Data Quality Reports for Session: 127713 User: renata Completed: 05/12/2010


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D021229.1TWP/SONDE/C1 - Broken Temperature SensortwpsondewnpnC1.a1, twpsondewnpnC1.b1
D021231.2TWP/Sonde/C1 - Bad RH ValuestwpsondewnpnC1.a1, twpsondewnpnC1.b1
D030427.2TWP/SONDE/C1 - Questionable RHtwpsondewnpnC1.a1, twpsondewnpnC1.b1
D030503.1TWP/SONDE/C1 - Bad RHtwpsondewnpnC1.a1, twpsondewnpnC1.b1
D030603.3TWP/SONDE/C1 - Broken temperature sensortwpsondewnpnC1.a1, twpsondewnpnC1.b1
D040228.1TWP/SONDE/C1 - Broken temperature sensortwpsondewnpnC1.b1
D040228.3TWP/SONDE/C1 - Bad RH sensorstwpsondewnpnC1.b1
D040530.1TWP/SONDE/C1 - Bad RH sensortwpsondewnpnC1.a1, twpsondewnpnC1.b1
D040606.1TWP/SONDE/C1 - Bad RH at beginning of soundingtwpsondewnpnC1.b1
D040627.1TWP/SONDE/C1 - Bad RH at beginning of soundingtwpsondewnpnC1.b1
D040627.2TWP/SONDE/C1 - Bad RH in short interval alofttwpsondewnpnC1.b1
D040711.1TWP/SONDE/C1 - Bad RH at beginning of soundingtwpsondewnpnC1.b1
D041011.3TWP/SONDE/C1 - Bad RH at beginning of soundingtwpsondewnpnC1.b1
D050105.1TWP/SONDE/C1 - Bad RH at beginning of soundingtwpsondewnpnC1.b1
D050201.2TWP/SONDE/C1 - Bad RH at beginning of soundingtwpsondewnpnC1.b1
D070313.3TWP/SONDE/C1 - Surface RH value incorrecttwpsondewnpnC1.b1


DQRID : D021229.1
Start DateStart TimeEnd DateEnd Time
09/18/2002233409/19/20020111
09/21/2002233009/22/20020059
10/01/2002234910/02/20020117
10/13/2002113910/13/20021303
10/18/2002235810/19/20020113
05/23/2003233005/24/20030135
06/14/2003233006/15/20030147
06/22/2003115906/22/20031332
06/22/2003234906/23/20030110
06/23/2003114606/23/20031323
06/25/2003234206/26/20030118
06/28/2003233006/29/20030056
06/29/2003113406/29/20031312
Subject:
TWP/SONDE/C1  - Broken Temperature Sensor
DataStreams:twpsondewnpnC1.a1, twpsondewnpnC1.b1
Description:
The temperature sensor on these radiosondes broke after launch.
All the temperature related data are incorrect.
Measurements:twpsondewnpnC1.a1:
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)
  • Dry bulb temperature(tdry)

twpsondewnpnC1.b1:
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)
  • Dry bulb temperature(tdry)


Back To Table of Contents

DQRID : D021231.2
Start DateStart TimeEnd DateEnd Time
12/21/2002233612/22/20020119
Subject:
TWP/Sonde/C1 - Bad RH Values
DataStreams:twpsondewnpnC1.a1, twpsondewnpnC1.b1
Description:
The RH values from the surface to approximately 11.5 km oscillate between reasonable 
values and zero.  This is apparently caused by a failure in the RH sensor heating process.  
Although the oscillation stops when the heating stops (at temperatures below ~ -45degC) I 
think the values must be viewed as suspect.
Measurements:twpsondewnpnC1.a1:
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)

twpsondewnpnC1.b1:
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)


Back To Table of Contents

DQRID : D030427.2
Start DateStart TimeEnd DateEnd Time
04/08/2003234704/09/20030003
Subject:
TWP/SONDE/C1 - Questionable RH
DataStreams:twpsondewnpnC1.a1, twpsondewnpnC1.b1
Description:
It appears that the RH sensor heating circuit failed during this portion of the sounding 
(approximately between 5km and 10km).  The RH oscillates rapidly between low (~5 %RH) and 
not so low (~50 %RH) and the oscillation seems to stop when the temperature is below ~35 
degC.  According to Vaisala these are symptoms of failure in the ASIC (application 
specific integrated circuit) that controls the heating.
Measurements:twpsondewnpnC1.a1:
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)

twpsondewnpnC1.b1:
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)


