • Right relative humidity(mv_hum_r)
  • Soil heat flow 1(mv_hft1)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil temperature 1(rr_ts1)
  • Soil heat flow 3(mv_hft3)
  • Atmospheric pressure(mv_pres)
  • Soil temperature 2(rr_ts2)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Soil moisture 5(r_sm5)
  • Soil temperature 4(rr_ts4)
  • Net radiation(mv_q)
  • scalar wind speed(wind_s)
  • Reference temperature(rr_tref)
  • Soil moisture 2(r_sm2)
  • Soil temperature 5(rr_ts5)
  • Left relative humidity(mv_hum_l)
  • Soil temperature 3(rr_ts3)
  • Home signal(mv_home)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 4(r_sm4)
  • Right air temperature(tair_r)
  • Soil moisture 3(r_sm3)
  • Soil heat flow 4(mv_hft4)
  • Soil heat flow 5(mv_hft5)
  • Soil moisture 1(r_sm1)

sgp5ebbrE19.a1:
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • Bottom humidity(hum_bot)
  • Top humidity(hum_top)
  • vector wind speed(res_ws)
  • wind direction (relative to true north)(wind_d)
  • bottom air temperature(tair_bot)
  • Reference Thermistor Temperature(tref)
  • top air temperature(tair_top)
  • scalar wind speed(wind_s)
  • Retrieved pressure profile(pres)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Home signal(home)
  • top vapor pressure(vp_top)
  • net radiation(q)

sgp30ebbrE19.a1:
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • vector wind speed(res_ws)
  • 5 cm soil heat flow, site 1(shf1)
  • soil heat flow, site 2(g2)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • soil heat flow, site 1(g1)
  • 5 cm soil heat flow, site 4(shf4)
  • net radiation(q)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • Reference Thermistor Temperature(tref)
  • volumetric soil moisture, site 3(sm3)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • 5 cm soil heat flow, site 5(shf5)
  • average surface soil heat flow(ave_shf)
  • top vapor pressure(vp_top)
  • volumetric soil moisture, site 4(sm4)
  • Soil heat capacity 3(cs3)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • Soil heat capacity 5(cs5)
  • Retrieved pressure profile(pres)
  • Temperature of the top humidity chamber(thum_top)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 2(shf2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Soil heat capacity 2(cs2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • volumetric soil moisture, site 2(sm2)
  • Soil heat capacity 1(cs1)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • Soil heat capacity 4(cs4)
  • h(h)
  • bottom vapor pressure(vp_bot)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 3(g3)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Top humidity(hum_top)
  • volumetric soil moisture, site 5(sm5)
  • Bottom humidity(hum_bot)
  • soil heat flow, site 5(g5)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 5 cm soil heat flow, site 3(shf3)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • bottom air temperature(tair_bot)
  • volumetric soil moisture, site 1(sm1)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)

sgp30ebbrE19.b1:
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • Reference Thermistor Temperature(tref)
  • 5 cm soil heat flow, site 2(shf2)
  • volumetric soil moisture, site 1(sm1)
  • volumetric soil moisture, site 2(sm2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • Soil heat capacity 5(cs5)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • volumetric soil moisture, site 4(sm4)
  • Soil heat capacity 3(cs3)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • h(h)
  • Temperature of the top humidity chamber(thum_top)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • Soil heat capacity 4(cs4)
  • Bottom humidity(hum_bot)
  • soil heat flow, site 4(g4)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • top vapor pressure(vp_top)
  • vector wind speed(res_ws)
  • average surface soil heat flow(ave_shf)
  • volumetric soil moisture, site 5(sm5)
  • latent heat flux(e)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 2(g2)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Soil heat capacity 1(cs1)
  • volumetric soil moisture, site 3(sm3)
  • soil heat flow, site 5(g5)
  • soil heat flow, site 1(g1)
  • soil heat flow, site 3(g3)
  • bottom vapor pressure(vp_bot)
  • 5 cm soil heat flow, site 1(shf1)
  • Soil heat capacity 2(cs2)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 5(shf5)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • Top humidity(hum_top)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • bottom air temperature(tair_bot)
  • net radiation(q)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 5 cm soil heat flow, site 3(shf3)
  • 5 cm soil heat flow, site 4(shf4)
  • Retrieved pressure profile(pres)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)


Back To Table of Contents

DQRID : D020717.40
Start DateStart TimeEnd DateEnd Time
11/28/2001204501/23/20021801
Subject:
SGP/EBBR/E20 - AEM Hung
DataStreams:sgp15ebbrE20.a1, sgp15ebbrE20.b1, sgp30ebbrE20.a1, sgp30ebbrE20.b1, sgp5ebbrE20.a1
Description:
The automatic exchange mechanism frequently hung in one position.  The home signals and 
automated qc in the netcdf data files indicate when this occurred.  Bowen ratio, sensible 
heat flux, and latent heat flux are incorrect for this condition.
Measurements:sgp30ebbrE20.b1:
  • Bottom humidity(hum_bot)
  • Top humidity(hum_top)
  • top vapor pressure(vp_top)
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • top air temperature(tair_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • latent heat flux(e)
  • h(h)
  • bottom air temperature(tair_bot)

sgp15ebbrE20.b1:
  • Left air temperature(tair_l)
  • Right air temperature(tair_r)

sgp5ebbrE20.a1:
  • Temperature of the top humidity chamber(thum_top)
  • top vapor pressure(vp_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • bottom vapor pressure(vp_bot)
  • top air temperature(tair_top)
  • bottom air temperature(tair_bot)
  • Top humidity(hum_top)
  • Bottom humidity(hum_bot)

sgp30ebbrE20.a1:
  • Bottom humidity(hum_bot)
  • Temperature of the top humidity chamber(thum_top)
  • bottom vapor pressure(vp_bot)
  • top air temperature(tair_top)
  • top vapor pressure(vp_top)
  • h(h)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • Top humidity(hum_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • latent heat flux(e)
  • bottom air temperature(tair_bot)

sgp15ebbrE20.a1:
  • Right air temperature(tair_r)
  • Left air temperature(tair_l)


Back To Table of Contents

DQRID : D020717.42
Start DateStart TimeEnd DateEnd Time
11/28/2001004511/28/20012031
02/06/2002213002/07/20020331
Subject:
SGP/EBBR/E20 - Battery Voltage Low
DataStreams:sgp15ebbrE20.a1, sgp15ebbrE20.b1, sgp30ebbrE20.a1, sgp30ebbrE20.b1, sgp5ebbrE20.00,
sgp5ebbrE20.a1
Description:
When the battery voltage is below 10.5 volts, most of the measurements are suspect.
Measurements:sgp30ebbrE20.b1:
  • 5 cm soil heat flow, site 2(shf2)
  • 5 cm soil heat flow, site 3(shf3)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • Reference Thermistor Temperature(tref)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • Soil heat capacity 3(cs3)
  • 5 cm soil heat flow, site 5(shf5)
  • average surface soil heat flow(ave_shf)
  • Bottom humidity(hum_bot)
  • Soil heat capacity 1(cs1)
  • volumetric soil moisture, site 2(sm2)
  • Soil heat capacity 4(cs4)
  • top vapor pressure(vp_top)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • scalar wind speed(wind_s)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • top air temperature(tair_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • net radiation(q)
  • bottom air temperature(tair_bot)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Soil heat capacity 2(cs2)
  • volumetric soil moisture, site 1(sm1)
  • soil heat flow, site 1(g1)
  • Top humidity(hum_top)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • volumetric soil moisture, site 5(sm5)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • volumetric soil moisture, site 4(sm4)
  • Temperature of the top humidity chamber(thum_top)
  • soil heat flow, site 5(g5)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • latent heat flux(e)
  • h(h)
  • wind direction (relative to true north)(wind_d)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • soil heat flow, site 4(g4)
  • Retrieved pressure profile(pres)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • soil heat flow, site 2(g2)
  • vector wind speed(res_ws)
  • 5 cm soil heat flow, site 4(shf4)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • volumetric soil moisture, site 3(sm3)
  • 5 cm soil heat flow, site 1(shf1)
  • soil heat flow, site 3(g3)
  • bottom vapor pressure(vp_bot)
  • Soil heat capacity 5(cs5)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)

sgp5ebbrE20.a1:
  • net radiation(q)
  • scalar wind speed(wind_s)
  • vector wind speed(res_ws)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • top air temperature(tair_top)
  • Top humidity(hum_top)
  • Bottom humidity(hum_bot)
  • Temperature of the top humidity chamber(thum_top)
  • top vapor pressure(vp_top)
  • wind direction (relative to true north)(wind_d)
  • bottom air temperature(tair_bot)
  • Reference Thermistor Temperature(tref)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Retrieved pressure profile(pres)
  • bottom vapor pressure(vp_bot)
  • Home signal(home)

sgp15ebbrE20.b1:
  • Soil heat flow 4(mv_hft4)
  • Soil temperature 2(rr_ts2)
  • Soil heat flow 5(mv_hft5)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil moisture 3(r_sm3)
  • Reference temperature(rr_tref)
  • Soil temperature 3(rr_ts3)
  • Soil temperature 5(rr_ts5)
  • Soil temperature 4(rr_ts4)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Left air temperature(tair_l)
  • Soil temperature 1(rr_ts1)
  • Net radiation(mv_q)
  • Home signal(mv_home)
  • Atmospheric pressure(mv_pres)
  • Soil moisture 4(r_sm4)
  • Soil moisture 5(r_sm5)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 2(mv_hft2)
  • Left relative humidity(mv_hum_l)
  • Soil heat flow 3(mv_hft3)
  • scalar wind speed(wind_s)
  • Right air temperature(tair_r)
  • Right relative humidity(mv_hum_r)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 2(r_sm2)
  • Soil moisture 1(r_sm1)

sgp5ebbrE20.00:
  • null(Raw data stream - documentation not supported)

sgp30ebbrE20.a1:
  • Bottom humidity(hum_bot)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • volumetric soil moisture, site 1(sm1)
  • 5 cm soil heat flow, site 1(shf1)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • volumetric soil moisture, site 5(sm5)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • net radiation(q)
  • 5 cm soil heat flow, site 2(shf2)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • Top humidity(hum_top)
  • 5 cm soil heat flow, site 5(shf5)
  • soil heat flow, site 5(g5)
  • Retrieved pressure profile(pres)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • volumetric soil moisture, site 3(sm3)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • volumetric soil moisture, site 2(sm2)
  • bottom vapor pressure(vp_bot)
  • soil heat flow, site 3(g3)
  • Soil heat capacity 5(cs5)
  • h(h)
  • wind direction (relative to true north)(wind_d)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • bottom air temperature(tair_bot)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • soil heat flow, site 1(g1)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • top air temperature(tair_top)
  • top vapor pressure(vp_top)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • average surface soil heat flow(ave_shf)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • soil heat flow, site 2(g2)
  • Soil heat capacity 4(cs4)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • vector wind speed(res_ws)
  • Reference Thermistor Temperature(tref)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • Soil heat capacity 3(cs3)
  • volumetric soil moisture, site 4(sm4)
  • Temperature of the top humidity chamber(thum_top)
  • Soil heat capacity 2(cs2)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • 5 cm soil heat flow, site 4(shf4)
  • latent heat flux(e)
  • Soil heat capacity 1(cs1)
  • scalar wind speed(wind_s)
  • 5 cm soil heat flow, site 3(shf3)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)

sgp15ebbrE20.a1:
  • Soil moisture 4(r_sm4)
  • Right air temperature(tair_r)
  • Soil temperature 5(rr_ts5)
  • Left air temperature(tair_l)
  • Soil moisture 3(r_sm3)
  • Right relative humidity(mv_hum_r)
  • Home signal(mv_home)
  • Soil heat flow 2(mv_hft2)
  • scalar wind speed(wind_s)
  • Left relative humidity(mv_hum_l)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 5(mv_hft5)
  • Soil moisture 1(r_sm1)
  • Reference temperature(rr_tref)
  • Soil heat flow 3(mv_hft3)
  • Soil moisture 1(rr_sm1)
  • Soil heat flow 4(mv_hft4)
  • Soil moisture 5(r_sm5)
  • Soil moisture 2(r_sm2)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Net radiation(mv_q)
  • Soil temperature 1(rr_ts1)
  • Soil moisture 2(rr_sm2)
  • Atmospheric pressure(mv_pres)
  • Soil moisture 5(rr_sm5)
  • Soil temperature 4(rr_ts4)
  • Soil temperature 3(rr_ts3)
  • Soil moisture 3(rr_sm3)
  • Soil temperature 2(rr_ts2)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 4(rr_sm4)
  • Temperature of left humidity sensor chamber(rr_thum_l)


Back To Table of Contents

DQRID : D020717.43
Start DateStart TimeEnd DateEnd Time
11/14/2001160511/28/20012031
Subject:
SGP/EBBR/E20 - Pressure Almost Constant
DataStreams:sgp30ebbrE20.a1, sgp30ebbrE20.b1, sgp5ebbrE20.a1
Description:
During the indicated period the pressure went to an almost constant 102 to 105 kPa.  This 
occurred after the replacement of another sensor.  The incorrect pressure values affect 
the sensible and latent heat flux calculations.
Measurements:sgp30ebbrE20.b1:
  • latent heat flux(e)
  • h(h)
  • Retrieved pressure profile(pres)

sgp5ebbrE20.a1:
  • Retrieved pressure profile(pres)

sgp30ebbrE20.a1:
  • latent heat flux(e)
  • h(h)
  • Retrieved pressure profile(pres)


Back To Table of Contents

DQRID : D020717.45
Start DateStart TimeEnd DateEnd Time
12/12/2001191512/12/20012031
Subject:
SGP/EBBR/E20 - Reprocess: Soil Heat Flow #3 Incorrect During Sensor Changeout
DataStreams:sgp30ebbrE20.a1, sgp30ebbrE20.b1
Description:
Soil heat flow #3 measurements were incorrect during it's replacement.

Average soil heat flow, sensible heat flux, and latent heat flux were affected; these can 
be recalculated using the remaining good sets of soil sensors as follows:
    ave_shf = (g1 + g5)/2
    e = -(q + ave_shf)/(1 + bowen)
    h = -(q + e + ave_shf)

Note: during the time period referenced here, soil heat flow #4 and #2 were also
incorrect.  This quality information is documented in D020717.46, D020717.47 and
D020726.1.
Measurements:sgp30ebbrE20.b1:
  • average surface soil heat flow(ave_shf)
  • 5 cm soil heat flow, site 3(shf3)
  • latent heat flux(e)
  • soil heat flow, site 3(g3)
  • h(h)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)

sgp30ebbrE20.a1:
  • latent heat flux(e)
  • soil heat flow, site 3(g3)
  • h(h)
  • 5 cm soil heat flow, site 3(shf3)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • average surface soil heat flow(ave_shf)


Back To Table of Contents

DQRID : D020717.46
Start DateStart TimeEnd DateEnd Time
12/04/2001000012/12/20011914
01/12/2002031501/13/20022131
01/17/2002091501/21/20020901
02/09/2002011502/12/20020731
02/20/2002164502/20/20021831
Subject:
SGP/EBBR/E20 - Reprocess: Soil Heat Flow #2 Has a Negative Offset
DataStreams:sgp30ebbrE20.a1, sgp30ebbrE20.b1
Description:
Soil heat flow #2 measurements had a negative offset.  Negative values were too small and 
negative values occurred when positive values should have.  This can occur when there is 
a poor wire connection or a J-panel over-voltage protection device has failed.

Average soil heat flow, sensible heat flux, and latent heat flux were affected; these can 
be recalculated using the remaining good sets of soil sensors as follows:
   ave_shf = (g1 + g3 + g4 + g5)/4
   e = -(q + ave_shf)/(1 + bowen)
   h = -(q + e + ave_shf)
Measurements:sgp30ebbrE20.b1:
  • 5 cm soil heat flow, site 2(shf2)
  • soil heat flow, site 2(g2)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • average surface soil heat flow(ave_shf)
  • latent heat flux(e)
  • h(h)

sgp30ebbrE20.a1:
  • latent heat flux(e)
  • 5 cm soil heat flow, site 2(shf2)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • soil heat flow, site 2(g2)
  • h(h)
  • average surface soil heat flow(ave_shf)


Back To Table of Contents

DQRID : D020717.47
Start DateStart TimeEnd DateEnd Time
12/24/2001053212/27/20011901
Subject:
SGP/EBBR/E20 - Reprocess: Soil Heat Flow #4 Incorrect
DataStreams:sgp30ebbrE20.a1, sgp30ebbrE20.b1
Description:
Soil heat flow #4 was incorrect because it's wires were loose from the connection 
terminals.

