Data Quality Reports for Session: 120081 User: saraht Completed: 06/08/2009


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D020909.1SGP/SONDE/C1 - Bad Temperature SensorsgplssondeC1.c1, sgpqmemwrcolC1.c1, sgpsondewnpnC1.a1, sgpsondewnpnC1.b1
D020909.2SGP/SONDE/C1 - Bad Temperature SensorsgplssondeC1.c1, sgpqmemwrcolC1.c1, sgpsondewnpnC1.a1, sgpsondewnpnC1.b1
D050510.1SGP/SONDE/C1 - Surface pressure bad and other data noisysgpsondewnpnC1.b1
D051121.1SGP/SONDE/C1 - Sonde RH values incorrectsgpsondewnpnC1.b1
D060928.3SGP/SONDE/C1 - Increased precision for alt, dp, pres, rh, tdrysgpsondewnpnC1.b1


DQRID : D020909.1
Start DateStart TimeEnd DateEnd Time
09/06/2002172709/06/20021854
Subject:
SGP/SONDE/C1 - Bad Temperature Sensor
DataStreams:sgplssondeC1.c1, sgpqmemwrcolC1.c1, sgpsondewnpnC1.a1, sgpsondewnpnC1.b1
Description:
It appears that the temperature sensor broke upon launch and therefore the reported 
temperature is approximately 40 degC too low.
Measurements:sgpsondewnpnC1.a1:
  • Dry bulb temperature(tdry)
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)

sgpsondewnpnC1.b1:
  • Surface dew point temperature(dp)
  • Dry bulb temperature(tdry)
  • Relative humidity inside the instrument enclosure(rh)

sgpqmemwrcolC1.c1:
  • Integrated vapor column from sonde using MWR Instrument Performance Model (IPM)(integ_vap_sonde)

sgplssondeC1.c1:
  • Relative humidity inside the instrument enclosure(rh)
  • Dry bulb temperature(tdry)


Back To Table of Contents

DQRID : D020909.2
Start DateStart TimeEnd DateEnd Time
09/06/2002052709/06/20020718
Subject:
SGP/SONDE/C1 - Bad Temperature Sensor
DataStreams:sgplssondeC1.c1, sgpqmemwrcolC1.c1, sgpsondewnpnC1.a1, sgpsondewnpnC1.b1
Description:
It appears that the temperature sensor broke shortly after launch. Temperatures are offset 
by approxmiately (but not constantly) 40 degC.
Measurements:sgpsondewnpnC1.a1:
  • Dry bulb temperature(tdry)
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)

sgpsondewnpnC1.b1:
  • Surface dew point temperature(dp)
  • Dry bulb temperature(tdry)
  • Relative humidity inside the instrument enclosure(rh)

sgpqmemwrcolC1.c1:
  • Integrated vapor column from sonde using MWR Instrument Performance Model (IPM)(integ_vap_sonde)

sgplssondeC1.c1:
  • Relative humidity inside the instrument enclosure(rh)
  • Dry bulb temperature(tdry)


Back To Table of Contents

DQRID : D050510.1
Start DateStart TimeEnd DateEnd Time
04/26/2005172604/26/20051910
Subject:
SGP/SONDE/C1 - Surface pressure bad and other data noisy
DataStreams:sgpsondewnpnC1.b1
Description:
I was checking this sounding because the surface pressure entered by the operator was 
considerably higher than that recorded by the TWHAPS and SMOS.  I'm not sure why, but the 
pressure data are very noisy while the sonde is on the ground (which accounts for the 
incorrect surface pressure) and the RH data aloft do not look terrible but do not look great 
either.
Measurements:sgpsondewnpnC1.b1:
  • Surface dew point temperature(dp)
  • base time(base_time)
  • Mean Wind Speed(wspd)
  • U-component(u_wind)
  • Relative humidity inside the instrument enclosure(rh)
  • lat(lat)
  • Ascent Rate(asc)
  • Retrieved pressure profile(pres)
  • Wind Direction(deg)
  • Wind Status(wstat)
  • lon(lon)
  • Dry bulb temperature(tdry)
  • Time offset of tweaks from base_time(time_offset)
  • V-component(v_wind)
  • Dummy altitude for Zeb(alt)


Back To Table of Contents

DQRID : D051121.1
Start DateStart TimeEnd DateEnd Time
04/13/2005234604/15/20051732
Subject:
SGP/SONDE/C1 - Sonde RH values incorrect
DataStreams:sgpsondewnpnC1.b1
Description:
RH values for all sondes launched during this time were incorrect due to a faulty Ground 
Check set which damaged or conditioned the RH sensor improperly.
Measurements:sgpsondewnpnC1.b1:
  • Surface dew point temperature(dp)
  • Relative humidity inside the instrument enclosure(rh)


Back To Table of Contents

DQRID : D060928.3
Start DateStart TimeEnd DateEnd Time
09/28/2006172812/31/99992359
Subject:
SGP/SONDE/C1 - Increased precision for alt, dp, pres, rh, tdry
DataStreams:sgpsondewnpnC1.b1
Description:
The output precision of the following variables was increased 20060928

alt  changed from 1m to 0.1m
dp   changed from 0.1degC to 0.01degC
pres changed from 0.1 hPa to 0.01 hPa
RH   changed from 1%RH to 0.01%RH
tdry changed from .1degC to 0.01degC

Note that the increase in precision does not necessarily imply an increase in accuracy.  
Users are cautioned to continue to use their best scientific judgement when using these or 
any other data.
Measurements:sgpsondewnpnC1.b1:
  • Surface dew point temperature(dp)
  • Dry bulb temperature(tdry)
  • Relative humidity inside the instrument enclosure(rh)
  • Retrieved pressure profile(pres)
  • Dummy altitude for Zeb(alt)


Back To Table of Contents



END OF DATA