DQR ID | Subject | Data Streams Affected |
---|---|---|
D041015.3 | TWP/GNDRAD/C2 - incorrect IRT calibration | twpgndrad60sC2.b1 |
D050504.2 | TWP/GNDRAD/C2 - IRT connector | twpgndrad60sC2.b1 |
D060630.35 | TWP/GNDRAD/C2 - Reprocess: Longwave Calibration error | twpgndrad60sC2.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/05/2004 | 0501 | 07/19/2004 | 0206 |
Subject: | TWP/GNDRAD/C2 - incorrect IRT calibration |
DataStreams: | twpgndrad60sC2.b1 |
Description: | When the GNDRAD data logger was upgraded, the calibration offset for the IRT was initially incorrect. |
Measurements: | twpgndrad60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/27/2005 | 2330 | 04/03/2005 | 0240 |
Subject: | TWP/GNDRAD/C2 - IRT connector |
DataStreams: | twpgndrad60sC2.b1 |
Description: | The IRT had a problem with the connector of either the signal or power cable which produced an output of 0 volts equivalent to a measurement of 223K. The problem was corrected when the connectors were cleaned and tightened. |
Measurements: | twpgndrad60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/05/2004 | 2202 | 12/21/2005 | 2149 |
Subject: | TWP/GNDRAD/C2 - Reprocess: Longwave Calibration error |
DataStreams: | twpgndrad60sC2.b1 |
Description: | Modified pyrgeometer calibration procedures were implemented beginning in December 2003. These modified procedures introduced a calibration bias in the longwave data. The previous procedures were re-implemented at all sites between December 2005 and February 2006 to restore proper calibrations. The data collected while the incorrect procedures were in place are being reprocessing to remove the calibration bias. Note, this also affected standard deviation, maximum and minimum data fields. |
Measurements: | twpgndrad60sC2.b1:
|