Data Quality Reports for Session: 114546 User: tra Completed: 10/04/2008


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D070313.2TWP/SONDE/C2 - Surface RH value incorrecttwpsondewnpnC2.b1
D070329.2TWP/SONDE/C2 - Surface RH Value incorrecttwpsondewnpnC2.b1


DQRID : D070313.2
Start DateStart TimeEnd DateEnd Time
03/08/2007235003/08/20072353
Subject:
TWP/SONDE/C2 - Surface RH value incorrect
DataStreams:twpsondewnpnC2.b1
Description:
The surface RH was reproted as 78%.  The value should have been in the low 50's.  It 
appears the observer did not allow the sonde to equilibrate to ambient conditions before 
taking the reading.  This error also caused an incorrect dew point.
Measurements:twpsondewnpnC2.b1:
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)


Back To Table of Contents

DQRID : D070329.2
Start DateStart TimeEnd DateEnd Time
03/28/2007231903/28/20072320
Subject:
TWP/SONDE/C2 - Surface RH Value incorrect
DataStreams:twpsondewnpnC2.b1
Description:
The surface value entered by the observers for RH (96%) is incorrect.  At the ascent time 
listed in the .ptu file the readings by the Sonde were around 65%.  This low RH value is 
supported by the nearby SMET tower RH reading of 71.8%.  This also caused the surface dew 
point to be incorrect.
Measurements:twpsondewnpnC2.b1:
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)


Back To Table of Contents



END OF DATA