Data Quality Reports for Session: 136393 User: rqm Completed: 07/25/2011


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D110118.1SBS/GNDRAD/M1 - Reprocess: IRT calibration errorsbsgndrad60sM1.b1


DQRID : D110118.1
Start DateStart TimeEnd DateEnd Time
09/23/2010170001/18/20111556
Subject:
SBS/GNDRAD/M1 - Reprocess: IRT calibration error
DataStreams:sbsgndrad60sM1.b1
Description:
The IRT surface temperature reported was incorrect because the analog output configuration 
did not agree with the data logger scale factors. The IRT configuration was corrected 
1/18.
Measurements:sbsgndrad60sM1.b1:
  • Effective Surface (Skin) Temperature(sfc_ir_temp)
  • Surface Infra-Red Temperature Minima(sfc_ir_temp_min)
  • Surface Infra-Red Temperature Maxima(sfc_ir_temp_max)


Back To Table of Contents