DQR ID | Subject | Data Streams Affected |
---|---|---|
D100129.1 | GRW/AOS/M1 - CPC TEC failure | grwaosM1.a1 |
D100224.6 | GRW/MWR/M1 - Reprocess: Biased LWP retrievals-Updated retrieval coefficients | grwmwrlosM1.b1, grwmwrtipM1.a1 |
D100402.3 | GRW/WACR/M1 - Coherence Filter Contamination | grwwacrM1.b1 |
D111005.3 | GRW/AOS/M1 - CCN signal decline | grwaosccnM1.a1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/19/2009 | 1844 | 01/15/2010 | 1820 |
Subject: | GRW/AOS/M1 - CPC TEC failure |
DataStreams: | grwaosM1.a1 |
Description: | The thermal electric cooler, which controls the temperature in the Condensation Particle Counter, failed. A new TEC was mailed to GRW. The TEC was replaced and the CPC was working again on Jan. 15th, 2010. |
Measurements: | grwaosM1.a1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/01/2009 | 0000 | 03/31/2010 | 1700 |
Subject: | GRW/MWR/M1 - Reprocess: Biased LWP retrievals-Updated retrieval coefficients |
DataStreams: | grwmwrlosM1.b1, grwmwrtipM1.a1 |
Description: | Retrieval coefficients were updated to screen for some bad soundings |
Measurements: | grwmwrtipM1.a1:
grwmwrlosM1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/09/2009 | 1600 | 04/02/2010 | 1200 |
Subject: | GRW/WACR/M1 - Coherence Filter Contamination |
DataStreams: | grwwacrM1.b1 |
Description: | The coherence filter (software) that was added to the WACR under BCR-01500 was causing occasional contamination to WACR data. This was exhibited in the first 2000 m of range by streaking in the data over a period of time. It didn't occur often and when it did, it is quite obvious looking at the plots. On 4/2/2010 at 12:00 UTC, the coherence filter was modified to preclude working on signals with a SNR greater than 60 dB. The vendor thinks this will solve the problem. |
Measurements: | grwwacrM1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/01/2010 | 0000 | 06/02/2010 | 1800 |
Subject: | GRW/AOS/M1 - CCN signal decline |
DataStreams: | grwaosccnM1.a1 |
Description: | There was a slow decline in the CCN signal from January to June of 2010 at GRW. The reason for this decline is unknown as all the instrument operating parameters appear normal. The decline was likely from a change in the CCN column thermal properties or a clogging of the detector inlet. |
Measurements: | grwaosccnM1.a1:
|