DQR ID | Subject | Data Streams Affected |
---|---|---|
D110628.2 | SGP/SIRS/C1 - NIP failure | sgpsirs20sC1.a0, sgpsirsC1.b1 |
D111202.1 | SGP/SIRS/C1 - NIP Failure | sgpsirs20sC1.a0, sgpsirsC1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/12/2011 | 0000 | 06/24/2011 | 1740 |
Subject: | SGP/SIRS/C1 - NIP failure |
DataStreams: | sgpsirs20sC1.a0, sgpsirsC1.b1 |
Description: | The short_direct_normal (NIP instrument) cable failed during this time period. Data are not recoverable. Users can derive an estimate of direct normal values using the following formula: short_direct_normal = (down_short_hemisp - down_short_diffuse_hemisp)/cos(zenith angle) |
Measurements: | sgpsirs20sC1.a0:
sgpsirsC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/16/2011 | 2144 | 08/25/2011 | 1435 |
Subject: | SGP/SIRS/C1 - NIP Failure |
DataStreams: | sgpsirs20sC1.a0, sgpsirsC1.b1 |
Description: | The short_direct_normal (NIP instrument) cable failed (broken wire in NIP connector) during this time period. Data are not recoverable. An estimate of direct normal values can be derived using the following formula: short_direct_normal = (down_short_hemisp - down_short_diffuse_hemisp)/cos(zenith angle) |
Measurements: | sgpsirs20sC1.a0:
sgpsirsC1.b1:
|