Average soil heat flow, sensible heat flux, and latent heat flux were affected; these can 
be recalculated using the remaining good sets of soil sensors as follows:
   ave_shf = (g1 + g2 + g3 + g5)/4
   e = -(q + ave_shf)/(1 + bowen)
   h = -(q + e + ave_shf)
Measurements:sgp30ebbrE20.b1:
  • 5 cm soil heat flow, site 4(shf4)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • average surface soil heat flow(ave_shf)
  • latent heat flux(e)
  • soil heat flow, site 4(g4)
  • h(h)

sgp30ebbrE20.a1:
  • latent heat flux(e)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • h(h)
  • soil heat flow, site 4(g4)
  • 5 cm soil heat flow, site 4(shf4)
  • average surface soil heat flow(ave_shf)


Back To Table of Contents

DQRID : D020717.48
Start DateStart TimeEnd DateEnd Time
11/14/2001160011/14/20011601
11/27/2001180011/27/20011801
05/15/2002170005/15/20021701
05/29/2002183005/29/20021901
Subject:
SGP/EBBR/E20 - Measurements Zero
DataStreams:sgp15ebbrE20.a1, sgp15ebbrE20.b1, sgp30ebbrE20.a1, sgp30ebbrE20.b1, sgp5ebbrE20.00,
sgp5ebbrE20.a1
Description:
Most measurements were zero during the indicated periods.  This often happens for the 
first reported data time after a CR10 datalogger restart.
Measurements:sgp30ebbrE20.b1:
  • 5 cm soil heat flow, site 2(shf2)
  • 5 cm soil heat flow, site 3(shf3)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • Reference Thermistor Temperature(tref)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • Soil heat capacity 3(cs3)
  • 5 cm soil heat flow, site 5(shf5)
  • average surface soil heat flow(ave_shf)
  • Bottom humidity(hum_bot)
  • Soil heat capacity 1(cs1)
  • volumetric soil moisture, site 2(sm2)
  • Soil heat capacity 4(cs4)
  • top vapor pressure(vp_top)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • scalar wind speed(wind_s)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • top air temperature(tair_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • net radiation(q)
  • bottom air temperature(tair_bot)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Soil heat capacity 2(cs2)
  • volumetric soil moisture, site 1(sm1)
  • soil heat flow, site 1(g1)
  • Top humidity(hum_top)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • volumetric soil moisture, site 5(sm5)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • volumetric soil moisture, site 4(sm4)
  • Temperature of the top humidity chamber(thum_top)
  • soil heat flow, site 5(g5)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • latent heat flux(e)
  • h(h)
  • wind direction (relative to true north)(wind_d)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • soil heat flow, site 4(g4)
  • Retrieved pressure profile(pres)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • soil heat flow, site 2(g2)
  • vector wind speed(res_ws)
  • 5 cm soil heat flow, site 4(shf4)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • volumetric soil moisture, site 3(sm3)
  • 5 cm soil heat flow, site 1(shf1)
  • soil heat flow, site 3(g3)
  • bottom vapor pressure(vp_bot)
  • Soil heat capacity 5(cs5)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)

sgp5ebbrE20.a1:
  • net radiation(q)
  • scalar wind speed(wind_s)
  • vector wind speed(res_ws)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • top air temperature(tair_top)
  • Top humidity(hum_top)
  • Bottom humidity(hum_bot)
  • Temperature of the top humidity chamber(thum_top)
  • top vapor pressure(vp_top)
  • wind direction (relative to true north)(wind_d)
  • bottom air temperature(tair_bot)
  • Reference Thermistor Temperature(tref)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Retrieved pressure profile(pres)
  • bottom vapor pressure(vp_bot)
  • Home signal(home)

sgp15ebbrE20.b1:
  • Soil heat flow 4(mv_hft4)
  • Soil temperature 2(rr_ts2)
  • Soil heat flow 5(mv_hft5)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil moisture 3(r_sm3)
  • Reference temperature(rr_tref)
  • Soil temperature 3(rr_ts3)
  • Soil temperature 5(rr_ts5)
  • Soil temperature 4(rr_ts4)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Left air temperature(tair_l)
  • Soil temperature 1(rr_ts1)
  • Battery(bat)
  • Net radiation(mv_q)
  • Home signal(mv_home)
  • Atmospheric pressure(mv_pres)
  • Soil moisture 4(r_sm4)
  • Soil moisture 5(r_sm5)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 2(mv_hft2)
  • Left relative humidity(mv_hum_l)
  • Soil heat flow 3(mv_hft3)
  • scalar wind speed(wind_s)
  • Right air temperature(tair_r)
  • Right relative humidity(mv_hum_r)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 2(r_sm2)
  • Soil moisture 1(r_sm1)

sgp5ebbrE20.00:
  • null(Raw data stream - documentation not supported)

sgp30ebbrE20.a1:
  • Bottom humidity(hum_bot)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • volumetric soil moisture, site 1(sm1)
  • 5 cm soil heat flow, site 1(shf1)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • volumetric soil moisture, site 5(sm5)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • net radiation(q)
  • 5 cm soil heat flow, site 2(shf2)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • Top humidity(hum_top)
  • 5 cm soil heat flow, site 5(shf5)
  • soil heat flow, site 5(g5)
  • Retrieved pressure profile(pres)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • volumetric soil moisture, site 3(sm3)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • volumetric soil moisture, site 2(sm2)
  • bottom vapor pressure(vp_bot)
  • soil heat flow, site 3(g3)
  • Soil heat capacity 5(cs5)
  • h(h)
  • wind direction (relative to true north)(wind_d)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • bottom air temperature(tair_bot)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • soil heat flow, site 1(g1)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • top air temperature(tair_top)
  • top vapor pressure(vp_top)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • average surface soil heat flow(ave_shf)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • soil heat flow, site 2(g2)
  • Soil heat capacity 4(cs4)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • vector wind speed(res_ws)
  • Reference Thermistor Temperature(tref)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • Soil heat capacity 3(cs3)
  • volumetric soil moisture, site 4(sm4)
  • Temperature of the top humidity chamber(thum_top)
  • Soil heat capacity 2(cs2)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • 5 cm soil heat flow, site 4(shf4)
  • latent heat flux(e)
  • Soil heat capacity 1(cs1)
  • scalar wind speed(wind_s)
  • 5 cm soil heat flow, site 3(shf3)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)

sgp15ebbrE20.a1:
  • Soil moisture 4(r_sm4)
  • Right air temperature(tair_r)
  • Soil temperature 5(rr_ts5)
  • Left air temperature(tair_l)
  • Soil moisture 3(r_sm3)
  • Right relative humidity(mv_hum_r)
  • Home signal(mv_home)
  • Soil heat flow 2(mv_hft2)
  • scalar wind speed(wind_s)
  • Left relative humidity(mv_hum_l)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 5(mv_hft5)
  • Soil moisture 1(r_sm1)
  • Reference temperature(rr_tref)
  • Soil heat flow 3(mv_hft3)
  • Soil moisture 1(rr_sm1)
  • Soil heat flow 4(mv_hft4)
  • Soil moisture 5(r_sm5)
  • Soil moisture 2(r_sm2)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Net radiation(mv_q)
  • Soil temperature 1(rr_ts1)
  • Soil moisture 2(rr_sm2)
  • Atmospheric pressure(mv_pres)
  • Soil moisture 5(rr_sm5)
  • Battery(bat)
  • Soil temperature 4(rr_ts4)
  • Soil temperature 3(rr_ts3)
  • Soil moisture 3(rr_sm3)
  • Soil temperature 2(rr_ts2)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 4(rr_sm4)
  • Temperature of left humidity sensor chamber(rr_thum_l)


Back To Table of Contents

DQRID : D020717.49
Start DateStart TimeEnd DateEnd Time
02/22/2002201502/22/20022101
02/23/2002174502/23/20021801
02/24/2002004502/25/20020201
03/05/2002201503/06/20020201
03/06/2002191503/07/20021901
Subject:
SGP/EBBR/E20 - Reprocess: Soil Temperature #3 Incorrect
DataStreams:sgp30ebbrE20.a1, sgp30ebbrE20.b1
Description:
Soil temperature 3 was incorrect during the indicated periods.

Average soil heat flow, sensible heat flux, and latent heat flux were affected; these can 
be recalculated using the remaining good sets of soil sensors as follows:

ave_shf = (g1 + g2 + g4 + g5)/4

e = -(q + ave_shf)/(1 + bowen)

h = -(q + e + ave_shf)
Measurements:sgp30ebbrE20.b1:
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • average surface soil heat flow(ave_shf)
  • latent heat flux(e)
  • soil heat flow, site 3(g3)
  • h(h)

sgp30ebbrE20.a1:
  • latent heat flux(e)
  • soil heat flow, site 3(g3)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • h(h)
  • average surface soil heat flow(ave_shf)
  • 0-5 cm integrated soil temperature, site 3(ts3)


Back To Table of Contents

DQRID : D020717.50
Start DateStart TimeEnd DateEnd Time
02/20/2002174502/20/20021901
Subject:
SGP/EBBR/E20 - Measurements Incorrect During Sensor Changeout
DataStreams:sgp30ebbrE20.a1, sgp30ebbrE20.b1
Description:
The replacement of soil heat flow 2 caused the loss of initial soil temperature values 
used in the calculation of CES, G, C, ave_shf, bowen, e, and h.
Measurements:sgp30ebbrE20.b1:
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • Soil heat capacity 1(cs1)
  • soil heat flow, site 2(g2)
  • soil heat flow, site 1(g1)
  • Soil heat capacity 4(cs4)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • soil heat flow, site 3(g3)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • Soil heat capacity 5(cs5)
  • soil heat flow, site 5(g5)
  • latent heat flux(e)
  • h(h)
  • Soil heat capacity 3(cs3)
  • Soil heat capacity 2(cs2)
  • average surface soil heat flow(ave_shf)
  • soil heat flow, site 4(g4)

sgp30ebbrE20.a1:
  • soil heat flow, site 1(g1)
  • Soil heat capacity 3(cs3)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • Soil heat capacity 2(cs2)
  • soil heat flow, site 3(g3)
  • h(h)
  • Soil heat capacity 5(cs5)
  • soil heat flow, site 4(g4)
  • average surface soil heat flow(ave_shf)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • soil heat flow, site 5(g5)
  • soil heat flow, site 2(g2)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • latent heat flux(e)
  • Soil heat capacity 4(cs4)
  • Soil heat capacity 1(cs1)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)


Back To Table of Contents

DQRID : D020717.51
Start DateStart TimeEnd DateEnd Time
12/27/2001184512/27/20011901
05/15/2002171505/15/20021731
Subject:
SGP/EBBR/E20 - Measurements Incorrect
DataStreams:sgp15ebbrE20.a1, sgp15ebbrE20.b1, sgp30ebbrE20.a1, sgp30ebbrE20.b1, sgp5ebbrE20.00,
sgp5ebbrE20.a1
Description:
Many measurements were incorrect because of the lack of an initial 15 minute value to use 
for gradient or soil change of energy storage calculations.
Measurements:sgp30ebbrE20.b1:
  • 5 cm soil heat flow, site 2(shf2)
  • 5 cm soil heat flow, site 3(shf3)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • Reference Thermistor Temperature(tref)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • Soil heat capacity 3(cs3)
  • 5 cm soil heat flow, site 5(shf5)
  • average surface soil heat flow(ave_shf)
  • Bottom humidity(hum_bot)
  • Soil heat capacity 1(cs1)
  • volumetric soil moisture, site 2(sm2)
  • Soil heat capacity 4(cs4)
  • top vapor pressure(vp_top)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • scalar wind speed(wind_s)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • top air temperature(tair_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • net radiation(q)
  • bottom air temperature(tair_bot)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Soil heat capacity 2(cs2)
  • volumetric soil moisture, site 1(sm1)
  • soil heat flow, site 1(g1)
  • Top humidity(hum_top)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • volumetric soil moisture, site 5(sm5)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • volumetric soil moisture, site 4(sm4)
  • Temperature of the top humidity chamber(thum_top)
  • soil heat flow, site 5(g5)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • latent heat flux(e)
  • h(h)
  • wind direction (relative to true north)(wind_d)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • soil heat flow, site 4(g4)
  • Retrieved pressure profile(pres)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • soil heat flow, site 2(g2)
  • vector wind speed(res_ws)
  • 5 cm soil heat flow, site 4(shf4)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • volumetric soil moisture, site 3(sm3)
  • 5 cm soil heat flow, site 1(shf1)
  • soil heat flow, site 3(g3)
  • bottom vapor pressure(vp_bot)
  • Soil heat capacity 5(cs5)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)

sgp5ebbrE20.a1:
  • net radiation(q)
  • scalar wind speed(wind_s)
  • vector wind speed(res_ws)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • top air temperature(tair_top)
  • Top humidity(hum_top)
  • Bottom humidity(hum_bot)
  • Temperature of the top humidity chamber(thum_top)
  • top vapor pressure(vp_top)
  • wind direction (relative to true north)(wind_d)
  • bottom air temperature(tair_bot)
  • Reference Thermistor Temperature(tref)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Retrieved pressure profile(pres)
  • bottom vapor pressure(vp_bot)
  • Home signal(home)

sgp15ebbrE20.b1:
  • Soil heat flow 4(mv_hft4)
  • Soil temperature 2(rr_ts2)
  • Soil heat flow 5(mv_hft5)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil moisture 3(r_sm3)
  • Reference temperature(rr_tref)
  • Soil temperature 3(rr_ts3)
  • Soil temperature 5(rr_ts5)
  • Soil temperature 4(rr_ts4)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Left air temperature(tair_l)
  • Soil temperature 1(rr_ts1)
  • Battery(bat)
  • Net radiation(mv_q)
  • Home signal(mv_home)
  • Atmospheric pressure(mv_pres)
  • Soil moisture 4(r_sm4)
  • Soil moisture 5(r_sm5)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 2(mv_hft2)
  • Left relative humidity(mv_hum_l)
  • Soil heat flow 3(mv_hft3)
  • scalar wind speed(wind_s)
  • Right air temperature(tair_r)
  • Right relative humidity(mv_hum_r)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 2(r_sm2)
  • Soil moisture 1(r_sm1)

sgp5ebbrE20.00:
  • null(Raw data stream - documentation not supported)

sgp30ebbrE20.a1:
  • Bottom humidity(hum_bot)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • volumetric soil moisture, site 1(sm1)
  • 5 cm soil heat flow, site 1(shf1)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • volumetric soil moisture, site 5(sm5)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • net radiation(q)
  • 5 cm soil heat flow, site 2(shf2)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • Top humidity(hum_top)
  • 5 cm soil heat flow, site 5(shf5)
  • soil heat flow, site 5(g5)
  • Retrieved pressure profile(pres)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • volumetric soil moisture, site 3(sm3)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • volumetric soil moisture, site 2(sm2)
  • bottom vapor pressure(vp_bot)
  • soil heat flow, site 3(g3)
  • Soil heat capacity 5(cs5)
  • h(h)
  • wind direction (relative to true north)(wind_d)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • bottom air temperature(tair_bot)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • soil heat flow, site 1(g1)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • top air temperature(tair_top)
  • top vapor pressure(vp_top)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • average surface soil heat flow(ave_shf)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • soil heat flow, site 2(g2)
  • Soil heat capacity 4(cs4)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • vector wind speed(res_ws)
  • Reference Thermistor Temperature(tref)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • Soil heat capacity 3(cs3)
  • volumetric soil moisture, site 4(sm4)
  • Temperature of the top humidity chamber(thum_top)
  • Soil heat capacity 2(cs2)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • 5 cm soil heat flow, site 4(shf4)
  • latent heat flux(e)
  • Soil heat capacity 1(cs1)
  • scalar wind speed(wind_s)
  • 5 cm soil heat flow, site 3(shf3)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)

sgp15ebbrE20.a1:
  • Soil moisture 4(r_sm4)
  • Right air temperature(tair_r)
  • Soil temperature 5(rr_ts5)
  • Left air temperature(tair_l)
  • Soil moisture 3(r_sm3)
  • Right relative humidity(mv_hum_r)
  • Home signal(mv_home)
  • Soil heat flow 2(mv_hft2)
  • scalar wind speed(wind_s)
  • Left relative humidity(mv_hum_l)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 5(mv_hft5)
  • Soil moisture 1(r_sm1)
  • Reference temperature(rr_tref)
  • Soil heat flow 3(mv_hft3)
  • Soil moisture 1(rr_sm1)
  • Soil heat flow 4(mv_hft4)
  • Soil moisture 5(r_sm5)
  • Soil moisture 2(r_sm2)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Net radiation(mv_q)
  • Soil temperature 1(rr_ts1)
  • Soil moisture 2(rr_sm2)
  • Atmospheric pressure(mv_pres)
  • Soil moisture 5(rr_sm5)
  • Battery(bat)
  • Soil temperature 4(rr_ts4)
  • Soil temperature 3(rr_ts3)
  • Soil moisture 3(rr_sm3)
  • Soil temperature 2(rr_ts2)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 4(rr_sm4)
  • Temperature of left humidity sensor chamber(rr_thum_l)


Back To Table of Contents

DQRID : D020726.1
Start DateStart TimeEnd DateEnd Time
12/12/2001203212/24/20010531
Subject:
SGP/EBBR/E20 - Reprocess: Soil Heat Flow #2 and #4 incorrect
DataStreams:sgp30ebbrE20.a1, sgp30ebbrE20.b1
Description:
Soil heat flow #2 measurements had a negative offset.  Negative values were too small and 
negative values occurred when positive values should have.  This can occur when there is 
a poor wire connection or a J-panel over-voltage protection device has failed.

Soil heat flow #4 was incorrect because it's wires were loose from the connection
terminals.

Average soil heat flow, sensible heat flux, and latent heat flux were affected;
these can be recalculated using the remaining good sets of soil sensors as follows: 
    ave_shf = (g1 + g3 + g5)/3
    e = -(q + ave_shf)/(1 + bowen)
    h = -(q + e + ave_shf)
Measurements:sgp30ebbrE20.b1:
  • 5 cm soil heat flow, site 2(shf2)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • soil heat flow, site 2(g2)
  • latent heat flux(e)
  • h(h)
  • 5 cm soil heat flow, site 4(shf4)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • average surface soil heat flow(ave_shf)
  • soil heat flow, site 4(g4)

sgp30ebbrE20.a1:
  • 5 cm soil heat flow, site 2(shf2)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • soil heat flow, site 2(g2)
  • 5 cm soil heat flow, site 4(shf4)
  • latent heat flux(e)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • h(h)
  • soil heat flow, site 4(g4)
  • average surface soil heat flow(ave_shf)


Back To Table of Contents

DQRID : D021113.4
Start DateStart TimeEnd DateEnd Time
12/01/1993000006/18/20032359
Subject:
SGP/EBBR - Incorrect Units for Soil Heat Capacity (cs1, cs2, cs3, cs4, cs5) in 30EBBR 
Header
DataStreams:sgp30ebbrE12.a1, sgp30ebbrE13.a1, sgp30ebbrE15.a1, sgp30ebbrE20.a1, sgp30ebbrE22.a1,
sgp30ebbrE26.a1, sgp30ebbrE4.a1, sgp30ebbrE7.a1, sgp30ebbrE8.a1, sgp30ebbrE9.a1
Description:
The units for Soil Heat Capacity (cs1, cs2, cs3, cs4,
cs5) in the EBBR 30 minute netcdf header have been 
incorrect from the beginning of processing the data
into netcdf files.  The correct units are MJ/m^3/C.
   
