Subject: | NSA/SKYRAD/C1 - Frosting/Condensation |
DataStreams: | nsaskyrad20sC1.a0, nsaskyrad60sC1.b1
|
Description: | The down_long_hemisp1 varied up to 40 W/m^2 higher than down_long_hemisp2 during the
affected time periods. Cold temperatures reported by the IRT suggest the issue is with PIR1.
Icing or frost on PIR1 or within the ventilator may be to blame but ops reports can't
substantiate this. Data from PIR1 should be avoided during these suspect time periods. |
Measurements: | nsaskyrad60sC1.b1: - down_long_hemisp1_min
- down_long_hemisp1_std
- Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer1(down_long_hemisp1)
- down_long_hemisp1_max
nsaskyrad20sC1.a0: - inst_down_long_hemisp1_tp
- inst_down_long_hemisp1_case_resist
- inst_down_long_hemisp1_dome_resist
|