DQR ID | Subject | Data Streams Affected |
---|---|---|
D071210.1 | SGP/MWR/B6 - Incorrect mixer kinetic temperature | sgpmwrlosB6.b1, sgpmwrtipB6.a1 |
D090225.11 | SGP/MWR/B6 - Loss of thermal stabilization | sgpmwrlosB6.b1, sgpmwrtipB6.a1 |
D090720.3 | SGP/MWR/B6 - Missing data | sgpmwrlosB6.b1, sgpmwrtipB6.a1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/01/2007 | 1800 | 01/12/2009 | 1700 |
Subject: | SGP/MWR/B6 - Incorrect mixer kinetic temperature |
DataStreams: | sgpmwrlosB6.b1, sgpmwrtipB6.a1 |
Description: | The mixer kinetic temperature values are incorrect. The mixer temp sensor was on an open circuit. This is a diagnostic variable. The noise diode mount temperature Tknd can be used to monitor the thermal stabilization of the instrument |
Measurements: | sgpmwrlosB6.b1:
sgpmwrtipB6.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/23/2008 | 1500 | 11/03/2008 | 1759 |
01/12/2009 | 1900 | 01/29/2009 | 1600 |
Subject: | SGP/MWR/B6 - Loss of thermal stabilization |
DataStreams: | sgpmwrlosB6.b1, sgpmwrtipB6.a1 |
Description: | Data are erratic due to loss of thermal stabilization of the instrument. |
Measurements: | sgpmwrlosB6.b1:
sgpmwrtipB6.a1:
|
Start Date | Start Time | End Date | End Time |
---|
Subject: | SGP/MWR/B6 - Missing data |
DataStreams: | sgpmwrlosB6.b1, sgpmwrtipB6.a1 |
Description: | Data are missing and unrecoverable. |
Measurements: | sgpmwrlosB6.b1:
sgpmwrtipB6.a1:
|