This does not affect the data quality.

(NOTE: The actual begin date of some of these data streams
 is as early as 7/4/93.  This DQR applies to all 30m EBBR
 data.)
Measurements:sgp30ebbrE4.a1:
  • Soil heat capacity 4(cs4)
  • Soil heat capacity 2(cs2)
  • Soil heat capacity 5(cs5)
  • Soil heat capacity 3(cs3)
  • Soil heat capacity 1(cs1)

sgp30ebbrE22.a1:
  • Soil heat capacity 1(cs1)
  • Soil heat capacity 5(cs5)
  • Soil heat capacity 4(cs4)
  • Soil heat capacity 3(cs3)
  • Soil heat capacity 2(cs2)

sgp30ebbrE12.a1:
  • Soil heat capacity 4(cs4)
  • Soil heat capacity 1(cs1)
  • Soil heat capacity 2(cs2)
  • Soil heat capacity 3(cs3)
  • Soil heat capacity 5(cs5)

sgp30ebbrE7.a1:
  • Soil heat capacity 2(cs2)
  • Soil heat capacity 5(cs5)
  • Soil heat capacity 3(cs3)
  • Soil heat capacity 1(cs1)
  • Soil heat capacity 4(cs4)

sgp30ebbrE20.a1:
  • Soil heat capacity 4(cs4)
  • Soil heat capacity 3(cs3)
  • Soil heat capacity 2(cs2)
  • Soil heat capacity 5(cs5)
  • Soil heat capacity 1(cs1)

sgp30ebbrE15.a1:
  • Soil heat capacity 4(cs4)
  • Soil heat capacity 5(cs5)
  • Soil heat capacity 1(cs1)
  • Soil heat capacity 3(cs3)
  • Soil heat capacity 2(cs2)

sgp30ebbrE9.a1:
  • Soil heat capacity 4(cs4)
  • Soil heat capacity 5(cs5)
  • Soil heat capacity 3(cs3)
  • Soil heat capacity 2(cs2)
  • Soil heat capacity 1(cs1)

sgp30ebbrE26.a1:
  • Soil heat capacity 5(cs5)
  • Soil heat capacity 3(cs3)
  • Soil heat capacity 2(cs2)
  • Soil heat capacity 1(cs1)
  • Soil heat capacity 4(cs4)

sgp30ebbrE8.a1:
  • Soil heat capacity 1(cs1)
  • Soil heat capacity 5(cs5)
  • Soil heat capacity 3(cs3)
  • Soil heat capacity 2(cs2)
  • Soil heat capacity 4(cs4)

sgp30ebbrE13.a1:
  • Soil heat capacity 3(cs3)
  • Soil heat capacity 4(cs4)
  • Soil heat capacity 5(cs5)
  • Soil heat capacity 2(cs2)
  • Soil heat capacity 1(cs1)


Back To Table of Contents

DQRID : D021113.5
Start DateStart TimeEnd DateEnd Time
10/22/1997000006/18/20032359
Subject:
SGP/EBBR - Incorrect Units for Soil Heat Capacity (cs1, cs2, cs3, cs4, cs5) in 30EBBR 
Header
DataStreams:sgp30baebbrE18.c1, sgp30baebbrE19.c1, sgp30baebbrE2.c1, sgp30ebbrE18.a1, sgp30ebbrE19.a1,
sgp30ebbrE2.a1
Description:
The units for Soil Heat Capacity (cs1, cs2, cs3, cs4,
   cs5) in the EBBR 30 minute netcdf header have been 
   incorrect from the beginning of processing the data
   into netcdf files.  The correct units are MJ/m^3/C.
   
   This does not affect the data quality.

(NOTE: The actual begin date of some of these data
streams is as early as 5/22/97.  This DQR applies to
all 30m EBBR data.)
Measurements:sgp30baebbrE2.c1:
  • Soil heat capacity 2(cs2)
  • Soil heat capacity 1(cs1)
  • Soil heat capacity 5(cs5)
  • Soil heat capacity 3(cs3)
  • Soil heat capacity 4(cs4)

sgp30baebbrE19.c1:
  • Soil heat capacity 5(cs5)
  • Soil heat capacity 3(cs3)
  • Soil heat capacity 4(cs4)
  • Soil heat capacity 2(cs2)
  • Soil heat capacity 1(cs1)

sgp30baebbrE18.c1:
  • Soil heat capacity 1(cs1)
  • Soil heat capacity 3(cs3)
  • Soil heat capacity 5(cs5)
  • Soil heat capacity 4(cs4)
  • Soil heat capacity 2(cs2)

sgp30ebbrE19.a1:
  • Soil heat capacity 3(cs3)
  • Soil heat capacity 4(cs4)
  • Soil heat capacity 5(cs5)
  • Soil heat capacity 2(cs2)
  • Soil heat capacity 1(cs1)

sgp30ebbrE2.a1:
  • Soil heat capacity 4(cs4)
  • Soil heat capacity 2(cs2)
  • Soil heat capacity 5(cs5)
  • Soil heat capacity 1(cs1)
  • Soil heat capacity 3(cs3)

sgp30ebbrE18.a1:
  • Soil heat capacity 1(cs1)
  • Soil heat capacity 3(cs3)
  • Soil heat capacity 5(cs5)
  • Soil heat capacity 2(cs2)
  • Soil heat capacity 4(cs4)


Back To Table of Contents

DQRID : D030110.24
Start DateStart TimeEnd DateEnd Time
11/14/2002160011/14/20021630
Subject:
SGP/EBBR/E19 - Data Incorrect during 6 month checks
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
Description:
Some of the measurements are incorrect during the 6 month checks.
Measurements:sgp15ebbrE19.a1:
  • Right air temperature(tair_r)
  • Left air temperature(tair_l)

sgp15ebbrE19.b1:
  • Right air temperature(tair_r)
  • Left air temperature(tair_l)

sgp5ebbrE19.a1:
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • bottom vapor pressure(vp_bot)
  • top vapor pressure(vp_top)
  • Temperature of the top humidity chamber(thum_top)
  • Bottom humidity(hum_bot)
  • bottom air temperature(tair_bot)
  • Top humidity(hum_top)
  • top air temperature(tair_top)

sgp30ebbrE19.a1:
  • top vapor pressure(vp_top)
  • latent heat flux(e)
  • h(h)
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • top air temperature(tair_top)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • bottom air temperature(tair_bot)
  • Top humidity(hum_top)
  • Bottom humidity(hum_bot)

sgp30ebbrE19.b1:
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • h(h)
  • Temperature of the top humidity chamber(thum_top)
  • top air temperature(tair_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Top humidity(hum_top)
  • Bottom humidity(hum_bot)
  • bottom air temperature(tair_bot)
  • top vapor pressure(vp_top)
  • bottom vapor pressure(vp_bot)
  • latent heat flux(e)


Back To Table of Contents

DQRID : D030110.25
Start DateStart TimeEnd DateEnd Time
11/26/2002190011/26/20021930
Subject:
SGP/EBBR/E20 - Data Incorrect during 6 month checks
DataStreams:sgp15ebbrE20.a1, sgp15ebbrE20.b1, sgp30ebbrE20.a1, sgp30ebbrE20.b1, sgp5ebbrE20.a1
Description:
Some of the measurements are incorrect during the 6 month checks.
Measurements:sgp30ebbrE20.b1:
  • Bottom humidity(hum_bot)
  • Top humidity(hum_top)
  • top vapor pressure(vp_top)
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • top air temperature(tair_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • latent heat flux(e)
  • h(h)
  • bottom air temperature(tair_bot)

sgp15ebbrE20.b1:
  • Left air temperature(tair_l)
  • Right air temperature(tair_r)

sgp5ebbrE20.a1:
  • Temperature of the top humidity chamber(thum_top)
  • top vapor pressure(vp_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • bottom vapor pressure(vp_bot)
  • top air temperature(tair_top)
  • bottom air temperature(tair_bot)
  • Top humidity(hum_top)
  • Bottom humidity(hum_bot)

sgp30ebbrE20.a1:
  • Bottom humidity(hum_bot)
  • Temperature of the top humidity chamber(thum_top)
  • bottom vapor pressure(vp_bot)
  • top air temperature(tair_top)
  • top vapor pressure(vp_top)
  • h(h)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • Top humidity(hum_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • latent heat flux(e)
  • bottom air temperature(tair_bot)

sgp15ebbrE20.a1:
  • Right air temperature(tair_r)
  • Left air temperature(tair_l)


Back To Table of Contents

DQRID : D030116.10
Start DateStart TimeEnd DateEnd Time
02/08/2002160004/18/20021700
Subject:
SGP/EBBR/E19 - REPROCESS: Soil Heat Flow #3 Incorrect Calibration
DataStreams:sgp30ebbrE19.a1, sgp30ebbrE19.b1
Description:
Soil Heat Flow #3 was incorrect during these periods.

To use the measurements, multiply Soil Heat Flow #3 by 0.726.
  
See below for a work-around.
Measurements:sgp30ebbrE19.a1:
  • latent heat flux(e)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • average surface soil heat flow(ave_shf)
  • 5 cm soil heat flow, site 3(shf3)
  • soil heat flow, site 3(g3)
  • h(h)

sgp30ebbrE19.b1:
  • soil heat flow, site 3(g3)
  • average surface soil heat flow(ave_shf)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • h(h)
  • latent heat flux(e)
  • 5 cm soil heat flow, site 3(shf3)


Back To Table of Contents

DQRID : D030116.11
Start DateStart TimeEnd DateEnd Time
08/25/2002210009/05/20021430
Subject:
SGP/EBBR/E19 - Water in Net Radiometer Dome
DataStreams:sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
Description:
The net radiation measurement is incorrect. Measurements 
calculated using net radiation are also incorrect.  

No work-around is possible for this situation.
Measurements:sgp5ebbrE19.a1:
  • net radiation(q)

sgp30ebbrE19.a1:
  • latent heat flux(e)
  • h(h)
  • net radiation(q)

sgp30ebbrE19.b1:
  • h(h)
  • latent heat flux(e)
  • net radiation(q)


Back To Table of Contents

DQRID : D030116.6
Start DateStart TimeEnd DateEnd Time
07/11/2002160007/11/20021629
Subject:
SGP/EBBR/E19 - Incorrect data after power interruption
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
Description:
Most measurements are suspect after a power outage, caused by
replacement of the solar panel regulator.
Measurements:sgp15ebbrE19.a1:
  • Right air temperature(tair_r)
  • Reference temperature(rr_tref)
  • Soil temperature 5(rr_ts5)
  • scalar wind speed(wind_s)
  • Soil temperature 4(rr_ts4)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 1(r_sm1)
  • Home signal(mv_home)
  • Soil moisture 3(r_sm3)
  • Soil moisture 2(r_sm2)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Left relative humidity(mv_hum_l)
  • Soil heat flow 4(mv_hft4)
  • Right relative humidity(mv_hum_r)
  • Soil heat flow 5(mv_hft5)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil heat flow 3(mv_hft3)
  • Soil moisture 4(r_sm4)
  • Soil temperature 1(rr_ts1)
  • Soil temperature 2(rr_ts2)
  • Soil moisture 5(r_sm5)
  • Soil temperature 3(rr_ts3)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Atmospheric pressure(mv_pres)
  • Net radiation(mv_q)

sgp15ebbrE19.b1:
  • Right relative humidity(mv_hum_r)
  • Soil heat flow 1(mv_hft1)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil temperature 1(rr_ts1)
  • Soil heat flow 3(mv_hft3)
  • Atmospheric pressure(mv_pres)
  • Soil temperature 2(rr_ts2)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Soil moisture 5(r_sm5)
  • Soil temperature 4(rr_ts4)
  • Net radiation(mv_q)
  • scalar wind speed(wind_s)
  • Reference temperature(rr_tref)
  • Soil moisture 2(r_sm2)
  • Soil temperature 5(rr_ts5)
  • Left relative humidity(mv_hum_l)
  • Soil temperature 3(rr_ts3)
  • Home signal(mv_home)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 4(r_sm4)
  • Right air temperature(tair_r)
  • Soil moisture 3(r_sm3)
  • Soil heat flow 4(mv_hft4)
  • Soil heat flow 5(mv_hft5)
  • Soil moisture 1(r_sm1)

sgp5ebbrE19.a1:
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • Bottom humidity(hum_bot)
  • Top humidity(hum_top)
  • vector wind speed(res_ws)
  • wind direction (relative to true north)(wind_d)
  • bottom air temperature(tair_bot)
  • Reference Thermistor Temperature(tref)
  • top air temperature(tair_top)
  • Retrieved pressure profile(pres)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Home signal(home)
  • top vapor pressure(vp_top)
  • net radiation(q)

sgp30ebbrE19.a1:
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • vector wind speed(res_ws)
  • 5 cm soil heat flow, site 1(shf1)
  • soil heat flow, site 2(g2)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • soil heat flow, site 1(g1)
  • 5 cm soil heat flow, site 4(shf4)
  • net radiation(q)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • Reference Thermistor Temperature(tref)
  • volumetric soil moisture, site 3(sm3)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • 5 cm soil heat flow, site 5(shf5)
  • average surface soil heat flow(ave_shf)
  • top vapor pressure(vp_top)
  • volumetric soil moisture, site 4(sm4)
  • Soil heat capacity 3(cs3)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • Soil heat capacity 5(cs5)
  • Retrieved pressure profile(pres)
  • Temperature of the top humidity chamber(thum_top)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 2(shf2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Soil heat capacity 2(cs2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • volumetric soil moisture, site 2(sm2)
  • Soil heat capacity 1(cs1)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • Soil heat capacity 4(cs4)
  • h(h)
  • bottom vapor pressure(vp_bot)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 3(g3)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Top humidity(hum_top)
  • volumetric soil moisture, site 5(sm5)
  • Bottom humidity(hum_bot)
  • soil heat flow, site 5(g5)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 5 cm soil heat flow, site 3(shf3)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • bottom air temperature(tair_bot)
  • volumetric soil moisture, site 1(sm1)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)

sgp30ebbrE19.b1:
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • Reference Thermistor Temperature(tref)
  • 5 cm soil heat flow, site 2(shf2)
  • volumetric soil moisture, site 1(sm1)
  • volumetric soil moisture, site 2(sm2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • Soil heat capacity 5(cs5)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • volumetric soil moisture, site 4(sm4)
  • Soil heat capacity 3(cs3)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • h(h)
  • Temperature of the top humidity chamber(thum_top)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • Soil heat capacity 4(cs4)
  • Bottom humidity(hum_bot)
  • soil heat flow, site 4(g4)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • top vapor pressure(vp_top)
  • vector wind speed(res_ws)
  • average surface soil heat flow(ave_shf)
  • volumetric soil moisture, site 5(sm5)
  • latent heat flux(e)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 2(g2)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Soil heat capacity 1(cs1)
  • volumetric soil moisture, site 3(sm3)
  • soil heat flow, site 5(g5)
  • soil heat flow, site 1(g1)
  • soil heat flow, site 3(g3)
  • bottom vapor pressure(vp_bot)
  • 5 cm soil heat flow, site 1(shf1)
  • Soil heat capacity 2(cs2)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 5(shf5)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • Top humidity(hum_top)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • bottom air temperature(tair_bot)
  • net radiation(q)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 5 cm soil heat flow, site 3(shf3)
  • 5 cm soil heat flow, site 4(shf4)
  • Retrieved pressure profile(pres)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)


