Data Quality Reports for Session: 117809 User: dmfoper Completed: 02/25/2009


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D011127.23SGP EBBR E7 - E7 EBBR Data Incorrect During 6 Month Checkssgp15ebbrE7.a1, sgp15ebbrE7.b1, sgp30ebbrE7.a1, sgp30ebbrE7.b1, sgp5ebbrE7.a1
D011127.33SGP EBBR E25 - E25 EBBR Data Incorrect During 6 Month Checkssgp15ebbrE25.a1, sgp15ebbrE25.b1, sgp30ebbrE25.a1, sgp30ebbrE25.b1, sgp5ebbrE25.a1
D011129.6SGP EBBR E7 - E7 EBBR AEM Position Caused Incorrect Datasgp15ebbrE7.a1, sgp15ebbrE7.b1, sgp30ebbrE7.a1, sgp30ebbrE7.b1, sgp5ebbrE7.a1
D011129.7SGP EBBR E7 - E7 EBBR All Data Incorrectsgp15ebbrE7.a1, sgp15ebbrE7.b1, sgp30ebbrE7.a1, sgp30ebbrE7.b1, sgp5ebbrE7.a1
D011129.8SGP EBBR E7 - E7 EBBR Data Incorrect After Recalibrationsgp15ebbrE7.a1, sgp15ebbrE7.b1, sgp30ebbrE7.a1, sgp30ebbrE7.b1
D011129.9SGP EBBR E7 - E7 EBBR Data Incorrect After Missing Data Periodsgp15ebbrE7.a1, sgp15ebbrE7.b1, sgp30ebbrE7.a1, sgp30ebbrE7.b1
D020716.26SGP/EBBR/E7 - Measurements Zerosgp15ebbrE7.a1, sgp15ebbrE7.b1, sgp30ebbrE7.a1, sgp30ebbrE7.b1, sgp5ebbrE7.00,
sgp5ebbrE7.a1
D020716.27SGP/EBBR/E7 - AEM Hungsgp15ebbrE7.a1, sgp15ebbrE7.b1, sgp30ebbrE7.a1, sgp30ebbrE7.b1, sgp5ebbrE7.a1
D020716.28SGP/EBBR/E7 - Wind Speed Incorrectsgp15ebbrE7.a1, sgp15ebbrE7.b1, sgp30ebbrE7.a1, sgp30ebbrE7.b1, sgp5ebbrE7.a1
D020716.8SGP/EBBR/E7 - 6 Month Checkssgp15ebbrE7.a1, sgp15ebbrE7.b1, sgp30ebbrE7.a1, sgp30ebbrE7.b1, sgp5ebbrE7.a1
D030110.30SGP/EBBR/E7 - Data Incorrect during 6 month checkssgp15ebbrE7.a1, sgp15ebbrE7.b1, sgp30ebbrE7.a1, sgp30ebbrE7.b1, sgp5ebbrE7.a1
D030113.1SGP/EBBR/E7 - Wind speed incorrectsgp15ebbrE7.a1, sgp15ebbrE7.b1, sgp30ebbrE7.a1, sgp30ebbrE7.b1, sgp5ebbrE7.a1
D030113.7SGP/EBBR/E7 - AEM Problemssgp15ebbrE7.a1, sgp15ebbrE7.b1, sgp30ebbrE7.a1, sgp30ebbrE7.b1, sgp5ebbrE7.a1
D030113.8SGP/EBBR/E7 - T/RH Ventilator Fans Stoppedsgp15ebbrE7.a1, sgp15ebbrE7.b1, sgp30ebbrE7.a1, sgp30ebbrE7.b1, sgp5ebbrE7.a1
D030114.2SGP/EBBR/E7 - Data Reported as Zero following Logger Replacementsgp15ebbrE7.a1, sgp15ebbrE7.b1, sgp30ebbrE7.a1, sgp30ebbrE7.b1, sgp5ebbrE7.a1
D030620.1SGP/EBBR/E7 - Soil Heat Flow #5 Incorrectsgp15ebbrE7.a1, sgp15ebbrE7.b1, sgp30ebbrE7.a1, sgp30ebbrE7.b1
D030712.13SGP/EBBR/E7 - 6 Month Checkssgp15ebbrE7.a1, sgp15ebbrE7.b1, sgp30ebbrE7.a1, sgp30ebbrE7.b1, sgp5ebbrE7.a1
D030712.9SGP/EBBR/E7 - AEM Hung Inbetween Home Positionsgp15ebbrE7.a1, sgp15ebbrE7.b1, sgp30ebbrE7.a1, sgp30ebbrE7.b1, sgp5ebbrE7.a1
D031211.6SGP/EBBR/E7 - Pressure Sensor Data Incorrectsgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.a1
D051115.10SGP/EBBR/E25 - metadata errorssgp15ebbrE25.b1, sgp30ebbrE25.b1


DQRID : D011127.23
Start DateStart TimeEnd DateEnd Time
05/08/2001180005/08/20011901
11/25/2001143511/25/20011601
Subject:
SGP EBBR E7 - E7 EBBR Data Incorrect During 6 Month Checks
DataStreams:sgp15ebbrE7.a1, sgp15ebbrE7.b1, sgp30ebbrE7.a1, sgp30ebbrE7.b1, sgp5ebbrE7.a1
Description:
The data was incorrect during the 6 month checks.
Measurements:sgp5ebbrE7.a1:
  • Temperature of the top humidity chamber(thum_top)
  • Reference Thermistor Temperature(tref)
  • Top humidity(hum_top)
  • Retrieved pressure profile(pres)
  • vector wind speed(res_ws)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • net radiation(q)
  • Bottom humidity(hum_bot)
  • bottom vapor pressure(vp_bot)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • top vapor pressure(vp_top)
  • wind direction (relative to true north)(wind_d)
  • top air temperature(tair_top)
  • scalar wind speed(wind_s)
  • bottom air temperature(tair_bot)

sgp30ebbrE7.a1:
  • wind direction (relative to true north)(wind_d)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • soil heat flow, site 2(g2)
  • top air temperature(tair_top)
  • average surface soil heat flow(ave_shf)
  • vector wind speed(res_ws)
  • volumetric soil moisture, site 5(sm5)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • bottom air temperature(tair_bot)
  • volumetric soil moisture, site 4(sm4)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • volumetric soil moisture, site 2(sm2)
  • Soil heat capacity 2(cs2)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • volumetric soil moisture, site 1(sm1)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • soil heat flow, site 4(g4)
  • soil heat flow, site 3(g3)
  • Temperature of the top humidity chamber(thum_top)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • Soil heat capacity 5(cs5)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • Soil heat capacity 3(cs3)
  • Retrieved pressure profile(pres)
  • Reference Thermistor Temperature(tref)
  • Soil heat capacity 1(cs1)
  • 5 cm soil heat flow, site 5(shf5)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • net radiation(q)
  • Soil heat capacity 4(cs4)
  • 5 cm soil heat flow, site 1(shf1)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • bottom vapor pressure(vp_bot)
  • h(h)
  • 5 cm soil heat flow, site 4(shf4)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • soil heat flow, site 1(g1)
  • Top humidity(hum_top)
  • soil heat flow, site 5(g5)
  • top vapor pressure(vp_top)
  • scalar wind speed(wind_s)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • 5 cm soil heat flow, site 2(shf2)
  • 5 cm soil heat flow, site 3(shf3)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • Bottom humidity(hum_bot)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • volumetric soil moisture, site 3(sm3)

