DQR ID | Subject | Data Streams Affected |
---|---|---|
D060522.1 | TWP/SMET/C1 - Upper wind sensor failure | twpsmet60sC1.b1 |
D061011.2 | TWP/SMET/C1 - Reprocess: Tipping bucket rain gauge added | twpsmet60sC1.b1 |
D070409.1 | TWP/SMET/C1 - Upper winds failure | twpsmet60sC1.b1 |
D080428.2 | TWP/SMET/C1 - Reprocess: ORG Data has very small rainrate values | twpsmet60sC1.b1 |
D080707.2 | TWP/SMET/C1 - ORG Precip Data Incorrect | twpsmet60sC1.b1 |
D081022.3 | TWP/SMET/C1 - Wind Speed data contain offset due to induced voltage | twpsmet60sC1.b1 |
D090304.1 | TWP/SMET/C1 - Optical Rain Gage Inoperative | twpsmet60sC1.b1 |
D090708.1 | TWP/SMET/C1 - Reprocess: Barometric pressure data scaled incorrectly | twpsmet60sC1.b1 |
D100603.1 | TWP/SMET/C1 - Missing RH, Barometer, and ORG data | twpsmet60sC1.b1 |
D110105.1 | TWP/SMET/C1 - Aux (lower) wind sensor failed | twpsmet60sC1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/09/2006 | 0000 | 10/14/2006 | 0237 |
Subject: | TWP/SMET/C1 - Upper wind sensor failure |
DataStreams: | twpsmet60sC1.b1 |
Description: | The upper wind speed sensor was binding at low wind speeds. Comparison to the lower wind sensor showed that the upper sensor flat-lines for long periods of time. The upper anemometer was replaced. |
Measurements: | twpsmet60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/09/1996 | 0000 | 10/17/2006 | 0000 |
Subject: | TWP/SMET/C1 - Reprocess: Tipping bucket rain gauge added |
DataStreams: | twpsmet60sC1.b1 |
Description: | A tipping bucket rain gauge was added to the TWP.C1 SMET suite of instruments on 20061017. Effective 20060926, two new variables (count and rain amount) were added to the end of the twpsmet60sC1.b1 datastream. Between 9/26 and 10/17, the data for these two variables are filled with zeros. Users should not use the data during this period as the zeros are not representative of the rain fall at the site. When reprocessing of this datastream occurs these variables will be added for the 19961009-20060926 and data from 19961009-20061017 will be filled with -9999 (missing). |
Measurements: | twpsmet60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/19/2007 | 0000 | 06/12/2007 | 0406 |
Subject: | TWP/SMET/C1 - Upper winds failure |
DataStreams: | twpsmet60sC1.b1 |
Description: | The upper wind instrument was inoperative resulting in constant wind direction and speed. A failed cable was replaced. |
Measurements: | twpsmet60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/01/2004 | 0000 | 06/30/2010 | 2359 |
Subject: | TWP/SMET/C1 - Reprocess: ORG Data has very small rainrate values |
DataStreams: | twpsmet60sC1.b1 |
Description: | The ORG data is collected via analog signal and not digital. Therefore, we continuously collect data even when no rain is occurring. We also capture small events such as leaves, large bugs, dust, etc blowing through the measurement path. Because this discrete event occurs quickly the value is usually captured for 1 or 2 seconds. This value is then averaged to the minute producing rainrate on the order of .009 mm/hr or less. These values are not indicative of rain and should be removed from the record or summing over long periods can bias the climate record. We are reprocessing the data to remove these small values and are adjusting the collection program to remove them before ingest. |
Measurements: | twpsmet60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/07/2008 | 1300 | 02/09/2008 | 0600 |
Subject: | TWP/SMET/C1 - ORG Precip Data Incorrect |
DataStreams: | twpsmet60sC1.b1 |
Description: | Moisture intrusion into the sensor body allowed condensation on the lens of the ORG. Continual and long-term precipitation was reported during this time. |
Measurements: | twpsmet60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/09/1996 | 0000 | 11/03/2008 | 0730 |
Subject: | TWP/SMET/C1 - Wind Speed data contain offset due to induced voltage |
DataStreams: | twpsmet60sC1.b1 |
Description: | The marine model of the RM Young wind sensor needs twisted pair shielded cable. The manufacturer did not originally supply the pigtail with twisted pair shielded cable. Use of twisted pair shielded cable for the rest of the cable run to the logger was also not specified. This caused an induced voltage in the wind speed line from the wind direction excitation. The induced voltage varied from sensor to sensor and probably changed anytime the cable was replaced or parts of the wind sensor were repaired or replaced. The induced voltage has been seen to vary from 0.15 m/s to as much as 1.08 m/s. |
Measurements: | twpsmet60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/07/2009 | 0100 | 04/03/2009 | 0225 |
Subject: | TWP/SMET/C1 - Optical Rain Gage Inoperative |
DataStreams: | twpsmet60sC1.b1 |
Description: | Beginning on 02/07/2009, the trending between the TBRG and ORG began to look poor, one recording precip when the other was not and vice versa. After a while the ORG quit registering any precip at all. ORG was replaced. |
Measurements: | twpsmet60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/06/2009 | 2319 | 07/07/2009 | 1348 |
Subject: | TWP/SMET/C1 - Reprocess: Barometric pressure data scaled incorrectly |
DataStreams: | twpsmet60sC1.b1 |
Description: | The barometer data was scaled incorrectly when a new program was loaded into the datalogger. The pressure data was reported as mb (hPa) instead of the standard kPa. So, in essence the value was a factor of ten to large. |
Measurements: | twpsmet60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/02/2010 | 0000 | 05/11/2010 | 0330 |
Subject: | TWP/SMET/C1 - Missing RH, Barometer, and ORG data |
DataStreams: | twpsmet60sC1.b1 |
Description: | Atmos_pressure data suddenly went missing on 03/02 at 07:34 UTC and consequently, the relative humidity data flatlined. Furthermore, after a sudden precipitation event when the max precip rate flag was tripped the TBRG total has not agreed with the comparison ORG or the precip rate plot. A loose wire connection was fixed 3/22 restoring RH and barometer data. A damaged cable was repaired 5/11 restoring ORG data. |
Measurements: | twpsmet60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/05/2010 | 0000 | 08/29/2010 | 0154 |
Subject: | TWP/SMET/C1 - Aux (lower) wind sensor failed |
DataStreams: | twpsmet60sC1.b1 |
Description: | The lower wind direction sensor on the C1 site was flatlining at 270 degrees throughout most of the daylight hours of the day. During the night the sensor appeared to be working fine. |
Measurements: | twpsmet60sC1.b1:
|