DQR ID | Subject | Data Streams Affected |
---|---|---|
D020509.2 | NSA/VCEIL/C1 - Occasional Data Transmission problems | nsavceil25kC1.00, nsavceil25kC1.a1, nsavceil25kC1.b1 |
D100215.3 | NSA/VCEIL/C1/C2 - Reprocess: Incorrect serial_number and sum_backscatter fields | nsavceil25kC1.b1, nsavceil25kC2.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/06/2000 | 0000 | 01/04/2001 | 0000 |
Subject: | NSA/VCEIL/C1 - Occasional Data Transmission problems |
DataStreams: | nsavceil25kC1.00, nsavceil25kC1.a1, nsavceil25kC1.b1 |
Description: | During this period, there were occasional data transmission problems that lead to cut-off data packets at times. These disappeared after ceilometer computer were replaced. |
Measurements: | nsavceil25kC1.00:
nsavceil25kC1.b1:
nsavceil25kC1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/22/1999 | 0000 | 10/31/2009 | 2359 |
Subject: | NSA/VCEIL/C1/C2 - Reprocess: Incorrect serial_number and sum_backscatter fields |
DataStreams: | nsavceil25kC1.b1, nsavceil25kC2.b1 |
Description: | Attribute serial_number: In most cases, this field is not representative of the actual serial number of the instrument because the value did not change when an instrument changed locations or was replaced with a spare. Variable sum_backscatter: An out-of-range condition for the "sum of detected and normalized backscatter" was reported as 0 instead of missing (-9999). The valid range is 0 - 0.0999 /srad. This DQR also applies to a1 and b1 level data collected prior to the begin date. |
Measurements: | nsavceil25kC2.b1:
nsavceil25kC1.b1:
|