DQR ID | Subject | Data Streams Affected |
---|---|---|
D100224.6 | GRW/MWR/M1 - Reprocess: Biased LWP retrievals-Updated retrieval coefficients | grwmwrlosM1.b1, grwmwrtipM1.a1 |
D100402.3 | GRW/WACR/M1 - Coherence Filter Contamination | grwwacrM1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/01/2009 | 0000 | 03/31/2010 | 1700 |
Subject: | GRW/MWR/M1 - Reprocess: Biased LWP retrievals-Updated retrieval coefficients |
DataStreams: | grwmwrlosM1.b1, grwmwrtipM1.a1 |
Description: | Retrieval coefficients were updated to screen for some bad soundings |
Measurements: | grwmwrtipM1.a1:
grwmwrlosM1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/09/2009 | 1600 | 04/02/2010 | 1200 |
Subject: | GRW/WACR/M1 - Coherence Filter Contamination |
DataStreams: | grwwacrM1.b1 |
Description: | The coherence filter (software) that was added to the WACR under BCR-01500 was causing occasional contamination to WACR data. This was exhibited in the first 2000 m of range by streaking in the data over a period of time. It didn't occur often and when it did, it is quite obvious looking at the plots. On 4/2/2010 at 12:00 UTC, the coherence filter was modified to preclude working on signals with a SNR greater than 60 dB. The vendor thinks this will solve the problem. |
Measurements: | grwwacrM1.b1:
|