Back To Table of Contents

DQRID : D030503.1
Start DateStart TimeEnd DateEnd Time
04/29/2003112304/29/20031205
Subject:
TWP/SONDE/C1 - Bad RH
DataStreams:twpsondewnpnC1.a1, twpsondewnpnC1.b1
Description:
The RH sensor heating circuit failed on this radiosonde.  The RH-related values obtained 
between 0 and 10000 meters oscillate between near-zero and reasonable ambient values.  
Above 10000 the RH values are likely correct.
Measurements:twpsondewnpnC1.a1:
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)

twpsondewnpnC1.b1:
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)


Back To Table of Contents

DQRID : D030603.3
Start DateStart TimeEnd DateEnd Time
05/23/2003233005/24/20030135
Subject:
TWP/SONDE/C1 - Broken temperature sensor
DataStreams:twpsondewnpnC1.a1, twpsondewnpnC1.b1
Description:
The temperature sensor on this radiosonde broke after launch.  All the temperature-related 
data are incorrect.
Measurements:twpsondewnpnC1.a1:
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)
  • Dry bulb temperature(tdry)

twpsondewnpnC1.b1:
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)
  • Dry bulb temperature(tdry)


Back To Table of Contents

DQRID : D040228.1
Start DateStart TimeEnd DateEnd Time
02/07/2004234202/08/20040115
02/09/2004113302/09/20041233
04/05/2004233204/06/20040105
04/22/2004115604/22/20041317
04/24/2004113004/24/20041312
04/29/2004113704/29/20041312
04/30/2004233605/01/20040112
05/03/2004233105/04/20040105
05/05/2004113505/05/20041319
05/07/2004113205/07/20041315
05/11/2004114205/11/20041311
05/12/2004114505/12/20041206
05/18/2004113005/18/20041321
07/10/2004113007/11/20041255
07/13/2004113907/13/20041320
07/16/2004234607/17/20040107
07/22/2004235507/23/20040140
07/26/2004234107/27/20040115
08/14/2004233008/15/20040104
08/22/2004233508/23/20040105
09/05/2004234009/06/20040119
09/14/2004232809/14/20042329
09/17/2004233209/18/20040102
10/28/2004233610/29/20040057
11/28/2004235211/29/20040044
Subject:
TWP/SONDE/C1 - Broken temperature sensor
DataStreams:twpsondewnpnC1.b1
Description:
The temperature sensor on these radiosondes broke after launch
All the temperature-related data are incorrect.
Measurements:twpsondewnpnC1.b1:
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)
  • Dry bulb temperature(tdry)


Back To Table of Contents

DQRID : D040228.3
Start DateStart TimeEnd DateEnd Time
02/06/2004234102/06/20040130
Subject:
TWP/SONDE/C1 - Bad RH sensors
DataStreams:twpsondewnpnC1.b1
Description:
It appears that the RH sensor heating circuit failed during this portion of the sounding 
(approximately between 0km and 12km).  The RH oscillates rapidly in a 20% range and the 
oscillation seems to stop when the temperature is below ~45 degC.  These are symptoms of 
failure in the ASIC (application specific integrated circuit) that controls the heating.
Measurements:twpsondewnpnC1.b1:
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)


Back To Table of Contents

DQRID : D040530.1
Start DateStart TimeEnd DateEnd Time
05/28/2003233005/29/20030010
Subject:
TWP/SONDE/C1 - Bad RH sensor
DataStreams:twpsondewnpnC1.a1, twpsondewnpnC1.b1
Description:
It appears that the RH sensor heating circuit failed during this portion of the sounding 
(approximately between the surface and 11km).  The RH oscillates rapidly between low (~0 
%RH) and high (~80 %RH) and the oscillation seems to stop when the temperature is below 
~40 degC.  According to Vaisala these are symptoms of failure in the ASIC (application 
specific integrated circuit) that controls the heating.
Measurements:twpsondewnpnC1.a1:
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)

twpsondewnpnC1.b1:
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)


Back To Table of Contents

DQRID : D040606.1
Start DateStart TimeEnd DateEnd Time
06/03/2004113206/03/20041212
Subject:
TWP/SONDE/C1 - Bad RH at beginning of sounding
DataStreams:twpsondewnpnC1.b1
Description:
It appears that the RH sensor heating circuit failed during this portion of the sounding 
(approximately between 0km and 11km).  The RH oscillates rapidly between low (~15 %RH) and 
not so low (~50 %RH) and the oscillation seems to stop when the temperature is below ~40 
degC.  According to Vaisala these are symptoms of failure in the ASIC (application 
specific integrated circuit) that controls the heating.
Measurements:twpsondewnpnC1.b1:
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)