sgp15ebbrE7.a1:
  • Atmospheric pressure(mv_pres)
  • Soil moisture 5(r_sm5)
  • Soil temperature 5(rr_ts5)
  • Left relative humidity(mv_hum_l)
  • Soil heat flow 5(mv_hft5)
  • Soil moisture 2(r_sm2)
  • Soil temperature 4(rr_ts4)
  • Soil heat flow 4(mv_hft4)
  • Soil moisture 3(r_sm3)
  • Wind direction (relative to true north)(mv_wind_d)
  • Left air temperature(tair_l)
  • Right air temperature(tair_r)
  • Soil moisture 2(rr_sm2)
  • Soil moisture 5(rr_sm5)
  • Soil moisture 1(r_sm1)
  • Soil moisture 3(rr_sm3)
  • Soil heat flow 1(mv_hft1)
  • Soil moisture 4(r_sm4)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil moisture 4(rr_sm4)
  • Right relative humidity(mv_hum_r)
  • Soil temperature 2(rr_ts2)
  • Soil moisture 1(rr_sm1)
  • Soil temperature 3(rr_ts3)
  • Soil heat flow 2(mv_hft2)
  • Soil heat flow 3(mv_hft3)
  • Net radiation(mv_q)
  • Soil temperature 1(rr_ts1)
  • Reference temperature(rr_tref)
  • scalar wind speed(wind_s)

sgp15ebbrE7.b1:
  • Net radiation(mv_q)
  • Soil moisture 2(r_sm2)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Soil moisture 3(r_sm3)
  • Soil heat flow 3(mv_hft3)
  • Soil heat flow 4(mv_hft4)
  • Soil moisture 5(r_sm5)
  • Home signal(mv_home)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil heat flow 5(mv_hft5)
  • Reference temperature(rr_tref)
  • Left relative humidity(mv_hum_l)
  • scalar wind speed(wind_s)
  • Soil heat flow 1(mv_hft1)
  • Right relative humidity(mv_hum_r)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil temperature 4(rr_ts4)
  • Left air temperature(tair_l)
  • Soil temperature 3(rr_ts3)
  • Soil moisture 4(r_sm4)
  • Soil temperature 5(rr_ts5)
  • Soil moisture 1(r_sm1)
  • Soil heat flow 2(mv_hft2)
  • Right air temperature(tair_r)
  • Soil temperature 1(rr_ts1)
  • Atmospheric pressure(mv_pres)
  • Soil temperature 2(rr_ts2)

sgp30ebbrE7.b1:
  • volumetric soil moisture, site 3(sm3)
  • wind direction (relative to true north)(wind_d)
  • volumetric soil moisture, site 2(sm2)
  • bottom air temperature(tair_bot)
  • volumetric soil moisture, site 4(sm4)
  • average surface soil heat flow(ave_shf)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • 5 cm soil heat flow, site 4(shf4)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • volumetric soil moisture, site 5(sm5)
  • Soil heat capacity 2(cs2)
  • 5 cm soil heat flow, site 3(shf3)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • top air temperature(tair_top)
  • volumetric soil moisture, site 1(sm1)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • top vapor pressure(vp_top)
  • scalar wind speed(wind_s)
  • latent heat flux(e)
  • vector wind speed(res_ws)
  • Top humidity(hum_top)
  • bottom vapor pressure(vp_bot)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • soil heat flow, site 3(g3)
  • 5 cm soil heat flow, site 5(shf5)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • soil heat flow, site 2(g2)
  • Soil heat capacity 1(cs1)
  • Bottom humidity(hum_bot)
  • Reference Thermistor Temperature(tref)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • h(h)
  • Soil heat capacity 4(cs4)
  • Temperature of the top humidity chamber(thum_top)
  • Soil heat capacity 5(cs5)
  • Retrieved pressure profile(pres)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • net radiation(q)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • soil heat flow, site 1(g1)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • soil heat flow, site 4(g4)
  • Soil heat capacity 3(cs3)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • soil heat flow, site 5(g5)
  • 5 cm soil heat flow, site 2(shf2)
  • 5 cm soil heat flow, site 1(shf1)


Back To Table of Contents

DQRID : D011127.33
Start DateStart TimeEnd DateEnd Time
04/04/2001153004/04/20011631
10/17/2001170510/17/20011811
Subject:
SGP EBBR E25 - E25 EBBR Data Incorrect During 6 Month Checks
DataStreams:sgp15ebbrE25.a1, sgp15ebbrE25.b1, sgp30ebbrE25.a1, sgp30ebbrE25.b1, sgp5ebbrE25.a1
Description:
The data was incorrect during the 6 month checks.
Measurements:sgp15ebbrE25.a1:
  • Soil heat flow 1(mv_hft1)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil heat flow 5(mv_hft5)
  • Left air temperature(tair_l)
  • Soil moisture 4(r_sm4)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Net radiation(mv_q)
  • Soil moisture 3(r_sm3)
  • Left relative humidity(mv_hum_l)
  • Soil temperature 1(rr_ts1)
  • Soil heat flow 4(mv_hft4)
  • Soil moisture 1(r_sm1)
  • Reference temperature(rr_tref)
  • Soil moisture 5(r_sm5)
  • Soil temperature 4(rr_ts4)
  • Atmospheric pressure(mv_pres)
  • Right air temperature(tair_r)
  • Right relative humidity(mv_hum_r)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil temperature 5(rr_ts5)
  • Soil moisture 2(r_sm2)
  • Soil heat flow 3(mv_hft3)
  • Soil temperature 3(rr_ts3)
  • Soil temperature 2(rr_ts2)
  • Soil heat flow 2(mv_hft2)
  • scalar wind speed(wind_s)

sgp30ebbrE25.b1:
  • volumetric soil moisture, site 4(sm4)
  • Soil heat capacity 3(cs3)
  • 5 cm soil heat flow, site 3(shf3)
  • latent heat flux(e)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • Soil heat capacity 5(cs5)
  • volumetric soil moisture, site 5(sm5)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • 5 cm soil heat flow, site 4(shf4)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • wind direction (relative to true north)(wind_d)
  • bottom air temperature(tair_bot)
  • 5 cm soil heat flow, site 5(shf5)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • Soil heat capacity 2(cs2)
  • h(h)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • top vapor pressure(vp_top)
  • soil heat flow, site 2(g2)
  • volumetric soil moisture, site 3(sm3)
  • net radiation(q)
  • 5 cm soil heat flow, site 2(shf2)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • 5 cm soil heat flow, site 1(shf1)
  • soil heat flow, site 1(g1)
  • volumetric soil moisture, site 2(sm2)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • Retrieved pressure profile(pres)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • Reference Thermistor Temperature(tref)
  • Soil heat capacity 4(cs4)
  • soil heat flow, site 3(g3)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • volumetric soil moisture, site 1(sm1)
  • soil heat flow, site 4(g4)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • average surface soil heat flow(ave_shf)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • Soil heat capacity 1(cs1)
  • Top humidity(hum_top)
  • scalar wind speed(wind_s)
  • soil heat flow, site 5(g5)
  • top air temperature(tair_top)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • Temperature of the top humidity chamber(thum_top)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • Bottom humidity(hum_bot)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • vector wind speed(res_ws)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • bottom vapor pressure(vp_bot)

sgp15ebbrE25.b1:
  • Left relative humidity(mv_hum_l)
  • Soil temperature 2(rr_ts2)
  • Atmospheric pressure(mv_pres)
  • Soil moisture 5(r_sm5)
  • Right relative humidity(mv_hum_r)
  • Reference temperature(rr_tref)
  • scalar wind speed(wind_s)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil heat flow 5(mv_hft5)
  • Soil temperature 4(rr_ts4)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Soil heat flow 3(mv_hft3)
  • Soil heat flow 4(mv_hft4)
  • Soil temperature 3(rr_ts3)
  • Right air temperature(tair_r)
  • Net radiation(mv_q)
  • Left air temperature(tair_l)
  • Soil moisture 3(r_sm3)
  • Soil moisture 4(r_sm4)
  • Soil temperature 1(rr_ts1)
  • Soil temperature 5(rr_ts5)
  • Soil heat flow 2(mv_hft2)
  • Soil moisture 2(r_sm2)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil heat flow 1(mv_hft1)
  • Soil moisture 1(r_sm1)