Back To Table of Contents

DQRID : D030116.7
Start DateStart TimeEnd DateEnd Time
07/11/2002150007/11/20021529
Subject:
SGP/EBBR/E19 - Most Measurements Zero following Restart
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
Description:
Most of the measurements were zero, which normally happens after a 
datalogger restart.
Measurements:sgp15ebbrE19.a1:
  • Right air temperature(tair_r)
  • Reference temperature(rr_tref)
  • Soil temperature 5(rr_ts5)
  • scalar wind speed(wind_s)
  • Soil temperature 4(rr_ts4)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 1(r_sm1)
  • Home signal(mv_home)
  • Soil moisture 3(r_sm3)
  • Soil moisture 2(r_sm2)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Left relative humidity(mv_hum_l)
  • Battery(bat)
  • Soil heat flow 4(mv_hft4)
  • Right relative humidity(mv_hum_r)
  • Soil heat flow 5(mv_hft5)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil heat flow 3(mv_hft3)
  • Soil moisture 4(r_sm4)
  • Soil temperature 1(rr_ts1)
  • Soil temperature 2(rr_ts2)
  • Soil moisture 5(r_sm5)
  • Soil temperature 3(rr_ts3)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Atmospheric pressure(mv_pres)
  • Net radiation(mv_q)

sgp15ebbrE19.b1:
  • Right relative humidity(mv_hum_r)
  • Soil heat flow 1(mv_hft1)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil temperature 1(rr_ts1)
  • Soil heat flow 3(mv_hft3)
  • Atmospheric pressure(mv_pres)
  • Soil temperature 2(rr_ts2)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Soil moisture 5(r_sm5)
  • Soil temperature 4(rr_ts4)
  • Net radiation(mv_q)
  • scalar wind speed(wind_s)
  • Reference temperature(rr_tref)
  • Soil moisture 2(r_sm2)
  • Soil temperature 5(rr_ts5)
  • Battery(bat)
  • Left relative humidity(mv_hum_l)
  • Soil temperature 3(rr_ts3)
  • Home signal(mv_home)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 4(r_sm4)
  • Right air temperature(tair_r)
  • Soil moisture 3(r_sm3)
  • Soil heat flow 4(mv_hft4)
  • Soil heat flow 5(mv_hft5)
  • Soil moisture 1(r_sm1)

sgp5ebbrE19.a1:
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • Bottom humidity(hum_bot)
  • Top humidity(hum_top)
  • vector wind speed(res_ws)
  • wind direction (relative to true north)(wind_d)
  • bottom air temperature(tair_bot)
  • Reference Thermistor Temperature(tref)
  • top air temperature(tair_top)
  • scalar wind speed(wind_s)
  • Retrieved pressure profile(pres)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Home signal(home)
  • top vapor pressure(vp_top)
  • net radiation(q)

sgp30ebbrE19.a1:
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • vector wind speed(res_ws)
  • 5 cm soil heat flow, site 1(shf1)
  • soil heat flow, site 2(g2)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • soil heat flow, site 1(g1)
  • 5 cm soil heat flow, site 4(shf4)
  • net radiation(q)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • Reference Thermistor Temperature(tref)
  • volumetric soil moisture, site 3(sm3)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • 5 cm soil heat flow, site 5(shf5)
  • average surface soil heat flow(ave_shf)
  • top vapor pressure(vp_top)
  • volumetric soil moisture, site 4(sm4)
  • Soil heat capacity 3(cs3)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • Soil heat capacity 5(cs5)
  • Retrieved pressure profile(pres)
  • Temperature of the top humidity chamber(thum_top)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 2(shf2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Soil heat capacity 2(cs2)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • volumetric soil moisture, site 2(sm2)
  • Soil heat capacity 1(cs1)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • Soil heat capacity 4(cs4)
  • h(h)
  • bottom vapor pressure(vp_bot)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 3(g3)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Top humidity(hum_top)
  • volumetric soil moisture, site 5(sm5)
  • Bottom humidity(hum_bot)
  • soil heat flow, site 5(g5)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 5 cm soil heat flow, site 3(shf3)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • bottom air temperature(tair_bot)
  • volumetric soil moisture, site 1(sm1)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)

sgp30ebbrE19.b1:
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • Reference Thermistor Temperature(tref)
  • 5 cm soil heat flow, site 2(shf2)
  • volumetric soil moisture, site 1(sm1)
  • volumetric soil moisture, site 2(sm2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • Soil heat capacity 5(cs5)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • volumetric soil moisture, site 4(sm4)
  • Soil heat capacity 3(cs3)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • h(h)
  • Temperature of the top humidity chamber(thum_top)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • Soil heat capacity 4(cs4)
  • Bottom humidity(hum_bot)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • soil heat flow, site 4(g4)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • top vapor pressure(vp_top)
  • vector wind speed(res_ws)
  • average surface soil heat flow(ave_shf)
  • volumetric soil moisture, site 5(sm5)
  • latent heat flux(e)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 2(g2)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Soil heat capacity 1(cs1)
  • volumetric soil moisture, site 3(sm3)
  • soil heat flow, site 5(g5)
  • soil heat flow, site 1(g1)
  • soil heat flow, site 3(g3)
  • bottom vapor pressure(vp_bot)
  • 5 cm soil heat flow, site 1(shf1)
  • Soil heat capacity 2(cs2)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 5(shf5)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • Top humidity(hum_top)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • bottom air temperature(tair_bot)
  • net radiation(q)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 5 cm soil heat flow, site 3(shf3)
  • 5 cm soil heat flow, site 4(shf4)
  • Retrieved pressure profile(pres)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)


Back To Table of Contents

DQRID : D030116.8
Start DateStart TimeEnd DateEnd Time
07/24/2002143007/24/20021500
Subject:
SGP/EBBR/E19 - REPROCESS: Soil Temperature #3 Incorrect
DataStreams:sgp30ebbrE19.a1, sgp30ebbrE19.b1
Description:
Soil Temperature #3 was incorrect while electronic connections
were being checked. The indicated measurements 
that are calculated using Soil Temperature #3 are 
also incorrect.

See a work-around below.
Measurements:sgp30ebbrE19.a1:
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • average surface soil heat flow(ave_shf)
  • soil heat flow, site 3(g3)
  • h(h)

sgp30ebbrE19.b1:
  • soil heat flow, site 3(g3)
  • average surface soil heat flow(ave_shf)
  • h(h)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 0-5 cm integrated soil temperature, site 3(ts3)


Back To Table of Contents

DQRID : D030116.9
Start DateStart TimeEnd DateEnd Time
07/25/2002093007/25/20021300
07/26/2002090007/26/20020900
07/26/2002113007/26/20021130
07/27/2002063007/27/20020830
07/28/2002030007/28/20020630
07/28/2002093007/28/20021230
07/29/2002080007/29/20020830
08/08/2002160008/08/20021600
08/09/2002023008/09/20020300
08/09/2002073008/09/20020730
08/09/2002093008/09/20021100
08/09/2002140008/09/20021600
08/10/2002013008/10/20020230
08/10/2002120008/10/20021330
08/13/2002060008/13/20020700
08/13/2002203008/13/20022030
08/13/2002213008/14/20020330
08/14/2002130008/14/20021330
08/21/2002163008/22/20021430
Subject:
SGP/EBBR/E19 - REPROCESS: Soil Heat Flow #1 Incorrect
DataStreams:sgp30ebbrE19.a1, sgp30ebbrE19.b1
Description:
Soil Heat Flow #1 was incorrect during these periods.  
See below for a work-around.
Measurements:sgp30ebbrE19.a1:
  • latent heat flux(e)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • average surface soil heat flow(ave_shf)
  • 5 cm soil heat flow, site 1(shf1)
  • h(h)
  • soil heat flow, site 1(g1)

sgp30ebbrE19.b1:
  • average surface soil heat flow(ave_shf)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • 5 cm soil heat flow, site 1(shf1)
  • h(h)
  • latent heat flux(e)
  • soil heat flow, site 1(g1)


Back To Table of Contents

DQRID : D030117.1
Start DateStart TimeEnd DateEnd Time
10/17/2002143010/31/20021600
Subject:
SGP/EBBR/E19 - Pressure Low
DataStreams:sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
Description:
The barometric pressure sensor read about 10 mb low during this
period because a lead from the sensor was loose or disconnected.

The error does not affect any other measurements.
Measurements:sgp5ebbrE19.a1:
  • Retrieved pressure profile(pres)

sgp30ebbrE19.a1:
  • Retrieved pressure profile(pres)

sgp30ebbrE19.b1:
  • Retrieved pressure profile(pres)


Back To Table of Contents

DQRID : D030117.10
Start DateStart TimeEnd DateEnd Time
11/13/2002183011/13/20021859
12/11/2002190012/11/20021929
Subject:
SGP/EBBR/E20 - Data Zero following logger restarts
DataStreams:sgp15ebbrE20.a1, sgp15ebbrE20.b1, sgp30ebbrE20.a1, sgp30ebbrE20.b1, sgp5ebbrE20.a1
Description:
Most of the measurements were zero, which normally happens after a 
datalogger restart.
Measurements:sgp30ebbrE20.b1:
  • 5 cm soil heat flow, site 2(shf2)
  • 5 cm soil heat flow, site 3(shf3)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • Reference Thermistor Temperature(tref)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • Soil heat capacity 3(cs3)
  • 5 cm soil heat flow, site 5(shf5)
  • average surface soil heat flow(ave_shf)
  • Bottom humidity(hum_bot)
  • Soil heat capacity 1(cs1)
  • volumetric soil moisture, site 2(sm2)
  • Soil heat capacity 4(cs4)
  • top vapor pressure(vp_top)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • scalar wind speed(wind_s)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • top air temperature(tair_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • net radiation(q)
  • bottom air temperature(tair_bot)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Soil heat capacity 2(cs2)
  • volumetric soil moisture, site 1(sm1)
  • soil heat flow, site 1(g1)
  • Top humidity(hum_top)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • volumetric soil moisture, site 5(sm5)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • volumetric soil moisture, site 4(sm4)
  • Temperature of the top humidity chamber(thum_top)
  • soil heat flow, site 5(g5)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • latent heat flux(e)
  • h(h)
  • wind direction (relative to true north)(wind_d)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • soil heat flow, site 4(g4)
  • Retrieved pressure profile(pres)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • soil heat flow, site 2(g2)
  • vector wind speed(res_ws)
  • 5 cm soil heat flow, site 4(shf4)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • volumetric soil moisture, site 3(sm3)
  • 5 cm soil heat flow, site 1(shf1)
  • soil heat flow, site 3(g3)
  • bottom vapor pressure(vp_bot)
  • Soil heat capacity 5(cs5)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)

sgp5ebbrE20.a1:
  • net radiation(q)
  • scalar wind speed(wind_s)
  • vector wind speed(res_ws)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • top air temperature(tair_top)
  • Top humidity(hum_top)
  • Bottom humidity(hum_bot)
  • Temperature of the top humidity chamber(thum_top)
  • top vapor pressure(vp_top)
  • wind direction (relative to true north)(wind_d)
  • bottom air temperature(tair_bot)
  • Reference Thermistor Temperature(tref)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Retrieved pressure profile(pres)
  • bottom vapor pressure(vp_bot)
  • Home signal(home)

sgp15ebbrE20.b1:
  • Soil heat flow 4(mv_hft4)
  • Soil temperature 2(rr_ts2)
  • Soil heat flow 5(mv_hft5)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil moisture 3(r_sm3)
  • Reference temperature(rr_tref)
  • Soil temperature 3(rr_ts3)
  • Soil temperature 5(rr_ts5)
  • Soil temperature 4(rr_ts4)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Left air temperature(tair_l)
  • Soil temperature 1(rr_ts1)
  • Net radiation(mv_q)
  • Home signal(mv_home)
  • Atmospheric pressure(mv_pres)
  • Soil moisture 4(r_sm4)
  • Soil moisture 5(r_sm5)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 2(mv_hft2)
  • Left relative humidity(mv_hum_l)
  • Soil heat flow 3(mv_hft3)
  • scalar wind speed(wind_s)
  • Right air temperature(tair_r)
  • Right relative humidity(mv_hum_r)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 2(r_sm2)
  • Soil moisture 1(r_sm1)

sgp30ebbrE20.a1:
  • Bottom humidity(hum_bot)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • volumetric soil moisture, site 1(sm1)
  • 5 cm soil heat flow, site 1(shf1)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • volumetric soil moisture, site 5(sm5)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • net radiation(q)
  • 5 cm soil heat flow, site 2(shf2)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • Top humidity(hum_top)
  • 5 cm soil heat flow, site 5(shf5)
  • soil heat flow, site 5(g5)
  • Retrieved pressure profile(pres)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • volumetric soil moisture, site 3(sm3)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • volumetric soil moisture, site 2(sm2)
  • bottom vapor pressure(vp_bot)
  • soil heat flow, site 3(g3)
  • Soil heat capacity 5(cs5)
  • h(h)
  • wind direction (relative to true north)(wind_d)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • bottom air temperature(tair_bot)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • soil heat flow, site 1(g1)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • top air temperature(tair_top)
  • top vapor pressure(vp_top)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • average surface soil heat flow(ave_shf)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • soil heat flow, site 2(g2)
  • Soil heat capacity 4(cs4)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • vector wind speed(res_ws)
  • Reference Thermistor Temperature(tref)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • Soil heat capacity 3(cs3)
  • volumetric soil moisture, site 4(sm4)
  • Temperature of the top humidity chamber(thum_top)
  • Soil heat capacity 2(cs2)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • 5 cm soil heat flow, site 4(shf4)
  • latent heat flux(e)
  • Soil heat capacity 1(cs1)
  • scalar wind speed(wind_s)
  • 5 cm soil heat flow, site 3(shf3)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)

sgp15ebbrE20.a1:
  • Soil moisture 4(r_sm4)
  • Right air temperature(tair_r)
  • Soil temperature 5(rr_ts5)
  • Left air temperature(tair_l)
  • Soil moisture 3(r_sm3)
  • Right relative humidity(mv_hum_r)
  • Home signal(mv_home)
  • Soil heat flow 2(mv_hft2)
  • scalar wind speed(wind_s)
  • Left relative humidity(mv_hum_l)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 5(mv_hft5)
  • Soil moisture 1(r_sm1)
  • Reference temperature(rr_tref)
  • Soil heat flow 3(mv_hft3)
  • Soil moisture 1(rr_sm1)
  • Soil heat flow 4(mv_hft4)
  • Soil moisture 5(r_sm5)
  • Soil moisture 2(r_sm2)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Net radiation(mv_q)
  • Soil temperature 1(rr_ts1)
  • Soil moisture 2(rr_sm2)
  • Atmospheric pressure(mv_pres)
  • Soil moisture 5(rr_sm5)
  • Soil temperature 4(rr_ts4)
  • Soil temperature 3(rr_ts3)
  • Soil moisture 3(rr_sm3)
  • Soil temperature 2(rr_ts2)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 4(rr_sm4)
  • Temperature of left humidity sensor chamber(rr_thum_l)


Back To Table of Contents

DQRID : D030117.11
Start DateStart TimeEnd DateEnd Time
08/09/2002170509/18/20021700
Subject:
SGP/EBBR/E20 - REPROCESS: Soil Temperature #2 Incorrect
DataStreams:sgp30ebbrE20.a1, sgp30ebbrE20.b1
Description:
Soil Temperature #2 was incorrect, apparently
from am loose wire after replacement of the EBBR unit.
The indicated measurements that are calculated using 
soil temperature are also incorrect.

See the work-around below.
Measurements:sgp30ebbrE20.b1:
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • soil heat flow, site 2(g2)
  • average surface soil heat flow(ave_shf)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • h(h)

sgp30ebbrE20.a1:
  • latent heat flux(e)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • soil heat flow, site 2(g2)
  • h(h)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • average surface soil heat flow(ave_shf)


Back To Table of Contents

DQRID : D030117.12
Start DateStart TimeEnd DateEnd Time
11/13/2002190011/13/20021929
12/11/2002200012/11/20022029
Subject:
SGP/EBBR/E20 - Change of Energy Storage Incorrect
DataStreams:sgp30ebbrE20.a1, sgp30ebbrE20.b1
Description:
Because soil temperature measurements were not available 
from the beginning of the half hour, the change in soil 
energy storage could not be calculated and thus is 
incorrect.

There is no fix to this problem. Average Soil Heat Flow, 
Sensible Heat Flux (h), and Latent Heat Flux (e) can not 
be calculated using the available, remaining measurements.
Measurements:sgp30ebbrE20.b1:
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • soil heat flow, site 2(g2)
  • soil heat flow, site 1(g1)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • soil heat flow, site 3(g3)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • soil heat flow, site 5(g5)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • latent heat flux(e)
  • average surface soil heat flow(ave_shf)
  • soil heat flow, site 4(g4)

sgp30ebbrE20.a1:
  • soil heat flow, site 1(g1)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • soil heat flow, site 3(g3)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • h(h)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • average surface soil heat flow(ave_shf)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • soil heat flow, site 5(g5)
  • soil heat flow, site 2(g2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • latent heat flux(e)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)


Back To Table of Contents

DQRID : D030117.13
Start DateStart TimeEnd DateEnd Time
10/23/2002021510/30/20022000
Subject:
SGP/EBBR/E20 - Battery voltage low
DataStreams:sgp15ebbrE20.a1, sgp15ebbrE20.b1, sgp30ebbrE20.a1, sgp30ebbrE20.b1, sgp5ebbrE20.a1
Description:
When Battery Voltage drops below 10.5 volts, most 
measurements are suspect and some are often incorrect.

