Data Quality Reports for Session: 134492 User: wlj Completed: 04/14/2011


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D090612.1GRW/SKYRAD/M1 - Reprocessed: IRT configurationgrwskyrad60sM1.b1


DQRID : D090612.1
Start DateStart TimeEnd DateEnd Time
05/20/2009224905/22/20091717
05/22/2009180506/11/20092001
Subject:
GRW/SKYRAD/M1 - Reprocessed: IRT configuration
DataStreams:grwskyrad60sM1.b1
Description:
The SKYRAD IRT was used in testing a revised serial communication program. During the 
testing, the analog output configuration was to changed to default values (used at TWP and 
NSA sites) that are different than those used at AMF. So, the IRT's analog output was set 
to a scale of 25K/V and an offset of 213.2K but the logger converted the signal using a 
scale of 26K/V and an offset of 173.2K. 

The sky temperature measurements were corrected using the following equation

        Tnew = (25/26)(Told-173.2K)+213.2K

and the data were reprocessed.  The reprocessed data were archived in April 2011.
Measurements:grwskyrad60sM1.b1:
  • Temperature, Sky Infra-Red(sky_ir_temp)
  • sky_ir_temp_min
  • sky_ir_temp_max


Back To Table of Contents