DQR ID | Subject | Data Streams Affected |
---|---|---|
D020214.7 | SGP/EBBR/E15 - Data Incorrect During 6 Month Checks | sgp15ebbrE15.a1, sgp15ebbrE15.b1, sgp30ebbrE15.a1, sgp30ebbrE15.b1, sgp5ebbrE15.a1 |
D020716.13 | SGP/EBBR/E15 - 6 Month Checks | sgp15ebbrE15.a1, sgp15ebbrE15.b1, sgp30ebbrE15.a1, sgp30ebbrE15.b1, sgp5ebbrE15.a1 |
D021113.1 | SGP/EBBR - Incorrect Units for Soil Heat Capacity (cs1, cs2, cs3, cs4, cs5) in 30EBBR Header | sgp30ebbrE12.b1, sgp30ebbrE13.b1, sgp30ebbrE15.b1, sgp30ebbrE18.b1, sgp30ebbrE19.b1, sgp30ebbrE2.b1, sgp30ebbrE20.b1, sgp30ebbrE22.b1, sgp30ebbrE26.b1, sgp30ebbrE4.b1, sgp30ebbrE7.b1, sgp30ebbrE8.b1, sgp30ebbrE9.b1 |
D030110.35 | SGP/EBBR/E15 - Data Incorrect during 6 month checks | sgp15ebbrE15.a1, sgp15ebbrE15.b1, sgp30ebbrE15.a1, sgp30ebbrE15.b1, sgp5ebbrE15.a1 |
D030115.7 | SGP/EBBR/E15 - REPROCESS: Soil Moisture #1 Offset | sgp30ebbrE15.a1, sgp30ebbrE15.b1 |
D030115.8 | SGP/EBBR/E15 - Data Mostly Zero Following Logger Restart | sgp15ebbrE15.a0, sgp15ebbrE15.a1, sgp30ebbrE15.a1, sgp30ebbrE15.b1, sgp5ebbrE15.a1 |
D030115.9 | SGP/EBBR/E15 - Pressure Low | sgp30ebbrE15.a1, sgp30ebbrE15.b1, sgp5ebbrE15.a1 |
D030712.18 | SGP/EBBR/E15 - 6 Month Checks | sgp15ebbrE15.a1, sgp15ebbrE15.b1, sgp30ebbrE15.a1, sgp30ebbrE15.b1, sgp5ebbrE15.a1 |
D030724.2 | SGP/EBBR/E15 - Wind Direction 10 Degrees Low | sgp15ebbrE15.a1, sgp15ebbrE15.b1, sgp30ebbrE15.a1, sgp30ebbrE15.b1, sgp5ebbrE15.a1 |
D030728.2 | SGP/EBBR/E15 - Top Humidity reported as zero in 30 minute data. | sgp30ebbrE15.a1, sgp30ebbrE15.b1 |
D031031.2 | SGP/EBBR/E15 - Reference Temperature Failed | sgp15ebbrE15.a1, sgp15ebbrE15.b1, sgp30ebbrE15.a1, sgp30ebbrE15.b1, sgp5ebbrE15.a1 |
D031031.6 | SGP/EBBR/E15 - Water In Net Radiometer, Surface Fluxes Incorrect | sgp15ebbrE15.a1, sgp15ebbrE15.b1, sgp30ebbrE15.a1, sgp30ebbrE15.b1, sgp5ebbrE15.a1 |
D031031.9 | SGP/EBBR/E15 - Soil Temperature #5 Incorrect, Surface Fluxes Incorrect | sgp15ebbrE15.a1, sgp15ebbrE15.b1, sgp30ebbrE15.a1, sgp30ebbrE15.b1 |
D031106.1 | SGP/EBBR/E15 - Soil Temperature #5 Incorrect, Surface Fluxes Incorrect | sgp15ebbrE15.b1, sgp30ebbrE15.b1 |
D031110.7 | SGP/EBBR/E15 - AEM Hung Between Home Positions, Surface Fluxes Incorrect | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D031204.11 | SGP/EBBR/E15 - 6 Month Checks | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D031222.18 | SGP/EBBR/E15 - Net Radiometer Iced Over During Storm | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D031222.19 | SGP/EBBR/E15 - Wind Speed Sensor Frozen By Ice Storm | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D040112.1 | SGP/EBBR/E15 - Wind Direction About 130 Degrees Low | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D040227.1 | SGP/EBBR/E15 - Net Radiometer not level | sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D040304.14 | SGP/EBBR/E15 - Wind Speed Frozen Stopped By Ice | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D040507.1 | SGP/EBBR/E15 - Wind Speed Incorrect | sgp15ebbrE15.b1, sgp30baebbrE15.c1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
D041004.4 | SGP/EBBR/E15 - Soil Moisture Coefficients Incorrect | sgp30ebbrE15.b1 |
D050719.10 | SGP/EBBR/E15 - Improved EBBR CR10 Program | sgp30ebbrE15.b1 |
D051111.12 | SGP/EBBR/E15 - metadata corrections | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/13/2001 | 1700 | 11/13/2001 | 1801 |
Subject: | SGP/EBBR/E15 - Data Incorrect During 6 Month Checks |
DataStreams: | sgp15ebbrE15.a1, sgp15ebbrE15.b1, sgp30ebbrE15.a1, sgp30ebbrE15.b1, sgp5ebbrE15.a1 |
