DQR ID | Subject | Data Streams Affected |
---|---|---|
D060707.1 | GEC/SUOMI/X1 - Temperature Sensor Failure at Site SA42 (Manus Island) | gec30suomigpsX1.c1 |
D060707.2 | GEC/SUOMI/X1 - Temperature, RH Sensor Failure at Site SG27 (Barrow) | gec30suomigpsX1.c1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/26/2005 | 0930 | 10/16/2006 | 0200 |
Subject: | GEC/SUOMI/X1 - Temperature Sensor Failure at Site SA42 (Manus Island) |
DataStreams: | gec30suomigpsX1.c1 |
Description: | The temperature sensor failed at this location causing it to read -51 degC constantly. This resulted in a bias in the retrieved precipitable water vapor (PWV) of approximately 1 cm (low). |
Measurements: | gec30suomigpsX1.c1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/26/2005 | 1600 | 07/03/2006 | 0000 |
Subject: | GEC/SUOMI/X1 - Temperature, RH Sensor Failure at Site SG27 (Barrow) |
DataStreams: | gec30suomigpsX1.c1 |
Description: | The temperature and relative humidity sensors failed at this location, causing the temperature to read -51 degC and the RH to read 100% constantly. This results in a bias in the retrieved precipitable water vapor (PWV). |
Measurements: | gec30suomigpsX1.c1:
|