The user will need to inspect the data carefully to 
determine the useability of the data.
Measurements:sgp30ebbrE20.b1:
  • 5 cm soil heat flow, site 2(shf2)
  • 5 cm soil heat flow, site 3(shf3)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • Reference Thermistor Temperature(tref)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • Soil heat capacity 3(cs3)
  • 5 cm soil heat flow, site 5(shf5)
  • average surface soil heat flow(ave_shf)
  • Bottom humidity(hum_bot)
  • Soil heat capacity 1(cs1)
  • volumetric soil moisture, site 2(sm2)
  • Soil heat capacity 4(cs4)
  • top vapor pressure(vp_top)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • scalar wind speed(wind_s)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • top air temperature(tair_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • net radiation(q)
  • bottom air temperature(tair_bot)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Soil heat capacity 2(cs2)
  • volumetric soil moisture, site 1(sm1)
  • soil heat flow, site 1(g1)
  • Top humidity(hum_top)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • volumetric soil moisture, site 5(sm5)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • volumetric soil moisture, site 4(sm4)
  • Temperature of the top humidity chamber(thum_top)
  • soil heat flow, site 5(g5)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • latent heat flux(e)
  • h(h)
  • wind direction (relative to true north)(wind_d)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • soil heat flow, site 4(g4)
  • Retrieved pressure profile(pres)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • soil heat flow, site 2(g2)
  • vector wind speed(res_ws)
  • 5 cm soil heat flow, site 4(shf4)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • volumetric soil moisture, site 3(sm3)
  • 5 cm soil heat flow, site 1(shf1)
  • soil heat flow, site 3(g3)
  • bottom vapor pressure(vp_bot)
  • Soil heat capacity 5(cs5)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)

sgp5ebbrE20.a1:
  • net radiation(q)
  • scalar wind speed(wind_s)
  • vector wind speed(res_ws)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • top air temperature(tair_top)
  • Top humidity(hum_top)
  • Bottom humidity(hum_bot)
  • Temperature of the top humidity chamber(thum_top)
  • top vapor pressure(vp_top)
  • wind direction (relative to true north)(wind_d)
  • bottom air temperature(tair_bot)
  • Reference Thermistor Temperature(tref)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Retrieved pressure profile(pres)
  • bottom vapor pressure(vp_bot)

sgp15ebbrE20.b1:
  • Soil heat flow 4(mv_hft4)
  • Soil temperature 2(rr_ts2)
  • Soil heat flow 5(mv_hft5)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil moisture 3(r_sm3)
  • Reference temperature(rr_tref)
  • Soil temperature 3(rr_ts3)
  • Soil temperature 5(rr_ts5)
  • Soil temperature 4(rr_ts4)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Left air temperature(tair_l)
  • Soil temperature 1(rr_ts1)
  • Net radiation(mv_q)
  • Home signal(mv_home)
  • Atmospheric pressure(mv_pres)
  • Soil moisture 4(r_sm4)
  • Soil moisture 5(r_sm5)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 2(mv_hft2)
  • Left relative humidity(mv_hum_l)
  • Soil heat flow 3(mv_hft3)
  • scalar wind speed(wind_s)
  • Right air temperature(tair_r)
  • Right relative humidity(mv_hum_r)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 2(r_sm2)
  • Soil moisture 1(r_sm1)

sgp30ebbrE20.a1:
  • Bottom humidity(hum_bot)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • volumetric soil moisture, site 1(sm1)
  • 5 cm soil heat flow, site 1(shf1)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • volumetric soil moisture, site 5(sm5)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • net radiation(q)
  • 5 cm soil heat flow, site 2(shf2)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • Top humidity(hum_top)
  • 5 cm soil heat flow, site 5(shf5)
  • soil heat flow, site 5(g5)
  • Retrieved pressure profile(pres)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • volumetric soil moisture, site 3(sm3)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • volumetric soil moisture, site 2(sm2)
  • bottom vapor pressure(vp_bot)
  • soil heat flow, site 3(g3)
  • Soil heat capacity 5(cs5)
  • h(h)
  • wind direction (relative to true north)(wind_d)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • bottom air temperature(tair_bot)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • soil heat flow, site 1(g1)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • top air temperature(tair_top)
  • top vapor pressure(vp_top)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • average surface soil heat flow(ave_shf)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • soil heat flow, site 2(g2)
  • Soil heat capacity 4(cs4)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • vector wind speed(res_ws)
  • Reference Thermistor Temperature(tref)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • Soil heat capacity 3(cs3)
  • volumetric soil moisture, site 4(sm4)
  • Temperature of the top humidity chamber(thum_top)
  • Soil heat capacity 2(cs2)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • 5 cm soil heat flow, site 4(shf4)
  • latent heat flux(e)
  • Soil heat capacity 1(cs1)
  • scalar wind speed(wind_s)
  • 5 cm soil heat flow, site 3(shf3)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)

sgp15ebbrE20.a1:
  • Soil moisture 4(r_sm4)
  • Right air temperature(tair_r)
  • Soil temperature 5(rr_ts5)
  • Left air temperature(tair_l)
  • Soil moisture 3(r_sm3)
  • Right relative humidity(mv_hum_r)
  • Home signal(mv_home)
  • Soil heat flow 2(mv_hft2)
  • scalar wind speed(wind_s)
  • Left relative humidity(mv_hum_l)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 5(mv_hft5)
  • Soil moisture 1(r_sm1)
  • Reference temperature(rr_tref)
  • Soil heat flow 3(mv_hft3)
  • Soil moisture 1(rr_sm1)
  • Soil heat flow 4(mv_hft4)
  • Soil moisture 5(r_sm5)
  • Soil moisture 2(r_sm2)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Net radiation(mv_q)
  • Soil temperature 1(rr_ts1)
  • Soil moisture 2(rr_sm2)
  • Atmospheric pressure(mv_pres)
  • Soil moisture 5(rr_sm5)
  • Soil temperature 4(rr_ts4)
  • Soil temperature 3(rr_ts3)
  • Soil moisture 3(rr_sm3)
  • Soil temperature 2(rr_ts2)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 4(rr_sm4)
  • Temperature of left humidity sensor chamber(rr_thum_l)


Back To Table of Contents

DQRID : D030117.14
Start DateStart TimeEnd DateEnd Time
12/04/2002120012/05/20021630
Subject:
SGP/EBBR/E20 - Wind speed and direction frozen
DataStreams:sgp15ebbrE20.a1, sgp15ebbrE20.b1, sgp30ebbrE20.a1, sgp30ebbrE20.b1, sgp5ebbrE2.a1
Description:
Wind Speed and Dierction were frozen in place by an ice storm.
Measurements:sgp30ebbrE20.b1:
  • vector wind speed(res_ws)
  • wind direction (relative to true north)(wind_d)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • scalar wind speed(wind_s)

sgp5ebbrE2.a1:
  • wind direction (relative to true north)(wind_d)
  • scalar wind speed(wind_s)
  • vector wind speed(res_ws)

sgp15ebbrE20.b1:
  • scalar wind speed(wind_s)

sgp30ebbrE20.a1:
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • wind direction (relative to true north)(wind_d)
  • vector wind speed(res_ws)
  • scalar wind speed(wind_s)

sgp15ebbrE20.a1:
  • scalar wind speed(wind_s)


Back To Table of Contents

DQRID : D030117.18
Start DateStart TimeEnd DateEnd Time
08/07/2002163008/07/20021730
Subject:
SGP/EBBR/E20 - AEM Hung between Home Positions
DataStreams:sgp15ebbrE20.a1, sgp15ebbrE20.b1, sgp30ebbrE20.a1, sgp30ebbrE20.b1, sgp5ebbrE20.a1
Description:
When the AEM (Automatic Exchange Mechanism) is hung 
inbetween home positions, the gradients of temperature and 
relative humidity, and thus sensible and latent heat fluxes, 
are incorrect.

There is no work-around to this problem.
Measurements:sgp30ebbrE20.b1:
  • Bottom humidity(hum_bot)
  • Top humidity(hum_top)
  • top vapor pressure(vp_top)
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • top air temperature(tair_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • latent heat flux(e)
  • h(h)
  • bottom air temperature(tair_bot)

sgp15ebbrE20.b1:
  • Left air temperature(tair_l)
  • Right air temperature(tair_r)

sgp5ebbrE20.a1:
  • Temperature of the top humidity chamber(thum_top)
  • top vapor pressure(vp_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • bottom vapor pressure(vp_bot)
  • top air temperature(tair_top)
  • bottom air temperature(tair_bot)
  • Top humidity(hum_top)
  • Bottom humidity(hum_bot)

sgp30ebbrE20.a1:
  • Bottom humidity(hum_bot)
  • Temperature of the top humidity chamber(thum_top)
  • bottom vapor pressure(vp_bot)
  • top air temperature(tair_top)
  • top vapor pressure(vp_top)
  • h(h)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • Top humidity(hum_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • latent heat flux(e)
  • bottom air temperature(tair_bot)

sgp15ebbrE20.a1:
  • Right air temperature(tair_r)
  • Left air temperature(tair_l)


Back To Table of Contents

DQRID : D030117.19
Start DateStart TimeEnd DateEnd Time
12/30/2002203001/08/20031730
Subject:
SGP/EBBR/E20 - REPROCESS: Heat Flow #3 and #5 Incorrect
DataStreams:sgp30ebbrE20.a1, sgp30ebbrE20.b1
Description:
Soil Heat Flow #3 was incorrect because part of the lead
wire was cut; apparent vandalism.

Soil Heat Flow #5 had been incorrect because of other
causes, but was replaced at the same time as #3.
Measurements:sgp30ebbrE20.b1:
  • soil heat flow, site 5(g5)
  • latent heat flux(e)
  • 5 cm soil heat flow, site 3(shf3)
  • h(h)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 5 cm soil heat flow, site 5(shf5)
  • average surface soil heat flow(ave_shf)
  • soil heat flow, site 3(g3)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)

sgp30ebbrE20.a1:
  • 5 cm soil heat flow, site 5(shf5)
  • soil heat flow, site 5(g5)
  • latent heat flux(e)
  • soil heat flow, site 3(g3)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • h(h)
  • 5 cm soil heat flow, site 3(shf3)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • average surface soil heat flow(ave_shf)


Back To Table of Contents

DQRID : D030117.2
Start DateStart TimeEnd DateEnd Time
08/09/2002170508/09/20021730
Subject:
SGP/EBBR/E20 - Data incorrect during instrument maintenance
DataStreams:sgp15ebbrE20.a1, sgp15ebbrE20.b1, sgp30ebbrE20.a1, sgp30ebbrE20.b1, sgp5ebbrE20.a1
Description:
Most of the data was incorrect just after the EBBR unit was replaced.
Measurements:sgp30ebbrE20.b1:
  • 5 cm soil heat flow, site 2(shf2)
  • 5 cm soil heat flow, site 3(shf3)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • Reference Thermistor Temperature(tref)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • Soil heat capacity 3(cs3)
  • 5 cm soil heat flow, site 5(shf5)
  • average surface soil heat flow(ave_shf)
  • Bottom humidity(hum_bot)
  • Soil heat capacity 1(cs1)
  • volumetric soil moisture, site 2(sm2)
  • Soil heat capacity 4(cs4)
  • top vapor pressure(vp_top)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • scalar wind speed(wind_s)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • top air temperature(tair_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • net radiation(q)
  • bottom air temperature(tair_bot)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Soil heat capacity 2(cs2)
  • volumetric soil moisture, site 1(sm1)
  • soil heat flow, site 1(g1)
  • Top humidity(hum_top)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • volumetric soil moisture, site 5(sm5)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • volumetric soil moisture, site 4(sm4)
  • Temperature of the top humidity chamber(thum_top)
  • soil heat flow, site 5(g5)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • latent heat flux(e)
  • h(h)
  • wind direction (relative to true north)(wind_d)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • soil heat flow, site 4(g4)
  • Retrieved pressure profile(pres)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • soil heat flow, site 2(g2)
  • vector wind speed(res_ws)
  • 5 cm soil heat flow, site 4(shf4)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • volumetric soil moisture, site 3(sm3)
  • 5 cm soil heat flow, site 1(shf1)
  • soil heat flow, site 3(g3)
  • bottom vapor pressure(vp_bot)
  • Soil heat capacity 5(cs5)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)

sgp5ebbrE20.a1:
  • net radiation(q)
  • scalar wind speed(wind_s)
  • vector wind speed(res_ws)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • top air temperature(tair_top)
  • Top humidity(hum_top)
  • Bottom humidity(hum_bot)
  • Temperature of the top humidity chamber(thum_top)
  • top vapor pressure(vp_top)
  • wind direction (relative to true north)(wind_d)
  • bottom air temperature(tair_bot)
  • Reference Thermistor Temperature(tref)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Retrieved pressure profile(pres)
  • bottom vapor pressure(vp_bot)
  • Home signal(home)

sgp15ebbrE20.b1:
  • Soil heat flow 4(mv_hft4)
  • Soil temperature 2(rr_ts2)
  • Soil heat flow 5(mv_hft5)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil moisture 3(r_sm3)
  • Reference temperature(rr_tref)
  • Soil temperature 3(rr_ts3)
  • Soil temperature 5(rr_ts5)
  • Soil temperature 4(rr_ts4)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Left air temperature(tair_l)
  • Soil temperature 1(rr_ts1)
  • Battery(bat)
  • Net radiation(mv_q)
  • Home signal(mv_home)
  • Atmospheric pressure(mv_pres)
  • Soil moisture 4(r_sm4)
  • Soil moisture 5(r_sm5)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 2(mv_hft2)
  • Left relative humidity(mv_hum_l)
  • Soil heat flow 3(mv_hft3)
  • scalar wind speed(wind_s)
  • Right air temperature(tair_r)
  • Right relative humidity(mv_hum_r)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 2(r_sm2)
  • Soil moisture 1(r_sm1)

sgp30ebbrE20.a1:
  • Bottom humidity(hum_bot)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • volumetric soil moisture, site 1(sm1)
  • 5 cm soil heat flow, site 1(shf1)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • volumetric soil moisture, site 5(sm5)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • net radiation(q)
  • 5 cm soil heat flow, site 2(shf2)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • Top humidity(hum_top)
  • 5 cm soil heat flow, site 5(shf5)
  • soil heat flow, site 5(g5)
  • Retrieved pressure profile(pres)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • volumetric soil moisture, site 3(sm3)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • volumetric soil moisture, site 2(sm2)
  • bottom vapor pressure(vp_bot)
  • soil heat flow, site 3(g3)
  • Soil heat capacity 5(cs5)
  • h(h)
  • wind direction (relative to true north)(wind_d)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • bottom air temperature(tair_bot)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • soil heat flow, site 1(g1)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • top air temperature(tair_top)
  • top vapor pressure(vp_top)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • average surface soil heat flow(ave_shf)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • soil heat flow, site 2(g2)
  • Soil heat capacity 4(cs4)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • vector wind speed(res_ws)
  • Reference Thermistor Temperature(tref)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • Soil heat capacity 3(cs3)
  • volumetric soil moisture, site 4(sm4)
  • Temperature of the top humidity chamber(thum_top)
  • Soil heat capacity 2(cs2)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • 5 cm soil heat flow, site 4(shf4)
  • latent heat flux(e)
  • Soil heat capacity 1(cs1)
  • scalar wind speed(wind_s)
  • 5 cm soil heat flow, site 3(shf3)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)

sgp15ebbrE20.a1:
  • Soil moisture 4(r_sm4)
  • Right air temperature(tair_r)
  • Soil temperature 5(rr_ts5)
  • Left air temperature(tair_l)
  • Soil moisture 3(r_sm3)
  • Right relative humidity(mv_hum_r)
  • Home signal(mv_home)
  • Soil heat flow 2(mv_hft2)
  • scalar wind speed(wind_s)
  • Left relative humidity(mv_hum_l)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 5(mv_hft5)
  • Soil moisture 1(r_sm1)
  • Reference temperature(rr_tref)
  • Soil heat flow 3(mv_hft3)
  • Soil moisture 1(rr_sm1)
  • Soil heat flow 4(mv_hft4)
  • Soil moisture 5(r_sm5)
  • Soil moisture 2(r_sm2)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Net radiation(mv_q)
  • Soil temperature 1(rr_ts1)
  • Soil moisture 2(rr_sm2)
  • Atmospheric pressure(mv_pres)
  • Soil moisture 5(rr_sm5)
  • Battery(bat)
  • Soil temperature 4(rr_ts4)
  • Soil temperature 3(rr_ts3)
  • Soil moisture 3(rr_sm3)
  • Soil temperature 2(rr_ts2)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 4(rr_sm4)
  • Temperature of left humidity sensor chamber(rr_thum_l)


Back To Table of Contents

DQRID : D030117.20
Start DateStart TimeEnd DateEnd Time
12/03/2002120012/04/20020400
12/06/2002200012/07/20020230
12/08/2002023012/09/20020500
12/09/2002093012/20/20021830
12/20/2002210012/21/20021630
12/21/2002173012/22/20020100
12/22/2002013012/30/20022000
Subject:
SGP/EBBR/E20 - REPROCESS: Soil Heat Flow #5 Incorrect
DataStreams:sgp30ebbrE20.a1, sgp30ebbrE20.b1
Description:
Soil Heat Flow #5 was incorrect during these periods. 
 
See below for a work-around.
Measurements:sgp30ebbrE20.b1:
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • soil heat flow, site 5(g5)
  • 5 cm soil heat flow, site 5(shf5)
  • average surface soil heat flow(ave_shf)
  • latent heat flux(e)
  • h(h)

sgp30ebbrE20.a1:
  • latent heat flux(e)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 5 cm soil heat flow, site 5(shf5)
  • soil heat flow, site 5(g5)
  • h(h)
  • average surface soil heat flow(ave_shf)


Back To Table of Contents

DQRID : D030117.22
Start DateStart TimeEnd DateEnd Time
11/18/2002190011/18/20022200
11/19/2002170011/19/20022200
11/20/2002143011/20/20022200
11/21/2002200011/21/20022030
11/22/2002163011/22/20022000
11/23/2002143011/23/20022230
11/24/2002170011/24/20022200
11/27/2002203011/27/20022300
11/28/2002170011/28/20022300
11/29/2002133011/29/20022130
12/01/2002193012/01/20022030
12/02/2002143012/02/20022230
Subject:
SGP/EBBR/E20 - REPROCESS: Soil Temperature #1 Incorrect
DataStreams:sgp30ebbrE20.a1, sgp30ebbrE20.b1
Description:
Soil Temperature #1 was intermittently incorrect.
The cause is unknown. The indicated measurements 
that are calculated using soil temperature are 
also incorrect.
Measurements:sgp30ebbrE20.b1:
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • soil heat flow, site 1(g1)
  • average surface soil heat flow(ave_shf)
  • latent heat flux(e)
  • h(h)

sgp30ebbrE20.a1:
  • latent heat flux(e)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • soil heat flow, site 1(g1)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • h(h)
  • average surface soil heat flow(ave_shf)


Back To Table of Contents

DQRID : D030117.25
Start DateStart TimeEnd DateEnd Time
12/30/2002203001/08/20031730
Subject:
SGP/EBBR/E20 - REPROCESS: Soil Temperature #3 Incorrect
DataStreams:sgp30ebbrE20.a1, sgp30ebbrE20.b1
Description:
Soil Temperature #3 was incorrect.
The cause is apparently vandalism; the probe leads were cut. 
The indicated measurements that are calculated using soil 
temperature are also incorrect.