Description: | The data was incorrect during the 6 month checks. |
Measurements: | sgp15ebbrE15.b1:
sgp5ebbrE15.a1:
sgp30ebbrE15.b1:
sgp30ebbrE15.a1:
sgp15ebbrE15.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/14/2002 | 1505 | 05/14/2002 | 1546 |
Subject: | SGP/EBBR/E15 - 6 Month Checks |
DataStreams: | sgp15ebbrE15.a1, sgp15ebbrE15.b1, sgp30ebbrE15.a1, sgp30ebbrE15.b1, sgp5ebbrE15.a1 |
Description: | The indicated measurements were incorrect during the 6 month checks. |
Measurements: | sgp15ebbrE15.b1:
sgp5ebbrE15.a1:
sgp30ebbrE15.b1:
sgp30ebbrE15.a1:
sgp15ebbrE15.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/01/2001 | 0000 | 06/18/2003 | 2359 |
Subject: | SGP/EBBR - Incorrect Units for Soil Heat Capacity (cs1, cs2, cs3, cs4, cs5) in 30EBBR |
DataStreams: | sgp30ebbrE12.b1, sgp30ebbrE13.b1, sgp30ebbrE15.b1, sgp30ebbrE18.b1, sgp30ebbrE19.b1, sgp30ebbrE2.b1, sgp30ebbrE20.b1, sgp30ebbrE22.b1, sgp30ebbrE26.b1, sgp30ebbrE4.b1, sgp30ebbrE7.b1, sgp30ebbrE8.b1, sgp30ebbrE9.b1 |
Description: | The units for Soil Heat Capacity (cs1, cs2, cs3, cs4, cs5) in the EBBR 30 minute netcdf header have been incorrect from the beginning of processing the data into netcdf files. The correct units are MJ/m^3/C. This does not affect the data quality. |
Measurements: | sgp30ebbrE8.b1:
sgp30ebbrE4.b1:
sgp30ebbrE9.b1:
sgp30ebbrE18.b1:
sgp30ebbrE26.b1:
sgp30ebbrE13.b1:
sgp30ebbrE7.b1:
sgp30ebbrE12.b1:
sgp30ebbrE19.b1:
sgp30ebbrE20.b1:
sgp30ebbrE22.b1:
sgp30ebbrE2.b1:
sgp30ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/12/2002 | 1710 | 11/12/2002 | 1740 |
Subject: | SGP/EBBR/E15 - Data Incorrect during 6 month checks |
DataStreams: | sgp15ebbrE15.a1, sgp15ebbrE15.b1, sgp30ebbrE15.a1, sgp30ebbrE15.b1, sgp5ebbrE15.a1 |
Description: | Some of the measurements are incorrect during the 6 month checks. |
Measurements: | sgp15ebbrE15.b1:
sgp5ebbrE15.a1:
sgp30ebbrE15.b1:
sgp30ebbrE15.a1:
sgp15ebbrE15.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/11/2001 | 1920 | 07/09/2002 | 1522 |
Subject: | SGP/EBBR/E15 - REPROCESS: Soil Moisture #1 Offset |
DataStreams: | sgp30ebbrE15.a1, sgp30ebbrE15.b1 |
Description: | The wrong offset (9.5264) was used in the calibration coefficients for Soil Moisture Probe #1. The correct offset is 0.95264. Sensible and Latent Heat Flux measurements are not significantly affected. |
Measurements: | sgp30ebbrE15.b1:
sgp30ebbrE15.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/22/2002 | 1630 | 07/22/2002 | 1659 |
Subject: | SGP/EBBR/E15 - Data Mostly Zero Following Logger Restart |
DataStreams: | sgp15ebbrE15.a0, sgp15ebbrE15.a1, sgp30ebbrE15.a1, sgp30ebbrE15.b1, sgp5ebbrE15.a1 |
Description: | Most of the measurements were zero, which normally happens after a datalogger restart. The restart on 07/22/02 occurred after a pressure sensor replacement. |
Measurements: | sgp5ebbrE15.a1:
sgp30ebbrE15.b1:
sgp15ebbrE15.a0:
sgp30ebbrE15.a1:
sgp15ebbrE15.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/10/2002 | 1300 | 12/23/2002 | 1600 |
Subject: | SGP/EBBR/E15 - Pressure Low |
DataStreams: | sgp30ebbrE15.a1, sgp30ebbrE15.b1, sgp5ebbrE15.a1 |
Description: | The pressure measurements were too low during this period, as much as 40 mb too low. This occurred after a restart of the EBBR CR10 datalogger, for unknown reasons. |
Measurements: | sgp5ebbrE15.a1:
sgp30ebbrE15.b1:
sgp30ebbrE15.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/29/2003 | 1540 | 04/29/2003 | 1655 |
Subject: | SGP/EBBR/E15 - 6 Month Checks |