sgp30ebbrE25.a1:
  • bottom vapor pressure(vp_bot)
  • average surface soil heat flow(ave_shf)
  • 5 cm soil heat flow, site 5(shf5)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • volumetric soil moisture, site 4(sm4)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • volumetric soil moisture, site 1(sm1)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • scalar wind speed(wind_s)
  • 5 cm soil heat flow, site 4(shf4)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • top vapor pressure(vp_top)
  • volumetric soil moisture, site 3(sm3)
  • Bottom humidity(hum_bot)
  • soil heat flow, site 4(g4)
  • soil heat flow, site 1(g1)
  • 5 cm soil heat flow, site 1(shf1)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • 5 cm soil heat flow, site 2(shf2)
  • 5 cm soil heat flow, site 3(shf3)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • Soil heat capacity 3(cs3)
  • soil heat flow, site 5(g5)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • volumetric soil moisture, site 2(sm2)
  • bottom air temperature(tair_bot)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • Soil heat capacity 4(cs4)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • vector wind speed(res_ws)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • wind direction (relative to true north)(wind_d)
  • Temperature of the top humidity chamber(thum_top)
  • h(h)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • top air temperature(tair_top)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • Retrieved pressure profile(pres)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • Reference Thermistor Temperature(tref)
  • Soil heat capacity 2(cs2)
  • Top humidity(hum_top)
  • soil heat flow, site 2(g2)
  • net radiation(q)
  • soil heat flow, site 3(g3)
  • Soil heat capacity 1(cs1)
  • volumetric soil moisture, site 5(sm5)
  • latent heat flux(e)
  • Soil heat capacity 5(cs5)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 0-5 cm integrated soil temperature, site 4(ts4)

sgp5ebbrE25.a1:
  • bottom vapor pressure(vp_bot)
  • vector wind speed(res_ws)
  • Retrieved pressure profile(pres)
  • Bottom humidity(hum_bot)
  • wind direction (relative to true north)(wind_d)
  • Top humidity(hum_top)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • bottom air temperature(tair_bot)
  • net radiation(q)
  • top air temperature(tair_top)
  • top vapor pressure(vp_top)
  • scalar wind speed(wind_s)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • Reference Thermistor Temperature(tref)
  • Temperature of the top humidity chamber(thum_top)


Back To Table of Contents

DQRID : D011129.6
Start DateStart TimeEnd DateEnd Time
04/10/2001194504/10/20012031
Subject:
SGP EBBR E7 - E7 EBBR AEM Position Caused Incorrect Data
DataStreams:sgp15ebbrE7.a1, sgp15ebbrE7.b1, sgp30ebbrE7.a1, sgp30ebbrE7.b1, sgp5ebbrE7.a1
Description:
The AEM stuck in one position, causing the indicated measurements and calculated
quantities to be incorrect.
Measurements:sgp5ebbrE7.a1:
  • top air temperature(tair_top)
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • top vapor pressure(vp_top)
  • Top humidity(hum_top)
  • bottom air temperature(tair_bot)
  • Bottom humidity(hum_bot)

sgp15ebbrE7.a1:
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Left relative humidity(mv_hum_l)
  • Left air temperature(tair_l)
  • Right air temperature(tair_r)
  • Right relative humidity(mv_hum_r)

sgp30ebbrE7.a1:
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • top air temperature(tair_top)
  • Top humidity(hum_top)
  • top vapor pressure(vp_top)
  • bottom air temperature(tair_bot)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • bottom vapor pressure(vp_bot)
  • latent heat flux(e)
  • h(h)
  • Bottom humidity(hum_bot)
  • Temperature of the top humidity chamber(thum_top)

sgp15ebbrE7.b1:
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Right relative humidity(mv_hum_r)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Left relative humidity(mv_hum_l)
  • Right air temperature(tair_r)
  • Left air temperature(tair_l)

sgp30ebbrE7.b1:
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • bottom vapor pressure(vp_bot)
  • Top humidity(hum_top)
  • top air temperature(tair_top)
  • bottom air temperature(tair_bot)
  • Bottom humidity(hum_bot)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • h(h)
  • top vapor pressure(vp_top)
  • latent heat flux(e)
  • Temperature of the top humidity chamber(thum_top)


Back To Table of Contents

DQRID : D011129.7
Start DateStart TimeEnd DateEnd Time
05/07/2001190005/07/20011901
Subject:
SGP EBBR E7 - E7 EBBR All Data Incorrect
DataStreams:sgp15ebbrE7.a1, sgp15ebbrE7.b1, sgp30ebbrE7.a1, sgp30ebbrE7.b1, sgp5ebbrE7.a1
Description:
The data was incorrect immediately after the re-calibrated system was 
installed and put into service.
Measurements:sgp5ebbrE7.a1:
  • Temperature of the top humidity chamber(thum_top)
  • Reference Thermistor Temperature(tref)
  • Top humidity(hum_top)
  • Retrieved pressure profile(pres)
  • vector wind speed(res_ws)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • net radiation(q)
  • Bottom humidity(hum_bot)
  • bottom vapor pressure(vp_bot)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • top vapor pressure(vp_top)
  • wind direction (relative to true north)(wind_d)
  • Home signal(home)
  • top air temperature(tair_top)
  • scalar wind speed(wind_s)
  • bottom air temperature(tair_bot)

sgp30ebbrE7.a1:
  • wind direction (relative to true north)(wind_d)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • soil heat flow, site 2(g2)
  • top air temperature(tair_top)
  • average surface soil heat flow(ave_shf)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • vector wind speed(res_ws)
  • volumetric soil moisture, site 5(sm5)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • bottom air temperature(tair_bot)
  • volumetric soil moisture, site 4(sm4)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • volumetric soil moisture, site 2(sm2)
  • Soil heat capacity 2(cs2)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • volumetric soil moisture, site 1(sm1)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • soil heat flow, site 4(g4)
  • soil heat flow, site 3(g3)
  • Temperature of the top humidity chamber(thum_top)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • Soil heat capacity 5(cs5)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • Soil heat capacity 3(cs3)
  • Retrieved pressure profile(pres)
  • Reference Thermistor Temperature(tref)
  • Soil heat capacity 1(cs1)
  • 5 cm soil heat flow, site 5(shf5)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • net radiation(q)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • Soil heat capacity 4(cs4)
  • 5 cm soil heat flow, site 1(shf1)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • bottom vapor pressure(vp_bot)
  • h(h)
  • 5 cm soil heat flow, site 4(shf4)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • soil heat flow, site 1(g1)
  • Top humidity(hum_top)
  • soil heat flow, site 5(g5)
  • top vapor pressure(vp_top)
  • scalar wind speed(wind_s)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • 5 cm soil heat flow, site 2(shf2)
  • 5 cm soil heat flow, site 3(shf3)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • Bottom humidity(hum_bot)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • volumetric soil moisture, site 3(sm3)

sgp15ebbrE7.a1:
  • Atmospheric pressure(mv_pres)
  • Soil moisture 5(r_sm5)
  • Soil temperature 5(rr_ts5)
  • Left relative humidity(mv_hum_l)
  • Soil heat flow 5(mv_hft5)
  • Soil moisture 2(r_sm2)
  • Soil temperature 4(rr_ts4)
  • Soil heat flow 4(mv_hft4)
  • Soil moisture 3(r_sm3)
  • Wind direction (relative to true north)(mv_wind_d)
  • Left air temperature(tair_l)
  • Right air temperature(tair_r)
  • Soil moisture 2(rr_sm2)
  • Soil moisture 5(rr_sm5)
  • Soil moisture 1(r_sm1)
  • Soil moisture 3(rr_sm3)
  • Soil heat flow 1(mv_hft1)
  • Soil moisture 4(r_sm4)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil moisture 4(rr_sm4)
  • Right relative humidity(mv_hum_r)
  • Soil temperature 2(rr_ts2)
  • Soil moisture 1(rr_sm1)
  • Soil temperature 3(rr_ts3)
  • Soil heat flow 2(mv_hft2)
  • Soil heat flow 3(mv_hft3)
  • Net radiation(mv_q)
  • Home signal(mv_home)
  • Soil temperature 1(rr_ts1)
  • Reference temperature(rr_tref)
  • scalar wind speed(wind_s)

