DQR ID | Subject | Data Streams Affected |
---|---|---|
D100224.3 | NSA/MMCR/C1 - Reflectivity bias | nsammcrmomC1.b1, nsammcrspecmomC1.a0 |
D100616.2 | NSA/VCEIL/C1 - Small gaps in data | nsavceil25kC1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/10/2010 | 0000 | 02/23/2010 | 0230 |
Subject: | NSA/MMCR/C1 - Reflectivity bias |
DataStreams: | nsammcrmomC1.b1, nsammcrspecmomC1.a0 |
Description: | The MMCR calibration was incorrect causing a very large positive shift in reflectivity values. |
Measurements: | nsammcrmomC1.b1:
nsammcrspecmomC1.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/03/2009 | 2107 | 06/15/2010 | 2143 |
Subject: | NSA/VCEIL/C1 - Small gaps in data |
DataStreams: | nsavceil25kC1.b1 |
Description: | When the ceilometer was replaced, small gaps in the netCDF files occurred due to the presence of '?' characters in the backscatter and header records of the raw data that could not be ingested. It is unknown whether this was caused by a poor CPU board or incompatibilities with the computer configuration (hyper-threading). The problem was corrected when the instrument and computer were both replaced. During this period, approximately 3% of data are missing each hour due to this problem. |
Measurements: | nsavceil25kC1.b1:
|