Data Quality Reports for Session: 139024 User: anderberg Completed: 11/29/2011


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D111017.1GAN/GNDRAD/M1 - IRT calibration errorgangndrad60sM1.b1


DQRID : D111017.1
Start DateStart TimeEnd DateEnd Time
09/23/2011074210/09/20110413
Subject:
GAN/GNDRAD/M1 - IRT calibration error
DataStreams:gangndrad60sM1.b1
Description:
The IRT surface temperature reported was incorrect because the analog output configuration 
did not agree with the data logger scale factors. The IRT configuration was corrected on 
10/9.
Measurements:gangndrad60sM1.b1:
  • Effective Surface (Skin) Temperature(sfc_ir_temp)
  • Surface Infra-Red Temperature Minima(sfc_ir_temp_min)
  • Surface Infra-Red Temperature Maxima(sfc_ir_temp_max)


Back To Table of Contents