Data Quality Reports for Session: 138500 User: j1a1l1mann Completed: 11/02/2011


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D100216.5HFE/VCEIL/M1 - Reprocess: Incorrect serial_number and sum_backscatter fieldshfevceil25kM1.b1


DQRID : D100216.5
Start DateStart TimeEnd DateEnd Time
05/08/2008000012/28/20082359
Subject:
HFE/VCEIL/M1 - Reprocess: Incorrect serial_number and sum_backscatter fields
DataStreams:hfevceil25kM1.b1
Description:
Attribute serial_number: In most cases, this field is not representative of the actual 
serial number of the instrument because the value did not change when an instrument changed 
locations or was replaced with a spare.
   
Variable sum_backscatter: An out-of-range condition for the "sum of detected and 
normalized backscatter" was reported as 0 instead of missing (-9999). The valid range is 0 - 
0.0999 /srad.
Measurements:hfevceil25kM1.b1:
  • sum_backscatter


Back To Table of Contents



END OF DATA