DQR ID | Subject | Data Streams Affected |
---|---|---|
D120416.16 | SGP/QCRAD/E19 - Met data bad or missing | sgpqcrad1longE19.c1, sgpqcrad1longE19.c2 |
D120416.17 | SGP/QCRAD/E19 - Met data bad or missing | sgpqcrad1longE19.c1, sgpqcrad1longE19.c2 |
D120419.2 | SGP/QCRAD/E19 - Ventilator failure | sgpqcrad1longE19.c1, sgpqcrad1longE19.c2, sgpqcrad1longE19.s1, sgpqcrad1longE19.s2 |
D120510.29 | SGP/QCRAD/E19 - noisy Tc/Td | sgpqcrad1longE19.c1, sgpqcrad1longE19.c2, sgpqcrad1longE19.s1, sgpqcrad1longE19.s2 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/03/2003 | 2000 | 11/13/2003 | 1615 |
Subject: | SGP/QCRAD/E19 - Met data bad or missing |
DataStreams: | sgpqcrad1longE19.c1, sgpqcrad1longE19.c2 |
Description: | Surface meteorological data were missing during these time periods. Met data is used in a variety of quality control checks described below. Air temperature data is needed for tests on LW up and LW down pyrgeometer case and dome temperatures, and tests on the irradiances from upwelling LW/SW and downwelling LW. The Rayleigh limit diffuse radiation test is based on surface pressure. Most periods with missing data described in instrument DQR D031110.8 |
Measurements: | sgpqcrad1longE19.c1:
sgpqcrad1longE19.c2:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/01/2007 | 1100 | 05/24/2007 | 1600 |
07/07/2009 | 0000 | 07/30/2009 | 1800 |
Subject: | SGP/QCRAD/E19 - Met data bad or missing |
DataStreams: | sgpqcrad1longE19.c1, sgpqcrad1longE19.c2 |
Description: | Surface meteorological data were missing during these time periods. Met data is used in a variety of quality control checks described below. Air temperature data is needed for tests on LW up and LW down pyrgeometer case and dome temperatures, and tests on the irradiances from upwelling LW/SW and downwelling LW. The Rayleigh limit diffuse radiation test is based on surface pressure. There were no instrument DQRs on this problem at the time this report was submitted. |
Measurements: | sgpqcrad1longE19.c1:
sgpqcrad1longE19.c2:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/11/2006 | 0000 | 04/13/2006 | 1522 |
Subject: | SGP/QCRAD/E19 - Ventilator failure |
DataStreams: | sgpqcrad1longE19.c1, sgpqcrad1longE19.c2, sgpqcrad1longE19.s1, sgpqcrad1longE19.s2 |
Description: | Dowenlling shortwave hemispheric instrument signal (Eppley PSP) is very erratic due to failed ventilator fan which was replaced during PM on 04/13/2006. Discussed in instrument DQR D060403.3. |
Measurements: | sgpqcrad1longE19.c1:
sgpqcrad1longE19.c2:
sgpqcrad1longE19.s1:
sgpqcrad1longE19.s2:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/25/2002 | 0000 | 10/11/2002 | 1800 |
09/22/2010 | 0000 | 10/10/2010 | 0100 |
Subject: | SGP/QCRAD/E19 - noisy Tc/Td |
DataStreams: | sgpqcrad1longE19.c1, sgpqcrad1longE19.c2, sgpqcrad1longE19.s1, sgpqcrad1longE19.s2 |
Description: | Case and dome temperatures for Longwave fluxes are sporadically noisy during these periods corresponding to additional noise in the longwave fluxes. No instrument DQRs exist documenting this problem. |
Measurements: | sgpqcrad1longE19.c1:
sgpqcrad1longE19.c2:
sgpqcrad1longE19.s1:
sgpqcrad1longE19.s2:
|