DQR ID | Subject | Data Streams Affected |
---|---|---|
D070529.2 | SGP/IRT/E16 - computer failure | sgpirt200msE16.a1, sgpirtE16.b1 |
D080424.3 | SGP/IRT/E16 - reference temperature incorrect | sgpirt200msE16.a1, sgpirtE16.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/14/2006 | 0600 | 04/11/2007 | 1559 |
Subject: | SGP/IRT/E16 - computer failure |
DataStreams: | sgpirt200msE16.a1, sgpirtE16.b1 |
Description: | The serial ports of the computer failed so no data were collected. The problem was solved when the computer was replaced. |
Measurements: | sgpirt200msE16.a1:
sgpirtE16.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/11/2007 | 1600 | 06/19/2008 | 2035 |
Subject: | SGP/IRT/E16 - reference temperature incorrect |
DataStreams: | sgpirt200msE16.a1, sgpirtE16.b1 |
Description: | The internal reference temperature was unrealistically high, ~360K (90C). The ambient temperature setting was checked. The problem was corrected by replacing the instrument. The sky temperature measurement appears to have been unaffected. |
Measurements: | sgpirt200msE16.a1:
sgpirtE16.b1:
|