sgp15ebbrE7.b1:
  • Net radiation(mv_q)
  • Soil moisture 2(r_sm2)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Soil moisture 3(r_sm3)
  • Soil heat flow 3(mv_hft3)
  • Soil heat flow 4(mv_hft4)
  • Soil moisture 5(r_sm5)
  • Home signal(mv_home)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil heat flow 5(mv_hft5)
  • Reference temperature(rr_tref)
  • Left relative humidity(mv_hum_l)
  • scalar wind speed(wind_s)
  • Soil heat flow 1(mv_hft1)
  • Right relative humidity(mv_hum_r)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil temperature 4(rr_ts4)
  • Left air temperature(tair_l)
  • Soil temperature 3(rr_ts3)
  • Soil moisture 4(r_sm4)
  • Soil temperature 5(rr_ts5)
  • Soil moisture 1(r_sm1)
  • Soil heat flow 2(mv_hft2)
  • Right air temperature(tair_r)
  • Soil temperature 1(rr_ts1)
  • Atmospheric pressure(mv_pres)
  • Soil temperature 2(rr_ts2)

sgp30ebbrE7.b1:
  • volumetric soil moisture, site 3(sm3)
  • wind direction (relative to true north)(wind_d)
  • volumetric soil moisture, site 2(sm2)
  • bottom air temperature(tair_bot)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • volumetric soil moisture, site 4(sm4)
  • average surface soil heat flow(ave_shf)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • 5 cm soil heat flow, site 4(shf4)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • volumetric soil moisture, site 5(sm5)
  • Soil heat capacity 2(cs2)
  • 5 cm soil heat flow, site 3(shf3)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • top air temperature(tair_top)
  • volumetric soil moisture, site 1(sm1)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • top vapor pressure(vp_top)
  • scalar wind speed(wind_s)
  • latent heat flux(e)
  • vector wind speed(res_ws)
  • Top humidity(hum_top)
  • bottom vapor pressure(vp_bot)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • soil heat flow, site 3(g3)
  • 5 cm soil heat flow, site 5(shf5)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • soil heat flow, site 2(g2)
  • Soil heat capacity 1(cs1)
  • Bottom humidity(hum_bot)
  • Reference Thermistor Temperature(tref)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • h(h)
  • Soil heat capacity 4(cs4)
  • Temperature of the top humidity chamber(thum_top)
  • Soil heat capacity 5(cs5)
  • Retrieved pressure profile(pres)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • net radiation(q)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • soil heat flow, site 1(g1)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • soil heat flow, site 4(g4)
  • Soil heat capacity 3(cs3)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • soil heat flow, site 5(g5)
  • 5 cm soil heat flow, site 2(shf2)
  • 5 cm soil heat flow, site 1(shf1)


Back To Table of Contents

DQRID : D011129.8
Start DateStart TimeEnd DateEnd Time
05/07/2001193005/07/20011931
Subject:
SGP EBBR E7 - E7 EBBR Data Incorrect After Recalibration
DataStreams:sgp15ebbrE7.a1, sgp15ebbrE7.b1, sgp30ebbrE7.a1, sgp30ebbrE7.b1
Description:
Soil temperature measurements and the half hour average were incorrect the half hour after 
starting the recalibrated system, causing other measurements and calculated quantities 
to be incorrect.
Measurements:sgp15ebbrE7.a1:
  • Soil temperature 5(rr_ts5)
  • Soil temperature 4(rr_ts4)
  • Soil temperature 2(rr_ts2)
  • Soil temperature 3(rr_ts3)
  • Soil temperature 1(rr_ts1)

sgp30ebbrE7.a1:
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • soil heat flow, site 2(g2)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • soil heat flow, site 4(g4)
  • h(h)
  • soil heat flow, site 3(g3)
  • soil heat flow, site 1(g1)
  • soil heat flow, site 5(g5)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • 0-5 cm integrated soil temperature, site 2(ts1)

sgp15ebbrE7.b1:
  • Soil temperature 4(rr_ts4)
  • Soil temperature 3(rr_ts3)
  • Soil temperature 5(rr_ts5)
  • Soil temperature 1(rr_ts1)
  • Soil temperature 2(rr_ts2)

sgp30ebbrE7.b1:
  • volumetric soil moisture, site 3(sm3)
  • volumetric soil moisture, site 2(sm2)
  • soil heat flow, site 3(g3)
  • soil heat flow, site 2(g2)
  • volumetric soil moisture, site 4(sm4)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • h(h)
  • volumetric soil moisture, site 5(sm5)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • volumetric soil moisture, site 1(sm1)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • soil heat flow, site 1(g1)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • latent heat flux(e)
  • soil heat flow, site 5(g5)


Back To Table of Contents

DQRID : D011129.9
Start DateStart TimeEnd DateEnd Time
08/28/2001190008/28/20011901
Subject:
SGP EBBR E7 - E7 EBBR Data Incorrect After Missing Data Period
DataStreams:sgp15ebbrE7.a1, sgp15ebbrE7.b1, sgp30ebbrE7.a1, sgp30ebbrE7.b1
Description:
Soil temperature measurements and the half hour average were incorrect the half hour after 
a missing data period, causing other measurements and calculated quantities to be 
incorrect.
Measurements:sgp15ebbrE7.a1:
  • Soil temperature 5(rr_ts5)
  • Soil temperature 4(rr_ts4)
  • Soil temperature 2(rr_ts2)
  • Soil temperature 3(rr_ts3)
  • Soil temperature 1(rr_ts1)

sgp30ebbrE7.a1:
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • soil heat flow, site 2(g2)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • soil heat flow, site 4(g4)
  • h(h)
  • soil heat flow, site 3(g3)
  • soil heat flow, site 1(g1)
  • soil heat flow, site 5(g5)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • 0-5 cm integrated soil temperature, site 2(ts1)

sgp15ebbrE7.b1:
  • Soil temperature 4(rr_ts4)
  • Soil temperature 3(rr_ts3)
  • Soil temperature 5(rr_ts5)
  • Soil temperature 1(rr_ts1)
  • Soil temperature 2(rr_ts2)

sgp30ebbrE7.b1:
  • volumetric soil moisture, site 3(sm3)
  • volumetric soil moisture, site 2(sm2)
  • soil heat flow, site 3(g3)
  • soil heat flow, site 2(g2)
  • volumetric soil moisture, site 4(sm4)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • h(h)
  • volumetric soil moisture, site 5(sm5)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • volumetric soil moisture, site 1(sm1)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • soil heat flow, site 1(g1)
  • soil heat flow, site 4(g4)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • latent heat flux(e)
  • soil heat flow, site 5(g5)


Back To Table of Contents

DQRID : D020716.26
Start DateStart TimeEnd DateEnd Time
03/12/2002210003/12/20022101
Subject:
SGP/EBBR/E7 - Measurements Zero
DataStreams:sgp15ebbrE7.a1, sgp15ebbrE7.b1, sgp30ebbrE7.a1, sgp30ebbrE7.b1, sgp5ebbrE7.00,
sgp5ebbrE7.a1
Description:
The indicated measurements were zero after interruption of the CR10 program.
Measurements:sgp5ebbrE7.00:
  • null(Raw data stream - documentation not supported)

sgp5ebbrE7.a1:
  • Temperature of the top humidity chamber(thum_top)
  • Reference Thermistor Temperature(tref)
  • Top humidity(hum_top)
  • Retrieved pressure profile(pres)
  • vector wind speed(res_ws)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • net radiation(q)
  • Bottom humidity(hum_bot)
  • bottom vapor pressure(vp_bot)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • top vapor pressure(vp_top)
  • wind direction (relative to true north)(wind_d)
  • Home signal(home)
  • top air temperature(tair_top)
  • scalar wind speed(wind_s)
  • bottom air temperature(tair_bot)