DataStreams: | sgp15ebbrE15.a1, sgp15ebbrE15.b1, sgp30ebbrE15.a1, sgp30ebbrE15.b1, sgp5ebbrE15.a1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp15ebbrE15.b1:
sgp5ebbrE15.a1:
sgp30ebbrE15.b1:
sgp30ebbrE15.a1:
sgp15ebbrE15.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/11/2001 | 1930 | 08/05/2003 | 1630 |
Subject: | SGP/EBBR/E15 - Wind Direction 10 Degrees Low |
DataStreams: | sgp15ebbrE15.a1, sgp15ebbrE15.b1, sgp30ebbrE15.a1, sgp30ebbrE15.b1, sgp5ebbrE15.a1 |
Description: | The wind direction at E15 EBBR was 10 degrees low for the period (since installation of the EBBR unit on 10/11/01). The wind direction sensor was apparently aligned to magnetic north instead of true north at installation. |
Measurements: | sgp15ebbrE15.b1:
sgp5ebbrE15.a1:
sgp30ebbrE15.b1:
sgp30ebbrE15.a1:
sgp15ebbrE15.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/19/2003 | 1900 | 06/19/2003 | 1900 |
Subject: | SGP/EBBR/E15 - Top Humidity reported as zero in 30 minute data. |
DataStreams: | sgp30ebbrE15.a1, sgp30ebbrE15.b1 |
Description: | The top relative humidity was reported as zero in the 30 minute data. This is an apparent misreporting by the CR10 datalogger or is caused by a communications problem in accessing the data from the datalogger. It is very infrequently seen; it's actual cause is unknown. No other data was affected. |
Measurements: | sgp30ebbrE15.b1:
sgp30ebbrE15.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/11/2003 | 2230 | 10/12/2003 | 2330 |
Subject: | SGP/EBBR/E15 - Reference Temperature Failed |
DataStreams: | sgp15ebbrE15.a1, sgp15ebbrE15.b1, sgp30ebbrE15.a1, sgp30ebbrE15.b1, sgp5ebbrE15.a1 |
Description: | The reference temperature failed, making air temperatures, e, h, and Bowen ratio incorrect. |
Measurements: | sgp15ebbrE15.b1:
sgp5ebbrE15.a1:
sgp30ebbrE15.b1:
sgp30ebbrE15.a1:
sgp15ebbrE15.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/29/2003 | 1630 | 09/30/2003 | 1600 |
Subject: | SGP/EBBR/E15 - Water In Net Radiometer, Surface Fluxes Incorrect |
DataStreams: | sgp15ebbrE15.a1, sgp15ebbrE15.b1, sgp30ebbrE15.a1, sgp30ebbrE15.b1, sgp5ebbrE15.a1 |
Description: | The net radiometer top dome was punctured by migrating birds. This alone does not cause incorrect net radiation measurements. However, when it subsequently rains, water collects in the bottom dome, causing the net radiation measurements to be too large during the day and too small at night. |
Measurements: | sgp15ebbrE15.b1:
sgp30ebbrE15.b1:
sgp5ebbrE15.a1:
sgp30ebbrE15.a1:
sgp15ebbrE15.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/11/2003 | 2230 | 10/12/2003 | 2359 |
Subject: | SGP/EBBR/E15 - Soil Temperature #5 Incorrect, Surface Fluxes Incorrect |
DataStreams: | sgp15ebbrE15.a1, sgp15ebbrE15.b1, sgp30ebbrE15.a1, sgp30ebbrE15.b1 |
Description: | Soil Temperature #5 was incorrect. This affected many measurements, including sensible and latent heat fluxes. |
Measurements: | sgp15ebbrE15.b1:
sgp30ebbrE15.b1:
sgp30ebbrE15.a1:
sgp15ebbrE15.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/13/2003 | 0000 | 10/14/2003 | 1830 |
Subject: | SGP/EBBR/E15 - Soil Temperature #5 Incorrect, Surface Fluxes Incorrect |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1 |
Description: | Soil Temperature #5 was incorrect. This affected many measurements, including sensible and latent heat fluxes. |
Measurements: | sgp15ebbrE15.b1:
sgp30ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/05/2003 | 1400 | 11/11/2003 | 1800 |
