DQR ID | Subject | Data Streams Affected |
---|---|---|
D111017.1 | GAN/GNDRAD/M1 - IRT calibration error | gangndrad60sM1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/23/2011 | 0742 | 10/09/2011 | 0413 |
Subject: | GAN/GNDRAD/M1 - IRT calibration error |
DataStreams: | gangndrad60sM1.b1 |
Description: | The IRT surface temperature reported was incorrect because the analog output configuration did not agree with the data logger scale factors. The IRT configuration was corrected on 10/9. |
Measurements: | gangndrad60sM1.b1:
|