DQR ID | Subject | Data Streams Affected |
---|---|---|
D080313.1 | SGP/ECOR/E24 - Water Vapor Density Incorrect | sgp30ecorE24.b1 |
D080526.1 | SGP/ECOR/E10 - Incorrect CO2 fluxes | sgp30ecorE10.b1 |
D080604.2 | SGP/ECOR/E14 - Sensor failure | sgp30ecorE14.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/01/2008 | 0000 | 04/12/2008 | 0230 |
Subject: | SGP/ECOR/E24 - Water Vapor Density Incorrect |
DataStreams: | sgp30ecorE24.b1 |
Description: | Mean Water Vapor Density (mean_q) often did not trend well with the co-located SMOS and sometimes jumped downwards approximately 200 to 300 mmol/m^3. Some of the instances were clearly a result of contamination by precipitation, but others were not; the latent heat flux was sometimes affected. It was determined that all instances of this behavior from the beginning of 2008 through the beginning date of this DQR were caused by precipitation/dew/frost. |
Measurements: | sgp30ecorE24.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/10/2008 | 1130 | 06/25/2008 | 1500 |
Subject: | SGP/ECOR/E10 - Incorrect CO2 fluxes |
DataStreams: | sgp30ecorE10.b1 |
Description: | The analog CO2 signal from the LI-7500 to the sonic anemometer became very noisy and incorrect CO2 fluxes resulted. Sonic anemometer was replaced. |
Measurements: | sgp30ecorE10.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/02/2008 | 1330 | 06/05/2008 | 1930 |
Subject: | SGP/ECOR/E14 - Sensor failure |
DataStreams: | sgp30ecorE14.b1 |
Description: | The water vapor output of the LI-7500 was too low, indicating that water vapor leaked into the sensor and that the chemicals are spent or there was an electronic failure. Latent heat flux was not normally affected by this problem. |
Measurements: | sgp30ecorE14.b1:
|