Subject: | SGP/EBBR/E15 - AEM Hung Between Home Positions, Surface Fluxes Incorrect |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | The AEM was hung inbetween home positions. Therefore the sensible and latent heat fluxes are also incorrect. |
Measurements: | sgp15ebbrE15.b1:
sgp30ebbrE15.b1:
sgp5ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/14/2003 | 1700 | 10/14/2003 | 1730 |
Subject: | SGP/EBBR/E15 - 6 Month Checks |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | Many measurements were incorrect while the AEM was positioned with both aspirators at the same height for comparison checks. |
Measurements: | sgp15ebbrE15.b1:
sgp30ebbrE15.b1:
sgp5ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/13/2003 | 1500 | 12/15/2003 | 0000 |
Subject: | SGP/EBBR/E15 - Net Radiometer Iced Over During Storm |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | The net radiometer was covered with ice from a storm. Net radiation values are incorrect, and are sometimes even negative during the daytime. Sensible and latent heat fluxes are therefore incorrect. |
Measurements: | sgp15ebbrE15.b1:
sgp30ebbrE15.b1:
sgp5ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/12/2003 | 1955 | 12/13/2003 | 1940 |
Subject: | SGP/EBBR/E15 - Wind Speed Sensor Frozen By Ice Storm |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | The wind speed sensor was frozen stopped by ice from a storm. The wind speed and residual wind speed measurements are incorrect. |
Measurements: | sgp15ebbrE15.b1:
sgp30ebbrE15.b1:
sgp5ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/25/2003 | 2150 | 01/20/2004 | 2030 |
Subject: | SGP/EBBR/E15 - Wind Direction About 130 Degrees Low |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | The EBBR wind direction sensor developed an approximately -130 degree offset. The pin holding the vane to the shaft had fallen out. |
Measurements: | sgp15ebbrE15.b1:
sgp30ebbrE15.b1:
sgp5ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/26/2004 | 0000 | 03/02/2004 | 1705 |
Subject: | SGP/EBBR/E15 - Net Radiometer not level |
DataStreams: | sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | The EBBR net radiometer at E15 was suspected to be off of level because the daily peak net radiation was recorded later than it was recorded by the SIRS. Site Ops found the EBBR net radiometer at E15 was significantly off of level (1/2 bubble tipped to the north). Adjustment of the level improved the comparison of peak time with the co-located SIRS. There was no significant effect on the sensible and latent heat fluxes. |
Measurements: | sgp30ebbrE15.b1:
sgp5ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/01/2004 | 1810 | 02/02/2004 | 2040 |
02/04/2004 | 2240 | 02/05/2004 | 1840 |
02/09/2004 | 1020 | 02/09/2004 | 1430 |
Subject: | SGP/EBBR/E15 - Wind Speed Frozen Stopped By Ice |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | An ice storm froze the wind speed sensor to a stopped condition. The wind speed and residual wind speed measurements are incorrect. This sensor failure does not affect the sensible and latent heat fluxes. |
Measurements: | sgp15ebbrE15.b1:
sgp30ebbrE15.b1:
sgp5ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/23/2004 | 2130 | 04/27/2004 | 1630 |
Subject: | SGP/EBBR/E15 - Wind Speed Incorrect |
DataStreams: | sgp15ebbrE15.b1, sgp30baebbrE15.c1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | The cups were broken off of the wind speed sensor by hail on 4/23/04. The cups were replaced on 4/27/04. Wind speed measurements were near zero and incorrect during the indicated period. |
Measurements: | sgp15ebbrE15.b1:
sgp30ebbrE15.b1:
sgp30baebbrE15.c1:
