DQR ID | Subject | Data Streams Affected |
---|---|---|
D061214.2 | TWP/SMET/C2 - Barometer failure | twpsmet60sC2.b1 |
D070110.1 | TWP/GNDRAD/C2 - IRT signal | twpgndrad60sC2.b1 |
D070206.1 | TWP/SKYRAD/C2 - Short direct normal | twpskyrad20sC2.a0, twpskyrad60sC2.b1 |
D070404.2 | TWP/SKYRAD/C2 - IRT measurement error | twpskyrad60sC2.b1 |
D070406.6 | TWP/GNDRAD/C2 - Reprocess: IRT calibration | twpgndrad60sC2.b1 |
D070510.5 | TWP/SMET/C2 - Lower windspeed sensor questionable | twpsmet60sC2.b1 |
D070520.1 | TWP/SMET/C2 - Lower Wind Sensor Direction Incorrect | twpsmet60sC2.b1 |
D071204.1 | TWP/GNDRAD/C2 - Reprocessed: Incorrect calibration | twpgndrad20sC2.a0, twpgndrad60sC2.b1 |
D071204.2 | TWP/SKYRAD/C2 - Reprocessed: Incorrect calibrations | twpskyrad20sC2.a0, twpskyrad60sC2.b1 |
D071219.1 | TWP/SMET/C2 - Lower Wind Speed Data Suspect | twpsmet60sC2.b1 |
D080313.2 | TWP/SKYRAD/C2 - Direct Normal reading too low | twpskyrad20sC2.a0, twpskyrad60sC2.b1, twpskyradC2.00 |
D080617.1 | TWP/SMET/C2 - ORG rainfall measurements too large | twpsmet60sC2.b1 |
D081022.5 | TWP/SMET/C2 - Wind Speed data contain offset due to induced voltage | twpsmet60sC2.b1 |
D081027.2 | TWP/SMET/C2 - ORG rainrate high due to calibration. | twpsmet60sC2.b1 |
D081222.3 | TWP/SKYRAD/C2 - Tracker failure | twpskyrad20sC2.a0, twpskyrad60sC2.b1 |
D090328.1 | TWP/SKYRAD/C2 - Tracker inoperative | twpskyrad20sC2.a0, twpskyrad60sC2.b1 |
D090605.2 | TWP/SKYRAD/C2 - Instrument grounding problem | twpskyrad20sC2.a0, twpskyrad60sC2.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/07/2006 | 1941 | 01/23/2007 | 0023 |
Subject: | TWP/SMET/C2 - Barometer failure |
DataStreams: | twpsmet60sC2.b1 |
Description: | The barometer began experiencing problems with noise resulting in incorrect data. The problem becames worse over time; after 12/19/2006 @0151 GMT the barometer data is completely missing. The barometer was replaced. |
Measurements: | twpsmet60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/03/2006 | 0400 | 02/04/2007 | 2316 |
Subject: | TWP/GNDRAD/C2 - IRT signal |
DataStreams: | twpgndrad60sC2.b1 |
Description: | When the other GNDRAD radiometers were replaced on 12/3/2006, the quality of the IRT signal began to degrade until 12/26 when it became unusable. The instrument was replaced on 2/4/2007. |
Measurements: | twpgndrad60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/31/2007 | 0223 | 02/10/2007 | 0055 |
Subject: | TWP/SKYRAD/C2 - Short direct normal |
DataStreams: | twpskyrad20sC2.a0, twpskyrad60sC2.b1 |
Description: | The short_direct_normal (NIP) data are missing or incorrect due to instrument cable failure. The cable was replaced. The short_direct_normal can be calculated using the following relationship for shortwave radiometers: global hemispheric = diffuse hemispheric + direct normal *cos (zenith angle) |
Measurements: | twpskyrad60sC2.b1:
twpskyrad20sC2.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/04/2006 | 2120 | 02/22/2007 | 0125 |
Subject: | TWP/SKYRAD/C2 - IRT measurement error |
DataStreams: | twpskyrad60sC2.b1 |
Description: | The IRT was not properly calibrated and was not able to measure sky temperatures below 273.2 K. (This instrument was the same one damaged by lightning in November 2005 in Darwin. It was returned for repair and received a new lens but was calibrated only down to 0 deg C, so the problem was neither detected nor repaired.) The IRT was replaced on 2/22/2007. |
Measurements: | twpskyrad60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/04/2007 | 2317 | 02/17/2007 | 0400 |
Subject: | TWP/GNDRAD/C2 - Reprocess: IRT calibration |
DataStreams: | twpgndrad60sC2.b1 |
Description: | The configuration of the IRT was set to that used at TWP/C3. This caused the calibration of the data to be wrong. It was corrected by reconfiguring the IRT to match the datalogger program. These data can be reprocessed and corrected by adding 10K, i.e. sfc_ir_temp(new) = sfc_ir_temp(old) + 10K. |
Measurements: | twpgndrad60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/05/2007 | 0000 | 08/19/2007 | 2135 |
Subject: | TWP/SMET/C2 - Lower windspeed sensor questionable |
DataStreams: | twpsmet60sC2.b1 |
Description: | Lower wind speed sensor had degraded bearings. This sensor was reporting as much as 0.4 m/s lower values than the upper sensors when the winds were 2 m/s or below. The sensor was replaced. |
Measurements: | twpsmet60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/11/2007 | 0000 | 05/22/2007 | 0047 |
Subject: | TWP/SMET/C2 - Lower Wind Sensor Direction Incorrect |
DataStreams: | twpsmet60sC2.b1 |
Description: | Wind directions between lower and upper anemometers were displaced by approximately 45 degrees when a new sensor was installed to replace the failing lower winde sensor. The new sensor direction data incorrect. |
Measurements: | twpsmet60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/25/2007 | 2202 | 11/30/2007 | 1724 |
Subject: | TWP/GNDRAD/C2 - Reprocessed: Incorrect calibration |
DataStreams: | twpgndrad20sC2.a0, twpgndrad60sC2.b1 |
