DQR ID | Subject | Data Streams Affected |
---|---|---|
D061214.3 | SGP/IRT/E19 - Reprocess: Calibration error | sgpirt200msE19.a1, sgpirtE19.b1 |
D070529.2 | SGP/IRT/E16 - computer failure | sgpirt200msE16.a1, sgpirtE16.b1 |
D080424.3 | SGP/IRT/E16 - reference temperature incorrect | sgpirt200msE16.a1, sgpirtE16.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/01/2005 | 0000 | 04/27/2007 | 1618 |
Subject: | SGP/IRT/E19 - Reprocess: Calibration error |
DataStreams: | sgpirt200msE19.a1, sgpirtE19.b1 |
Description: | The IRT was measuring ~20K higher sky temperatures than at other nearby extended facilies. The calibration of the instrument was incorrect. The problem was corrected by replacing the IRT. The offset in the calibration will be determined and the data will be corrected. |
Measurements: | sgpirt200msE19.a1:
sgpirtE19.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/14/2006 | 0600 | 04/11/2007 | 1559 |
Subject: | SGP/IRT/E16 - computer failure |
DataStreams: | sgpirt200msE16.a1, sgpirtE16.b1 |
Description: | The serial ports of the computer failed so no data were collected. The problem was solved when the computer was replaced. |
Measurements: | sgpirt200msE16.a1:
sgpirtE16.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/11/2007 | 1600 | 06/19/2008 | 2035 |
Subject: | SGP/IRT/E16 - reference temperature incorrect |
DataStreams: | sgpirt200msE16.a1, sgpirtE16.b1 |
Description: | The internal reference temperature was unrealistically high, ~360K (90C). The ambient temperature setting was checked. The problem was corrected by replacing the instrument. The sky temperature measurement appears to have been unaffected. |
Measurements: | sgpirt200msE16.a1:
sgpirtE16.b1:
|