DQR ID | Subject | Data Streams Affected |
---|---|---|
D050912.3 | PYE/SONDE/M1 - Bad RH sensor | pyesondewnpnM1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/31/2005 | 1130 | 08/31/2005 | 1140 |
Subject: | PYE/SONDE/M1 - Bad RH sensor |
DataStreams: | pyesondewnpnM1.b1 |
Description: | The RH sensor heating circuit failed on this sounding. RH values during the first part of the sounding oscillate between a reasonable value and a very low value. It appears that (a) the heating was out of phase, and (b) incorrect values were being reported during the first 10 minutes or so of flight. |
Measurements: | pyesondewnpnM1.b1:
|