DQR D030117.19 already provides the work-around for
calculating correct sensible (h) and latent heat (e) fluxes,
as soil heat flow #3 and soil heat flow #5 were also incorrect
during this period.  That work-around is repeated below.
Measurements:sgp30ebbrE20.b1:
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • average surface soil heat flow(ave_shf)
  • latent heat flux(e)
  • soil heat flow, site 3(g3)
  • h(h)

sgp30ebbrE20.a1:
  • latent heat flux(e)
  • soil heat flow, site 3(g3)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • h(h)
  • average surface soil heat flow(ave_shf)
  • 0-5 cm integrated soil temperature, site 3(ts3)


Back To Table of Contents

DQRID : D030117.3
Start DateStart TimeEnd DateEnd Time
10/20/2002130010/20/20021500
Subject:
SGP/EBBR/E20 - Resistance-based Measurements Incorrect
DataStreams:sgp15ebbrE20.a1, sgp15ebbrE20.b1, sgp30ebbrE20.a1, sgp30ebbrE20.b1, sgp5ebbrE20.a1
Description:
When the Reference Temperature in the EBBR 
enclosure is not reported correctly, it
affects all resistance-based measurements
that are adjusted with the temperature of 
the reference junction.

No work-around is possible for this situation.
Measurements:sgp30ebbrE20.b1:
  • soil heat flow, site 1(g1)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • Reference Thermistor Temperature(tref)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • soil heat flow, site 5(g5)
  • latent heat flux(e)
  • h(h)
  • average surface soil heat flow(ave_shf)
  • soil heat flow, site 4(g4)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • soil heat flow, site 2(g2)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • soil heat flow, site 3(g3)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • top air temperature(tair_top)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • bottom air temperature(tair_bot)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)

sgp15ebbrE20.b1:
  • Left air temperature(tair_l)
  • Right air temperature(tair_r)

sgp5ebbrE20.a1:
  • top air temperature(tair_top)
  • bottom air temperature(tair_bot)

sgp30ebbrE20.a1:
  • soil heat flow, site 1(g1)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • top air temperature(tair_top)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • soil heat flow, site 5(g5)
  • soil heat flow, site 2(g2)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • Reference Thermistor Temperature(tref)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • soil heat flow, site 3(g3)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • latent heat flux(e)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • bottom air temperature(tair_bot)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)

sgp15ebbrE20.a1:
  • Right air temperature(tair_r)
  • Left air temperature(tair_l)


Back To Table of Contents

DQRID : D030117.4
Start DateStart TimeEnd DateEnd Time
07/14/2002143001/09/20031600
Subject:
SGP/EBBR/E19 - Soil Temperatures #2 and #3 Intermittently Bad
DataStreams:sgp30ebbrE19.a1, sgp30ebbrE19.b1
Description:
Soil Temperatures #2 and #3 were intermittently incorrect.
#3 appeared to be okay after the J-panel was replaced. 
#2 continues to be flaky, suggesting that the sensor
itself has a problem.  A DQPR has been submitted to
replace #2.  The indicated measurements 
that are calculated using soil temperature are 
also suspect.  However, the intermittent problems normally
occurred during night times and were fairly small in magnitude,
thereby having little effect on the sensible and latent heat fluxes.
Measurements:sgp30ebbrE19.a1:
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • average surface soil heat flow(ave_shf)
  • soil heat flow, site 3(g3)
  • soil heat flow, site 2(g2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • latent heat flux(e)
  • h(h)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)

sgp30ebbrE19.b1:
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • h(h)
  • soil heat flow, site 2(g2)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • soil heat flow, site 3(g3)
  • average surface soil heat flow(ave_shf)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • latent heat flux(e)


Back To Table of Contents

DQRID : D030117.6
Start DateStart TimeEnd DateEnd Time
01/09/2003170001/09/20031729
Subject:
SGP/EBBR/E19 - Measurements zero following logger restart
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
Description:
Most of the measurements were zero, which normally happens after a 
datalogger restart.  The restart on 01/09/03 occurred after J-panel
replacements.
Measurements:sgp15ebbrE19.a1:
  • Right air temperature(tair_r)
  • Reference temperature(rr_tref)
  • Soil temperature 5(rr_ts5)
  • scalar wind speed(wind_s)
  • Soil temperature 4(rr_ts4)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 1(r_sm1)
  • Home signal(mv_home)
  • Soil moisture 3(r_sm3)
  • Soil moisture 2(r_sm2)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Left relative humidity(mv_hum_l)
  • Soil heat flow 4(mv_hft4)
  • Right relative humidity(mv_hum_r)
  • Soil heat flow 5(mv_hft5)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil heat flow 3(mv_hft3)
  • Soil moisture 4(r_sm4)
  • Soil temperature 1(rr_ts1)
  • Soil temperature 2(rr_ts2)
  • Soil moisture 5(r_sm5)
  • Soil temperature 3(rr_ts3)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Atmospheric pressure(mv_pres)
  • Net radiation(mv_q)

sgp15ebbrE19.b1:
  • Right relative humidity(mv_hum_r)
  • Soil heat flow 1(mv_hft1)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil temperature 1(rr_ts1)
  • Soil heat flow 3(mv_hft3)
  • Atmospheric pressure(mv_pres)
  • Soil temperature 2(rr_ts2)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Soil moisture 5(r_sm5)
  • Soil temperature 4(rr_ts4)
  • Net radiation(mv_q)
  • scalar wind speed(wind_s)
  • Reference temperature(rr_tref)
  • Soil moisture 2(r_sm2)
  • Soil temperature 5(rr_ts5)
  • Left relative humidity(mv_hum_l)
  • Soil temperature 3(rr_ts3)
  • Home signal(mv_home)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 4(r_sm4)
  • Right air temperature(tair_r)
  • Soil moisture 3(r_sm3)
  • Soil heat flow 4(mv_hft4)
  • Soil heat flow 5(mv_hft5)
  • Soil moisture 1(r_sm1)

sgp5ebbrE19.a1:
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • Bottom humidity(hum_bot)
  • Top humidity(hum_top)
  • vector wind speed(res_ws)
  • wind direction (relative to true north)(wind_d)
  • bottom air temperature(tair_bot)
  • Reference Thermistor Temperature(tref)
  • top air temperature(tair_top)
  • scalar wind speed(wind_s)
  • Retrieved pressure profile(pres)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Home signal(home)
  • top vapor pressure(vp_top)
  • net radiation(q)

sgp30ebbrE19.a1:
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • vector wind speed(res_ws)
  • 5 cm soil heat flow, site 1(shf1)
  • soil heat flow, site 2(g2)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • soil heat flow, site 1(g1)
  • 5 cm soil heat flow, site 4(shf4)
  • net radiation(q)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • Reference Thermistor Temperature(tref)
  • volumetric soil moisture, site 3(sm3)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • 5 cm soil heat flow, site 5(shf5)
  • average surface soil heat flow(ave_shf)
  • top vapor pressure(vp_top)
  • volumetric soil moisture, site 4(sm4)
  • Soil heat capacity 3(cs3)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • Soil heat capacity 5(cs5)
  • Retrieved pressure profile(pres)
  • Temperature of the top humidity chamber(thum_top)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 2(shf2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Soil heat capacity 2(cs2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • volumetric soil moisture, site 2(sm2)
  • Soil heat capacity 1(cs1)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • Soil heat capacity 4(cs4)
  • h(h)
  • bottom vapor pressure(vp_bot)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 3(g3)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Top humidity(hum_top)
  • volumetric soil moisture, site 5(sm5)
  • Bottom humidity(hum_bot)
  • soil heat flow, site 5(g5)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 5 cm soil heat flow, site 3(shf3)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • bottom air temperature(tair_bot)
  • volumetric soil moisture, site 1(sm1)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)

sgp30ebbrE19.b1:
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • Reference Thermistor Temperature(tref)
  • 5 cm soil heat flow, site 2(shf2)
  • volumetric soil moisture, site 1(sm1)
  • volumetric soil moisture, site 2(sm2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • Soil heat capacity 5(cs5)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • volumetric soil moisture, site 4(sm4)
  • Soil heat capacity 3(cs3)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • h(h)
  • Temperature of the top humidity chamber(thum_top)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • Soil heat capacity 4(cs4)
  • Bottom humidity(hum_bot)
  • soil heat flow, site 4(g4)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • top vapor pressure(vp_top)
  • vector wind speed(res_ws)
  • average surface soil heat flow(ave_shf)
  • volumetric soil moisture, site 5(sm5)
  • latent heat flux(e)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 2(g2)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Soil heat capacity 1(cs1)
  • volumetric soil moisture, site 3(sm3)
  • soil heat flow, site 5(g5)
  • soil heat flow, site 1(g1)
  • soil heat flow, site 3(g3)
  • bottom vapor pressure(vp_bot)
  • 5 cm soil heat flow, site 1(shf1)
  • Soil heat capacity 2(cs2)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 5(shf5)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • Top humidity(hum_top)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • bottom air temperature(tair_bot)
  • net radiation(q)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 5 cm soil heat flow, site 3(shf3)
  • 5 cm soil heat flow, site 4(shf4)
  • Retrieved pressure profile(pres)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)


Back To Table of Contents

DQRID : D030117.7
Start DateStart TimeEnd DateEnd Time
01/09/2003173001/09/20031759
Subject:
SGP/EBBR/E19 - Change of Energy Storage Incorrect
DataStreams:sgp30ebbrE19.a1, sgp30ebbrE19.b1
Description:
Because soil temperature measurements were not available 
from the beginning of the half hour, the change in soil 
energy storage could not be calculated and thus is 
incorrect.

There is no fix to this problem. Average Soil Heat Flow, 
Sensible Heat Flux (h), and Latent Heat Flux (e) can not 
be calculated using the available, remaining measurements.
Measurements:sgp30ebbrE19.a1:
  • average surface soil heat flow(ave_shf)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • latent heat flux(e)
  • soil heat flow, site 5(g5)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • h(h)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • soil heat flow, site 3(g3)
  • soil heat flow, site 2(g2)
  • soil heat flow, site 1(g1)
  • soil heat flow, site 4(g4)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)

sgp30ebbrE19.b1:
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • h(h)
  • soil heat flow, site 2(g2)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • soil heat flow, site 4(g4)
  • soil heat flow, site 5(g5)
  • soil heat flow, site 1(g1)
  • soil heat flow, site 3(g3)
  • average surface soil heat flow(ave_shf)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)


Back To Table of Contents

DQRID : D030117.8
Start DateStart TimeEnd DateEnd Time
08/09/2002173012/11/20021930
Subject:
SGP/EBBR/E20 - REPROCESS: Net Rad Coefficients Incorrect
DataStreams:sgp30ebbrE20.a1, sgp30ebbrE20.b1, sgp5ebbrE20.a1
Description:
The Net Radiometer coefficients were incorrect in the CR10
program when the replacement EBBR unit was installed on
08/09/02.

The data can be easily corrected; see the Suggestion area below.
Measurements:sgp30ebbrE20.b1:
  • latent heat flux(e)
  • net radiation(q)
  • h(h)

sgp5ebbrE20.a1:
  • net radiation(q)

sgp30ebbrE20.a1:
  • net radiation(q)
  • latent heat flux(e)
  • h(h)


Back To Table of Contents

DQRID : D030121.1
Start DateStart TimeEnd DateEnd Time
12/20/2002183012/20/20022100
12/21/2002163012/21/20021730
12/22/2002010012/22/20020130
Subject:
SGP/EBBR/E20 - REPROCESS: Soil Temperature #1 and Soil Heat Flow #5 Incorrect
DataStreams:sgp30ebbrE20.a1, sgp30ebbrE20.b1
Description:
Soil Temperature #1 and Soil Heat Flow #5 were incorrect
during these periods. The indicated calculated measurements
are also incorrect.
Measurements:sgp30ebbrE20.b1:
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • soil heat flow, site 5(g5)
  • soil heat flow, site 1(g1)
  • latent heat flux(e)
  • h(h)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 5 cm soil heat flow, site 5(shf5)
  • average surface soil heat flow(ave_shf)

sgp30ebbrE20.a1:
  • soil heat flow, site 1(g1)
  • 5 cm soil heat flow, site 5(shf5)
  • soil heat flow, site 5(g5)
  • latent heat flux(e)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • h(h)
  • average surface soil heat flow(ave_shf)


Back To Table of Contents

DQRID : D030711.10
Start DateStart TimeEnd DateEnd Time
02/01/2003100002/01/20031000
Subject:
SGP/EBBR/E20 - RHtop Zero
DataStreams:sgp30ebbrE20.a1, sgp30ebbrE20.b1
Description:
The RHtop value in the 30 minute file is zero.  5 and 15 minute files show RHtop 
correctly. This behavior has been seen infrequently and can probably be attributed to a CR10 
communication error.

No other data, including the fluxes, are affected.
Measurements:sgp30ebbrE20.b1:
  • Top humidity(hum_top)

sgp30ebbrE20.a1:
  • Top humidity(hum_top)


Back To Table of Contents

DQRID : D030711.11
Start DateStart TimeEnd DateEnd Time
01/09/2003163001/09/20031700
Subject:
SGP/EBBR/E19 - Soil Temperature #2 Incorrect
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1
Description:
Soil Temperature #2 was incorrect during J-panel replacements.
Measurements:sgp15ebbrE19.a1:
  • Soil temperature 2(rr_ts2)

sgp15ebbrE19.b1:
  • Soil temperature 2(rr_ts2)

sgp30ebbrE19.a1:
  • average surface soil heat flow(ave_shf)
  • soil heat flow, site 2(g2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • latent heat flux(e)
  • h(h)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)

sgp30ebbrE19.b1:
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • h(h)
  • soil heat flow, site 2(g2)
  • average surface soil heat flow(ave_shf)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • latent heat flux(e)


Back To Table of Contents

DQRID : D030711.12
Start DateStart TimeEnd DateEnd Time
01/09/2003170001/09/20031700
02/06/2003173002/06/20031730
04/03/2003183004/03/20031830
Subject:
SGP/EBBR/E19 - All Data Zeros
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
Description:
All data (5, 15, and 30 minute) was zero after replacement of sensors and restart of the 
CR10.

This is a common occurrence.
Measurements:sgp15ebbrE19.a1:
  • Right air temperature(tair_r)
  • Reference temperature(rr_tref)
  • Soil temperature 5(rr_ts5)
  • scalar wind speed(wind_s)
  • Soil temperature 4(rr_ts4)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 1(r_sm1)
  • Home signal(mv_home)
  • Soil moisture 3(r_sm3)
  • Soil moisture 2(r_sm2)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Left relative humidity(mv_hum_l)
  • Battery(bat)
  • Soil heat flow 4(mv_hft4)
  • Right relative humidity(mv_hum_r)
  • Soil heat flow 5(mv_hft5)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil heat flow 3(mv_hft3)
  • Soil moisture 4(r_sm4)
  • Soil temperature 1(rr_ts1)
  • Soil temperature 2(rr_ts2)
  • Soil moisture 5(r_sm5)
  • Soil temperature 3(rr_ts3)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Atmospheric pressure(mv_pres)
  • Net radiation(mv_q)

sgp15ebbrE19.b1:
  • Right relative humidity(mv_hum_r)
  • Soil heat flow 1(mv_hft1)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil temperature 1(rr_ts1)
  • Soil heat flow 3(mv_hft3)
  • Atmospheric pressure(mv_pres)
  • Soil temperature 2(rr_ts2)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Soil moisture 5(r_sm5)
  • Soil temperature 4(rr_ts4)
  • Net radiation(mv_q)
  • scalar wind speed(wind_s)
  • Reference temperature(rr_tref)
  • Soil moisture 2(r_sm2)
  • Soil temperature 5(rr_ts5)
  • Battery(bat)
  • Left relative humidity(mv_hum_l)
  • Soil temperature 3(rr_ts3)
  • Home signal(mv_home)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 4(r_sm4)
  • Right air temperature(tair_r)
  • Soil moisture 3(r_sm3)
  • Soil heat flow 4(mv_hft4)
  • Soil heat flow 5(mv_hft5)
  • Soil moisture 1(r_sm1)

sgp5ebbrE19.a1:
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • Bottom humidity(hum_bot)
  • Top humidity(hum_top)
  • vector wind speed(res_ws)
  • wind direction (relative to true north)(wind_d)
  • bottom air temperature(tair_bot)
  • Reference Thermistor Temperature(tref)
  • top air temperature(tair_top)
  • scalar wind speed(wind_s)
  • Retrieved pressure profile(pres)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Home signal(home)
  • top vapor pressure(vp_top)
  • net radiation(q)

sgp30ebbrE19.a1:
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • vector wind speed(res_ws)
  • 5 cm soil heat flow, site 1(shf1)
  • soil heat flow, site 2(g2)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • soil heat flow, site 1(g1)
  • 5 cm soil heat flow, site 4(shf4)
  • net radiation(q)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • Reference Thermistor Temperature(tref)
  • volumetric soil moisture, site 3(sm3)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • 5 cm soil heat flow, site 5(shf5)
  • average surface soil heat flow(ave_shf)
  • top vapor pressure(vp_top)
  • volumetric soil moisture, site 4(sm4)
  • Soil heat capacity 3(cs3)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • Soil heat capacity 5(cs5)
  • Retrieved pressure profile(pres)
  • Temperature of the top humidity chamber(thum_top)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 2(shf2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Soil heat capacity 2(cs2)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • volumetric soil moisture, site 2(sm2)
  • Soil heat capacity 1(cs1)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • Soil heat capacity 4(cs4)
  • h(h)
  • bottom vapor pressure(vp_bot)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 3(g3)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Top humidity(hum_top)
  • volumetric soil moisture, site 5(sm5)
  • Bottom humidity(hum_bot)
  • soil heat flow, site 5(g5)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 5 cm soil heat flow, site 3(shf3)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • bottom air temperature(tair_bot)
  • volumetric soil moisture, site 1(sm1)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)

sgp30ebbrE19.b1:
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • Reference Thermistor Temperature(tref)
  • 5 cm soil heat flow, site 2(shf2)
  • volumetric soil moisture, site 1(sm1)
  • volumetric soil moisture, site 2(sm2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • Soil heat capacity 5(cs5)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • volumetric soil moisture, site 4(sm4)
  • Soil heat capacity 3(cs3)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • h(h)
  • Temperature of the top humidity chamber(thum_top)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • Soil heat capacity 4(cs4)
  • Bottom humidity(hum_bot)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • soil heat flow, site 4(g4)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • top vapor pressure(vp_top)
  • vector wind speed(res_ws)
  • average surface soil heat flow(ave_shf)
  • volumetric soil moisture, site 5(sm5)
  • latent heat flux(e)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 2(g2)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Soil heat capacity 1(cs1)
  • volumetric soil moisture, site 3(sm3)
  • soil heat flow, site 5(g5)
  • soil heat flow, site 1(g1)
  • soil heat flow, site 3(g3)
  • bottom vapor pressure(vp_bot)
  • 5 cm soil heat flow, site 1(shf1)
  • Soil heat capacity 2(cs2)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 5(shf5)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • Top humidity(hum_top)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • bottom air temperature(tair_bot)
  • net radiation(q)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 5 cm soil heat flow, site 3(shf3)
  • 5 cm soil heat flow, site 4(shf4)
  • Retrieved pressure profile(pres)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)


