DQR ID | Subject | Data Streams Affected |
---|---|---|
D071002.1 | NSA/SKYRAD/C1 - Preventive maintenance | nsaskyrad60sC1.b1, nsaskyrad20sC1.a0 |
D071025.1 | NSA/SKYRAD/C1 - Reprocessed: Incorrect calibration coefficient | nsaskyrad20sC1.a0, nsaskyrad60sC1.b1 |
D080107.2 | NSA/SKYRAD/C1 - Instrument cleaning effects | nsaskyrad20sC1.a0, nsaskyrad60sC1.b1 |
D080222.3 | NSA/SKYRAD/C1 - IRT mirror heater failure | nsaskyrad20sC1.a0, nsaskyrad60sC1.b1 |
D080418.1 | NSA/SKYRAD/C1 - Direct and Diffuse incorrect | nsaskyrad20sC1.a0, nsaskyrad60sC1.b1 |
D080630.1 | NSA/SKYRAD/C1 - Tracker inoperative | nsaskyrad20sC1.a0, nsaskyrad60sC1.b1 |
D090318.1 | NSA/SKYRAD/C1 - IRT covered with snow | nsaskyrad60sC1.b1 |
D100211.1 | NSA/SKYRAD/C1 - Frosting/Condensation | nsaskyrad20sC1.a0, nsaskyrad60sC1.b1 |
D110817.6 | NSA/SKYRAD/C1 - Reprocessed: Incorrect calibration coefficients | nsaskyrad20sC1.a0, nsaskyrad60sC1.b1 |
D110902.1 | NSA/SKYRAD/C1 - Incorrect data during power outage | nsaskyrad20sC1.a0, nsaskyrad60sC1.b1 |
D111010.1 | NSA/SKYRAD/C1 - Noisy data due to grounding issue | nsaqcrad1longC1.c1, nsaskyrad60sC1.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/30/2007 | 1937 | 10/03/2007 | 0300 |
Subject: | NSA/SKYRAD/C1 - Preventive maintenance |
DataStreams: | nsaskyrad60sC1.b1, nsaskyrad20sC1.a0 |
Description: | Preventive maintenance to rebuild ventilator fan and exchange instruments during this time removed Skyrad instruments from operation. Most data collected are not scientifically relevant. |
Measurements: | nsaskyrad60sC1.b1:
nsaskyrad20sC1.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/03/2007 | 0359 | 10/25/2007 | 2148 |
Subject: | NSA/SKYRAD/C1 - Reprocessed: Incorrect calibration coefficient |
DataStreams: | nsaskyrad20sC1.a0, nsaskyrad60sC1.b1 |
Description: | Installed Skyrad PSP (downwelling shortwave hemispheric irradiance) had calibration coefficient Rs applied when it should have used Rsc. ARM is using IR corrected Rsc and CFc coefficients for PSPs deployed at all facilities. PSP #32026 should use RSc@45=8.7578 ?V/W/m? and not Rs@45=8.6335. Reprocessed to correct downwelling shortwave hemispheric irradiance (DS) by: DS(new) = (DS(old) / 115.83 ) * 114.18 Reprocessing completed March 2008. |
Measurements: | nsaskyrad60sC1.b1:
nsaskyrad20sC1.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/16/2007 | 0000 | 12/24/2007 | 2359 |
Subject: | NSA/SKYRAD/C1 - Instrument cleaning effects |
DataStreams: | nsaskyrad20sC1.a0, nsaskyrad60sC1.b1 |
Description: | Starting Dec 16th the Skyrad PIR1 and PIR2 (down_long_hemispheric)started differing by greater than 30W/m2 at times. This is much greater than expected for these instruments (typical is <10-15 W/m2). NSA OMIS reports and observer comments reveal that PM periodically clean the snow from the IRT and ventilator fans on Skyrad and Gndrad. In the case of the Skyrad PIRs, one lies in the wind shadow of the other and builds up more snow which is periodically cleaned out and then builds up again causing the PIR1 PIR2 discrepancies. Disregard down_long_hemispheric during this time period. However, data suggests PIR2 is most likely at fault - potentially use PIR1 data during this time period. |
Measurements: | nsaskyrad60sC1.b1:
nsaskyrad20sC1.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/16/2007 | 0000 | 02/11/2008 | 1135 |
Subject: | NSA/SKYRAD/C1 - IRT mirror heater failure |
DataStreams: | nsaskyrad20sC1.a0, nsaskyrad60sC1.b1 |
Description: | The heater of the IRT gold mirror failed due to a broken wire (that probably occurred on 12/13 during maintenance of the solar tracker) allowing snow and frost to accumulate on the mirror until it was cleared during daily preventative maintenance. The problem was corrected when the heater wires were replaced. |
Measurements: | nsaskyrad60sC1.b1:
