DQR ID | Subject | Data Streams Affected |
---|---|---|
D060717.1 | SGP/THWAPS/B4 - Temp and RH data incorrect | sgpthwapsB4.b1 |
D060830.1 | SGP/THWAPS/B4 - Reprocess: RH% values 9% High | sgpthwapsB4.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/05/2006 | 1800 | 07/18/2006 | 1647 |
Subject: | SGP/THWAPS/B4 - Temp and RH data incorrect |
DataStreams: | sgpthwapsB4.b1 |
Description: | After sensor replacement the temperature values are too low. The range of the sensor was incorrectly set at -40 to 80 instead of -30 to 70. RH values are also suspect. |
Measurements: | sgpthwapsB4.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/05/2006 | 1825 | 09/26/2006 | 0000 |
Subject: | SGP/THWAPS/B4 - Reprocess: RH% values 9% High |
DataStreams: | sgpthwapsB4.b1 |
Description: | After replacement of the THWAPS T/RH sensor the span was incorrectly set to 0-100 instead of 0-110. This caused the RH values to be 9% higher than actual values. |
Measurements: | sgpthwapsB4.b1:
|