Back To Table of Contents

DQRID : D040627.1
Start DateStart TimeEnd DateEnd Time
06/04/2004001006/23/20042330
Subject:
TWP/SONDE/C1 - Bad RH at beginning of sounding
DataStreams:twpsondewnpnC1.b1
Description:
It appears that the RH sensor heating circuit failed during this portion of the sounding 
(approximately between 0km and 11km).  The RH oscillates rapidly between low (~0 %RH) and 
not so low (~80 %RH) and the oscillation seems to stop when the temperature is below ~40 
degC.  According to Vaisala these are symptoms of failure in the ASIC (application 
specific integrated circuit) that controls the heating.
Measurements:twpsondewnpnC1.b1:
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)


Back To Table of Contents

DQRID : D040627.2
Start DateStart TimeEnd DateEnd Time
06/21/2004001006/21/20040011
Subject:
TWP/SONDE/C1  - Bad RH in short interval aloft
DataStreams:twpsondewnpnC1.b1
Description:
It appears that the RH sensor heating circuit on this sonde failed about 31 minutes into 
the flight.  The RH values look reasonable from the surface to approximately 8.25 km at 
which point they begin to oscillate between 0 and 40%.  This behavior stops when the sonde 
reaches approximately 11.0km when the air temperature drops below -40 degC.  The 
oscillation stops at that point, though the RH remains higher than might be expected even under 
supersaturated conditions.
Measurements:twpsondewnpnC1.b1:
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)


Back To Table of Contents

DQRID : D040711.1
Start DateStart TimeEnd DateEnd Time
06/30/2004113006/30/20041204
Subject:
TWP/SONDE/C1 - Bad RH at beginning of sounding
DataStreams:twpsondewnpnC1.b1
Description:
It appears that the RH sensor heating circuit failed during this portion of the sounding 
(approximately between 0km and 11km).  The RH oscillates rapidly between low (~2 %RH) and 
not so low (~90 %RH) and the oscillation seems to stop when the temperature is below ~40 
degC.  According to Vaisala these are symptoms of failure in the ASIC (application 
specific integrated circuit) that controls the heating.
Measurements:twpsondewnpnC1.b1:
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)


Back To Table of Contents

DQRID : D041011.3
Start DateStart TimeEnd DateEnd Time
10/08/2004113110/08/20041200
Subject:
TWP/SONDE/C1 - Bad RH at beginning of sounding
DataStreams:twpsondewnpnC1.b1
Description:
It appears that the RH sensor heating circuit failed during this portion of the sounding 
(approximately between 0km and 11km).  The RH oscillates rapidly between low (~1 %RH) and 
high (~80 %RH).  According to Vaisala these are symptoms of failure in the ASIC 
(application specific integrated circuit) that controls the heating.
Measurements:twpsondewnpnC1.b1:
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)


Back To Table of Contents

DQRID : D050105.1
Start DateStart TimeEnd DateEnd Time
12/30/2004233012/31/20040108
Subject:
TWP/SONDE/C1 - Bad RH at beginning of sounding
DataStreams:twpsondewnpnC1.b1
Description:
It appears that the RH sensor heating circuit failed during this portion of the sounding 
(approximately between 0km and 11km).  The RH oscillates rapidly between low (~5 %RH) and 
not so low (~70 %RH) and the oscillation seems to stop when the temperature is below ~40 
degC.  According to Vaisala these are symptoms of failure in the ASIC (application 
specific integrated circuit) that controls the heating.
Measurements:twpsondewnpnC1.b1:
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)


Back To Table of Contents

DQRID : D050201.2
Start DateStart TimeEnd DateEnd Time
01/07/2005233001/08/20050102
Subject:
TWP/SONDE/C1 - Bad RH at beginning of sounding
DataStreams:twpsondewnpnC1.b1
Description:
It appears that the RH sensor heating circuit failed during this portion of the sounding 
(approximately between 0km and 12km).  The RH oscillates rapidly between low (~2 %RH) and 
not so low (~80 %RH) and the oscillation seems to stop when the temperature is below ~40 
degC.  According to Vaisala these are symptoms of failure in the ASIC (application 
specific integrated circuit) that controls the heating.
Measurements:twpsondewnpnC1.b1:
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)


Back To Table of Contents

DQRID : D070313.3
Start DateStart TimeEnd DateEnd Time
03/08/2007000003/08/20070001
Subject:
TWP/SONDE/C1 - Surface RH value incorrect
DataStreams:twpsondewnpnC1.b1
Description:
The surface RH value of 77% when the value should have been in the low 60's.  It appears 
the observers did not allow the sonde to equilibrate to ambient conditions before 
collecting the surface data.  This error also caused the dew point to be incorrect.
Measurements:twpsondewnpnC1.b1:
  • Relative humidity inside the instrument enclosure(rh)
  • Surface dew point temperature(dp)


Back To Table of Contents