Data Quality Reports for Session: 119530 User: lovejoon82 Completed: 05/09/2009


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D050912.3PYE/SONDE/M1 - Bad RH sensorpyesondewnpnM1.b1


DQRID : D050912.3
Start DateStart TimeEnd DateEnd Time
08/31/2005113008/31/20051140
Subject:
PYE/SONDE/M1 - Bad RH sensor
DataStreams:pyesondewnpnM1.b1
Description:
The RH sensor heating circuit failed on this sounding.  RH values during the first part of 
the sounding oscillate between a reasonable value and a very low value.  It appears that 
(a) the heating was out of phase, and (b) incorrect values were being reported during 
the first 10 minutes or so of flight.
Measurements:pyesondewnpnM1.b1:
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)


Back To Table of Contents



END OF DATA