Description: | The Gndrad PSP was changed out 11/25 but the datalogger program was not updated until 11/30. PSP #31293 installed 20071125.2202. Correct calibration coefficient = 123.46. Reprocessed to correct upwelling shortwave hemispheric irradiance (US) by: US(new) = (US(old) / 123.56 ) * 123.46 Reprocessing completed March 2008. |
Measurements: | twpgndrad60sC2.b1:
twpgndrad20sC2.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/25/2007 | 2220 | 11/30/2007 | 1726 |
Subject: | TWP/SKYRAD/C2 - Reprocessed: Incorrect calibrations |
DataStreams: | twpskyrad20sC2.a0, twpskyrad60sC2.b1 |
Description: | The SKYRAD PSPs were changed out 11/25 but the datalogger program was not updated until 11/30. Reprocessed to correct calibration per: down_short_diffuse(corrected) = down_short_diffuse(orig) * (118.18/112.42) short_direct_normal(corrected)= short_direct_normal(orig) * (122.20/118.11) down_short_hemips(corrected) = down_short_hemisp(orig) * (133.16/137.57) Reprocessing completed March 2008. |
Measurements: | twpskyrad60sC2.b1:
twpskyrad20sC2.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/22/2007 | 0048 | 08/19/2007 | 2150 |
Subject: | TWP/SMET/C2 - Lower Wind Speed Data Suspect |
DataStreams: | twpsmet60sC2.b1 |
Description: | The lower wind speed bearings appeared to be binding at speeds below 2 m/s as it did not compare well with the upper wind sensor. |
Measurements: | twpsmet60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/02/2008 | 2230 | 03/10/2008 | 2130 |
Subject: | TWP/SKYRAD/C2 - Direct Normal reading too low |
DataStreams: | twpskyrad20sC2.a0, twpskyrad60sC2.b1, twpskyradC2.00 |
Description: | The NIP (direct_normal shortwave irradiance) measurement was reading too low during this time period. No TWP Ops documentation could be found to determine the cause but data review indicate that the NIP appeared to be reading low by about 25W/m2 during this period. The NIP data had well defined steps in irradiance when compared to three component checks indicating the problem period. |
Measurements: | twpskyradC2.00:
twpskyrad60sC2.b1:
twpskyrad20sC2.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/08/2008 | 0000 | 08/07/2008 | 2300 |
Subject: | TWP/SMET/C2 - ORG rainfall measurements too large |
DataStreams: | twpsmet60sC2.b1 |
Description: | The ORG was collecting large amounts of rainfall compared to the co-located tipping bucket raingauge. The cable was chewed by mice causing the wires to make contact intermittently mimicking rainfall. |
Measurements: | twpsmet60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/01/1998 | 0000 | 02/23/2009 | 0410 |
Subject: | TWP/SMET/C2 - Wind Speed data contain offset due to induced voltage |
DataStreams: | twpsmet60sC2.b1 |
Description: | The marine model of the RM Young wind sensor needs twisted pair shielded cable. The manufacturer did not originally supply the pigtail with twisted pair shielded cable. Use of twisted pair shielded cable for the rest of the cable run to the logger was also not specified. This caused an induced voltage in the wind speed line from the wind direction excitation. The induced voltage varied from sensor to sensor and probably changed anytime the cable was replaced or parts of the wind sensor were repaired or replaced. The induced voltage has been seen to vary from 0.15 m/s to as much as 1.08 m/s. |
Measurements: | twpsmet60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/05/2008 | 2311 | 10/05/2008 | 2344 |
Subject: | TWP/SMET/C2 - ORG rainrate high due to calibration. |
DataStreams: | twpsmet60sC2.b1 |
Description: | The rainrate from the ORG was high due to calibration. This data is not real. |
Measurements: | twpsmet60sC2.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/17/2008 | 1942 | 12/21/2008 | 2215 |
Subject: | TWP/SKYRAD/C2 - Tracker failure |
DataStreams: | twpskyrad20sC2.a0, twpskyrad60sC2.b1 |
Description: | Tracker failed to wake up on 12/17. Ops reset tracker on 12/21. The affected instruments are the 8-48 (down_short_diffuse) and NIP (direct_normal) shortwave instruments. |
Measurements: | twpskyrad60sC2.b1:
twpskyrad20sC2.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/25/2009 | 2000 | 05/18/2009 | 0442 |
Subject: | TWP/SKYRAD/C2 - Tracker inoperative |
DataStreams: | twpskyrad20sC2.a0, twpskyrad60sC2.b1 |
Description: | The tracker failed in park position. Tracker was replaced 5/14 but did not resume fully normal operation until 5/18. The down_short_diffuse and short_direct_normal data are directly affected by this failure. These data are not recoverable. |
Measurements: | twpskyrad60sC2.b1:
twpskyrad20sC2.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/21/2009 | 1855 | 05/25/2009 | 2100 |
Subject: | TWP/SKYRAD/C2 - Instrument grounding problem |
DataStreams: | twpskyrad20sC2.a0, twpskyrad60sC2.b1 |
Description: | During this time period the IRT analog output line appears to have adversely affected the skyrad datalogger grounding. This resulted in intermittent anomalous readings across all skyrad datalogger channels including logger voltage. |
Measurements: | twpskyrad60sC2.b1:
twpskyrad20sC2.a0:
|