Subject: | SGP/SMOS/E13 - RH sensor failure |
DataStreams: | sgp1440smosE13.b1, sgp1smosE13.b1, sgp30smosE13.b1
|
Description: | The T/RH sensor relative humidity data is suspect during this time. Values from this
sensor failed the 6-month check and subsequent bi-weekly checks. The RH values were lower by
larger than the manufacturer suggested uncertainty of 3%. |
Measurements: | sgp1440smosE13.b1: - Time of Maximum Relative Humidity(time_max_rh)
- Time of Maximum Vapor Pressure(time_max_vap_pres)
- Minimum Relative Humidity(min_rh)
- Time of Minimum Vapor Pressure(time_min_vap_pres)
- Maximum Relative Humidity(max_rh)
- Time of Minimum Relative Humidity(time_min_rh)
- Maximum Vapor Pressure(max_vap_pres)
- Minimum Vapor Pressure(min_vap_pres)
sgp30smosE13.b1: - Standard Deviation of Relative Humidity(sd_rh)
- Standard Deviation of Vapor Pressure(sd_vap_pres)
- Vapor Pressure (kiloPascals)(vap_pres)
- Relative humidity inside the instrument enclosure(rh)
sgp1smosE13.b1: - Relative humidity inside the instrument enclosure(rh)
- Vapor Pressure (kiloPascals)(vap_pres)
|