Back To Table of Contents

DQRID : D030711.13
Start DateStart TimeEnd DateEnd Time
01/09/2003173001/09/20031730
Subject:
SGP/EBBR/E19 - Several Soil Measurements Incorrect
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1
Description:
During replacement of a J-panel, several soil measurements were incorrect.
Measurements:sgp15ebbrE19.a1:
  • Soil temperature 3(rr_ts3)

sgp15ebbrE19.b1:
  • Soil temperature 3(rr_ts3)

sgp30ebbrE19.a1:
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • average surface soil heat flow(ave_shf)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • latent heat flux(e)
  • soil heat flow, site 5(g5)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • h(h)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • soil heat flow, site 3(g3)
  • soil heat flow, site 2(g2)
  • soil heat flow, site 1(g1)
  • soil heat flow, site 4(g4)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)

sgp30ebbrE19.b1:
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • h(h)
  • soil heat flow, site 2(g2)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • soil heat flow, site 4(g4)
  • soil heat flow, site 5(g5)
  • soil heat flow, site 1(g1)
  • soil heat flow, site 3(g3)
  • average surface soil heat flow(ave_shf)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)


Back To Table of Contents

DQRID : D030711.14
Start DateStart TimeEnd DateEnd Time
01/26/2003170001/26/20031700
01/26/2003180001/26/20031800
Subject:
SGP/EBBR/E19 -Several Soil Measurements Were Incorrect
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1
Description:
Several soil measurements were incorrect during replacement of soil temperature probe #2.
Measurements:sgp15ebbrE19.a1:
  • Soil temperature 2(rr_ts2)

sgp15ebbrE19.b1:
  • Soil temperature 2(rr_ts2)

sgp30ebbrE19.a1:
  • average surface soil heat flow(ave_shf)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • latent heat flux(e)
  • soil heat flow, site 5(g5)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • h(h)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • soil heat flow, site 3(g3)
  • soil heat flow, site 2(g2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • soil heat flow, site 1(g1)
  • soil heat flow, site 4(g4)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)

sgp30ebbrE19.b1:
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • h(h)
  • soil heat flow, site 2(g2)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • soil heat flow, site 4(g4)
  • soil heat flow, site 5(g5)
  • soil heat flow, site 1(g1)
  • soil heat flow, site 3(g3)
  • average surface soil heat flow(ave_shf)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)


Back To Table of Contents

DQRID : D030711.16
Start DateStart TimeEnd DateEnd Time
02/28/2003133003/02/20030530
03/02/2003173003/03/20030030
Subject:
SGP/EBBR/E19 - AEM Hung in home30 position
DataStreams:sgp30ebbrE19.a1, sgp30ebbrE19.b1
Description:
The automatic exchange mechanism (AEM) hung in the home30 position, resulting in sensible 
and latent heat fluxes being incorrect.
Measurements:sgp30ebbrE19.a1:
  • Temperature of the top humidity chamber(thum_top)
  • top air temperature(tair_top)
  • top vapor pressure(vp_top)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • latent heat flux(e)
  • bottom air temperature(tair_bot)
  • Top humidity(hum_top)
  • h(h)
  • bottom vapor pressure(vp_bot)
  • Bottom humidity(hum_bot)

sgp30ebbrE19.b1:
  • top vapor pressure(vp_top)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • h(h)
  • Temperature of the top humidity chamber(thum_top)
  • top air temperature(tair_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Top humidity(hum_top)
  • bottom vapor pressure(vp_bot)
  • latent heat flux(e)
  • Bottom humidity(hum_bot)
  • bottom air temperature(tair_bot)


Back To Table of Contents

DQRID : D030711.17
Start DateStart TimeEnd DateEnd Time
01/09/2003173004/03/20031900
Subject:
SGP/EBBR/E19 - Soil Heat Flow #3 Incorrect
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1
Description:
Soil Heat Flow #3 was a flat -24 W/m2 for most of this period.
Measurements:sgp15ebbrE19.a1:
  • Soil heat flow 3(mv_hft3)

sgp15ebbrE19.b1:
  • Soil heat flow 3(mv_hft3)

sgp30ebbrE19.a1:
  • average surface soil heat flow(ave_shf)
  • soil heat flow, site 3(g3)
  • latent heat flux(e)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • 5 cm soil heat flow, site 3(shf3)
  • h(h)

sgp30ebbrE19.b1:
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • h(h)
  • 5 cm soil heat flow, site 3(shf3)
  • soil heat flow, site 3(g3)
  • average surface soil heat flow(ave_shf)
  • latent heat flux(e)


Back To Table of Contents

DQRID : D030711.19
Start DateStart TimeEnd DateEnd Time
04/03/2003190004/03/20031900
Subject:
SGP/EBBR/E19 - Soil Heat Flow#1 Incorrect
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1
Description:
Soil Heat Flow #1 was incorrect during replacement of some other sensors.
Measurements:sgp15ebbrE19.a1:
  • Soil heat flow 1(mv_hft1)

sgp15ebbrE19.b1:
  • Soil heat flow 1(mv_hft1)

sgp30ebbrE19.a1:
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • average surface soil heat flow(ave_shf)
  • 5 cm soil heat flow, site 1(shf1)
  • soil heat flow, site 1(g1)
  • latent heat flux(e)
  • h(h)

sgp30ebbrE19.b1:
  • h(h)
  • soil heat flow, site 1(g1)
  • average surface soil heat flow(ave_shf)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • 5 cm soil heat flow, site 1(shf1)
  • latent heat flux(e)


Back To Table of Contents

DQRID : D030711.20
Start DateStart TimeEnd DateEnd Time
05/15/2003160005/15/20031700
Subject:
SGP/EBBR/E19 - Many Measurements Incorrect
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
Description:
Many measurements were incorrect during replacement of a multiplexer.
Measurements:sgp15ebbrE19.a1:
  • Right air temperature(tair_r)
  • Reference temperature(rr_tref)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil temperature 1(rr_ts1)
  • Soil temperature 5(rr_ts5)
  • Soil temperature 2(rr_ts2)
  • Soil temperature 4(rr_ts4)
  • Soil temperature 3(rr_ts3)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Left air temperature(tair_l)

sgp15ebbrE19.b1:
  • Soil temperature 3(rr_ts3)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil temperature 1(rr_ts1)
  • Soil temperature 2(rr_ts2)
  • Left air temperature(tair_l)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Soil temperature 4(rr_ts4)
  • Right air temperature(tair_r)
  • Reference temperature(rr_tref)
  • Soil temperature 5(rr_ts5)

sgp5ebbrE19.a1:
  • Temperature of the top humidity chamber(thum_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • bottom air temperature(tair_bot)
  • Reference Thermistor Temperature(tref)
  • top air temperature(tair_top)

sgp30ebbrE19.a1:
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • h(h)
  • soil heat flow, site 3(g3)
  • soil heat flow, site 2(g2)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • soil heat flow, site 1(g1)
  • average surface soil heat flow(ave_shf)
  • soil heat flow, site 5(g5)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • Temperature of the top humidity chamber(thum_top)
  • top air temperature(tair_top)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • bottom air temperature(tair_bot)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)

sgp30ebbrE19.b1:
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • soil heat flow, site 2(g2)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • soil heat flow, site 5(g5)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • soil heat flow, site 1(g1)
  • soil heat flow, site 3(g3)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • h(h)
  • Temperature of the top humidity chamber(thum_top)
  • top air temperature(tair_top)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • bottom air temperature(tair_bot)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • average surface soil heat flow(ave_shf)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)


Back To Table of Contents

DQRID : D030711.8
Start DateStart TimeEnd DateEnd Time
01/01/2003000001/08/20031730
Subject:
SGP/EBBR/E20 - Soil Heat Flow#5 Incorrect
DataStreams:sgp15ebbrE20.a1, sgp15ebbrE20.b1, sgp30ebbrE20.a1, sgp30ebbrE20.b1
Description:
Soil Heat Flow #5 was producing incorrect values.  Average soil heat flow, sensible heat 
flux, and latent heat flux were affected.
Measurements:sgp30ebbrE20.b1:
  • soil heat flow, site 5(g5)
  • latent heat flux(e)
  • h(h)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 5 cm soil heat flow, site 5(shf5)
  • average surface soil heat flow(ave_shf)

sgp15ebbrE20.b1:
  • Soil heat flow 5(mv_hft5)

sgp30ebbrE20.a1:
  • 5 cm soil heat flow, site 5(shf5)
  • soil heat flow, site 5(g5)
  • latent heat flux(e)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • h(h)
  • average surface soil heat flow(ave_shf)

sgp15ebbrE20.a1:
  • Soil heat flow 5(mv_hft5)


Back To Table of Contents

DQRID : D030711.9
Start DateStart TimeEnd DateEnd Time
01/08/2003210001/08/20032100
Subject:
SGP/EBBR/E20 - Soil Storage Incorrect After Startup
DataStreams:sgp30ebbrE20.a1, sgp30ebbrE20.b1
Description:
For the half hour data after CR10 datalogger startup, stored soil temperature values do 
not exist from the beginning of the half hour.  Therefore soil heat storage can not be 
calculated correctly.

This affects numerous measurements that are calculated using soil heat storage.
Measurements:sgp30ebbrE20.b1:
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • soil heat flow, site 2(g2)
  • soil heat flow, site 1(g1)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • soil heat flow, site 3(g3)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • soil heat flow, site 5(g5)
  • latent heat flux(e)
  • h(h)
  • average surface soil heat flow(ave_shf)
  • soil heat flow, site 4(g4)

sgp30ebbrE20.a1:
  • soil heat flow, site 1(g1)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • soil heat flow, site 3(g3)
  • h(h)
  • soil heat flow, site 4(g4)
  • average surface soil heat flow(ave_shf)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • soil heat flow, site 5(g5)
  • soil heat flow, site 2(g2)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)


Back To Table of Contents

DQRID : D030712.20
Start DateStart TimeEnd DateEnd Time
05/01/2003144005/01/20031515
Subject:
SGP/EBBR/E19 -  6 Month Checks
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
Description:
Many measurements were incorrect while the AEM was positioned with both aspirators at the 
same height for comparison checks.
Measurements:sgp15ebbrE19.a1:
  • Right relative humidity(mv_hum_r)
  • Right air temperature(tair_r)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Left air temperature(tair_l)
  • Left relative humidity(mv_hum_l)

sgp15ebbrE19.b1:
  • Right air temperature(tair_r)
  • Right relative humidity(mv_hum_r)
  • Left relative humidity(mv_hum_l)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Left air temperature(tair_l)
  • Temperature of left humidity sensor chamber(rr_thum_l)

sgp5ebbrE19.a1:
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • bottom vapor pressure(vp_bot)
  • top vapor pressure(vp_top)
  • Temperature of the top humidity chamber(thum_top)
  • Bottom humidity(hum_bot)
  • bottom air temperature(tair_bot)
  • Top humidity(hum_top)
  • top air temperature(tair_top)

sgp30ebbrE19.a1:
  • top vapor pressure(vp_top)
  • latent heat flux(e)
  • h(h)
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • top air temperature(tair_top)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • bottom air temperature(tair_bot)
  • Top humidity(hum_top)
  • Bottom humidity(hum_bot)

sgp30ebbrE19.b1:
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • h(h)
  • Temperature of the top humidity chamber(thum_top)
  • top air temperature(tair_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Top humidity(hum_top)
  • Bottom humidity(hum_bot)
  • bottom air temperature(tair_bot)
  • top vapor pressure(vp_top)
  • bottom vapor pressure(vp_bot)
  • latent heat flux(e)


Back To Table of Contents

DQRID : D030712.21
Start DateStart TimeEnd DateEnd Time
04/16/2003195004/16/20032045
Subject:
SGP/EBBR/E20 -  6 Month Checks
DataStreams:sgp15ebbrE20.a1, sgp15ebbrE20.b1, sgp30ebbrE20.a1, sgp30ebbrE20.b1, sgp5ebbrE20.a1
Description:
Many measurements were incorrect while the AEM was positioned with both aspirators at the 
same height for comparison checks.
Measurements:sgp30ebbrE20.b1:
  • Bottom humidity(hum_bot)
  • Top humidity(hum_top)
  • top vapor pressure(vp_top)
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • top air temperature(tair_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • latent heat flux(e)
  • h(h)
  • bottom air temperature(tair_bot)

sgp5ebbrE20.a1:
  • Temperature of the top humidity chamber(thum_top)
  • top vapor pressure(vp_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • bottom vapor pressure(vp_bot)
  • top air temperature(tair_top)
  • bottom air temperature(tair_bot)
  • Top humidity(hum_top)
  • Bottom humidity(hum_bot)

sgp15ebbrE20.b1:
  • Left relative humidity(mv_hum_l)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Left air temperature(tair_l)
  • Right air temperature(tair_r)
  • Right relative humidity(mv_hum_r)

sgp30ebbrE20.a1:
  • Bottom humidity(hum_bot)
  • Temperature of the top humidity chamber(thum_top)
  • bottom vapor pressure(vp_bot)
  • top air temperature(tair_top)
  • top vapor pressure(vp_top)
  • h(h)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • Top humidity(hum_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • latent heat flux(e)
  • bottom air temperature(tair_bot)

sgp15ebbrE20.a1:
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Right air temperature(tair_r)
  • Left air temperature(tair_l)
  • Left relative humidity(mv_hum_l)
  • Right relative humidity(mv_hum_r)
  • Temperature of left humidity sensor chamber(rr_thum_l)


Back To Table of Contents

DQRID : D030919.10
Start DateStart TimeEnd DateEnd Time
08/30/2003140009/04/20031445
Subject:
SGP/EBBR/E19  - Water in Net Radiometer; Surface Fluxes are Incorrect
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
Description:
The net radiometer top dome was punctured by migrating birds.  This alone does not cause 
incorrect net radiation measurements.  However, when it subsequently rains, water collects 
in the bottom dome, causing the net radiation measurements to be too large during the 
day and too small at night.
Measurements:sgp15ebbrE19.a1:
  • Net radiation(mv_q)

sgp15ebbrE19.b1:
  • Net radiation(mv_q)

sgp5ebbrE19.a1:
  • net radiation(q)

sgp30ebbrE19.a1:
  • latent heat flux(e)
  • h(h)
  • net radiation(q)

sgp30ebbrE19.b1:
  • h(h)
  • latent heat flux(e)
  • net radiation(q)


Back To Table of Contents

DQRID : D030922.6
Start DateStart TimeEnd DateEnd Time
09/17/2003183009/17/20032000
09/17/2003223009/18/20030100
Subject:
SGP/EBBR/E19 - Tair Right Offscale
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
Description:
The right air temperature data were going offscale.  This 
                  affected the calculation of sensible and latent heat 
                  fluxes.
Measurements:sgp15ebbrE19.a1:
  • Right air temperature(tair_r)

sgp15ebbrE19.b1:
  • Right air temperature(tair_r)

sgp5ebbrE19.a1:
  • bottom air temperature(tair_bot)
  • top air temperature(tair_top)

sgp30ebbrE19.a1:
  • top air temperature(tair_top)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • latent heat flux(e)
  • bottom air temperature(tair_bot)
  • h(h)

sgp30ebbrE19.b1:
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • h(h)
  • top air temperature(tair_top)
  • latent heat flux(e)
  • bottom air temperature(tair_bot)


Back To Table of Contents

DQRID : D031031.7
Start DateStart TimeEnd DateEnd Time
09/30/2003000010/01/20032100
Subject:
SGP/EBBR/E20  - Water in Net Radiometer, Surface Fluxes Incorrect
DataStreams:sgp15ebbrE20.a1, sgp15ebbrE20.b1, sgp30ebbrE20.a1, sgp30ebbrE20.b1, sgp5ebbrE20.a1
Description:
The net radiometer top dome was punctured by migrating birds.  This alone does not cause 
incorrect net radiation measurements.  However, when it subsequently rains, water collects 
in the bottom dome, causing the net radiation measurements to be too large during the 
day and too small at night.
Measurements:sgp30ebbrE20.b1:
  • latent heat flux(e)
  • net radiation(q)
  • h(h)

sgp5ebbrE20.a1:
  • net radiation(q)

sgp15ebbrE20.b1:
  • Net radiation(mv_q)

sgp30ebbrE20.a1:
  • net radiation(q)
  • latent heat flux(e)
  • h(h)

sgp15ebbrE20.a1:
  • Net radiation(mv_q)