nsaskyrad20sC1.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/17/2008 | 0632 | 04/23/2008 | 1815 |
Subject: | NSA/SKYRAD/C1 - Direct and Diffuse incorrect |
DataStreams: | nsaskyrad20sC1.a0, nsaskyrad60sC1.b1 |
Description: | The short_direct_normal (NIP) instrument and cable were damaged by tracker rewind malfunction. Tracker and NIP replaced on 4/22 2200GMT but tracker not operational until 4/23 1815GMT. Thus, down_short_diffuse_hemisp (8-48) was not shaded 4/22 to 4/23. Do not use 8-48 data from 4/22 to 4/23. NIP data is not recoverable during entire time but can be estimated using component sum for each 1 minute data: Direct = (global - diffuse)/cos(zenith angle) |
Measurements: | nsaskyrad60sC1.b1:
nsaskyrad20sC1.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/27/2008 | 2000 | 06/30/2008 | 1740 |
Subject: | NSA/SKYRAD/C1 - Tracker inoperative |
DataStreams: | nsaskyrad20sC1.a0, nsaskyrad60sC1.b1 |
Description: | Tracker stopped working at this site during the time range specified. Reason for failure is unknown. Tracker power cycled and restarted by maintenance 06/30 2008 1740 GMT. The short_direct_normal (NIP) and down_short_diffuse_hemisp (8-48) data are bad during this time. Data are not recoverable. |
Measurements: | nsaskyrad60sC1.b1:
nsaskyrad20sC1.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/20/2009 | 1900 | 02/24/2009 | 0130 |
Subject: | NSA/SKYRAD/C1 - IRT covered with snow |
DataStreams: | nsaskyrad60sC1.b1 |
Description: | The IRT mirror was buried with snow by the dew-line snow-blower. |
Measurements: | nsaskyrad60sC1.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/09/2010 | 1500 | 02/09/2010 | 2300 |
01/18/2010 | 0345 | 01/21/2010 | 0030 |
01/24/2010 | 1930 | 01/26/2010 | 0800 |
Subject: | NSA/SKYRAD/C1 - Frosting/Condensation |
DataStreams: | nsaskyrad20sC1.a0, nsaskyrad60sC1.b1 |
Description: | The down_long_hemisp1 varied up to 40 W/m^2 higher than down_long_hemisp2 during the affected time periods. Cold temperatures reported by the IRT suggest the issue is with PIR1. Icing or frost on PIR1 or within the ventilator may be to blame but ops reports can't substantiate this. Data from PIR1 should be avoided during these suspect time periods. |
Measurements: | nsaskyrad60sC1.b1:
nsaskyrad20sC1.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
07/20/2011 | 2229 | 07/25/2011 | 1843 |
Subject: | NSA/SKYRAD/C1 - Reprocessed: Incorrect calibration coefficients |
DataStreams: | nsaskyrad20sC1.a0, nsaskyrad60sC1.b1 |
Description: | On 7/20/2011 starting 2229GMT ops replaced the Skyrad shortwave instruments (PSP/8-48/NIP) corresponding to the datastreams (down_short_hemisp, down_short_diffuse_hemisp, short_direct_normal) but the calibration coefficients weren't changed until 07/25/2011 1842GMT. Data were reprocessed and archived in September 2011. |
Measurements: | nsaskyrad60sC1.b1:
nsaskyrad20sC1.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/27/2011 | 1700 | 05/02/2011 | 0100 |
Subject: | NSA/SKYRAD/C1 - Incorrect data during power outage |
DataStreams: | nsaskyrad20sC1.a0, nsaskyrad60sC1.b1 |
Description: | During a power outage the downwelling diffuse 8-48 and the downwelling shortwave direct were incorrect. The IRT was also affected during this time period. |
Measurements: | nsaskyrad60sC1.b1:
nsaskyrad20sC1.a0:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/02/2011 | 0022 | 09/24/2011 | 0145 |
Subject: | NSA/SKYRAD/C1 - Noisy data due to grounding issue |
DataStreams: | nsaqcrad1longC1.c1, nsaskyrad60sC1.b1 |
Description: | Improper grounding of the IRT serial cable connection caused significant noise in the signals of the other SKYRAD instruments. It was corrected by disconnecting the serial cable from the RocketPort. |
Measurements: | nsaskyrad60sC1.b1:
nsaqcrad1longC1.c1:
|