sgp30ebbrE7.a1:
  • wind direction (relative to true north)(wind_d)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • soil heat flow, site 2(g2)
  • top air temperature(tair_top)
  • average surface soil heat flow(ave_shf)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • vector wind speed(res_ws)
  • volumetric soil moisture, site 5(sm5)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • bottom air temperature(tair_bot)
  • volumetric soil moisture, site 4(sm4)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • volumetric soil moisture, site 2(sm2)
  • Soil heat capacity 2(cs2)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • volumetric soil moisture, site 1(sm1)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • soil heat flow, site 4(g4)
  • soil heat flow, site 3(g3)
  • Temperature of the top humidity chamber(thum_top)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • Soil heat capacity 5(cs5)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • Soil heat capacity 3(cs3)
  • Retrieved pressure profile(pres)
  • Reference Thermistor Temperature(tref)
  • Soil heat capacity 1(cs1)
  • 5 cm soil heat flow, site 5(shf5)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • net radiation(q)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • Soil heat capacity 4(cs4)
  • 5 cm soil heat flow, site 1(shf1)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • bottom vapor pressure(vp_bot)
  • h(h)
  • 5 cm soil heat flow, site 4(shf4)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • soil heat flow, site 1(g1)
  • Top humidity(hum_top)
  • soil heat flow, site 5(g5)
  • top vapor pressure(vp_top)
  • scalar wind speed(wind_s)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • 5 cm soil heat flow, site 2(shf2)
  • 5 cm soil heat flow, site 3(shf3)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • Bottom humidity(hum_bot)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • volumetric soil moisture, site 3(sm3)

sgp15ebbrE7.a1:
  • Atmospheric pressure(mv_pres)
  • Soil moisture 5(r_sm5)
  • Soil temperature 5(rr_ts5)
  • Left relative humidity(mv_hum_l)
  • Soil heat flow 5(mv_hft5)
  • Soil moisture 2(r_sm2)
  • Soil temperature 4(rr_ts4)
  • Soil heat flow 4(mv_hft4)
  • Soil moisture 3(r_sm3)
  • Battery(bat)
  • Wind direction (relative to true north)(mv_wind_d)
  • Left air temperature(tair_l)
  • Right air temperature(tair_r)
  • Soil moisture 2(rr_sm2)
  • Soil moisture 5(rr_sm5)
  • Soil moisture 1(r_sm1)
  • Soil moisture 3(rr_sm3)
  • Soil heat flow 1(mv_hft1)
  • Soil moisture 4(r_sm4)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil moisture 4(rr_sm4)
  • Right relative humidity(mv_hum_r)
  • Soil temperature 2(rr_ts2)
  • Soil moisture 1(rr_sm1)
  • Soil temperature 3(rr_ts3)
  • Soil heat flow 2(mv_hft2)
  • Soil heat flow 3(mv_hft3)
  • Net radiation(mv_q)
  • Home signal(mv_home)
  • Soil temperature 1(rr_ts1)
  • Reference temperature(rr_tref)
  • scalar wind speed(wind_s)

sgp15ebbrE7.b1:
  • Net radiation(mv_q)
  • Soil moisture 2(r_sm2)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Soil moisture 3(r_sm3)
  • Soil heat flow 3(mv_hft3)
  • Soil heat flow 4(mv_hft4)
  • Soil moisture 5(r_sm5)
  • Home signal(mv_home)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil heat flow 5(mv_hft5)
  • Reference temperature(rr_tref)
  • Left relative humidity(mv_hum_l)
  • scalar wind speed(wind_s)
  • Battery(bat)
  • Soil heat flow 1(mv_hft1)
  • Right relative humidity(mv_hum_r)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil temperature 4(rr_ts4)
  • Left air temperature(tair_l)
  • Soil temperature 3(rr_ts3)
  • Soil moisture 4(r_sm4)
  • Soil temperature 5(rr_ts5)
  • Soil moisture 1(r_sm1)
  • Soil heat flow 2(mv_hft2)
  • Right air temperature(tair_r)
  • Soil temperature 1(rr_ts1)
  • Atmospheric pressure(mv_pres)
  • Soil temperature 2(rr_ts2)

sgp30ebbrE7.b1:
  • volumetric soil moisture, site 3(sm3)
  • wind direction (relative to true north)(wind_d)
  • volumetric soil moisture, site 2(sm2)
  • bottom air temperature(tair_bot)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • volumetric soil moisture, site 4(sm4)
  • average surface soil heat flow(ave_shf)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • 5 cm soil heat flow, site 4(shf4)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • volumetric soil moisture, site 5(sm5)
  • Soil heat capacity 2(cs2)
  • 5 cm soil heat flow, site 3(shf3)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • top air temperature(tair_top)
  • volumetric soil moisture, site 1(sm1)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • top vapor pressure(vp_top)
  • scalar wind speed(wind_s)
  • latent heat flux(e)
  • vector wind speed(res_ws)
  • Top humidity(hum_top)
  • bottom vapor pressure(vp_bot)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • soil heat flow, site 3(g3)
  • 5 cm soil heat flow, site 5(shf5)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • soil heat flow, site 2(g2)
  • Soil heat capacity 1(cs1)
  • Bottom humidity(hum_bot)
  • Reference Thermistor Temperature(tref)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • Soil heat capacity 4(cs4)
  • h(h)
  • Temperature of the top humidity chamber(thum_top)
  • Soil heat capacity 5(cs5)
  • Retrieved pressure profile(pres)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • net radiation(q)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • soil heat flow, site 1(g1)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • soil heat flow, site 4(g4)
  • Soil heat capacity 3(cs3)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • soil heat flow, site 5(g5)
  • 5 cm soil heat flow, site 2(shf2)
  • 5 cm soil heat flow, site 1(shf1)


Back To Table of Contents

DQRID : D020716.27
Start DateStart TimeEnd DateEnd Time
04/14/2002220004/23/20021531
05/20/2002120006/06/20021501
Subject:
SGP/EBBR/E7 - AEM Hung
DataStreams:sgp15ebbrE7.a1, sgp15ebbrE7.b1, sgp30ebbrE7.a1, sgp30ebbrE7.b1, sgp5ebbrE7.a1
Description:
The indicated measurements were incorrect because the automatic exchange mechanism hung in 
one position.
Measurements:sgp5ebbrE7.a1:
  • top air temperature(tair_top)
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • top vapor pressure(vp_top)
  • Top humidity(hum_top)
  • bottom air temperature(tair_bot)
  • Bottom humidity(hum_bot)

sgp15ebbrE7.a1:
  • Left air temperature(tair_l)
  • Right air temperature(tair_r)

sgp30ebbrE7.a1:
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • top air temperature(tair_top)
  • Top humidity(hum_top)
  • top vapor pressure(vp_top)
  • bottom air temperature(tair_bot)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • bottom vapor pressure(vp_bot)
  • latent heat flux(e)
  • h(h)
  • Bottom humidity(hum_bot)
  • Temperature of the top humidity chamber(thum_top)

sgp15ebbrE7.b1:
  • Right air temperature(tair_r)
  • Left air temperature(tair_l)

sgp30ebbrE7.b1:
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • bottom vapor pressure(vp_bot)
  • Top humidity(hum_top)
  • top air temperature(tair_top)
  • bottom air temperature(tair_bot)
  • Bottom humidity(hum_bot)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • h(h)
  • top vapor pressure(vp_top)
  • latent heat flux(e)
  • Temperature of the top humidity chamber(thum_top)


Back To Table of Contents

DQRID : D020716.28
Start DateStart TimeEnd DateEnd Time
05/20/2002120005/21/20021601
Subject:
SGP/EBBR/E7 - Wind Speed Incorrect
DataStreams:sgp15ebbrE7.a1, sgp15ebbrE7.b1, sgp30ebbrE7.a1, sgp30ebbrE7.b1, sgp5ebbrE7.a1
Description:
The wind speed and resultant wind speed are incorrect for the indicated times because a 
cup had broken off of the cup anemometer.
Measurements:sgp5ebbrE7.a1:
  • vector wind speed(res_ws)
  • scalar wind speed(wind_s)

sgp15ebbrE7.a1:
  • scalar wind speed(wind_s)

sgp30ebbrE7.a1:
  • scalar wind speed(wind_s)
  • vector wind speed(res_ws)

sgp30ebbrE7.b1:
  • vector wind speed(res_ws)
  • scalar wind speed(wind_s)

sgp15ebbrE7.b1:
  • scalar wind speed(wind_s)


