DQR ID | Subject | Data Streams Affected |
---|---|---|
D100216.5 | HFE/VCEIL/M1 - Reprocess: Incorrect serial_number and sum_backscatter fields | hfevceil25kM1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/08/2008 | 0000 | 12/28/2008 | 2359 |
Subject: | HFE/VCEIL/M1 - Reprocess: Incorrect serial_number and sum_backscatter fields |
DataStreams: | hfevceil25kM1.b1 |
Description: | Attribute serial_number: In most cases, this field is not representative of the actual serial number of the instrument because the value did not change when an instrument changed locations or was replaced with a spare. Variable sum_backscatter: An out-of-range condition for the "sum of detected and normalized backscatter" was reported as 0 instead of missing (-9999). The valid range is 0 - 0.0999 /srad. |
Measurements: | hfevceil25kM1.b1:
|