DQR ID | Subject | Data Streams Affected |
---|---|---|
D031217.4 | SGP/EBBR/E12 - Pressure Sensor Measurement Scales With Battery Voltage | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.a1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/03/2003 | 1500 | 10/12/2003 | 2355 |
Subject: | SGP/EBBR/E12 - Pressure Sensor Measurement Scales With Battery Voltage |
DataStreams: | sgp15ebbrE12.b1, sgp30ebbrE12.b1, sgp5ebbrE12.a1 |
Description: | The barometer output scaled with supply voltage (it followed the charging of the battery by the solar panel). The problem did not affect the sensible and latent heat flux calculations beyond the expected system error. |
Measurements: | sgp15ebbrE12.b1:
sgp5ebbrE12.a1:
sgp30ebbrE12.b1:
|