Back To Table of Contents

DQRID : D020716.8
Start DateStart TimeEnd DateEnd Time
05/07/2002164505/07/20021716
Subject:
SGP/EBBR/E7 - 6 Month Checks
DataStreams:sgp15ebbrE7.a1, sgp15ebbrE7.b1, sgp30ebbrE7.a1, sgp30ebbrE7.b1, sgp5ebbrE7.a1
Description:
The indicated measurements were incorrect during the 6 month checks.
Measurements:sgp5ebbrE7.a1:
  • top air temperature(tair_top)
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • top vapor pressure(vp_top)
  • Top humidity(hum_top)
  • bottom air temperature(tair_bot)
  • Bottom humidity(hum_bot)

sgp15ebbrE7.a1:
  • Left air temperature(tair_l)
  • Right air temperature(tair_r)

sgp30ebbrE7.a1:
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • top air temperature(tair_top)
  • Top humidity(hum_top)
  • top vapor pressure(vp_top)
  • bottom air temperature(tair_bot)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • bottom vapor pressure(vp_bot)
  • latent heat flux(e)
  • h(h)
  • Bottom humidity(hum_bot)
  • Temperature of the top humidity chamber(thum_top)

sgp15ebbrE7.b1:
  • Right air temperature(tair_r)
  • Left air temperature(tair_l)

sgp30ebbrE7.b1:
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • bottom vapor pressure(vp_bot)
  • Top humidity(hum_top)
  • top air temperature(tair_top)
  • bottom air temperature(tair_bot)
  • Bottom humidity(hum_bot)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • h(h)
  • top vapor pressure(vp_top)
  • latent heat flux(e)
  • Temperature of the top humidity chamber(thum_top)


Back To Table of Contents

DQRID : D030110.30
Start DateStart TimeEnd DateEnd Time
11/19/2002200011/19/20022030
Subject:
SGP/EBBR/E7 - Data Incorrect during 6 month checks
DataStreams:sgp15ebbrE7.a1, sgp15ebbrE7.b1, sgp30ebbrE7.a1, sgp30ebbrE7.b1, sgp5ebbrE7.a1
Description:
Some of the measurements are incorrect during the 6 month checks.
Measurements:sgp5ebbrE7.a1:
  • top air temperature(tair_top)
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • top vapor pressure(vp_top)
  • Top humidity(hum_top)
  • bottom air temperature(tair_bot)
  • Bottom humidity(hum_bot)

sgp15ebbrE7.a1:
  • Left air temperature(tair_l)
  • Right air temperature(tair_r)

sgp30ebbrE7.a1:
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • top air temperature(tair_top)
  • Top humidity(hum_top)
  • top vapor pressure(vp_top)
  • bottom air temperature(tair_bot)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • bottom vapor pressure(vp_bot)
  • latent heat flux(e)
  • h(h)
  • Bottom humidity(hum_bot)
  • Temperature of the top humidity chamber(thum_top)

sgp15ebbrE7.b1:
  • Right air temperature(tair_r)
  • Left air temperature(tair_l)

sgp30ebbrE7.b1:
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • bottom vapor pressure(vp_bot)
  • Top humidity(hum_top)
  • top air temperature(tair_top)
  • bottom air temperature(tair_bot)
  • Bottom humidity(hum_bot)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • h(h)
  • top vapor pressure(vp_top)
  • latent heat flux(e)
  • Temperature of the top humidity chamber(thum_top)


Back To Table of Contents

DQRID : D030113.1
Start DateStart TimeEnd DateEnd Time
07/11/2002210007/18/20021700
Subject:
SGP/EBBR/E7 - Wind speed incorrect
DataStreams:sgp15ebbrE7.a1, sgp15ebbrE7.b1, sgp30ebbrE7.a1, sgp30ebbrE7.b1, sgp5ebbrE7.a1
Description:
A wind speed sensor electronics problem limited the
measurement to the sensor offset.  This data was incorrect.
Measurements:sgp5ebbrE7.a1:
  • vector wind speed(res_ws)
  • scalar wind speed(wind_s)

sgp15ebbrE7.a1:
  • scalar wind speed(wind_s)

sgp30ebbrE7.a1:
  • scalar wind speed(wind_s)
  • vector wind speed(res_ws)

sgp30ebbrE7.b1:
  • vector wind speed(res_ws)
  • scalar wind speed(wind_s)

sgp15ebbrE7.b1:
  • scalar wind speed(wind_s)


Back To Table of Contents

DQRID : D030113.7
Start DateStart TimeEnd DateEnd Time
07/10/2002080007/16/20021900
07/22/2002163007/30/20021630
08/24/2002023008/27/20021800
Subject:
SGP/EBBR/E7 - AEM Problems
DataStreams:sgp15ebbrE7.a1, sgp15ebbrE7.b1, sgp30ebbrE7.a1, sgp30ebbrE7.b1, sgp5ebbrE7.a1
Description:
The AEM (Automatic Exchange Mechanism) hung in a non-home position for
a few lengthy periods, during which the indicated measurements are incorrect.
Measurements:sgp5ebbrE7.a1:
  • top air temperature(tair_top)
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • top vapor pressure(vp_top)
  • Top humidity(hum_top)
  • bottom air temperature(tair_bot)
  • Bottom humidity(hum_bot)

sgp15ebbrE7.a1:
  • Left air temperature(tair_l)
  • Right air temperature(tair_r)

sgp30ebbrE7.a1:
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • top air temperature(tair_top)
  • Top humidity(hum_top)
  • top vapor pressure(vp_top)
  • bottom air temperature(tair_bot)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • bottom vapor pressure(vp_bot)
  • latent heat flux(e)
  • h(h)
  • Bottom humidity(hum_bot)
  • Temperature of the top humidity chamber(thum_top)

sgp15ebbrE7.b1:
  • Right air temperature(tair_r)
  • Left air temperature(tair_l)

sgp30ebbrE7.b1:
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • bottom vapor pressure(vp_bot)
  • Top humidity(hum_top)
  • top air temperature(tair_top)
  • bottom air temperature(tair_bot)
  • Bottom humidity(hum_bot)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • h(h)
  • top vapor pressure(vp_top)
  • latent heat flux(e)
  • Temperature of the top humidity chamber(thum_top)


Back To Table of Contents

DQRID : D030113.8
Start DateStart TimeEnd DateEnd Time
10/01/2002170010/08/20021645
Subject:
SGP/EBBR/E7 - T/RH Ventilator Fans Stopped
DataStreams:sgp15ebbrE7.a1, sgp15ebbrE7.b1, sgp30ebbrE7.a1, sgp30ebbrE7.b1, sgp5ebbrE7.a1
Description:
The temperature and relative humidity gradients are suspect because the 
ventilator fans were not operating during this time period. The data needs
to be carefully inspected to determine when measurements, if any, are correct.
Measurements:sgp5ebbrE7.a1:
  • top air temperature(tair_top)
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • top vapor pressure(vp_top)
  • Top humidity(hum_top)
  • bottom air temperature(tair_bot)
  • Bottom humidity(hum_bot)

sgp15ebbrE7.a1:
  • Left air temperature(tair_l)
  • Right air temperature(tair_r)

sgp30ebbrE7.a1:
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • top air temperature(tair_top)
  • Top humidity(hum_top)
  • top vapor pressure(vp_top)
  • bottom air temperature(tair_bot)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • bottom vapor pressure(vp_bot)
  • latent heat flux(e)
  • h(h)
  • Bottom humidity(hum_bot)
  • Temperature of the top humidity chamber(thum_top)

sgp15ebbrE7.b1:
  • Right air temperature(tair_r)
  • Left air temperature(tair_l)

sgp30ebbrE7.b1:
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • bottom vapor pressure(vp_bot)
  • Top humidity(hum_top)
  • top air temperature(tair_top)
  • bottom air temperature(tair_bot)
  • Bottom humidity(hum_bot)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • h(h)
  • top vapor pressure(vp_top)
  • latent heat flux(e)
  • Temperature of the top humidity chamber(thum_top)


