DQR ID | Subject | Data Streams Affected |
---|---|---|
D020905.4 | SGP/IRT10M/C1 - IRT not insulated | sgpirt10mC1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/19/1997 | 0000 | 12/24/1998 | 1845 |
Subject: | SGP/IRT10M/C1 - IRT not insulated |
DataStreams: | sgpirt10mC1.b1 |
Description: | The upwelling IRT on the 10m tower was insufficiently insulated to maintain an internal reference temperature above 0 degrees C. Measurements of surface temperature were over-estimated when instrument was below freezing. |
Measurements: | sgpirt10mC1.b1:
|