sgp5ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/11/2001 | 1500 | 09/28/2004 | 1509 |
Subject: | SGP/EBBR/E15 - Soil Moisture Coefficients Incorrect |
DataStreams: | sgp30ebbrE15.b1 |
Description: | The soil Moisture coefficients used for E15 were incorrect from the time of installation of the system in program version 8. On 9 July 02 parameter 4 (an offset value) of program step 136 was changed from 9.5264 to the correct value 0.95264 (this error had been made by the vendor). The error masked the real problem (wrong soil type) and after 9 July 02 resulted in very low soil moisture measurements. The coefficients used were for coarse sand. On 28 Sep 04 I visited the E15 site and found the soil type a fine sandy clay or fine sandy loam. I changed the coefficients at 1509 GMT to those for a very fine sandy loam; this produced reasonable soil moistures for the soil type. The old coefficients in parameters 4 through 9 for step 136 of the program were: 0.95264, -6.7859, 36.224, -58.619, 25.073, 4.2278 The new coefficients are: 10.241, -27.943, 49.523, -12.648, -52.643, 38.054 |
Measurements: | sgp30ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/01/2001 | 0000 | 03/29/2005 | 1900 |
Subject: | SGP/EBBR/E15 - Improved EBBR CR10 Program |
DataStreams: | sgp30ebbrE15.b1 |
Description: | Effective 20050329.1900, the EBBR.E15 CR10 program was revised to improve the quality of the primary measurements as follows: 1) An RMS procedure is used to determine max and min limits of acceptable soil heat flow. This is applied to the individual soil sets. Measurements outside the limits are rejected and the measurements within the limits are used to calculate the average soil heat flow. 2) Max and min limits are used to determine incorrect values of net radiation, sensible heat flux (h), latent heat flux (e), and automatic exchange mechanism (AEM) signal. If AEM signal is outside the limits, h and e are set to 999s. If net radiation is outside the limits, h and e are set to 999s. If h or e are outside the limits, h and e are set to 999s. Virtually no incorrect soil measurement will affect the primary measurements of h and e. By setting h and e to 999s, it can be easily seen that the primary variables are incorrect; no other interpretation is possible. Prior to 20050329, the improved CR10 program was NOT in effect. DQRs have been submitted for known instances of incorrect soil measurements which affected the quality of the primary measurements of h and e. Note: the DQR begin date is the begin date of the sgp30ebbrE15.b1 data stream. The earlier version of the CR10 program was also used on previous EBBR datastream names (e.g. sgp30ebbrE15.a1). |
Measurements: | sgp30ebbrE15.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/13/2003 | 0000 | 07/22/2005 | 2359 |
Subject: | SGP/EBBR/E15 - metadata corrections |
DataStreams: | sgp15ebbrE15.b1, sgp30ebbrE15.b1, sgp5ebbrE15.b1 |
Description: | On 20050723, a series of metadata corrections and additions were completed. These metadata changes apply to all EBBR.E15 data collected by ARM back to the installation of the instrument in September 1992. Please see the current metadata for correct information. These changes do not affect data values or quality. The changes are summarized below: 1) Update of "sensor location" information 2) Addition of installation dates for the systems 3) Correction of soil type 4) Correction of soil moisture units (from "by volume" to "gravimetric") |
Measurements: | sgp15ebbrE15.b1:
sgp30ebbrE15.b1:
sgp5ebbrE15.b1:
|