DQR ID | Subject | Data Streams Affected |
---|---|---|
D080710.2 | NSA/MWRP/C1 - Reprocess: LN2 calibration needed | nsamwrpC1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/01/2008 | 0000 | 07/18/2008 | 1950 |
Subject: | NSA/MWRP/C1 - Reprocess: LN2 calibration needed |
DataStreams: | nsamwrpC1.b1 |
Description: | The V-band measurements are incorrect. The quantities affected are the temperature and humidity profiles. V-band data should not be used until the instrument is calibrated and data are reprocessed. The K-band data are acceptable. |
Measurements: | nsamwrpC1.b1:
|