Back To Table of Contents

DQRID : D031124.11
Start DateStart TimeEnd DateEnd Time
10/10/2003150510/10/20031755
Subject:
SGP/EBBR/E19 - Data Missing During Recalibration Changeout
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30baebbrE19.c1, sgp30ebbrE19.a1, sgp30ebbrE19.b1,
sgp5ebbrE19.a1, sgpebbrE19.00, sgpebbrE19.01
Description:
Data was missing during the recalibration changeout.
Measurements:sgp15ebbrE19.a1:
  • Reference temperature(rr_tref)
  • Soil temperature 5(rr_ts5)
  • scalar wind speed(wind_s)
  • Soil temperature 4(rr_ts4)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 1(r_sm1)
  • Home signal(mv_home)
  • Soil moisture 3(r_sm3)
  • Soil moisture 2(r_sm2)
  • Left air temperature(tair_l)
  • Left relative humidity(mv_hum_l)
  • Soil heat flow 4(mv_hft4)
  • Right relative humidity(mv_hum_r)
  • Soil heat flow 5(mv_hft5)
  • Soil heat flow 3(mv_hft3)
  • Soil moisture 4(r_sm4)
  • Soil moisture 5(r_sm5)
  • Signature(signature)
  • Dummy altitude for Zeb(alt)
  • base time(base_time)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Atmospheric pressure(mv_pres)
  • lat(lat)
  • Right air temperature(tair_r)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 2(mv_hft2)
  • Battery(bat)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil temperature 1(rr_ts1)
  • Soil temperature 2(rr_ts2)
  • Time offset of tweaks from base_time(time_offset)
  • lon(lon)
  • Soil temperature 3(rr_ts3)
  • Net radiation(mv_q)

sgp15ebbrE19.b1:
  • Right relative humidity(mv_hum_r)
  • Soil heat flow 1(mv_hft1)
  • Soil temperature 1(rr_ts1)
  • Soil heat flow 3(mv_hft3)
  • Soil temperature 2(rr_ts2)
  • lon(lon)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Reference temperature(rr_tref)
  • Battery(bat)
  • Soil temperature 3(rr_ts3)
  • Home signal(mv_home)
  • Right air temperature(tair_r)
  • Dummy altitude for Zeb(alt)
  • Soil heat flow 4(mv_hft4)
  • Soil heat flow 5(mv_hft5)
  • lat(lat)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Atmospheric pressure(mv_pres)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Soil moisture 5(r_sm5)
  • Soil temperature 4(rr_ts4)
  • Net radiation(mv_q)
  • scalar wind speed(wind_s)
  • Time offset of tweaks from base_time(time_offset)
  • Soil moisture 2(r_sm2)
  • Soil temperature 5(rr_ts5)
  • Left relative humidity(mv_hum_l)
  • Wind direction (relative to true north)(mv_wind_d)
  • Signature(signature)
  • Soil moisture 4(r_sm4)
  • Soil moisture 3(r_sm3)
  • base time(base_time)
  • Soil moisture 1(r_sm1)

sgp30baebbrE19.c1:
  • 5 cm soil heat flow, site 3(shf3)
  • top vapor pressure(vp_top)
  • 5 cm soil heat flow, site 1(shf1)
  • Dummy altitude for Zeb(alt)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • Retrieved pressure profile(pres)
  • lon(lon)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • wind direction (relative to true north)(wind_d)
  • bottom vapor pressure(vp_bot)
  • Best estimate Latent heat flux(ecomb)
  • 5 cm soil heat flow, site 4(shf4)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • h(h)
  • net radiation(q)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 5(shf5)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • 5 cm soil heat flow, site 2(shf2)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • base time(base_time)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • Aerodynamic Latent heat flux(ale)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • average surface soil heat flow(ave_shf)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • lat(lat)
  • bottom air temperature(tair_bot)
  • soil heat flow, site 4(g4)
  • soil heat flow, site 5(g5)
  • Best estimate Sensible heat flux(hcomb)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • Temperature of the top humidity chamber(thum_top)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • soil heat flow, site 1(g1)
  • Aerodynamic Sensible heat flux(ah)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • Reference Thermistor Temperature(tref)
  • top air temperature(tair_top)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • volumetric soil moisture, site 5(sm5)
  • soil heat flow, site 3(g3)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • latent heat flux(e)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • soil heat flow, site 2(g2)
  • Bottom humidity(hum_bot)
  • Estimated height of the vegetation.(vegetation_height)
  • volumetric soil moisture, site 1(sm1)
  • Friction velocity(austar)
  • scalar wind speed(wind_s)
  • Soil heat capacity 1(cs1)
  • Soil heat capacity 5(cs5)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Soil heat capacity 3(cs3)
  • volumetric soil moisture, site 3(sm3)
  • vector wind speed(res_ws)
  • Soil heat capacity 4(cs4)
  • Time offset of tweaks from base_time(time_offset)
  • Soil heat capacity 2(cs2)
  • Top humidity(hum_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • volumetric soil moisture, site 2(sm2)
  • volumetric soil moisture, site 4(sm4)

sgpebbrE19.00:
  • null(Raw data stream - documentation not supported)

sgp5ebbrE19.a1:
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • Bottom humidity(hum_bot)
  • base time(base_time)
  • wind direction (relative to true north)(wind_d)
  • bottom air temperature(tair_bot)
  • top air temperature(tair_top)
  • scalar wind speed(wind_s)
  • top vapor pressure(vp_top)
  • lat(lat)
  • lon(lon)
  • vector wind speed(res_ws)
  • Top humidity(hum_top)
  • Dummy altitude for Zeb(alt)
  • Reference Thermistor Temperature(tref)
  • Retrieved pressure profile(pres)
  • Time offset of tweaks from base_time(time_offset)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Home signal(home)
  • net radiation(q)

sgp30ebbrE19.a1:
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • vector wind speed(res_ws)
  • soil heat flow, site 2(g2)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • soil heat flow, site 1(g1)
  • net radiation(q)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • volumetric soil moisture, site 3(sm3)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • average surface soil heat flow(ave_shf)
  • volumetric soil moisture, site 4(sm4)
  • Soil heat capacity 3(cs3)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • Soil heat capacity 5(cs5)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • base time(base_time)
  • Soil heat capacity 2(cs2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • volumetric soil moisture, site 2(sm2)
  • Soil heat capacity 1(cs1)
  • latent heat flux(e)
  • bottom vapor pressure(vp_bot)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 3(g3)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Top humidity(hum_top)
  • lat(lat)
  • soil heat flow, site 5(g5)
  • scalar wind speed(wind_s)
  • bottom air temperature(tair_bot)
  • volumetric soil moisture, site 1(sm1)
  • Dummy altitude for Zeb(alt)
  • 5 cm soil heat flow, site 1(shf1)
  • 5 cm soil heat flow, site 4(shf4)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • Reference Thermistor Temperature(tref)
  • 5 cm soil heat flow, site 5(shf5)
  • top vapor pressure(vp_top)
  • lon(lon)
  • Retrieved pressure profile(pres)
  • Temperature of the top humidity chamber(thum_top)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 2(shf2)
  • soil heat flow, site 4(g4)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • Time offset of tweaks from base_time(time_offset)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • Soil heat capacity 4(cs4)
  • h(h)
  • volumetric soil moisture, site 5(sm5)
  • Bottom humidity(hum_bot)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 5 cm soil heat flow, site 3(shf3)
  • top air temperature(tair_top)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)

sgpebbrE19.01:
  • null(Raw data stream - documentation not supported)

sgp30ebbrE19.b1:
  • Dummy altitude for Zeb(alt)
  • Reference Thermistor Temperature(tref)
  • 5 cm soil heat flow, site 2(shf2)
  • volumetric soil moisture, site 1(sm1)
  • volumetric soil moisture, site 2(sm2)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • volumetric soil moisture, site 4(sm4)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • h(h)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • top vapor pressure(vp_top)
  • average surface soil heat flow(ave_shf)
  • volumetric soil moisture, site 5(sm5)
  • latent heat flux(e)
  • lat(lat)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 2(g2)
  • Soil heat capacity 1(cs1)
  • base time(base_time)
  • soil heat flow, site 5(g5)
  • soil heat flow, site 3(g3)
  • Soil heat capacity 2(cs2)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 5(shf5)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • bottom air temperature(tair_bot)
  • net radiation(q)
  • 5 cm soil heat flow, site 4(shf4)
  • Retrieved pressure profile(pres)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Soil heat capacity 5(cs5)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • Soil heat capacity 3(cs3)
  • Temperature of the top humidity chamber(thum_top)
  • Soil heat capacity 4(cs4)
  • Bottom humidity(hum_bot)
  • soil heat flow, site 4(g4)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • lon(lon)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • vector wind speed(res_ws)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Time offset of tweaks from base_time(time_offset)
  • volumetric soil moisture, site 3(sm3)
  • soil heat flow, site 1(g1)
  • 5 cm soil heat flow, site 1(shf1)
  • bottom vapor pressure(vp_bot)
  • Top humidity(hum_top)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 5 cm soil heat flow, site 3(shf3)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)


Back To Table of Contents

DQRID : D031204.17
Start DateStart TimeEnd DateEnd Time
10/10/2003153010/10/20031600
Subject:
SGP/EBBR/E19 - 6 Month Checks
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
Description:
Many measurements were incorrect while the AEM was positioned with both aspirators at the 
same height for comparison checks.
Measurements:sgp15ebbrE19.a1:
  • Right relative humidity(mv_hum_r)
  • Right air temperature(tair_r)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Left air temperature(tair_l)
  • Left relative humidity(mv_hum_l)

sgp15ebbrE19.b1:
  • Right air temperature(tair_r)
  • Right relative humidity(mv_hum_r)
  • Left relative humidity(mv_hum_l)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Left air temperature(tair_l)
  • Temperature of left humidity sensor chamber(rr_thum_l)

sgp5ebbrE19.a1:
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • bottom vapor pressure(vp_bot)
  • top vapor pressure(vp_top)
  • Temperature of the top humidity chamber(thum_top)
  • Bottom humidity(hum_bot)
  • bottom air temperature(tair_bot)
  • Top humidity(hum_top)
  • top air temperature(tair_top)

sgp30ebbrE19.a1:
  • top vapor pressure(vp_top)
  • latent heat flux(e)
  • h(h)
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • top air temperature(tair_top)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • bottom air temperature(tair_bot)
  • Top humidity(hum_top)
  • Bottom humidity(hum_bot)

sgp30ebbrE19.b1:
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • h(h)
  • Temperature of the top humidity chamber(thum_top)
  • top air temperature(tair_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Top humidity(hum_top)
  • Bottom humidity(hum_bot)
  • bottom air temperature(tair_bot)
  • top vapor pressure(vp_top)
  • bottom vapor pressure(vp_bot)
  • latent heat flux(e)


Back To Table of Contents

DQRID : D031204.2
Start DateStart TimeEnd DateEnd Time
10/10/2003180010/10/20031800
Subject:
SGP/EBBR/E19 - Data All Zeros
DataStreams:sgp15ebbrE19.a1, sgp15ebbrE19.b1, sgp30ebbrE19.a1, sgp30ebbrE19.b1, sgp5ebbrE19.a1
Description:
The time period after restart of the datalogger produced all
zero data.
Measurements:sgp15ebbrE19.a1:
  • Reference temperature(rr_tref)
  • Soil temperature 5(rr_ts5)
  • scalar wind speed(wind_s)
  • Soil temperature 4(rr_ts4)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 1(r_sm1)
  • Home signal(mv_home)
  • Soil moisture 3(r_sm3)
  • Soil moisture 2(r_sm2)
  • Left air temperature(tair_l)
  • Left relative humidity(mv_hum_l)
  • Soil heat flow 4(mv_hft4)
  • Soil heat flow 5(mv_hft5)
  • Right relative humidity(mv_hum_r)
  • Soil moisture 4(r_sm4)
  • Soil heat flow 3(mv_hft3)
  • Soil moisture 5(r_sm5)
  • Signature(signature)
  • Dummy altitude for Zeb(alt)
  • base time(base_time)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Atmospheric pressure(mv_pres)
  • lat(lat)
  • Right air temperature(tair_r)
  • Soil heat flow 1(mv_hft1)
  • Soil heat flow 2(mv_hft2)
  • Battery(bat)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil temperature 1(rr_ts1)
  • Soil temperature 2(rr_ts2)
  • Time offset of tweaks from base_time(time_offset)
  • lon(lon)
  • Soil temperature 3(rr_ts3)
  • Net radiation(mv_q)

sgp15ebbrE19.b1:
  • Right relative humidity(mv_hum_r)
  • Soil heat flow 1(mv_hft1)
  • Soil temperature 1(rr_ts1)
  • Soil heat flow 3(mv_hft3)
  • Soil temperature 2(rr_ts2)
  • Soil heat flow 2(mv_hft2)
  • Left air temperature(tair_l)
  • Reference temperature(rr_tref)
  • Battery(bat)
  • Soil temperature 3(rr_ts3)
  • Home signal(mv_home)
  • Right air temperature(tair_r)
  • Soil heat flow 4(mv_hft4)
  • Soil heat flow 5(mv_hft5)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Atmospheric pressure(mv_pres)
  • Soil moisture 5(r_sm5)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Soil temperature 4(rr_ts4)
  • Net radiation(mv_q)
  • scalar wind speed(wind_s)
  • Soil moisture 2(r_sm2)
  • Soil temperature 5(rr_ts5)
  • Left relative humidity(mv_hum_l)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil moisture 4(r_sm4)
  • Soil moisture 3(r_sm3)
  • Soil moisture 1(r_sm1)

sgp5ebbrE19.a1:
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • Bottom humidity(hum_bot)
  • wind direction (relative to true north)(wind_d)
  • bottom air temperature(tair_bot)
  • top air temperature(tair_top)
  • scalar wind speed(wind_s)
  • top vapor pressure(vp_top)
  • vector wind speed(res_ws)
  • Top humidity(hum_top)
  • Reference Thermistor Temperature(tref)
  • Retrieved pressure profile(pres)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Home signal(home)
  • net radiation(q)

sgp30ebbrE19.a1:
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • vector wind speed(res_ws)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • soil heat flow, site 2(g2)
  • soil heat flow, site 1(g1)
  • net radiation(q)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • volumetric soil moisture, site 3(sm3)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • average surface soil heat flow(ave_shf)
  • Soil heat capacity 3(cs3)
  • volumetric soil moisture, site 4(sm4)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • Soil heat capacity 5(cs5)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • base time(base_time)
  • Soil heat capacity 2(cs2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • volumetric soil moisture, site 2(sm2)
  • Soil heat capacity 1(cs1)
  • latent heat flux(e)
  • bottom vapor pressure(vp_bot)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 3(g3)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Top humidity(hum_top)
  • lat(lat)
  • soil heat flow, site 5(g5)
  • scalar wind speed(wind_s)
  • bottom air temperature(tair_bot)
  • volumetric soil moisture, site 1(sm1)
  • Dummy altitude for Zeb(alt)
  • 5 cm soil heat flow, site 1(shf1)
  • 5 cm soil heat flow, site 4(shf4)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • Reference Thermistor Temperature(tref)
  • 5 cm soil heat flow, site 5(shf5)
  • top vapor pressure(vp_top)
  • lon(lon)
  • Retrieved pressure profile(pres)
  • Temperature of the top humidity chamber(thum_top)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 2(shf2)
  • soil heat flow, site 4(g4)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • Time offset of tweaks from base_time(time_offset)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • Soil heat capacity 4(cs4)
  • h(h)
  • volumetric soil moisture, site 5(sm5)
  • Bottom humidity(hum_bot)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 5 cm soil heat flow, site 3(shf3)
  • top air temperature(tair_top)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)

sgp30ebbrE19.b1:
  • Reference Thermistor Temperature(tref)
  • 5 cm soil heat flow, site 2(shf2)
  • volumetric soil moisture, site 1(sm1)
  • volumetric soil moisture, site 2(sm2)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • volumetric soil moisture, site 4(sm4)
  • h(h)
  • scalar wind speed(wind_s)
  • top air temperature(tair_top)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • top vapor pressure(vp_top)
  • average surface soil heat flow(ave_shf)
  • volumetric soil moisture, site 5(sm5)
  • latent heat flux(e)
  • wind direction (relative to true north)(wind_d)
  • soil heat flow, site 2(g2)
  • Soil heat capacity 1(cs1)
  • soil heat flow, site 5(g5)
  • soil heat flow, site 3(g3)
  • Soil heat capacity 2(cs2)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow, site 5(shf5)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • bottom air temperature(tair_bot)
  • net radiation(q)
  • 5 cm soil heat flow, site 4(shf4)
  • Retrieved pressure profile(pres)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Soil heat capacity 5(cs5)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • Soil heat capacity 3(cs3)
  • Temperature of the top humidity chamber(thum_top)
  • Soil heat capacity 4(cs4)
  • Bottom humidity(hum_bot)
  • soil heat flow, site 4(g4)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • vector wind speed(res_ws)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • volumetric soil moisture, site 3(sm3)
  • soil heat flow, site 1(g1)
  • 5 cm soil heat flow, site 1(shf1)
  • bottom vapor pressure(vp_bot)
  • Top humidity(hum_top)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 5 cm soil heat flow, site 3(shf3)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)


Back To Table of Contents

DQRID : D040525.3
Start DateStart TimeEnd DateEnd Time
05/30/1997000005/06/20032359
Subject:
SGP/EBBR/E19 - Metadata errors
DataStreams:sgp15ebbrE19.a1, sgp30baebbrE19.c1, sgp30ebbrE19.a1
Description:
The latitude, longitude and altitude of the El Reno (E19) site were
incorrectly entered into the ARM database.  The correct location
of the EBBR.E19 instrument is:
 Lat: 35.557N
 Lon: 98.017W
 Alt:    421m
Measurements:sgp15ebbrE19.a1:
  • Dummy altitude for Zeb(alt)
  • lon(lon)
  • lat(lat)

sgp30baebbrE19.c1:
  • lat(lat)
  • lon(lon)
  • Dummy altitude for Zeb(alt)

sgp30ebbrE19.a1:
  • lon(lon)
  • lat(lat)
  • Dummy altitude for Zeb(alt)


Back To Table of Contents