Subject: | HFE/MWR/M1 - Insufficient calibration stability |
DataStreams: | hfemwrlosM1.b1, hfemwrtipM1.a1
|
Description: | The noise diode temperature suddenly jumped to > 60 C after a power outage. Following
additional power problems the noise diode injection temperature drifted fast and the
radiometer was not able to calibrate properly.
Data during the indicated time periods are poorly calibrated and sometimes erratic. Users
should be aware that the accuracy of the measurements is dramatically degraded.
Users can use the microwave profiler (MWRP) data to supplement the MWR data. |