DQR ID | Subject | Data Streams Affected |
---|---|---|
D070521.1 | SGP/SMOS/E7 - Wind sensor binding at low wind speeds | sgp1440smosE7.b1, sgp1smosE7.b1, sgp30smosE7.b1 |
D070618.1 | SGP/SMOS/E1 - Failed temperature probe | sgp1440smosE1.b1, sgp1smosE1.b1, sgp30smosE1.b1 |
D070618.2 | SGP/SMOS/E13 - RH sensor failure | sgp1440smosE13.b1, sgp1smosE13.b1, sgp30smosE13.b1 |
D080617.5 | SGP/SMOS/E1 - Missing data | sgp1440smosE1.b1, sgp1smosE1.b1, sgp30smosE1.b1 |
D080703.12 | SGP/SMOS/E7 - Missing data | sgp1440smosE7.b1, sgp1smosE7.b1, sgp30smosE7.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/21/2007 | 0000 | 10/30/2007 | 1715 |
Subject: | SGP/SMOS/E7 - Wind sensor binding at low wind speeds |
DataStreams: | sgp1440smosE7.b1, sgp1smosE7.b1, sgp30smosE7.b1 |
Description: | The wind sensor bearings were failing causing the wind speed sensor to report lower than actual wind speeds when the winds were below 2 m/s. The wind monitor was replaced on 05/29/2007 but this failed to correct the problem. Subsequently, the site was flooded and it took some time to re-establish normal operations. The wind sensor was again replaced on 10/30/2007. After replacement the wind data from the SMOS compared more favorably to the co-located EBBR data. |
Measurements: | sgp1440smosE7.b1:
sgp30smosE7.b1:
sgp1smosE7.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/03/2007 | 0000 | 06/26/2007 | 2025 |
Subject: | SGP/SMOS/E1 - Failed temperature probe |
DataStreams: | sgp1440smosE1.b1, sgp1smosE1.b1, sgp30smosE1.b1 |
Description: | The temperature probe was out of tolerance as it failed the 6-month calibration checks and subsequent bi-weekly checks. The temperature values consistently were outside the manufacturers suggested uncertainty of 0.2 C. |
Measurements: | sgp1smosE1.b1:
sgp30smosE1.b1:
sgp1440smosE1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/16/2007 | 0000 | 06/18/2007 | 2000 |
Subject: | SGP/SMOS/E13 - RH sensor failure |
DataStreams: | sgp1440smosE13.b1, sgp1smosE13.b1, sgp30smosE13.b1 |
Description: | The T/RH sensor relative humidity data is suspect during this time. Values from this sensor failed the 6-month check and subsequent bi-weekly checks. The RH values were lower by larger than the manufacturer suggested uncertainty of 3%. |
Measurements: | sgp1440smosE13.b1:
sgp30smosE13.b1:
sgp1smosE13.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/25/2006 | 1800 | 05/02/2006 | 1100 |
05/10/2006 | 1400 | 05/15/2006 | 2300 |
06/23/2007 | 2300 | 06/26/2007 | 2200 |
08/08/2007 | 0800 | 08/15/2007 | 0630 |
12/11/2007 | 2000 | 12/17/2007 | 2130 |
12/22/2007 | 1500 | 12/26/2007 | 1600 |
05/23/2008 | 0700 | 05/27/2008 | 0200 |
Subject: | SGP/SMOS/E1 - Missing data |
DataStreams: | sgp1440smosE1.b1, sgp1smosE1.b1, sgp30smosE1.b1 |
Description: | Data are missing and unrecoverable |
Measurements: | sgp1smosE1.b1:
sgp30smosE1.b1:
sgp1440smosE1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/20/2003 | 2100 | 12/02/2003 | 2000 |
04/24/2004 | 0430 | 04/30/2004 | 1600 |
07/04/2004 | 1330 | 07/09/2004 | 1730 |
10/31/2004 | 2230 | 11/02/2004 | 2000 |
01/05/2005 | 0130 | 01/11/2005 | 2000 |
06/16/2005 | 0000 | 06/28/2005 | 1930 |
03/30/2006 | 2200 | 04/04/2006 | 1930 |
08/28/2006 | 0300 | 09/02/2006 | 2300 |
06/30/2007 | 0800 | 08/10/2007 | 1930 |
08/21/2007 | 2000 | 09/04/2007 | 1900 |
Subject: | SGP/SMOS/E7 - Missing data |
DataStreams: | sgp1440smosE7.b1, sgp1smosE7.b1, sgp30smosE7.b1 |
Description: | Data are missing and unrecoverable |
Measurements: | sgp1440smosE7.b1:
sgp30smosE7.b1:
sgp1smosE7.b1:
|