Back To Table of Contents

DQRID : D030114.2
Start DateStart TimeEnd DateEnd Time
07/30/2002180007/30/20021829
Subject:
SGP/EBBR/E7 - Data Reported as Zero following Logger Replacement
DataStreams:sgp15ebbrE7.a1, sgp15ebbrE7.b1, sgp30ebbrE7.a1, sgp30ebbrE7.b1, sgp5ebbrE7.a1
Description:
The data was reported as zero after a data logger replacement.
Measurements:sgp5ebbrE7.a1:
  • Temperature of the top humidity chamber(thum_top)
  • Reference Thermistor Temperature(tref)
  • Top humidity(hum_top)
  • Retrieved pressure profile(pres)
  • vector wind speed(res_ws)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • net radiation(q)
  • Bottom humidity(hum_bot)
  • bottom vapor pressure(vp_bot)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • top vapor pressure(vp_top)
  • wind direction (relative to true north)(wind_d)
  • Home signal(home)
  • top air temperature(tair_top)
  • scalar wind speed(wind_s)
  • bottom air temperature(tair_bot)

sgp30ebbrE7.a1:
  • wind direction (relative to true north)(wind_d)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • soil heat flow, site 2(g2)
  • top air temperature(tair_top)
  • average surface soil heat flow(ave_shf)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • vector wind speed(res_ws)
  • volumetric soil moisture, site 5(sm5)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • bottom air temperature(tair_bot)
  • volumetric soil moisture, site 4(sm4)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • volumetric soil moisture, site 2(sm2)
  • Soil heat capacity 2(cs2)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • volumetric soil moisture, site 1(sm1)
  • latent heat flux(e)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • soil heat flow, site 4(g4)
  • soil heat flow, site 3(g3)
  • Temperature of the top humidity chamber(thum_top)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • Soil heat capacity 5(cs5)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • Soil heat capacity 3(cs3)
  • Retrieved pressure profile(pres)
  • Reference Thermistor Temperature(tref)
  • Soil heat capacity 1(cs1)
  • 5 cm soil heat flow, site 5(shf5)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • net radiation(q)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • Soil heat capacity 4(cs4)
  • 5 cm soil heat flow, site 1(shf1)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • bottom vapor pressure(vp_bot)
  • h(h)
  • 5 cm soil heat flow, site 4(shf4)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • soil heat flow, site 1(g1)
  • Top humidity(hum_top)
  • soil heat flow, site 5(g5)
  • top vapor pressure(vp_top)
  • scalar wind speed(wind_s)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • 5 cm soil heat flow, site 2(shf2)
  • 5 cm soil heat flow, site 3(shf3)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • Bottom humidity(hum_bot)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • volumetric soil moisture, site 3(sm3)

sgp15ebbrE7.a1:
  • Atmospheric pressure(mv_pres)
  • Soil moisture 5(r_sm5)
  • Soil temperature 5(rr_ts5)
  • Left relative humidity(mv_hum_l)
  • Soil heat flow 5(mv_hft5)
  • Soil moisture 2(r_sm2)
  • Soil temperature 4(rr_ts4)
  • Soil heat flow 4(mv_hft4)
  • Soil moisture 3(r_sm3)
  • Battery(bat)
  • Wind direction (relative to true north)(mv_wind_d)
  • Left air temperature(tair_l)
  • Right air temperature(tair_r)
  • Soil moisture 2(rr_sm2)
  • Soil moisture 5(rr_sm5)
  • Soil moisture 1(r_sm1)
  • Soil moisture 3(rr_sm3)
  • Soil heat flow 1(mv_hft1)
  • Soil moisture 4(r_sm4)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil moisture 4(rr_sm4)
  • Right relative humidity(mv_hum_r)
  • Soil temperature 2(rr_ts2)
  • Soil moisture 1(rr_sm1)
  • Soil temperature 3(rr_ts3)
  • Soil heat flow 2(mv_hft2)
  • Soil heat flow 3(mv_hft3)
  • Net radiation(mv_q)
  • Home signal(mv_home)
  • Soil temperature 1(rr_ts1)
  • Reference temperature(rr_tref)
  • scalar wind speed(wind_s)

sgp15ebbrE7.b1:
  • Net radiation(mv_q)
  • Soil moisture 2(r_sm2)
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Soil moisture 3(r_sm3)
  • Soil heat flow 3(mv_hft3)
  • Soil heat flow 4(mv_hft4)
  • Soil moisture 5(r_sm5)
  • Home signal(mv_home)
  • Wind direction (relative to true north)(mv_wind_d)
  • Soil heat flow 5(mv_hft5)
  • Reference temperature(rr_tref)
  • Left relative humidity(mv_hum_l)
  • scalar wind speed(wind_s)
  • Battery(bat)
  • Soil heat flow 1(mv_hft1)
  • Right relative humidity(mv_hum_r)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Soil temperature 4(rr_ts4)
  • Left air temperature(tair_l)
  • Soil temperature 3(rr_ts3)
  • Soil moisture 4(r_sm4)
  • Soil temperature 5(rr_ts5)
  • Soil moisture 1(r_sm1)
  • Soil heat flow 2(mv_hft2)
  • Right air temperature(tair_r)
  • Soil temperature 1(rr_ts1)
  • Atmospheric pressure(mv_pres)
  • Soil temperature 2(rr_ts2)

sgp30ebbrE7.b1:
  • volumetric soil moisture, site 3(sm3)
  • wind direction (relative to true north)(wind_d)
  • volumetric soil moisture, site 2(sm2)
  • bottom air temperature(tair_bot)
  • Exchange mechanism position indicator (0 to 15 min)(home_15)
  • volumetric soil moisture, site 4(sm4)
  • average surface soil heat flow(ave_shf)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • 5 cm soil heat flow, site 4(shf4)
  • 0-5 cm integrated soil temperature, site 4(ts4)
  • volumetric soil moisture, site 5(sm5)
  • Soil heat capacity 2(cs2)
  • 5 cm soil heat flow, site 3(shf3)
  • 0-5 cm integrated soil temperature, site 5(ts5)
  • top air temperature(tair_top)
  • volumetric soil moisture, site 1(sm1)
  • 0-5 cm change in soil heat storage with time, site 2(ces2)
  • 0-5 cm integrated soil temperature, site 2(ts1)
  • top vapor pressure(vp_top)
  • scalar wind speed(wind_s)
  • latent heat flux(e)
  • vector wind speed(res_ws)
  • Top humidity(hum_top)
  • bottom vapor pressure(vp_bot)
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • standard deviation of wind direction (sigma theta)(sigma_wd)
  • soil heat flow, site 3(g3)
  • 5 cm soil heat flow, site 5(shf5)
  • 5 cm soil heat flow corrected for soil moisture content, site 3(c_shf3)
  • soil heat flow, site 2(g2)
  • Soil heat capacity 1(cs1)
  • Bottom humidity(hum_bot)
  • Reference Thermistor Temperature(tref)
  • 0-5 cm integrated soil temperature, site 2(ts2)
  • Soil heat capacity 4(cs4)
  • h(h)
  • Temperature of the top humidity chamber(thum_top)
  • Soil heat capacity 5(cs5)
  • Retrieved pressure profile(pres)
  • 0-5 cm change in soil heat storage with time, site 5(ces5)
  • Exchange mechanism position indicator (15 to 30 min)(home_30)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • 0-5 cm change in soil heat storage with time, site 4(ces4)
  • net radiation(q)
  • 0-5 cm change in soil heat storage with time, site 2(ces1)
  • 0-5 cm integrated soil temperature, site 3(ts3)
  • soil heat flow, site 1(g1)
  • 5 cm soil heat flow corrected for soil moisture content, site 1(c_shf1)
  • soil heat flow, site 4(g4)
  • Soil heat capacity 3(cs3)
  • 0-5 cm change in soil heat storage with time, site 3(ces3)
  • 5 cm soil heat flow corrected for soil moisture content, site 2(c_shf2)
  • 5 cm soil heat flow corrected for soil moisture content, site 4(c_shf4)
  • soil heat flow, site 5(g5)
  • 5 cm soil heat flow, site 2(shf2)
  • 5 cm soil heat flow, site 1(shf1)


