DQR ID | Subject | Data Streams Affected |
---|---|---|
D040526.2 | SGP/AERI/C1 - Metadata errors | sgpaeri01ch1C1.a1, sgpaeri01ch2C1.a1, sgpaeri01engineerC1.a1, sgpaeri01summaryC1.a1, sgpaerilblcloudsC1.c1, sgpaerilbldiffC1.c1, sgpaerilbldifflsC1.c1, sgpqmeaerilblC1.c1, sgpqmeaerilbllsC1.c1, sgpqmeaerimeansC1.c1 |
D970627.1 | Software update on the AERI | sgpaeri01ch1C1.a1, sgpaeri01ch2C1.a1, sgpaeri01engineerC1.a1, sgpaeri01summaryC1.a1, sgpaerilblcloudsC1.c1, sgpaerilbldiffC1.c1, sgpaerilbldifflsC1.c1 |
D990920.1 | SGP/AERI/C1 - Reprocessed: radiance data | sgpaeri01ch1C1.a1, sgpaeri01ch2C1.a1, sgpaeri01summaryC1.a1, sgpaerilbldiffC1.c1, sgpaerilbldifflsC1.c1, sgpaeriprofC1.c1, sgplblch1C1.c1, sgplblch1lsC1.c1, sgplblch2C1.c1, sgplblch2lsC1.c1, sgpqmeaerilblC1.c1, sgpqmeaerilbllsC1.c1, sgpqmeaerimeansC1.c1, sgpaeri01C1.00 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/22/1995 | 0012 | 05/14/2003 | 2359 |
Subject: | SGP/AERI/C1 - Metadata errors |
DataStreams: | sgpaeri01ch1C1.a1, sgpaeri01ch2C1.a1, sgpaeri01engineerC1.a1, sgpaeri01summaryC1.a1, sgpaerilblcloudsC1.c1, sgpaerilbldiffC1.c1, sgpaerilbldifflsC1.c1, sgpqmeaerilblC1.c1, sgpqmeaerilbllsC1.c1, sgpqmeaerimeansC1.c1 |
Description: | The latitude, longitude and altitude of the AERI were incorrectly entered into the ARM database. The correct location of the SGP.C1 AERI is: Lat: 36.606N Lon: 97.485W Alt: 316m |
Measurements: | sgpaeri01ch1C1.a1:
sgpaerilbldifflsC1.c1:
sgpaeri01summaryC1.a1:
sgpaerilblcloudsC1.c1:
sgpaeri01engineerC1.a1:
sgpaeri01ch2C1.a1:
sgpaerilbldiffC1.c1:
sgpqmeaerimeansC1.c1:
sgpqmeaerilblC1.c1:
sgpqmeaerilbllsC1.c1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/10/1997 | 0000 | 06/23/1997 | 2359 |
Subject: | Software update on the AERI |
DataStreams: | sgpaeri01ch1C1.a1, sgpaeri01ch2C1.a1, sgpaeri01engineerC1.a1, sgpaeri01summaryC1.a1, sgpaerilblcloudsC1.c1, sgpaerilbldiffC1.c1, sgpaerilbldifflsC1.c1 |
Description: | On June 10th, staff from the University of Wisconsin-Madison visited the SGP CART site to upgrade the software running on the AERI instrument. While there, they took the opportunity to perform some routine checks of the instrument. Due to these tests, data from 6/10/97 - 6/11/97 should not be ingested or used for scientific studies. The software installed was the same software as will be used on the next generation AERIs (AERI-Nexts). The AERI-Nexts have some improved hardware capabilities, which is taken advantage of by the new software. Unfortunately, as the AERI 01 at the CART site does not have these capabilities, NaNs resulted in a couple of fields in the sgpaeri01summaryC1.a1 platform. These NaNs are placeholders in arrays that were too long for the AERI01 but the correct length for the AERI-Nexts. All valid data (non-NaN data) is correct, and hence these NaNs should be ignored. A routine was modified on 6/24 which effectively fixes the problem (the size of the arrays written out is now determined by the hardware's capabilities, versus outputting a fixed length vector). |
Measurements: | sgpaeri01ch1C1.a1:
sgpaerilbldifflsC1.c1:
sgpaerilblcloudsC1.c1:
sgpaeri01summaryC1.a1:
sgpaeri01engineerC1.a1:
sgpaeri01ch2C1.a1:
sgpaerilbldiffC1.c1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/01/1996 | 0000 | 06/30/1997 | 2359 |
Subject: | SGP/AERI/C1 - Reprocessed: radiance data |
DataStreams: | sgpaeri01ch1C1.a1, sgpaeri01ch2C1.a1, sgpaeri01summaryC1.a1, sgpaerilbldiffC1.c1, sgpaerilbldifflsC1.c1, sgpaeriprofC1.c1, sgplblch1C1.c1, sgplblch1lsC1.c1, sgplblch2C1.c1, sgplblch2lsC1.c1, sgpqmeaerilblC1.c1, sgpqmeaerilbllsC1.c1, sgpqmeaerimeansC1.c1, sgpaeri01C1.00 |
Description: | Due to a software error in the real-time calibration software, the AERI-01 radiance data between 10/96 and 6/97 were incorrectly calibrated. All of the AERI-01 radiance data collected at the SGP central facility during this time period was corrupted by this software error and were reprocessed to correct the radiance data. The error also affected all derived products of the AERI radiance data. AERI data collected prior to and after this time period was unaffected by this programming error. The proper calibrated radiance spectra was recovered from the raw archived dataset through post-processing. This reprocessing of the AERI data was carried out by the instrument mentor at PNNL and confirmed independently by Univ. of Wisconsin. The reprocessed AERI radiance data during the period 10/96 to 6/97 supercedes all previous versions of the AERI data in all instances. reprocessed, archived 9/99: sgpaeri01C1.00 reprocessed, archived 10/99: sgpaeri01ch1C1.a1 sgpaeri01ch2C1.a1 sgpaeri01engineerC1.a1 sgpaeri01summaryC1.a1 sgpaerilbldiffC1.c1 sgpaerilbldifflsC1.c1 sgplblch1C1.c0 sgplblch1C1.c1 sgplblch1lsC1.c0 sgplblch1lsC1.c1 sgplblch2C1.c0 sgplblch2C1.c1 sgplblch2lsC1.c0 sgplblch2lsC1.c1 sgpqmeaerilblC1.c1 sgpqmeaerilbllsC1.c1 sgpqmeaerimeansC1.c1 reprocessed, archived 11/99: sgpaeriprofC1.c1 sgpqmeaeriprofC1.c1 not reprocessed: sgpaerilblcloudsC1.c1 sgpaerilbldiffssC1.c1 sgpqmeaerilblssC1.c1 sgpqmeaerimeansssC1.c1 |
Measurements: | sgpaeri01ch1C1.a1:
sgplblch2lsC1.c1:
sgpaeriprofC1.c1:
sgpqmeaerimeansC1.c1:
sgplblch1lsC1.c1:
sgpaerilbldifflsC1.c1:
sgpaeri01summaryC1.a1:
sgpaeri01C1.00:
sgpaeri01ch2C1.a1:
sgpaerilbldiffC1.c1:
sgplblch2C1.c1:
sgpqmeaerilblC1.c1:
sgplblch1C1.c1: