DQR ID | Subject | Data Streams Affected |
---|---|---|
D041027.1 | NSA/MWRP/C1 - Missing data | nsamwrpC1.b1 |
D070709.1 | NSA/MWRP/C1 - Reprocessed: 6-channel retrieval replaced by 5-channel retrieval | nsamwrpC1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/21/2004 | 0925 | 02/23/2004 | 2238 |
10/06/2004 | 0008 | 10/27/2004 | 1237 |
11/20/2004 | 0039 | 11/22/2004 | 1851 |
04/10/2005 | 0000 | 04/11/2005 | 1622 |
09/04/2005 | 0531 | 09/06/2005 | 1857 |
09/10/2005 | 0052 | 09/12/2005 | 1830 |
09/21/2005 | 1805 | 10/12/2005 | 2311 |
06/23/2006 | 1800 | 08/08/2006 | 2250 |
08/11/2006 | 2359 | 08/14/2006 | 1454 |
10/09/2006 | 2101 | 10/11/2006 | 1733 |
01/05/2007 | 2358 | 01/08/2007 | 2019 |
02/28/2007 | 2309 | 03/05/2007 | 0027 |
07/13/2007 | 1838 | 12/08/2007 | 0125 |
Subject: | NSA/MWRP/C1 - Missing data |
DataStreams: | nsamwrpC1.b1 |
Description: | Data are missing and unrecoverable. |
Measurements: | nsamwrpC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/19/2004 | 0214 | 06/02/2007 | 2359 |
Subject: | NSA/MWRP/C1 - Reprocessed: 6-channel retrieval replaced by 5-channel retrieval |
DataStreams: | nsamwrpC1.b1 |
Description: | Due to repeated failure of the V-band receiver and calibration problems, the 6-channel retrieval was retired and a new 5-channel retrieval which uses only the K-band channels was implemented. All data previously processed with the 6-channel retrieval have been reprocessed. Please note that the 7 channels (brightness temperatures) in the V-band range (51 to 60 GHz) should not be used. See D080331.1 and D080331.2. |
Measurements: | nsamwrpC1.b1:
|