Back To Table of Contents

DQRID : D030620.1
Start DateStart TimeEnd DateEnd Time
06/11/2003173007/15/20032000
Subject:
SGP/EBBR/E7 - Soil Heat Flow #5 Incorrect
DataStreams:sgp15ebbrE7.a1, sgp15ebbrE7.b1, sgp30ebbrE7.a1, sgp30ebbrE7.b1
Description:
Soil Heat Flow Plate #5 data was intermittently incorrect at first 
(values too small negative), then was almost constantly so during the 
last half of the period.

Comparison of the individual soil heat flows will quickly show the times
that soil heat flow #5 was incorrect.
Measurements:sgp15ebbrE7.a1:
  • Soil heat flow 5(mv_hft5)

sgp30ebbrE7.a1:
  • average surface soil heat flow(ave_shf)
  • soil heat flow, site 5(g5)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • latent heat flux(e)
  • 5 cm soil heat flow, site 5(shf5)
  • h(h)

sgp15ebbrE7.b1:
  • Soil heat flow 5(mv_hft5)

sgp30ebbrE7.b1:
  • average surface soil heat flow(ave_shf)
  • 5 cm soil heat flow corrected for soil moisture content, site 5(c_shf5)
  • h(h)
  • latent heat flux(e)
  • 5 cm soil heat flow, site 5(shf5)
  • soil heat flow, site 5(g5)


Back To Table of Contents

DQRID : D030712.13
Start DateStart TimeEnd DateEnd Time
04/22/2003165004/22/20031720
Subject:
SGP/EBBR/E7  - 6 Month Checks
DataStreams:sgp15ebbrE7.a1, sgp15ebbrE7.b1, sgp30ebbrE7.a1, sgp30ebbrE7.b1, sgp5ebbrE7.a1
Description:
Many measurements were incorrect while the AEM was positioned with both aspirators at the 
same height for comparison checks.
Measurements:sgp5ebbrE7.a1:
  • top air temperature(tair_top)
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • top vapor pressure(vp_top)
  • Top humidity(hum_top)
  • bottom air temperature(tair_bot)
  • Bottom humidity(hum_bot)

sgp15ebbrE7.a1:
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Left relative humidity(mv_hum_l)
  • Left air temperature(tair_l)
  • Right air temperature(tair_r)
  • Right relative humidity(mv_hum_r)

sgp30ebbrE7.a1:
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • top air temperature(tair_top)
  • Top humidity(hum_top)
  • top vapor pressure(vp_top)
  • bottom air temperature(tair_bot)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • bottom vapor pressure(vp_bot)
  • latent heat flux(e)
  • h(h)
  • Bottom humidity(hum_bot)
  • Temperature of the top humidity chamber(thum_top)

sgp15ebbrE7.b1:
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Right relative humidity(mv_hum_r)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Left relative humidity(mv_hum_l)
  • Right air temperature(tair_r)
  • Left air temperature(tair_l)

sgp30ebbrE7.b1:
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • bottom vapor pressure(vp_bot)
  • Top humidity(hum_top)
  • top air temperature(tair_top)
  • bottom air temperature(tair_bot)
  • Bottom humidity(hum_bot)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • h(h)
  • top vapor pressure(vp_top)
  • latent heat flux(e)
  • Temperature of the top humidity chamber(thum_top)


Back To Table of Contents

DQRID : D030712.9
Start DateStart TimeEnd DateEnd Time
02/24/2003090003/11/20031800
Subject:
SGP/EBBR/E7  - AEM Hung Inbetween Home Position
DataStreams:sgp15ebbrE7.a1, sgp15ebbrE7.b1, sgp30ebbrE7.a1, sgp30ebbrE7.b1, sgp5ebbrE7.a1
Description:
The AEM was hung inbetween home positions, causing many measurements to be incorrect.
Measurements:sgp5ebbrE7.a1:
  • top air temperature(tair_top)
  • bottom vapor pressure(vp_bot)
  • Temperature of the top humidity chamber(thum_top)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • top vapor pressure(vp_top)
  • Top humidity(hum_top)
  • bottom air temperature(tair_bot)
  • Bottom humidity(hum_bot)

sgp15ebbrE7.a1:
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Left relative humidity(mv_hum_l)
  • Left air temperature(tair_l)
  • Right air temperature(tair_r)
  • Right relative humidity(mv_hum_r)

sgp30ebbrE7.a1:
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • top air temperature(tair_top)
  • Top humidity(hum_top)
  • top vapor pressure(vp_top)
  • bottom air temperature(tair_bot)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • bottom vapor pressure(vp_bot)
  • latent heat flux(e)
  • h(h)
  • Bottom humidity(hum_bot)
  • Temperature of the top humidity chamber(thum_top)

sgp15ebbrE7.b1:
  • Temperature of left humidity sensor chamber(rr_thum_l)
  • Right relative humidity(mv_hum_r)
  • Temperature of right humidity sensor chamber(rr_thum_r)
  • Left relative humidity(mv_hum_l)
  • Right air temperature(tair_r)
  • Left air temperature(tair_l)

sgp30ebbrE7.b1:
  • ratio of sensible/latent heat fluxes (Bowen ratio)(bowen)
  • bottom vapor pressure(vp_bot)
  • Top humidity(hum_top)
  • top air temperature(tair_top)
  • bottom air temperature(tair_bot)
  • Bottom humidity(hum_bot)
  • Temperature of bottom humidity sensor chamber(thum_bot)
  • h(h)
  • top vapor pressure(vp_top)
  • latent heat flux(e)
  • Temperature of the top humidity chamber(thum_top)


Back To Table of Contents

DQRID : D031211.6
Start DateStart TimeEnd DateEnd Time
03/07/2003150010/12/20032355
Subject:
SGP/EBBR/E7 - Pressure Sensor Data Incorrect
DataStreams:sgp15ebbrE7.b1, sgp30ebbrE7.b1, sgp5ebbrE7.a1
Description:
The E7 pressure sensor suffered
from spikes in pressure, which were related to 
increases in voltage during the day as the 
solar panels charged the battery.  These spikes
were not of a magnitude that would affect the 
30 minute surface fluxes.

Replacement of the pressure sensor solved the problem.
Measurements:sgp5ebbrE7.a1:
  • Retrieved pressure profile(pres)

sgp15ebbrE7.b1:
  • Atmospheric pressure(mv_pres)

sgp30ebbrE7.b1:
  • Retrieved pressure profile(pres)


Back To Table of Contents

DQRID : D051115.10
Start DateStart TimeEnd DateEnd Time
04/01/2001000004/08/20022359
Subject:
SGP/EBBR/E25 - metadata errors
DataStreams:sgp15ebbrE25.b1, sgp30ebbrE25.b1
Description:
During the entire span of EBBR.E25 data, the cdf data files contained a few minor global 
metadata errors. These errors do not affect the data values or quality.

The errors include:
* incomplete sensor location information
* incomplete installation date information
* soil moisture documented as "by volume" should have read "gravimetric"
Measurements:sgp30ebbrE25.b1:
  • base time(base_time)

sgp15ebbrE25.b1:
  • base time(base_time)


Back To Table of Contents



END OF DATA