Data Quality Reports for Session: 125858 User: kjohnson Completed: 02/03/2010


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D090515.1TWP/SMET/C3 - Reprocess: Temperature Biased low by 1.06 Degreestwpsmet60sC3.b1


DQRID : D090515.1
Start DateStart TimeEnd DateEnd Time
04/19/2005042004/04/20090115
Subject:
TWP/SMET/C3 - Reprocess: Temperature Biased low by 1.06 Degrees
DataStreams:twpsmet60sC3.b1
Description:
The temperature of the SMET system was biased low by 1.06 Deg C due to a non-precision 
resistor being used int he measurement circuit.  The data will be reprocessed.  Saturation 
Vapor pressure and Vapor pressure will have to be reprocessed also.

Add 1.06 degrees to the temperature data.  

Correct Sat Vap Press using the Buck approximation:
   Es = 6.1121exp((17.502*T)/(240.97 + T) 
where T is the temperature in C.  

Correct Vapor pressure using: E = (RH*Es)/100
Measurements:twpsmet60sC3.b1:
  • Vapor pressure standard deviation(vappress_sd)
  • Mean Air Temperature or Hardware Error(temp_mean)
  • Vapor pressure mean(vappress_mean)
  • Std Dev of Air Temperature(temp_sd)


Back To Table of Contents



END OF DATA